STERRAD CYCLESURE 24 Biological Indicator

K151971 · Advanced Sterilization Products · FRC · Oct 29, 2015 · General Hospital

Device Facts

Record IDK151971
Device NameSTERRAD CYCLESURE 24 Biological Indicator
ApplicantAdvanced Sterilization Products
Product CodeFRC · General Hospital
Decision DateOct 29, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2800
Device ClassClass 2

Indications for Use

The STERRAD CYCLESURE 24 Biological Indicator is intended to be used as a standard method for frequent monitoring of the following STERRAD Sterilization Systems and has been validated for use in health care facilities: - STERRAD 100S - STERRAD 50 - STERRAD 200 - STERRAD NX® - STERRAD 100NX® (Standard, Flex and Express Cycles) - For STERRAD 100NX DUO Cycle in the United States, the STERRAD CYCLESURE 24 Biological Indicator should only be used in a test pack configuration (REF 20243).

Device Story

STERRAD CYCLESURE 24 Biological Indicator (BI) is a self-contained, stand-alone monitor for STERRAD sterilization processes. Device contains glass fiber disc with minimum 1x10^6 Geobacillus stearothermophilus spores and glass ampoule of nutrient growth medium. Cap features two small circular openings allowing hydrogen peroxide vapor diffusion into vial; openings restrict diffusion to simulate sterilization challenge. Used in healthcare facilities to verify sterilization efficacy. After sterilization cycle, BI is activated to release growth medium; incubation at 55-60°C for 24-72 hours allows spore growth if sterilization failed. Chemical indicator on cap changes color (red to golden yellow/bronze) to confirm exposure to hydrogen peroxide. Provides healthcare providers with objective verification of sterilization cycle success; ensures patient safety by confirming instruments are sterile before reuse.

Clinical Evidence

Bench testing only. Stability studies verified that spore population and resistance characteristics (D-value, survival time, kill time) meet specifications throughout the extended 9-month shelf life.

Technological Characteristics

Self-contained biological indicator. Materials: glass fiber disc carrier, glass ampoule with nutrient growth medium, cap with diffusion-restricting openings. Biological agent: Geobacillus stearothermophilus (ATCC 7953). Resistance: D-value ≥ ~1 second at 2.5 mg/L H2O2. Incubation: 55-60°C for 24-72 hours. Chemical indicator: color-changing (red to golden yellow/bronze) upon H2O2 exposure.

Indications for Use

Indicated for routine monitoring of STERRAD sterilization systems (100S, 50, 200, NX, 100NX) in healthcare facilities. For STERRAD 100NX DUO cycle, must be used in test pack configuration.

Regulatory Classification

Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 October 29, 2015 Advanced Sterilization Products Ms. Sun Choi Regulatory Affairs Specialist IV 33 Technology Dr. Irvine, CA 92618 Re: K151971 Trade/Device Name: STERRAD® CYCLESURE® 24 Biological Indicator Regulation Number: 21 CFR 880.2800 Regulation Name: Biological Sterilization Process Indicator Regulatory Class: II Product Code: FRC Dated: October 13, 2015 Received: October 14, 2015 Dear Ms. Choi: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Page 2 - Ms. Choi Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, **Tejashri Purohit-Sheth, M.D.** Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) ## K151971 Device Name STERRAD® CYCLESURE® 24 Biological Indicator Indications for Use (Describe) The STERRAD CYCLESURE 24 Biological Indicator is intended to be used as a standard method for frequent montoring of the following STERRAD Sterlization Systems and has been validated for use in health care facilities: - . STERRAD 100S - o STERRAD 50 - . STERRAD 200 - o STERRAD NX® - . STERRAD 100NX® (Standard, Flex and Express Cycles) - For STERRAD 100NX DUO Cycle in the United States, the STERRAD CYCLESURE 24 Biological 0 Indicator should only be used in a test pack configuration (REF 20243). | Type of Use (Select one or both, as applicable) | | |---------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------| | <div style="display:inline-block"><input type="checkbox"/></div> Prescription Use (Part 21 CFR 801 Subpart D) | <div style="display:inline-block"><input checked="" type="checkbox"/></div> Over-The-Counter Use (21 CFR 801 Subpart C) | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." 20 {3}------------------------------------------------ # 510(k) Summary #### I. SUBMITTER Advanced Sterilization Products 33 Technology Drive Irvine, CA 92618 Contact Person: Sun Choi Regulatory Affairs Specialist IV Telephone: (949) 453-6378 Fax: (949) 798-3900 Date Prepared: July 13, 2015 #### II. DEVICE Name of Device: STERRAD® CYCLESURE® 24 Biological Indicator Common or Usual Name: Biological Indicator Classification Name: Biological Sterilization Process Indicator (21 CFR 880.2800) Regulatory Class: II Product Code: FRC #### III. PREDICATE DEVICE STERRAD CYCLESURE 24 Biological Indicator K123017 No reference devices were used in this submission. #### IV. DEVICE DESCRIPTION The STERRAD CYCLESURE 24 Biological Indicator (BI) is a self-contained stand- alone biological monitor intended for the routine monitoring of the STERRAD Sterilization Process. It consists of a glass fiber disc containing a minimum of 1x106 Geobacillus stearothermophilus spores, a glass ampoule containing nutrient growth medium, a cap and liner closing the vial and a chemical indicator on top of the cap. The cap contains two small circular openings that allow for diffusion of hydrogen peroxide vapor into the vial. The relatively small size of the circular openings serves as a restriction to this diffusion. {4}------------------------------------------------ #### V. INDICATIONS FOR USE The STERRAD CYCLESURE 24 Biological Indicator is intended to be used as a standard method for frequent monitoring of the following STERRAD Sterilization Systems and has been validated for use in health care facilities: - STERRAD 100S - STERRAD 50 - o STERRAD 200 - STERRAD NX® ● - STERRAD 100NX® (Standard, Flex and Express Cycles) ● - For STERRAD 100NX DUO Cycle in the United States, the STERRAD O CYCLESURE 24 Biological Indicator should only be used in a test pack configuration (REF 20243). ### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The proposed modification is to extend the labeled shelf life of the STERRAD CYCLESURE 24 BI. The intended use/indications for use, technological characteristics, design, materials, and principles of operation have not changed. Refer to Table 1 for comparison between modified and predicate devices for the STERRAD CYCLESURE 24 BI. | | Predicate Device:<br>STERRAD CYCLESURE 24 BI<br>(K123017) | Modified Device:<br>STERRAD<br>CYCLESURE 24 BI | |---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------| | Intended Use /<br>Indications for Use | The STERRAD CYCLESURE 24 Biological<br>Indicator is intended to be used as a standard<br>method for frequent monitoring of the following<br>STERRAD Sterilization Systems:<br>• STERRAD 100S<br>• STERRAD 50<br>• STERRAD 200<br>• STERRAD NX<br>• STERRAD 100NX (Standard, Flex and<br>Express Cycles)<br>o For STERRAD 100NX DUO Cycle in<br>the United States, the STERRAD<br>CYCLESURE 24 Biological Indicator<br>should only be used in a test pack<br>configuration (REF 20243). | Same as predicate | Table 1: Comparison between Modified and Predicate Devices {5}------------------------------------------------ | | Predicate Device:<br>STERRAD CYCLESURE 24 BI<br>(K123017) | Modified Device:<br>STERRAD<br>CYCLESURE 24 BI | |----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------| | Technological<br>Characteristics | The STERRAD CYCLESURE 24 BI is a self-<br>contained stand- alone biological monitor<br>intended for the routine monitoring of the<br>STERRAD Sterilization Process. It consists of<br>a glass fiber disc containing a minimum of<br>1x10⁶ Geobacillus stearothermophilus spores, a<br>glass ampoule containing nutrient growth<br>medium, a cap and liner closing the vial and a<br>chemical indicator on top of the cap. The cap<br>contains two small circular openings that allow<br>for diffusion of hydrogen peroxide vapor into<br>the vial. The relatively small size of the circular<br>openings serves as a restriction to this diffusion. | Same as predicate | | Organism | Geobacillus stearothermophilus ATCC 7953 | Same as predicate | | Viable Spore Population | Minimum of 1 x 10⁶ spores/BI | Same as predicate | | Carrier Material | Glass Fiber | Same as predicate | | Resistance Characteristics | D-value at 2.5 mg/L of hydrogen peroxide:<br>≥ ~ 1 second<br>Survival Time: calculated based on D-value<br>Kill Time: 60 seconds | Same as predicate | | Incubation Temperature | 55 - 60 °C | Same as predicate | | Incubation Time | 24 hours to 3 days (72 hours) | Same as predicate | | Chemical Indicator | A throughput indicator, changes color from red<br>to golden yellow or bronze to indicate exposure<br>to hydrogen peroxide | Same as predicate | | Shelf-life | 6 months | 9 months | {6}------------------------------------------------ #### VII. PERFORMANCE DATA The modified and predicate devices have the same intended use/indications for use, technological characteristics, design, materials, and principles of operation. The proposed modification is to extend the labeled shelf life of the STERRAD CYCLESURE 24 BI from 6 months to 9 months. Stability studies were conducted to demonstrate that the specifications (D-value, survival time, kill time, spore population, and positive BI color test) are maintained throughout the proposed labeled shelf life of the modified device. Refer to Table 2 for the summary of stability studies. | Study Type | Description | Results | |------------|---------------------------------------------------------------------------------------------------------------------------------|---------| | Stability | Verification of spore population for 9 month labeled<br>product shelf life | Passed | | Stability | Verification of resistance characteristics (D-value,<br>survival time, and kill time) for 9 month labeled<br>product shelf life | Passed | Table 2: Summary of Performance Testing # VIII. CONCLUSIONS Stability study results demonstrate that the specifications (D-value, survival time, kill time, spore population, and positive BI color test) are maintained throughout the proposed labeled shelf life (9 months) of the STERRAD CYCLESURE 24 BI. Modified and predicate devices have the same intended use/indications for use, technological characteristics, design, materials, and principles of operation. Therefore, Advanced Sterilization Products considers the modified device with the extended shelf life to be substantially equivalent to its predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...