Infrared Thermometer

K182133 · Shenzhen AOJ Medical Technology Co., Ltd. · FLL · Nov 20, 2018 · General Hospital

Device Facts

Record IDK182133
Device NameInfrared Thermometer
ApplicantShenzhen AOJ Medical Technology Co., Ltd.
Product CodeFLL · General Hospital
Decision DateNov 20, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2910
Device ClassClass 2
AttributesPediatric

Indications for Use

The Infrared thermometers (AOJ-20A and AOJ-20B) take human body temperature via the eardrum or forehead. They apply to all age groups except for babies under three months. Both devices apply to both professional use and home use.

Device Story

Handheld infrared thermometer; models AOJ-20A and AOJ-20B; measures human body temperature via eardrum or forehead. Device uses thermopile sensor to detect infrared thermal energy emitted from skin or eardrum; converts energy into temperature reading in Celsius or Fahrenheit. Powered by two 1.5V AAA alkaline batteries; results displayed on LCD. Used in clinical or home settings by professionals or patients. Device includes memory for 20 measurements. Provides non-invasive temperature monitoring; aids healthcare providers in clinical decision-making regarding patient health status.

Clinical Evidence

Clinical study conducted per ASTM E1965-98 using a randomized, single-blind, homologous control, pairing design. Study included at least 90 subjects (1/3 children, 1/3 adults). Primary endpoint: clinical accuracy compared to reference device. Acceptance criteria: accuracy ±0.2°C within 32°C–42.9°C range, 99% confidence. All testing passed pre-specified criteria.

Technological Characteristics

Handheld infrared thermometer; PC+ABS patient-contacting materials; thermopile sensor; powered by 2x AAA 1.5V alkaline batteries. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ASTM E1965-98, ISO 10993-5/10. Biocompatibility: In vitro cytotoxicity, skin sensitization, irritation. Connectivity: None. Software: Embedded firmware.

Indications for Use

Indicated for human body temperature measurement via eardrum or forehead in all age groups except infants under three months. Suitable for professional and home use.

Regulatory Classification

Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; (2) Device is not a clinical thermometer with continuous temperature measurement functions; and (3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. November 20, 2018 Shenzhen AOJ Medical Technology Co., Ltd. Qihuan Zhao General Manager Room 202, HaoGu Industry Park, 2037 Guanguang Road, Guangming District, Shenzhen Guangdong, 518105 China Re: K182133 Trade/Device Name: Infrared Thermometer, models AOJ-20A and AOJ-20B Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: October 12, 2018 Received: October 22, 2018 Dear Oihuan Zhao: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Geeta K. Pamidimukkala -S for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K182133 Device Name Infrared thermometer, models AOJ-20A and AOJ-20B Indications for Use (Describe) The Infrared thermometers (AOJ-20A and AOJ-20B) take human body temperature via the eardrum or forehead. They apply to all age groups except for babies under three months. Both devices apply to both professional use and home use. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| Prescription Use (Part 21 CFR 801 Subpart D) |X | Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary # K182133 # Prepared in accordance with the requirements of 21 CFR Part 807.92 | 1. Submitter: | Shenzhen AOJ Medical Technology Co., Ltd. | |---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Room 202, HaoGu industry park, 2037 Guanguang road, Guangming district,<br>518105, Shenzhen, P.R. China | | | Tel.: +86 18603031299 | | Contact Person: | Qihuan Zhao | | Prepare date: | 2018-11-15 | | 2. Device name<br>and classification: | Device Name: Infrared Thermometer<br>Models: AOJ-20A, AOJ-20B | | | Regulation Number: 21 CFR 880.2910 | | | Regulation Name: Clinical Electronic Thermometer | | | Product code: FLL | | | Regulatory Class: Class II | | 3. Predicate<br>Device(s): | Shenzhen Brav Electronic Technologies Co., Ltd., EFT-165 Infrared Thermometer<br>/K171214 | | 4. Device<br>Description: | The infrared thermometers (AOJ-20A and AOJ-20B) are handheld instruments,<br>which can measure human body's temperature either via the eardrum or the<br>forehead for clinical or home use. The intended patient population is all age<br>groups except for babies under three months. | | | The thermometers are powered by AAA 1.5V×2 alkaline batteries. The results<br>can be displayed on LCD. | | | A thermopile sensor is employed to detect or monitor the infrared thermal energy<br>emitted from the eardrum or the surface of the skin of the forehead, which is<br>converted into temperature measurement with the unit of °C or °F. | | 5. Indications for<br>Use: | The Infrared thermometers (AOJ-20Aand AOJ-20B) take human body temperature<br>via the eardrum or forehead. They apply to all age groups except for babies under<br>three months. Both devices apply to both professional use and home use. | {4}------------------------------------------------ # 6.Predicate Device Comparison | ITEM | Proposed Device<br>AOJ-20A&AOJ-20B | Predicate Device<br>EFT-165/K171214 | Comparison<br>Result | |------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------| | Manufacture | Shenzhen AOJ Medical<br>Technology Co., Ltd. | Brav Electronic Technologies<br>Co., Ltd. | --- | | Indications for<br>Use | The Infrared thermometers<br>(AOJ-20Aand AOJ-20B) take<br>human body temperature via the<br>eardrum or forehead. They apply<br>to all age groups except for<br>babies under three months. Both<br>devices apply to both<br>professional use and home use. | The infrared thermometer is<br>intended for the measurement<br>and monitoring of human body<br>temperature by doctor or<br>customers in the hospital or<br>home. | Different 1 | | Operational Specifications | | | | | Operational<br>Principle | Infrared radiation detection | Infrared radiation detection | Same | | Measuring Mode | Forehead and ear | Forehead and ear | Same | | Measurement<br>Range | 32.0°C~42.9°C<br>(89.6°F~109.2°F) | 32.0°C~42.9°C<br>(89.6°F~109.2°F) | Same | | Measurement<br>Distance | 0 cm | 0 cm | Same | | Accuracy | ±0.2°C(0.4°F) | ±0.2°C(0.4°F) | Same | | Memory Data<br>Limit | memorize 20 measurements<br>automatically | memorize 20 measurements<br>automatically | Same | | Product<br>configuration | It is mainly composed with<br>infrared sensor, signal receiving<br>processor, buttons, buzzer, LCD<br>display, battery and etc. | It is mainly composed with<br>infrared sensor, signal receiving<br>processor, buttons, buzzer, LCD<br>display, battery and etc. | Same | | Temperature unit<br>and conversion | Dual temperature units “°C” and<br>“°F” optional, and the two units<br>can convert by the conversion<br>key automatically | Dual temperature units “°C” and<br>“°F” optional, and the two units<br>can convert by the conversion<br>key automatically | Same | | Applicable<br>Standards | IEC 60601-1, IEC 60601-1-2,<br>IEC 60601-1-11, and ASTM<br>E1965-98 | AAMI ANSI ES60601-1, IEC<br>60601-1-2, IEC 60601-1-11, and<br>ASTM E1965-98 | Same | | Display | 0.1°C/°F, LCD | 0.1°C/°F, LCD | Same | | Operating<br>Environment | Temperature: 10°C~40°C<br>(50°F-104°F)<br>Humidity: 15%-95% RH,<br>non-condensing<br>Atmospheric pressure: 70–106<br>kPa | 15°C ~ 35°C (59°F to 95 °F)<br><85% moisture condensation | Different 2 | | Storage<br>Environment | Ambient Temperature: -20°C to<br>55°C (-4°F-131°F)<br>Relative Humidity: 0-95% RH,<br>non-condensing<br>Atmospheric pressure: 50kPa to<br>106kPa | -20°C~55°C(-4°F~131°F)≤90%<br>moisture condensation | Same | Table 1 Comparison between main predicate EFT-165 and the subject device {5}------------------------------------------------ | Physical Specifications | | | | |------------------------------------|-----------------------------------------------------------|-----------------------------------------------------------|-------------| | Weight | 90g (battery included) | 72 g (without battery) | Different 3 | | Dimensions | 146mm *52 mm *40 mm | 152mm * 44 mm *30mm | | | Biological Specifications | | | | | Patient<br>Contacting<br>Materials | PC+ABS | PC+ABS | Same | | Patient<br>Contacting | Surface-contacting,<br>Less than 24 h | Surface-contacting,<br>Less than 24 h | Same | | Biocompatibility<br>Standard | ISO 10993-5,<br>ISO 10993-10 | ISO 10993-5,<br>ISO 10993-10 | Same | | Biocompatibility<br>Testing Items | In vitro Cytotoxicity<br>Skin Sensitization<br>Irritation | In vitro Cytotoxicity<br>Skin Sensitization<br>Irritation | Same | #### Justification for the differences: - 1) Different Indications for Use As indicated in the comparison table, the subject device and the predicate device have similar indications for use, the subject device includes the patient population and areas to take temperature. The intended use is the same. - 2) Different Operation&Storage&Transport Environments Minor difference to operation environments (including Temperature and Relative Humidity) for the subject device, but the system has demonstrated to be as safe and effective as the predicate device since the safety testing was conducted under the suggested environment; Moreover, environment testing data shows the device can work as intended under the suggested conditions. So those changes do not raise any safety and effectiveness concerns. - 3) Different Physical Specifications The subject and the predicate are of different size but proximity. Moreover, such engineering design has been verified during development by compliance with the applicable international standards, so such minor different will not raise any new problems. As seen in the comparison tables, the subject and predicate devices have almost the same design features and performance specifications. Moreover, as demonstrated in the non-clinical and clinical testing, AOJ-20A&AOJ-20B Infrared thermometer system can perform as intended. ## 7.Performance Testing ## 7.1 Non-Clinical Data The following safety standards are conducted on the subject device support of the substantial equivalence determination, including performance testing and bench testing. ## Performance Testing Information: - (1) ISO 10993-5 Biological evaluation for medical device Part 5 Test for In Vitro Cytotoxicity - (2) ISO 10993-10 Biological evaluation for medical device Part 10 Test for irritation and skin sensitization - (3) AAMI TIR12:2004 Designing, testing, and labeling reusable medical devices for reprocessing in health care facilities: A guide for medical device manufacturers, Approved 23 December 2004. {6}------------------------------------------------ ## Bench Testing Information: - (1) ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009((R)2012 and A2:2010/(R)2012 Medical electrical equipment Part 1: General requirements for basic safety and essential performance. - (2) IEC 60601-1-11: 2010 MEDICAL ELECTRICAL EQUIPMENT -Part 1-11: General requirements for basic safety and essential performance –Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment. - (3) IEC 60601-1-11: 2010 MEDICAL ELECTRICAL EQUIPMENT -Part 1-11: General requirements for basic safety and essential performance –Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment. - (4) IEC 60601-1-2: 2014 Medical electrical equipment =Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests. - (5) ISO 80601-2-56: 2009 Medical electrical equipment Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement for performance effectiveness. - (6) Software verification and validation testing per FDA's Guidance for Industry and FDA Staff, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, issued on May 11, 2005. # 7.2 Clinical Data Clinical testing is conducted per ASTM E 1965-98 (Reapproved 2009) Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature. This clinical study is a randomization, simple blind homologous control, pairing design of clinical investigation, consists of a minimum of 90 subjects, of which 1/3 are children and the rest 1/3 are adults (NOTE: Infants---newborn to one year; Children--- greater than one to five years; Adults--- greater than five years old.). with no single measurement error exceeding the allowable limit, provides a 99% confidence. A reference device is introduced as required by the standard, and the acceptance criteria is that clinical accuracy of the subject device is at least the same as that of the reference, and the accuracy is ±0.2℃ during the claimed measurement range 32℃ - 42.9℃. # 8.Conclusion Verification and validation testing was conducted on the subject device and all testing passed pre-specified criteria. This premarket notification submission demonstrates that the AOJ-20A & AOJ-20B Infrared Thermometer is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...