Clever Forehead Thermometer / Clever Medical Forehead Thermometer (Model TD-1241)

K200946 · Taidoc Technology Corporation · FLL · Apr 5, 2022 · General Hospital

Device Facts

Record IDK200946
Device NameClever Forehead Thermometer / Clever Medical Forehead Thermometer (Model TD-1241)
ApplicantTaidoc Technology Corporation
Product CodeFLL · General Hospital
Decision DateApr 5, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2910
Device ClassClass 2
AttributesPediatric

Indications for Use

The Clever Forehead Thermometer/Clever Medical Forehead Thermometer (TD-1241) is an infrared thermometer intended for the non-contact measurement of human body temperature in people of all ages and may be used by medical professionals or by consumers in a home environment.

Device Story

Handheld, battery-powered infrared thermometer; detects infrared energy emitted from forehead (1.2–5 cm distance); converts energy to body temperature via thermopile/thermistor sensing and microcontroller processing; displays results on LCD. Features Bluetooth connectivity for data transmission to mobile apps and PC-based health management systems for trend tracking. Used by medical professionals or consumers in home environments. Output provides temperature readings to assist in clinical decision-making and patient health monitoring.

Clinical Evidence

Clinical accuracy study conducted on 250 subjects across three age groups (infants, children, adults) per ASTM E1965-98. Compared against WelchAllyn SureTemp PLUS model 690. Results demonstrated high correlation with comparator; clinical bias and repeatability were within acceptable limits defined by ASTM standards.

Technological Characteristics

Handheld infrared thermometer; ABS housing/buttons, PC lens/cap. Sensing: IR thermopile and thermistor. MCU: MSP430FG42x0. Connectivity: Bluetooth. Power: 2x 1.5V AA batteries. Standards: ASTM E1965-98 (accuracy), IEC 60601-1 (safety), IEC 60601-1-2 (EMC), ISO 10993-1 (biocompatibility).

Indications for Use

Indicated for non-contact human body temperature measurement in all ages; for use by medical professionals or consumers in home environments.

Regulatory Classification

Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; (2) Device is not a clinical thermometer with continuous temperature measurement functions; and (3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. TaiDoc Technology Corporation Jacky Chou Regulatory Affairs Director 6F, No.127, Wugong 2nd Rd., Wugu District New Taipei City, 24888 TWN April 5, 2022 ### Re: K200946 Trade/Device Name: Clever Forehead Thermometer / Clever Medical Forehead Thermometer (TD-1241) Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: March 4, 2022 Received: March 7, 2022 Dear Jacky Chou: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Payal Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K200946 Device Name Clever Forehead Thermometer/Clever Medical Forehead Thermometer (TD-1241) Indications for Use (Describe) The Clever Forehead Thermometer/Clever Medical Forehead Thermometer (TD-1241) is an infrared thermometer intended for the non-contact measurement of human body temperature in people of all ages and may be used by medical professionals or by consumers in a home environment. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary K200946 | Manufacturer | TaiDoc Technology Corporation | |--------------------------------|-------------------------------------------------------------------------| | Address | 6F, No. 127, Wugong 2nd Rd. Wugu Dist.<br>New Taipei City, Taiwan 24888 | | Establishment Registration No. | 3004145393 | | Date Prepared | March 23, 2022 | | Correspondent | TaiDoc Technology Corporation | | Correspondent Contact | Jacky Chou | | Title | Regulatory Affairs Director | | Phone | +886-2-6625-8188 | # Submitter information # Proposed Device Information | Proprietary name | Clever Forehead Thermometer/Clever Medical Forehead<br>Thermometer (TD-1241) | |----------------------|------------------------------------------------------------------------------| | Common name | Clinical Electronic Thermometer | | Product code | FLL | | Classification panel | General Hospital | | Classification | II | | Regulation Number | 21 CFR §880.2910 | # Predicate Device Information | Manufacturer | TaiDoc Technology Corporation | |------------------|---------------------------------------------------------------------------------| | Proprietary Name | Caregiver Professional Clinical Thermometer, PRO-TF series<br>(Model PRO-TF300) | | Common Name | Clinical Electronic Thermometer | | 510(k) Number | K131771 | {4}------------------------------------------------ ### Indications for use The Clever Forehead Thermometer/Clever Medical Forehead Thermometer (TD-1241) is an infrared thermometer intended for the non-contact measurement of human body temperature in people of all ages and may be used by medical professionals or by consumers in a home environment. ### Device Description The Clever Forehead Thermometer/Clever Medical Forehead Thermometer (TD-1241) is a handheld, battery powered, infrared forehead thermometer with a non-contact body temperature measuring function. It detects infrared energy emitted from forehead within 1.2 - 5 cm and converts to a body temperature. This device can be used on people of all ages by medical professionals and consumers in a home environment. The final measured temperature will be displayed on the LCD screen of the device. The users can transmit temperature results from the Clever Forehead Thermometer/Clever Medical Forehead Thermometer to Healthy Check on their mobile devices and Health care management system on pc via Bluetooth. The Healthy check and Health care management system provide an overview of users record with historical data and trend graph. # Principle of Operation The thermometer measures temperature by reading infrared radiation emitting from the forehead and converts it into a temperature value. # Comparison to the Predicate The similarities and differences between the predicate and proposed devices are summarized in Table 1 below. {5}------------------------------------------------ | Characteristic | Predicate device: | Proposed device: | Comparison | | | |-----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|------------| | Indications for use | Caregiver Professional<br>Clinical Thermometer,<br>PRO-TF series (Model<br>PRO-TF300) (K131771) | Clever Forehead<br>Thermometer/Clever<br>Medical Forehead<br>Thermometer (K200946) | Similar | | | | | Caregiver Professional<br>Clinical Thermometer is an<br>infrared thermometer<br>intended for the<br>measurement of human<br>body temperature in people<br>of all ages without contact<br>to the body and may be<br>used by medical<br>professionals or by<br>consumers in a home<br>environment. | The Clever Forehead<br>Thermometer/Clever<br>Medical Forehead<br>Thermometer (TD-1241) is<br>an infrared thermometer<br>intended for the non-contact<br>measurement of human<br>body temperature in people<br>of all ages and may be used<br>by medical professionals or<br>by consumers in a home<br>environment. | | | | | | Components | Power button, temperature<br>measurement button,<br>scanner, 4 side buttons,<br>protective cap,<br>microcontroller, and LCD | | Power button, temperature<br>measurement button,<br>scanner, 4 side buttons,<br>protective cap,<br>microcontroller, LCD, and<br>Bluetooth module | Different | | | Features | Temperatrue measurement<br><br>Memory feature (10 sets)<br><br>3 Measurement modes | | Temperatrue measurement<br><br>Memory feature (30 sets)<br><br>3 Measurement modes<br><br>Bluetooth<br><br>Apps for data transfer | Different | | | Characteristic | Predicate device: | | Proposed device: | Comparison | | | | Caregiver Professional<br>Clinical Thermometer,<br>PRO-TF series (Model<br>PRO-TF300) (K131771) | | Clever Forehead<br>Thermometer/Clever<br>Medical Forehead<br>Thermometer (K200946) | | | Principles of<br>operation | Temperatures of a subject<br>and environment are sensed<br>by the thermopile and<br>thermistor respectively, then<br>temperature is transduced<br>into electrical signal and<br>feed to analog signal<br>amplification circuit. The<br>micro-controler consists of<br>ADC (analog to digital<br>signal converter) to process<br>the signals from the<br>amplification circuit, Then,<br>through mathematical<br>process, and the LCD<br>displays the temperature<br>reading on a screen. | Temperatures of a subject<br>and environment are sensed<br>by the thermopile and<br>thermistor respectively, then<br>temperature is transduced<br>into electrical signal and<br>feed to analog signal<br>amplification circuit. The<br>micro-controler consists of<br>ADC (analog to digital<br>signal converter) to process<br>the signals from the<br>amplification circuit, Then,<br>through mathematical<br>process, and the LCD<br>displays the temperature<br>reading on a screen. | Identical | | | | MCU | The MSP430FG42x0 is a<br>microcontroller<br>configuration with a 16-bit<br>timer, a high-performance<br>16-bit sigma-delta A/D<br>converter, 12-bit D/A<br>converter, two configurable<br>operational amplifiers, 32<br>I/O pins, and a liquid crystal<br>display driver. | The MSP430FG42x0 is a<br>microcontroller<br>configuration with a 16-bit<br>timer, a high-performance<br>16-bit sigma-delta A/D<br>converter, 12-bit D/A<br>converter, two configurable<br>operational amplifiers, 32<br>I/O pins, and a liquid crystal<br>display driver. | Identical | | | | Characteristic | Predicate device: | Proposed device: | Comparison | | | | Materials | Caregiver Professional<br>Clinical Thermometer,<br>PRO-TF series (Model<br>PRO-TF300) (K131771) | Clever Forehead<br>Thermometer/Clever<br>Medical Forehead<br>Thermometer (K200946) | Identical | | | | | User contacting materials<br>include ABS (device<br>housing, power button,<br>swithes), PC(lens &<br>protective head cap). | User contacting materials<br>include ABS (device<br>housing, power button,<br>swithes), PC(lens,<br>protective head cap). | | | | | | | | | | | | | | | | | | | | | | | | | | Performance | Meets ASTM E 1965:2016 | Meets ASTM E 1965 | Identical | | | | Electrical Safety | Meets EN 60601-1:2014 | Meets IEC 60601-1 | Identical | | | | EMC | Meets IEC 60601-1-2:2014 | Meets IEC 60601-1-2 | Identical | | | | Measurement time | 1 second | 1 second | Identical | | | | Measurement distance | 1.2 cm – 5 cm | 1.2 cm – 5 cm | Identical | | | | Reference body site | Rectal for < 3 years old | Rectal for < 3 years old | Identical | | | | | Oral for > 3 years old | Oral for > 3 years old | | | | | Mode of operation | Adjusted mode | Adjusted mode | Identical | | | | | | | | | | | Weight | 83 g | 83 g | Identical | | | | Memory/storage | 32kb +256b flash memory | 32kb +256b flash memory | Identical | | | | | 256b ram | 256b ram | | | | | Biocompatibility | ISO 10993-1 | ISO 10993-1 | Identical surface material | | | | Probe cover | Yes | Yes | Identical | | | | Sensor | IR Thermopile sensor | IR Thermopile sensor | Identical | | | | Dimension (mm) | 150 (L) x 48.48 (W) x<br>55.16 (H) | 150 (L) x 48.48 (W) x<br>55.16 (H) | Identical | | | | | | | | | | | Characteristic | Predicate device:<br>Caregiver Professional<br>Clinical Thermometer,<br>PRO-TF series (Model<br>PRO-TF300) (K131771) | Proposed device:<br>Clever Forehead<br>Thermometer/Clever<br>Medical Forehead<br>Thermometer (K200946) | Comparison | | | | Display<br>Resolution | 0.1 °F /0.1°C | 0.1 °F /0.1°C | Identical | | | | Measurement<br>unit | °F or °C | °F or °C | Identical | | | | Ambient<br>operation<br>condition | 50°F to 104°F (10°C to<br>40°C) | 50°F to 104°F (10°C to<br>40°C) | Identical | | | | Meter storage/<br>transportation<br>condition | -4°F to 140°F (-20°C to<br>60°C) | -4°F to 140°F (-20°C to<br>60°C) | Identical | | | | Measurement<br>range | Forehead:<br>94°F to 108°F (34.4°C to<br>42.2°C) | Forehead:<br>71.6°F to 111.2°F (22°C to<br>44°C) | Different | | | | | Ambient:<br>50°F to 104°F (10°C to 40<br>°C) | Ambient:<br>50°F to 104°F (10°C to 40<br>°C) | Identical | | | | HI/LO Indicator | Forehead:<br>• ≤94°F (34.4°C) LCD<br>Display LO<br>• ≥108°F(42.2°C) LCD<br>Display HI<br>Surface: | Forehead:<br>• ≤94°F (34.4°C) LCD<br>Display LO<br>• ≥108°F(42.2°C) LCD<br>Display HI<br>Surface: | Identical | | | | | • < 32°F (0°C) LCD<br>Display LO<br>• > 140°F (60°C) LCD<br>Display HI | • < 32°F (0°C) LCD<br>Display LO<br>• > 140°F (60°C) LCD<br>Display HI | Identical | | | | Characteristic | Predicate device:<br>Caregiver Professional<br>Clinical Thermometer,<br>PRO-TF series (Model<br>PRO-TF300) (K131771) | Proposed device:<br>Clever Forehead<br>Thermometer/Clever<br>Medical Forehead<br>Thermometer (K200946) | Comparison | | | | Accuracy<br>(Forehead) | $±0.4°F (±0.2°C) from96.8°F to 102.2°F (36°C to39°C)$ $±0.5°F (±0.3°C) for therange of71.6°F to 96.7°F (22°C to35.9°C) or 102.3°F to108.5°F (39.1°C to 42.5°C)$ | $±0.4°F (±0.2°C) from96.8°F to 102.2°F (36°C to39°C)$ $±0.5°F (±0.3°C) for therange ofless than 36 °C (96.8°F) or greater than 39 °C(102.2 °F)$ | Similar | | | | Power source | 2 x 1.5V AA batteries | 2 x 1.5V AA batteries | Identical | | | | Time to Power<br>saving | 30 Seconds | 30 Seconds | Identical | | | | Signal output<br>and display | LCD, Buzzer | LCD, Buzzer, Bluetooth | Different | | | # Table 1: Comparison between the Predicate and Proposed Device {6}------------------------------------------------ {7}------------------------------------------------ {8}------------------------------------------------ {9}------------------------------------------------ {10}------------------------------------------------ # Substantial Equivalent Discussion #### . Indications for use Indications for use of the Clever Forehead Thermometer/Clever Medical Forehead Thermometer (TD-1241) is similar to its predicate device. The without contact is rephrased to non-contact for more precise wording. No new safety and effectiveness questions were raised. #### . Components The modified device has an added Bluetooth module. ES and EMC testing as well as function validation has been implemented to ensure the safety or effectiveness of the device is unaffected. The difference does not raise new safety or effectiveness questions. #### . Features The Clever Forehead Thermometer/Clever Medical Forehead Thermometer (TD-1241) increases the memory storage limit to 30 temperature readings and added a Bluetooth transfer function. The Clever Forehead Thermometer/Clever Medical Forehead Thermometer (TD-1241) uses the identical memory component, 32kb +256b flash memory 256b ram, as its predicate device. Both the Bluetooth and the memory features have been validated and tested in ES and EMC to ensure the safety or effectiveness of the device. An additional application feature is added for Bluetooth data transfer. The application has been verified and validated to ensure the safety or effectiveness of the device is unaffected. #### . Measurement range and Accuracy The measurement range and accuracy comply with the ASTM E1965-98 (2016) standards. Laboratory test was conducted, and the test results met the requirements. The difference does not raise new safety and effectiveness questions. #### . Signal output and display Bluetooth signal output is added to the Clever Forehead Thermometer/Clever Medical Forehead Thermometer (TD-1241). The device has been validated and tested in ES and EMC to ensure the Bluetooth signal output does not raised safety or effectiveness issue for the device. {11}------------------------------------------------ # Non-clinical and performance testing The entire Hazard Analysis for the Clever Forehead Thermometer/Clever Medical Forehead Thermometer was evaluated to identify all the risks / hazards that could be affected by the modifications to the Clever Forehead Thermometer/Clever Medical Forehead Thermometer. These risks were mitigated using planned measures that included testing to recognized FDA consensus standards. Changes in software were verified and validated using the software development process. The following table, which includes a summary of non-clinical testing data conducted according to FDA recognized consensus standards, is provided in support of the substantial equivalence determination: | Test report | Performance Standard | |----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Electromagnetic<br>compatibility (EMC) | • IEC 60601-1-2:2014, Medical electrical equipment -- Part 1-2: General<br>requirements for basic safety and essential performance -- Collateral<br>Standard: Electromagnetic disturbances -- Requirements and tests | | Electrical Safety | • IEC 60601-1:2014, Medical electrical equipment - Part 1: General<br>requirements for basic safety and essential performance<br>• IEC 60601-1-11:2015, Medical electrical equipment - Part 1-11:<br>General requirements for basic safety and essential performance -<br>Collateral Standard: Requirements for medical electrical equipment and<br>medical electrical systems used in the home healthcare environment | | Laboratory accuracy | • ASTM E1965-98:2016, Standard Specification for Infrared<br>Thermometers for Intermittent Determination of Patient Temperature | | Clinical accuracy | • ASTM E1965-98:2016, Standard Specification for Infrared<br>Thermometers for Intermittent Determination of Patient Temperature | | Storage stability | • ASTM E1965-98:2016, Standard Specification for Infrared<br>Thermometers for Intermittent Determination of Patient Temperature | | Shock test | • ASTM E1965-98:2016, Standard Specification for Infrared<br>Thermometers for Intermittent Determination of Patient Temperature | | Cleaning procedure | • ASTM E1965-98:2016, Standard Specification for Infrared<br>Thermometers for Intermittent Determination of Patient Temperature | | Risk management<br>report | • ISO 14971, Medical devices - Applications of risk management to<br>medical devices | {12}------------------------------------------------ | Biocompatibility | ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process | |-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Software and<br>Cybersecurity | IEC 62304, Medical device software - Software life cycle processes ISO/IEC 12207, Systems and software engineering Software life cycle processes Software verification and validation test according to the requirements of the FDA "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005)" Guidance for Industry and FDA Staff: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, October 2018. | # Clinical Testing A clinical accuracy testing evaluated 250 subjects which were performed on the following three age groups: infants (newborn to one vear), children (greater than one to five years), and adults (greater than five years old) in accordance with ASTM standard (E1965-98) to compare the Clever Forehead Thermometer/Clever Medical Forehead Thermometer with the comparator method (WelchAllyn SureTemp PLUS model 690 Thermometer). This clinical accuracy study demonstrated that the temperatures obtained with the Clever Forehead Thermometer/Clever Medical Forehead Thermometer were highly related when compared to the comparator method. The clinical bias with stated uncertainty and clinical repeatability as defined in the ASTM standard (E1965-98) were within clinical acceptability. # Conclusion The intended use of the subject device, the Clever Forehead Thermometer/Clever Medical Forehead Thermometer, is identical to that of the predicate device, the Caregiver Professional Clinical Thermometer, PRO-TF series (Model PRO-TF300), and a risk analysis was performed to identify risks associated with the device modifications. Verification tests have been performed to demonstrate that the identified risks have been mitigated. The testing and analysis demonstrate that the subject device is substantially equivalent to the predicate device
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...