K173048 · Dongguan Simzo Electronic Technology Co., Ltd. · FLL · Dec 6, 2018 · General Hospital
Device Facts
Record ID
K173048
Device Name
No-contact Forehead Thermometer
Applicant
Dongguan Simzo Electronic Technology Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
Dec 6, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Non-contact Forehead Thermometer is an infrared thermometer intended for the intermittent measurement of human body temperature in people of all ages.
Device Story
Hand-held, battery-powered infrared thermometer; measures human body temperature via infrared energy emitted from forehead. Device utilizes thermopile sensor, integrated thermistor for target reading, ambient thermistor, and parabolic mirror for focusing. Infrared distance sensor detects distance (5-8cm or 1-2cm depending on model) to provide temperature compensation. Used in home or clinical settings by patients or healthcare providers. Output displayed on LCD; provides fever alarm and memory storage. Enables non-invasive, rapid temperature monitoring to assist in clinical decision-making regarding patient health status.
Clinical Evidence
Clinical study conducted on 96 subjects across three age groups: infants (newborn to 1 year), children (>1 to 5 years), and adults (>5 years), with 32 patients per group. 30% of subjects exhibited temperatures ≥38.0°C. Results confirmed compliance with ASTM E1965-98 (2016) performance requirements.
Technological Characteristics
Infrared thermopile sensor; parabolic mirror; ambient thermistor; LCD display. Powered by 2x AAA batteries. Conforms to EN60601-1 (safety), IEC60601-1-2 (EMC), and ASTM E1965-98 (performance). Materials evaluated per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (sensitization/irritation).
Indications for Use
Indicated for intermittent measurement of human body temperature in people of all ages.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
Non-contact Clinical Thermometer, Model THB0F (K121428)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 6, 2018
Dongguan SIMZO Electronic Technology Co. Ltd. Mei Tan OA Manager No.81, Tianxin Street, Chongkong, Shijie Town Dongguan, Guangdong, China
Re: K173048
Trade/Device Name: Non-contact Forehead Thermometer, Model: HW-2/HW-2S/HW-3/HW-4 /HW-4S/HW-302/HW-303 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: October 22, 2018 Received: October 25, 2018
Dear Mei Tan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Geeta K. Pamidimukkala -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173048
#### Device Name
Non-contact Forehead Thermometer, Model: HW-2/HW-2S/HW-3/HW-4/HW-4S/HW-302/HW-303
Indications for Use (Describe)
The Non-contact Forehead Thermometer is an infrared thermometer intended for the intermittent measurement of human body temperature in people of all ages.
Type of Use (Select one or both, as applicable)Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K)Summary
### K173048
# Prepared Date: 2018/12/4 Submitter Information
Name : Dongquan SIMZO Electronic Technology Co.Ltd. Address: No.81,Tianxin Street,Chongkong,Shijie Town,Dongguan,Guangdong,China Tel.+0086-769-86339010 Fax.+0086-769-86387762 Contact person: Mei Tan E-mail: Mei_FDA@foxmail.com
## Subject Device Information
Trade name: Non-contact Forehead Thermometer Model: HW-2/HW-2S/HW-3/HW-4 /HW-4S/HW-302/HW-303 Common name: Non-contact Clinical Thermometer Classification name: Clinical electronic thermometer Production requlation: 21 CFR 880.2910 Product code: FLL
# Predicate Device
Non-contact Clinical Thermometer, Model THB0F. (Radiant Innovation Inc,K121428)
# Description
The Non-contact Forehead Thermometers are hand-held, battery powered devices designed to measure human body temperature. The subject devices are Infrared thermometers that convert a user's forehead temperature using the infrared energy emitted in the area around the user's forehead to an oral equivalent temperature when placed in the distance from subject's forehead up to 5-8cm(HW-2,HW-2S,HW-3,HW-302 and HW-303) or 1-2cm(HW-4 and HW-4S).
The thermometers use a thermopile sensor with integrated thermistor for the target reading, a thermistor mounted in the head of the thermometer for ambient temperature readings, a parabolic mirror to help focus the infrared energy emitted from the forehead, and an infrared distance sensor for detection with the distance from forehead up to 5~8cm(HW-2,HW-2S,HW-3,HW-302 and HW-303) or 1-2cm(HW-4 and HW-4S) and compensation of the temperature reading.
{4}------------------------------------------------
### Indications for Use
The Non-contact Forehead Thermometer is an infrared thermometer intended for the intermittent measurement of human body temperature in people of all ages.
# Summary of technological characteristics of device compared to the predicate devices (K121428), see the table1
A comparison of key technological characteristics between the subject devices and predicate device was listed as below:
| | Subject device(SD) | Predicate device(PD) | Note |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| | SIMZO HW series | Rdiant THB0F | |
| 510K number | K173048 | K121428 | -- |
| Intended Use&<br>Indication for<br>use | The Non-contact Forehead<br>Thermometer is an infrared<br>thermometer intended for the<br>intermittent measurement of<br>human body temperature in<br>people of all ages. | The Non-contact Clinical<br>Thermometer, Model<br>THBOF is an infrared<br>thermometer intended for<br>the intermittent<br>measurement of human<br>body temperature in<br>people of all ages. | Same |
| Measurement<br>Method | Infrared radiation detection | Infrared radiation<br>detection | Same |
| Measurement<br>Range | Forehead mode:<br>35.5°C ~42.9°C<br>(95.9°F~109.2°F) | Forehead mode:<br>34.0°C ~42.2°C<br>(93.2°F ~108.0°F) | Similar |
| Accuracy | Forehead mode:<br>±0.2°C (0.4°F) within<br>35.5°C~42.9°C<br>(95.9°F ~109.2°F),<br>±0.3°C(0.5°F) other range | Forehead mode:<br>±0.2°C (0.4°F) within<br>36~39°C(96.8~102°F),<br>±0.3°C(0.5°F) other<br>range | Similar |
| Display | 0.1°C(0.1°F) | 0.1°C(0.1°F) | Same |
| Measurement<br>distance | HW-2/HW-2S/HW-3/HW-302/<br>HW-303:5-8cm<br>HW-4/HW-4S:1-2cm | 2-3cm | Similar |
| Measurement<br>place | Forehead | Forehead | same |
| Response time | 1S | 1S | same |
Table 1
{5}------------------------------------------------
## DONGGUAN SIMZO ELECTRONIC TECHNOLOGY CO.,LTD.
| Sensor type | Thermopile | Thermopile | Same |
|-----------------------------------------|----------------------------------------------------------------------|----------------------------------------------------------|---------|
| Fever alarm | Yes | Yes | Same |
| Scale selection | °C / °F | °C/°F | same |
| Display screen | LCD | LCD | same |
| Memory | HW-2/2S:32 sets<br>HW-3/4/4S: 1 set<br>HW-302/303:64 Sets | 60sets | Similar |
| Fever alarm | Yes | Yes | same |
| Buzzer | Yes | Yes | same |
| Auto power-off<br>while no<br>operation | Yes | Yes | Same |
| Power supply | 2 x AAA | 2 x AAA | Same |
| Operation | 10~40°C(50°F ~104 °F) | 10~40°C(50°F ~104 °F) | Same |
| Environment | RH < 85% | 15-85% RH | Similar |
| Storage | -25℃-+55℃(-13°F-+131°F) | -20℃-+55(-4°F-+131°F) | Similar |
| Environment | RH≤90% | RH≤90% | --- |
| Dimension | 93*153*41mm | 180.3*47.5*29.2mm | --- |
| Weight | 90-125 g | 104.7g | --- |
| Conformance standard | EN60601-1(Safety),<br>IEC60601-1-2(EMC)<br>ASTME1965-98(performance) | EN60601-1(Safety),<br>IEC60601-1-2(EMC)<br>ASTM E1965-98 | Same |
#### Analysis
From the comparison table1, the subject devices and predicate device have the same Intended use & Indications for Use, Measurement place, Scale selection, Display screen, Auto power-off while no operation & Conformance standard. There are slightly differences between the devices and predicate device as follows.
| Difference clause | Discussion |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Measurement<br>Range&Accuracy | Both devices have different measurement range, but they have<br>the same accuracy and the measurement range of subject<br>devices meet the requirements of ASTME1965-98 |
| Measurement distance | Measurement distance of the subject devices is 5-8cm or<br>1-2cm, the predicate device's will be in the range of 2-3cm. But<br>the performance test result of subject device shows the<br>accuracy meets the requirements within the distance range. |
| Memory | The memory capacity of predicate devices in K121428 is<br>difference from subject devices ,but that does not impact the |
{6}------------------------------------------------
| | performance of subject devices. | |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------|--|
| Operation &Storage<br>Environment | Both devices have slightly different Operation &Storage<br>Environment, but the subject devices meet the requirements | |
| | of IEC60601-1 | |
#### Non-Clinical Data
The Non-contact Forehead Thermometer has been tested according to the following applicable standards:
| Test name | Report description | Result |
|-------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Electrical<br>Safety | Test report / IEC 60606-1:Medical Electrical Equipment -<br>Part 1: General Requirements for Safety | Pass |
| EMC | Test report / IEC 60601-1-2 Medical Electrical<br>Equipment-Part 1-2: General Requirements for Safety -<br>Collateral Standard: Electromagnetic Compatibility -<br>Requirements and Tests | Pass |
| Cytotoxicity | Cytotoxicity test report /ISO 10993-5 Biological evaluation of<br>medical devices -- Part 5: Tests for in vitro cytotoxicity | Pass |
| Sensitization | Sensitization test report /ISO 10993-10 Biological<br>evaluation of medical devices -- Part 10: Tests for irritation<br>and skin sensitization | Pass |
| Irritation | Irritation test report /ISO10993-10 Biological evaluation of<br>medical devices -- Part 10: Tests for irritation and skin<br>sensitization | Pass |
| Performance<br>test in home<br>healthcare<br>environment | Test report/IEC60601-1-11 Medical electrical equipment<br>-Part 1-11: General requirements for basic safety and<br>essential performance - Collateral Standard: Requirements<br>for medical electrical equipment and medical electrical<br>systems used in the home healthcare environment | Pass |
| Performance<br>(Shock test,<br>Storage test,<br>Laboratory<br>accuracy test ) | Test report/ASTM E 1965-98 Standard Specification for<br>Infrared Thermometers for Intermittent Determination of<br>Patient Temperature | Pass |
| Cleaning and<br>Disinfection | Cleaning and Disinfection validation report /ASTM E 1837 – 96<br>(Reapproved 2014), Standard Test Method to Determine<br>Efficacy of Disinfection Processes for Reusable Medical Devices<br>(Simulated Use Test) | Pass |
| Shelf life | Shelf life test report /scanner switch on/off cycles 136875<br>times(use life 5years) | Pass |
| Software | Guidance for the Content of Premarket Submissions for | Pass |
| validation and | Software Contained in Medical Devices,dated May 11, 2005 | |
| verification | | |
{7}------------------------------------------------
#### Clinical data
Clinical tests were conducted on HW-2,HW-2S,HW-3,HW-4,HW-4S,HW-302 and HW-303. The clinical tests evaluated 96 subjects which were divided into three group age ranges-infant (new born to 1 year), children (greater than 1 to 5 years old) and adult(greater than5 years old). Each group has 32 patients and 30% of them got temperature equaling or exceeding 38.0℃.
Based on the result, the clinical performance of the subject device complied with the requirement of ASTM E1965-98 (2016).
### Conclusion
Based on the performance testing and compliance with acceptable voluntary standards, we believe that the Non-contact Forehead Thermometer is substantially equivalent to its predicate device in K121428.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.