Use the 3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD in conjunction with the 3M™ Attest™ Auto-reader 490 having software version 4.0.0 or greater or a 3M™ Attest™ Mini Auto-reader 490M or a 3M™ Attest™ Auto-reader 490H as a standard method of routine monitoring and performance qualification of vaporized hydrogen peroxide sterilization processes in the follow systems: STERRAD 100S® Sterilization System, STERRAD NX® Sterilization System (Standard and Advanced cycles), STERRAD 100NX® Sterilization System (Standard, Flex, Express, and Duo cycles), STERRAD NX® with ALLClear® Technology Sterilization System (Standard and Advanced cycles), STERRAD 100NX® with ALLClear® Technology Sterilization System (Standard, Flex, Express, and Duo cycles), V-PRO® 1 Low Temperature Sterilization System (Lumen cycle), V-PRO® 1 Plus Low Temperature Sterilization System (Lumen and Non Lumen cycles), V-PRO® maX Low Temperature Sterilization System (Lumen, and Flexible cycles), V-PRO® 60 Low Temperature Sterilization System (Lumen, and Flexible cycles), V-PRO® maX 2 Low Temperature Sterilization System (Lumen, Flexible, and Fast Non Lumen cycles), V-PRO® s2 Low Temperature Sterilization System (Lumen, Flexible, and Fast cycles) SteroScope® Sterilization System
Device Story
Single-use challenge pack for monitoring vaporized hydrogen peroxide (VH2O2) sterilization; contains Geobacillus stearothermophilus biological indicator (BI) and Tri-Metric chemical indicator (CI). Pack design limits air removal and sterilant penetration to challenge sterilization efficacy. Used in clinical sterilization departments; processed in compatible low-temperature sterilizers. After cycle, BI is incubated in 3M Attest Auto-reader (490, 490M, or 490H); reader detects fluorescence indicating sterilization failure. CI provides visual ACCEPT/REJECT status based on exposure time, temperature, and concentration. Enables rapid verification of sterilization process, ensuring medical instruments are safe for patient use.
Clinical Evidence
Bench testing only. Performance verified in SteroScope® Sterilization System using full and incomplete cycles. Endpoints included CI color change (b* value) and BI fluorescence response. All tests met acceptance criteria.
Technological Characteristics
Clear plastic shell with two channels and foil lid. Contains Geobacillus stearothermophilus spores (≥ 1x10^6) and Tri-Metric chemical indicator. Connectivity via 3M Attest Auto-reader (490, 490M, 490H) with software version 4.0.0+. Complies with ANSI/AAMI/ISO 11138-1:2017.
Indications for Use
Indicated for routine monitoring and performance qualification of vaporized hydrogen peroxide sterilization processes in specified STERRAD, V-PRO, and SteroScope systems. Used in clinical settings by healthcare personnel.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
K173435 — 3M Attest Rapid Readout Biological Indicator · 3M Company · Feb 9, 2018
K160546 — 3M Attest Rapid Readout Biological Indicator · 3M Company · Jun 9, 2016
K211705 — 3M Attest Rapid Readout Biological Indicator, 1295 · 3M Company · Jun 28, 2022
K233814 — 3M Attest Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack (1295PCD); 3M Attest Auto-reader (490); 3M Attest Auto-reader (490H); 3M Attest Mini Auto-reader (490M) · 3M Company · Mar 15, 2024
K210277 — 3M Attest Rapid Readout Biological Indicator, 1295 · 3M Company · Mar 3, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 27, 2024
3M Company Samantha Miller Regulatory Affairs Associate 3M Center 2510 Conway Avenue, Building 275-5W-06 St. Paul, Minnesota 55144
Re: K241542
Trade/Device Name: 3MTM Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD (1295PCD) Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: FRC Dated: May 30, 2024 Received: May 31, 2024
Dear Samantha Miller:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Christopher K. Dugard -S
Christopher K. Dugard, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices
{2}------------------------------------------------
OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K241542
Device Name
3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD (1295PCD)
Indications for Use (Describe)
3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack (1295PCD): Use the 3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD in conjunction with the 3M™ Attest™ Auto-reader 490 having software version 4.0.0 or greater or a 3M™ Attest™ Mini Auto-reader 490M or a 3M™ Attest™ Auto-reader 490H as a standard method of routine monitoring and performance qualification of vaporized hydrogen peroxide sterilization processes in the follow systems:
STERRAD 100S® Sterilization System,
STERRAD NX® Sterilization System (Standard and Advanced cycles),
STERRAD 100NX® Sterilization System (Standard, Flex, Express, and Duo cycles),
STERRAD NX® with ALLClear® Technology Sterilization System (Standard and Advanced cycles), STERRAD 100NX® with ALLClear® Technology Sterilization System (Standard, Flex, Express, and Duo cycles),
V-PRO® 1 Low Temperature Sterilization System (Lumen cycle),
V-PRO® 1 Plus Low Temperature Sterilization System (Lumen and Non Lumen cycles),
V-PRO® maX Low Temperature Sterilization System (Lumen, and Flexible cvcles),
V-PRO® 60 Low Temperature Sterilization System (Lumen, and Flexible cycles),
V-PRO® maX 2 Low Temperature Sterilization System (Lumen, Flexible, and Fast Non Lumen cycles),
V-PRO® s2 Low Temperature Sterilization System (Lumen, Flexible, and Fast cycles) SteroScope® Sterilization System
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the 3M logo. The logo is red and features the number "3" followed by the letter "M". The font is bold and sans-serif. The logo is simple and recognizable.
# 510(k) Summary for 3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD
Sponsor Information:
3M Company 3M Center, Bldg. 275-5W-06 St. Paul, MN 55144-1000
Contact: Samantha Miller Regulatory Affairs Associate
Date of Summary: May 30, 2024
510(k) Reference: K241542
{5}------------------------------------------------
#### PREMARKET NOTIFICATION [510(k)] 3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD
#### 1. Device Name and Classification:
| Common or Usual Name: | Sterilization Biological Indicator |
|------------------------|-------------------------------------------------------------------------------------|
| Proprietary Name: | 3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide<br>Clear Challenge Pack 1295PCD |
| Classification Name: | Indicator, biological sterilization process |
| Device Classification: | Class II, 21 CFR § 880.2800(a) |
| Product Code: | FRC |
#### 2. Predicate Device:
K233814, 3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD
#### 3. Description of Device:
The 3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD is designed as a standard method of rapid and reliable routine monitoring and performance qualification of vaporized hydrogen peroxide sterilization processes.
The 1295PCD Challenge Pack consists of a clear plastic shell, with two individual channels allowing for air removal and sterilant penetration. Those two individual channels connect to a cavity containing the monitoring products and all are covered by a foil lid. The 1295PCD is a single-use device.
Each 1295PCD Challenge Pack contains a 3M™ Attest™ Rapid Readout Biological Indicator (BI) 1295 (pink cap) and a 3MTM Attest™ Vaporized Hydrogen Peroxide Tri-Metric Chemical Indicator (CI) 1348. The 1348 CI verifies that the stated values for the three critical parameters of exposure time, temperature, and concentration of vaporized hydrogen peroxide have been achieved and offers an ACCEPT or REJECT reading. The 1295 BI is specifically designed for rapid and reliable routine monitoring of vaporized hydrogen peroxide sterilization processes when used in conjunction with the 3MTM Attest™ Auto-reader 490 having software version 4.0.0 or greater or a 3MTM Attest™ Mini Auto-reader 490M or a 3M™ Attest™ Auto-reader 490H. The presence of fluorescence within the specified incubation time for the 1295 BI in the 490, 490M, or the 490H Auto-reader indicates a sterilization process failure. A printed chemical process indicator is present on the cap of the 1295 BI and is visible through the clear plastic shell of the challenge pack. The process indicator turns color upon exposure to vaporized hydrogen peroxide and is used by the customer to verify that the challenge pack was exposed to vaporized hydrogen peroxide.
{6}------------------------------------------------
#### 4. Indications for Use
Use the 3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD in conjunction with the 3M™ Attest™ Auto-reader 490 having software version 4.0.0 or greater or a 3M™ Attest™ Mini Auto-reader 490M or a 3M™ Attest™ Auto-reader 490H as a standard method of routine monitoring and performance qualification of vaporized hydrogen peroxide sterilization processes in the follow systems:
STERRAD 100S® Sterilization System
STERRAD NX® Sterilization System (Standard and Advanced cycles)
STERRAD 100NX® Sterilization System (Standard, Flex, Express, and Duo cycles)
STERRAD NX® with ALLClear® Technology Sterilization System (Standard and Advanced cycles)
STERRAD 100NX® with ALLClear® Technology Sterilization System (Standard, Flex, Express, and Duo cycles)
V-PRO® 1 Low Temperature Sterilization System (Lumen cycle)
V-PRO® 1 Plus Low Temperature Sterilization System (Lumen and Non Lumen cvcles)
V-PRO® maX Low Temperature Sterilization System (Lumen, Non Lumen, and Flexible cycles)
V-PRO® 60 Low Temperature Sterilization System (Lumen, Non Lumen, and Flexible cycles)
V-PRO® maX 2 Low Temperature Sterilization System (Lumen, Non Lumen, Flexible, and Fast Non Lumen cycles)
V-PRO® s2 Low Temperature Sterilization System (Lumen, Flexible, and Fast cycles)
SteroScope® Sterilization System
{7}------------------------------------------------
| Feature | Submission Device:<br>3MTM AttestTM Super Rapid<br>Vaporized Hydrogen Peroxide<br>Clear Challenge Pack 1295PCD | Predicate Device (K233814):<br>3MTM AttestTM Super Rapid<br>Vaporized Hydrogen Peroxide<br>Clear Challenge Pack 1295PCD | Comparison |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|
| Device<br>Models | 1295PCD | 1295PCD | Identical |
| Intended Use | Single Use sterilization process<br>indicator | Single Use sterilization process<br>indicator | Identical |
| Indications<br>for use | Use the 3M™ Attest™ Super<br>Rapid Vaporized Hydrogen<br>Peroxide Clear Challenge Pack<br>1295PCD in conjunction with the<br>3MTM Attest™ Auto-reader 490<br>having software version 4.0.0 or<br>greater or a 3M™ Attest™ Mini<br>Auto-reader 490M or a 3MTM<br>Attest™ Auto-reader 490H as a<br>standard method of routine<br>monitoring and performance<br>qualification of vaporized<br>hydrogen peroxide sterilization<br>processes in the follow systems:<br>STERRAD 100S® Sterilization<br>System<br>STERRAD NX® Sterilization<br>System (Standard and Advanced<br>cycles)<br>STERRAD 100NX® Sterilization<br>System (Standard, Flex, Express,<br>and Duo cycles)<br>STERRAD NX® with<br>ALLClear® Technology<br>Sterilization System (Standard<br>and Advanced cycles)<br>STERRAD 100NX® with<br>ALLClear® Technology<br>Sterilization System (Standard,<br>Flex, Express, and Duo cycles)<br>V-PRO® 1 Low Temperature<br>Sterilization System (Lumen<br>cycle)<br>V-PRO® 1 Plus Low<br>Temperature Sterilization System<br>(Lumen and Non Lumen cycles)<br>V-PRO® maX Low Temperature<br>Sterilization System (Lumen,<br>Non Lumen, and Flexible cycles) | Use the 3M™ Attest™ Super<br>Rapid Vaporized Hydrogen<br>Peroxide Clear Challenge Pack<br>1295PCD in conjunction with the<br>3MTM AttestTM Auto-reader 490<br>having software version 4.0.0 or<br>greater or a 3M™ Attest™ Mini<br>Auto-reader 490M or a 3MTM<br>Attest™ Auto-reader 490H as a<br>standard method of routine<br>monitoring and performance<br>qualification of vaporized<br>hydrogen peroxide sterilization<br>processes in the follow systems:<br>STERRAD 100S® Sterilization<br>System<br>STERRAD NX® Sterilization<br>System (Standard and Advanced<br>cycles)<br>STERRAD 100NX® Sterilization<br>System (Standard, Flex, Express,<br>and Duo cycles)<br>STERRAD NX® with<br>ALLClear® Technology<br>Sterilization System (Standard<br>and Advanced cycles)<br>STERRAD 100NX® with<br>ALLClear® Technology<br>Sterilization System (Standard,<br>Flex, Express, and Duo cycles)<br>V-PRO® 1 Low Temperature<br>Sterilization System (Lumen<br>cycle)<br>V-PRO® 1 Plus Low<br>Temperature Sterilization System<br>(Lumen and Non Lumen cycles)<br>V-PRO® maX Low Temperature<br>Sterilization System (Lumen,<br>Non Lumen, and Flexible cycles) | Addition of<br>SteroScope®<br>Sterilization<br>System |
| Feature | Submission Device:<br>3M™ Attest™ Super Rapid<br>Vaporized Hydrogen Peroxide<br>Clear Challenge Pack 1295PCD | Predicate Device (K233814):<br>3M™ Attest™ Super Rapid<br>Vaporized Hydrogen Peroxide<br>Clear Challenge Pack 1295PCD | Comparison |
| | V-PRO® 60 Low Temperature<br>Sterilization System (Lumen,<br>Non Lumen, and Flexible cycles)<br>V-PRO® maX 2 Low<br>Temperature Sterilization System<br>(Lumen, Non Lumen, Flexible,<br>and Fast Non Lumen cycles)<br>V-PRO® s2 Low Temperature<br>Sterilization System (Lumen,<br>Non Lumen, Flexible, and Fast<br>cycles)<br>SteroScope® Sterilization<br>System | V-PRO® 60 Low Temperature<br>Sterilization System (Lumen,<br>Non Lumen, and Flexible cycles)<br>V-PRO® maX 2 Low<br>Temperature Sterilization System<br>(Lumen, Non Lumen, Flexible,<br>and Fast Non Lumen cycles)<br>V-PRO® s2 Low Temperature<br>Sterilization System (Lumen,<br>Non Lumen, Flexible, and Fast<br>cycles) | |
| Biological<br>Indicator | 3M™ Attest™ Rapid Readout<br>Biological Indicator 1295 with ≥<br>1 x106 Geobacillus<br>stearothermophilus spores | 3M™ Attest™ Rapid Readout<br>Biological Indicator 1295 with ≥<br>1 x106 Geobacillus<br>stearothermophilus spores | Identical |
| Chemical<br>Indicator | 3M™ Attest™ Vaporized<br>Hydrogen Peroxide Tri- Metric<br>Chemical Indicator 1348. | 3M™ Attest™ Vaporized<br>Hydrogen Peroxide Tri- Metric<br>Chemical Indicator 1348. | Identical |
| General<br>Design /<br>Construction | Clear plastic tray with channels<br>to the cavity containing the<br>indicators, acting as a challenge<br>to limit air removal and VH2O2<br>penetration, sealed with a foil lid. | Clear plastic tray with channels<br>to the cavity containing the<br>indicators, acting as a challenge<br>to limit air removal and VH2O2<br>penetration, sealed with a foil lid. | Identical |
| Process<br>Indicator | Process indicator present on<br>1295 BI cap is visible through<br>PCD and turns color from blue<br>towards pink upon vaporized<br>hydrogen peroxide exposure. | Process indicator present on 1295<br>BI cap is visible through PCD<br>and turns color from blue towards<br>pink upon vaporized hydrogen<br>peroxide exposure. | Identical |
## 5. Technological Characteristic Comparison Table
{8}------------------------------------------------
## PREMARKET NOTIFICATION [510(k)] 3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD
{9}------------------------------------------------
## 6. Nonclinical Comparison to the Predicate Device
The 3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD is identical to the 3M-owned predicate device (K233814) with respect to design, materials, performance specifications and fundamental technology. The key difference between the submission device and the predicate device is that the submission device adds an indication for pack control monitoring of the SteroScope® Sterilization System.
To demonstrate performance in the newly claimed sterilizer and cycles, nonclinical testing was performed in accordance with the FDA Guidance for Industry and FDA Staff: Biological Indicator (BI) Premarket Notification [510(k)] Submissions, issued October 04, 2007, and ANSI/AAMI/ISO 11138-1:2017 Sterilization of health care products-Biological Indicators- Part 1: General requirements (FDA Recognition Number 14-502).
Reference Table 6.1 for testing completed in SteroScope® Sterilization System.
| Test Name | Purpose | Acceptance Criteria | Result |
|------------------------------------|------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Full Cycle<br>Performance | Verify performance<br>in the full normal<br>cycle of a loaded<br>SteroScope®<br>Sterilization System. | Chemical Indicator will change color in<br>the ACCEPT region of the indicator<br>window to a b* value equal to or<br>greater than -18 (PASS) | Pass |
| | | Biological Indicator will result in 100%<br>florescence NEGATIVE response | Pass |
| Incomplete<br>Cycle<br>Performance | Verify performance<br>in an incomplete<br>normal cycle of a<br>loaded SteroScope®<br>Sterilization System. | Chemical Indicator will change color in<br>the ACCEPT region of the indicator<br>window to a b* value less than -20<br>(FAIL) | Pass |
| | | Biological Indicator will result in 100%<br>florescence POSITIVE response | Pass |
Table 6.1 Summary of Tri-Metric Chemical Indicator Nonclinical Testing
# 7. Conclusion
Based on the non-clinical performance data, the 3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD is as safe, as effective, and performs as well as the legally marketed predicate, the 3MTM Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD cleared under K233814, Class II (21 CFR 880.2800), product code FRC.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.