STERRAD CYCLESURE 24 BIOLOGICAL INDICATOR MODEL 14324
K123017 · Advanced Sterilization Products · FRC · Dec 26, 2012 · General Hospital
Device Facts
Record ID
K123017
Device Name
STERRAD CYCLESURE 24 BIOLOGICAL INDICATOR MODEL 14324
Applicant
Advanced Sterilization Products
Product Code
FRC · General Hospital
Decision Date
Dec 26, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
The STERRAD® CYCLESURE® 24 Biological Indicator is intended to be used as a standard method for frequent monitoring of the following STERRAD® Sterilization Systems and Cycles: STERRAD® 100S . STERRAD® 50 . STERRAD® 200 ◆ STERRAD® NX® . . STERRAD® 100NX® (Standard, Flex and Express Cycles) 0 For STERRAD® 100NX® DUO Cycle in the United States, the STERRAD® CYCLESURE® 24 Biological Indicator should only be used in a test pack configuration (REF 20243).
Device Story
Self-contained biological monitor for STERRAD® sterilization systems; contains glass fiber disc with Geobacillus stearothermophilus spores, nutrient growth medium ampoule, and chemical indicator. Hydrogen peroxide vapor diffuses into vial through restricted circular openings in cap. Device monitors sterilization cycle efficacy; if spores survive, growth medium indicates failure. Used in clinical settings by healthcare staff to ensure instrument sterility. Provides objective verification of sterilization process; helps prevent use of non-sterile instruments; improves patient safety by reducing infection risk.
Clinical Evidence
Bench testing only. Studies included evaporation, positive BI color verification, bacteriostasis, BI validation in STERRAD® systems (dose response), minimum incubation time verification, growth promotion, and spore resistance testing. All tests passed.
Technological Characteristics
Self-contained biological indicator. Components: glass fiber disc with Geobacillus stearothermophilus spores (min 1x10^6), nutrient growth medium in glass ampoule, cap with restricted diffusion openings, chemical indicator. Sterilant: hydrogen peroxide. No electronic components or software.
Indications for Use
Indicated for routine monitoring of STERRAD® sterilization systems (100S, 50, 200, NX, 100NX) to verify sterilization efficacy. For 100NX DUO cycle, must be used in test pack configuration (REF 20243). No specific patient population restrictions.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
K071014 — MODIFICATION TO STERRAD CYCLESURE BIOLOGICAL INDICATOR · Advanced Sterilization Products · May 24, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
K123017
# 510(k) Summary
# DEC 2 6 2012
Applicant's Name, Address, Telephone, FAX, Contact Person Advanced Sterilization Products, Division of Ethicon, Inc. 33 Technology Drive Irvine, CA 92618
## Contact Person
Nancy Chu Manager. Regulatory Affairs Tel: (949) 453-6435 Fax: (949) 789-3900
Summary Date: November 14, 2012
# 1. CLASSIFICATION, COMMON OR USUAL NAME, DEVICE NAME
| Classification Name: | Biological Sterilization Process Indicator |
|----------------------------|---------------------------------------------|
| Common/Usual Name: | Biological Indicator |
| Product Classification: | II |
| Product Code: | FRC |
| Classification Regulation: | 21 CFR 880.2800 |
| Proprietary Name: | STERRAD® CYCLESURE® 24 Biological Indicator |
#### 2. PREDICATE DEVICES
STERRAD® CYCLESURE® 24 Biological Indicator, K111375, 9/14/2012
### 3. INDICATIONS FOR USE
The STERRAD® CYCLESURE® 24 Biological Indicator is intended to be used as a standard method for frequent monitoring of the following STERRAD® Sterilization Systems and Cycles:
- STERRAD® 100S .
- STERRAD® 50 .
- STERRAD® 200 ◆
- STERRAD® NX® .
- .
- STERRAD® 100NX® (Standard, Flex and Express Cycles)
0 For STERRAD® 100NX® DUO Cycle in the United States, the STERRAD® CYCLESURE® 24 Biological Indicator should only be used in a test pack configuration (REF 20243).
{1}------------------------------------------------
# 4. DESCRIPTION OF DEVICE
The STERRAD® CYCLESURE® 24 Biological Indicator is a self-contained standalone biological monitor intended for the routine monitoring of the STERRAD® Sterilization Process. It consists of a glass fiber disc containing a minimum of 1x10° Geobacillus stearothermophilus spores, a glass ampoule containing nutrient growth medium, a cap and liner closing the vial and a chemical indicator on top of the cap. The cap contains two small circular openings that allow for diffusion of hydrogen peroxide vapor into the vial. The relatively small size of the circular openings serves as a restriction to this diffusion.
## 5. SUMMARY OF NONCLINICAL TESTS
Both the modified and predicate devices have the same intended use, the same technological characteristics, the same operating principles, and utilize the same sterilant (hydrogen peroxide).
Testing was conducted to confirm that the STERRAD® CYCLESURE® 24 Biological Indicator performs as intended when used with the STERRRAD® Sterilization Systems. The following tests were performed:
| Studies Performed | Results |
|-----------------------------------------------------|---------|
| Evaporation | Passed |
| Verification of Positive BI Color | Passed |
| Bacteriostasis | Passed |
| BI Validation in the STERRAD® Sterilization Systems | Passed |
| (Dose Response) | |
| Verification of Minimum Incubation Time | Passed |
| Growth Promotion | Passed |
| Spore Resistance Testing | Passed |
## 6. DESCRIPTION OF CHANGE
The change of the material grade of the outer vial resin (relative to the predicate device, K111375) is shown in this submission to work as intended with the STERRAD® Sterilization Systems.
## 7. OVERALL PERFORMANCE CONCLUSIONS
Performance testing demonstrated that the modified STERRAD® CYCLESURE® 24 Biological Indicator is substantially equivalent to the predicate device (K111375) because they have the same intended use, the same technological characteristics, the same operating principles, and are exposed to the same sterilant (hydrogen peroxide).
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 26, 2012
Ms. Nancy Chu Manager, Regulatory Affairs Advanced Sterilization Products 33 Technology Drive IRVINE CA 92618
Re: K123017
Trade/Device Name: STERRAD® CYCLESURE® 24 Biological Indicator Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: II Product Code: FRC Dated: November 28, 2012 Received: November 29, 2012
Dear Ms. Chu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kwame O. Ulmer
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
K123017 510(k) Number (if known): TBD
Device Name: STERRAD® CYCLESURE® 24 Biological Indicator
#### Indication for Use:
The STERRAD® CYCLESURE® 24 Biological Indicator is intended to be used as standard method for frequent monitoring of the following STERRAD® a Sterilization Systems and Cycles:
- STERRAD® 100S ●
- STERRAD® 50
- STERRAD® 200
- STERRAD® NX®
- STERRAD® 100NX® (Standard, Flex and Express Cycles)
- o For STERRAD® 100NX® DUO Cycle, the STERRAD® CYCLESURE® 24 Biological Indicator should only be used in a test pack configuration (REF 20243).
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
**Elizabeth F. Claverie**
2012.12.19 20:42:14 -05'00'
KI23012
Page 1 of 1
(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital
Division of Anesthesiology, General Devices Division of Anesthestology, Dental Devices
510(k) Number,
18
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.