K150645 · Andersen Sterilizers, Inc. · JOJ · Nov 19, 2015 · General Hospital
Device Facts
Record ID
K150645
Device Name
AN1087 Dosimeter
Applicant
Andersen Sterilizers, Inc.
Product Code
JOJ · General Hospital
Decision Date
Nov 19, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
AN1087 Dosimeter is a single-use color change chemical indicator that is callbrated for sterlization temperature and used to verify adequate cumulative ethylene oxide exposure in the 3 hour sterilization cycle at 50°C in the Andersen Sterilizers EOGas 4 Ethylene Oxide Gas Sterilizer.
Device Story
AN1087 Dosimeter is a single-use chemical indicator for ethylene oxide (EO) sterilization. It consists of a pH-sensitive indicator ink within a glass capillary tube mounted on a plastic tray. During the 3-hour, 50°C sterilization cycle in the EOGas 4 Sterilizer, EO gas reacts with the indicator, causing a color change from yellow-orange to dark blue. The length of the blue color migration along the tube is proportional to the cumulative EO exposure. The device is calibrated so that the color reaching a specific mark indicates exposure sufficient to inactivate a 6-Log Bacillus atrophaeus biological indicator. It is used by sterilization technicians to verify cycle adequacy. The output is a visual color change observed by the operator after the cycle completes. It does not replace biological indicators but provides immediate visual confirmation of process parameters.
Clinical Evidence
Bench testing only. Performance characterized in a Chemical Indicator Evaluator Resistometer and EOGas 4 Sterilizer. Testing confirmed functionality (correlation between color change and cumulative EO exposure), endpoint color stability (stable 28 days at 20-25°C; 3 days at 50°C), and 3-year shelf life. Biocompatibility confirmed for non-patient-contacting device.
Technological Characteristics
Single-use chemical indicator; glass capillary tube with pH-sensitive indicator ink on plastic tray. Principle: chemical reaction with EO gas changes ink pH and color. Calibrated for 50°C ± 3°C, 35-90% relative humidity, 3-hour exposure. Non-patient contacting.
Indications for Use
Indicated for use as a single-use chemical indicator to verify adequate cumulative ethylene oxide exposure during 3-hour sterilization cycles at 50°C in Andersen Sterilizers EOGas 4 systems.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
Predicate Devices
AN87 Dosimeter (Pre-amendment device)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the most complete and the bottom profile being the least complete.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-000
November 19, 2015
Andersen Sterilizers, Inc. William K. Andersen, BE, MD, FAAOS President 3154 Caroline Dr Haw River, NC 27258
Re: K150645
Trade/Device Name: AN1087 Dosimeter Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: II Product Code: JOJ Dated: October 20, 2015 Received: October 22, 2015
Dear Dr. William Andersen,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{1}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Erin I. Keith -S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K150645
Device Name AN1087 Dosimeter
Indications for Use (Describe)
AN1087 Dosimeter is a single-use color change chemical indicator that is callbrated for sterlization temperature and used to verify adequate cumulative ethylene oxide exposure in the 3 hour sterilization cycle at 50°C in the Andersen Sterilizers EOGas 4 Ethylene Oxide Gas Sterilizer.
Critical process parameters for the cycle are summarized in Table 1.
Table 1. Critical parameters for the 3 hour cycle in the EOGas 4 Ethylene Oxide Gas Sterilizer
| Ethylene Oxide | Temperature | Relative Humidity | Ethylene Oxide Exposure Time | Total Cycle Time |
|----------------|-------------|-------------------|------------------------------|------------------|
| 17.6 g ± 5% | 50°C ± 3°C | 35-90% | 3 hours | 3.5 hours |
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|× Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary
| 5.1 Applicant's Name and Address | Andersen Sterilizers, Inc.<br>3154 Caroline Drive<br>Haw River, NC 27258 | |
|----------------------------------|-------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| 5.2 Contact Person | William K. Andersen, BE, MD, FAAOS<br>President<br>Phone: 336-376-8622; Fax: 336-376-5428 | |
| 5.3 Date of Preparation | October 20, 2015 | |
| 5.4 Device | Proprietary Name<br>Common Name<br>Classification<br>Product Code | AN1087 Dosimeter®<br>Indicator, Physical/Chemical Sterilization Process<br>Class II (21 CFR 880.2800) Chemical Indicator<br>JOJ |
The AN1087 Dosimeter is part of the refill kit for the EOGas 4 Ethylene Oxide Gas Sterilizer registered with the US Environmental Protection Agency (EPA #69340-7).
#### 5.5 Pre-amendment Device
AN87 Dosimeter® Device Name 510(k) number NA Manufacturer Andersen Sterilizers, Inc.
The AN87 Dosimeter was legally marketed prior to the Medical Device Act, and has been in continuous use with Andersen ethylene oxide sterilization systems since the early 1970's. No 510(k) number exists for the AN87 Dosimeter.
The purpose for this submission is to modify the sterilization temperature claim of the preamendment device - AN87 Dosimeter. No modifications were made to the technology or intended use.
#### 5.6 Device Description
The AN1087 Dosimeter is a single-use chemical indicator for cumulative ethylene oxide exposure. It is an accessory for the EOGas 40 Ethylene Oxide Gas Sterilizer.
The AN1087 Dosimeter contains a proprietary pH indicator in a glass capillary tube that is sealed on one end and mounted on a plastic tray. It is calibrated for a 50°C sterilization temperature, and responds to ethylene oxide concentration and sterilization time. With exposure to ethylene oxide, the indicator turns from yellow-orange to a dark blue color from the open end toward the closed end. The extent of the color change is proportional to the cumulative ethylene oxide exposure. The calibration mark represents adequate cumulative ethylene oxide
{5}------------------------------------------------
exposure to inactivate a 6-Log Bacillus atrophaeus biological indicator at the location of the AN1087 Dosimeter. The AN1087 Dosimeter is not a replacement for a biological indicator.
5.7 Indications for Use
AN1087 Dosimeter is a single-use color change chemical indicator that is calibrated for sterilization temperature and used to verify adequate cumulative ethylene oxide exposure in the 3 hour sterilization cycle at 50°C in the Andersen Sterilizers EOGas 4 Ethylene Oxide Gas Sterilizer.
Critical process parameters for the cycle are summarized in Table 5-1.
| Ethylene Oxide | Temperature | Relative<br>Humidity | Ethylene Oxide<br>Exposure Time | Total Cycle<br>Time |
|----------------|-------------|----------------------|---------------------------------|---------------------|
| 17.6 g ± 5% | 50°C ± 3°C | 35-90% | 3 hours | 3.5 hours |
Table 5-1. Critical parameters for the 3 hour cvcle in the EOGas 4 Ethylene Oxide Gas Sterilizer
#### 5.8 Substantial Equivalence Comparison
The AN1087 Dosimeter is substantially equivalent to the pre-amendment device - AN87 Dosimeter, because there is no difference in intended use, design, technical characteristics, and performance between them. The slight difference between two Dosimeters is the sterilization temperature indicated for use, but the difference raises no issues related to safety or effectiveness because they are calibrated and validated at the sterilization temperatures claimed. A comparison between the indicators is listed in Table 5-2.
| | AN87 Dosimeter | AN1087 Dosimeter | Comparison |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Intended<br>Use | Chemical indicator for<br>cumulative EO exposure | Chemical indicator for<br>cumulative EO exposure | Equivalent |
| Design | A pH indicator changes color<br>when exposed to EO;<br>The extent of the color changes<br>is proportional to cumulative<br>EO exposure;<br>Calibrated for use at 23±3°C. | A pH indicator changes color<br>when exposed to EO;<br>The extent of the color changes<br>is proportional to cumulative<br>EO exposure;<br>Calibrated for use at 50±3°C. | Equivalent |
| Technology | Chemical reactions with EO<br>changes the pH and thus the<br>color of the indicator ink | Chemical reactions with EO<br>changes the pH and thus the<br>color of the indicator ink | Equivalent |
| Performance | Accurately indicated EO<br>exposure; | Accurately indicated EO<br>exposure; | Equivalent |
Table 5-2. Comparison between AN1087 Dosimeter and the pre-amendment device
{6}------------------------------------------------
### 5.9 Performance Testing
The performance of the AN1087 Dosimeter was characterized in an Andersen Chemical Indicator Evaluator Resistometer as well as in an EOGas 4 Ethylene Oxide Gas Sterilizer using the 3 hour cycle at 50 C. Performance testing is summarized in Table 5-3.
| | | | Table 5-3. Summary of bench tests performed to demonstrate safety and effectiveness of the | |
|------------------|--|--|--------------------------------------------------------------------------------------------|--|
| AN1087 Dosimeter | | | | |
| Test | Description | Result |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Functionality | 1) Critical parameters include temperature, time, and gas<br>concentration under a relative humidity of 35-90%;<br>2) Performance parallels that of a biological indicator;<br>3) A correlation between the extent of the blue color formed and<br>the cumulative EO exposure;<br>4) A correlation between the measured vs. calculated EO exposure. | Pass |
| Biocompatibility | Not direct or indirect patient-contacting devices;<br>Non-toxic ingredients;<br>Provides reasonable assurance for safety | Pass |
| Endpoint Color<br>Stability | Stable for at least 28 days at 20-25°C and 3 days at 50°C;<br>Stability demonstrates reasonable assurance for effectiveness | Pass |
| Shelf Life | Maintains performance specifications throughout the stated shelf<br>life of 3 years | Pass |
In conclusion, the bench performance testing demonstrates that the AN1087 Dosimeter performs as intended to indicate cumulative EO exposure, and provides reasonable assurance of safety and effectiveness, equivalent to the legally marketed pre-amendment device, the AN87 Dosimeter.