STERITEC LEAD FREE STERILIZATION PROCESS INDICATOR TAPE
K122555 · Steritec Products Mfg Co, Inc. · JOJ · Dec 12, 2012 · General Hospital
Device Facts
Record ID
K122555
Device Name
STERITEC LEAD FREE STERILIZATION PROCESS INDICATOR TAPE
Applicant
Steritec Products Mfg Co, Inc.
Product Code
JOJ · General Hospital
Decision Date
Dec 12, 2012
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
SteriTec Green Tape can be used in 121°C gravity sterilizers for 30 minutes or pre-vacuum sterilizers operating at 132°C for 4 minutes and 135°C for 3 minutes. Stripes turn black after exposure to steam sterlization conditions, thus providing identification of processed items.
Device Story
SteriTec Green Tape is a chemical sterilization process indicator tape. It is applied to items prior to steam sterilization. Upon exposure to specified steam sterilization conditions (121°C gravity or 132°C/135°C pre-vacuum), the indicator stripes on the tape undergo a color change from green to black. This visual change allows healthcare personnel to distinguish between processed and unprocessed items. The device is used in clinical settings where steam sterilization is performed.
Clinical Evidence
Bench testing only.
Technological Characteristics
Chemical indicator tape; color-changing ink (green to black) responsive to steam sterilization parameters (temperature/time); dimensions vary by model (CI 159/1, CI 159/.5, CI 159/.75).
Indications for Use
Indicated for use as a steam sterilization process indicator to identify processed items in 121°C gravity sterilizers or 132°C/135°C pre-vacuum sterilizers.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
K040901 — STERITEC EMU-GRAPH, MODEL NO. CI 131/US · Steritec Products, Inc. · Sep 9, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with its wings spread, and three wavy lines below it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
### December 12, 2012
Mr. Jonathan Rutigliano Director, Regulatory Affairs SteriTec Products Manufacturing Company, Incorporated 74 Inverness Drive East Englewood, Colorado 80112
Re: K122555
Trade/Device Name: SteriTec Green Tape, Model No: Cl 159/1, CI 159/.5, CI 159/.75 Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: II Product Code: JOJ Dated: November 20, 2012 Received: November 21, 2012
Dear Mr. Rutigliano:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.
{1}------------------------------------------------
## Page 2 .- Mr. Rutigliano
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known): __
K122555
Device Name: SteriTec Green Tape, Model No .: CI 159/1, CI 159/.5, CI 159/.75
### Indications For Use:
SteriTec Green Tape can be used in 121°C gravity sterilizers for 30 minutes or pre-vacuum sterilizers operating at 132°C for 4 minutes and 135°C for 3 minutes. Stripes turn black after exposure to steam sterlization conditions, thus providing identification of processed items.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use X (21 CFR 801 Subpart C)
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE : . . . IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Elizabeth F. Claverie 2012.12.11,16:41:22 -05'00'
(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital
Division of Anesthesiology, General Devices Division of Anesthesions gyl
Infection Control, Dental Devices
..........
510(k) Number:_
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.