K171214 · Shenzhen Brav Electronic Technologies Co., Ltd. · FLL · Mar 2, 2018 · General Hospital
Device Facts
Record ID
K171214
Device Name
Infrared Thermometer
Applicant
Shenzhen Brav Electronic Technologies Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
Mar 2, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Indications for Use
The infrared thermometer is intended for the measurement and monitoring of human body temperature by doctors or customers in the hospital or at home.
Device Story
EFT-16 series infrared thermometers measure human body temperature via infrared radiation emitted from the forehead or ear. The device uses a Fresnel lens to focus infrared energy onto an internal sensor; a signal processor converts this input into a temperature reading, which is displayed on an LCD screen. Operated by clinicians or patients in clinical or home settings, the device provides intermittent temperature monitoring. It includes a buzzer for alerts and memory storage for 20 measurements. The device aids in clinical decision-making by providing rapid, non-invasive temperature data, potentially benefiting patients through convenient, accessible monitoring.
Clinical Evidence
Clinical testing was conducted per ASTM E 1965-98 Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature. Non-clinical data included biocompatibility (ISO 10993-1, -5, -10), electrical safety (ANSI/AAMI ES60601-1), EMC (IEC 60601-1-2), and performance testing (ISO 80601-2-56). All tests passed, supporting substantial equivalence.
Technological Characteristics
Infrared sensor-based thermometer; forehead and ear measurement modes. Materials: ABS plastic and PE lens filters. Power: 3Vdc battery. Connectivity: None. Display: LCD with 0.1°C/°F resolution. Standards: ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 80601-2-56, ASTM E1965-98.
Indications for Use
Indicated for the measurement and monitoring of human body temperature in patients by clinicians in hospitals or by consumers at home.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
March 2, 2018
Shenzhen Brav Electronic Technologies Co., Ltd. Wenhui Cha QA Manager 4/F, Block 11, Tongfuyu Industrial District Lezhujiao, Huangmabu Community, Xixiang, Baoan Shenzhen, Guandong, 518102 China
Re: K171214
Trade/Device Name: Infrared Thermometer, Models EFT-162, EFT-163, EFT-164, EFT-165, EFT-166 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: January 24, 2018 Received: January 31, 2018
Dear Wenhui Cha:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
for
Image /page/1/Picture/5 description: This image shows a formal letter. The letter is signed by Alan M. Stevens and Tina Kiang, Ph.D., Acting Director of the Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices, Office of Device Evaluation, Center for Devices and Radiological Health. There is also a digital signature from Alan M. Stevens with the date 2018.03.02 07:49:17-05'00'.
Enclosure
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K171214
Device Name
Infrared thermometer, models EFT-161, EFT-162, EFT-164, EFT-165, EFT-166
Indications for Use (Describe)
The infrared thermometer is intended for the measurement and monitoring of human body temperature by doctors or customers in the hospital or at home.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K171214
### 510(k) Summary
### 1. Submitter:
#### Submitter (Owner) and Contact Information
| Company / Institution Name | | | |
|---------------------------------------------------------------------------------------------|----------------------------------|-----------------|---------|
| Shenzhen Brav Electronic Technologies Co., Ltd. | | | |
| Phone Number (including area code) | FAX Number (including area code) | | |
| +86 0755 29781489 | +86 0755 29569480 | | |
| Street Address | | | |
| 4/F, Block 11, Tongfuyu Industrial District, Lezhujiao, Huangmabu Community, Xixiang, Baoan | | | |
| City | State / Province | ZIP/Postal Code | Country |
| Shenzhen | Guangdong | 518102 | China |
| Contact Name | | | |
| Wenhui Cha | | | |
| Contact Title | Contact E-mail Address | | |
| QA Manager | 1281130583@qq.com | | |
Date Prepared: 2018-01-24
This summary of 510(k) information is being submitted in accordance with the requirements of 21 CFR §807.92.
### 2. Device Information:
Type of 510(k) submission: Traditional Trade Name: Infrared Thermometer, Models EFT-161, EFT-162, EFT-163, EFT-164, EFT-165, EFT-166. Classification name: thermometer, electronic, clinical Review Panel: General Hospital Product Code: FLL Device Class: II Regulation Number: 880.2910
### 3. Predicate device:
Sponsor: Famidoc Technology Co., Ltd. Device: Famidoc Infrared Thermometer Model: FDIR-V2, FDIR-V4, FDIR-V5, FDIR-V6, FDIR-V7, FDIR-V9, FDIR-V9-3, FDIR-V10, FDIR-V15 510(K) Number: K113761
{5}------------------------------------------------
## 4. Indications for Use
The infrared thermometer is intended for the measurement and monitoring of human body temperature by doctors or customers in the hospital or at home.
## 5. Basic principle
Any object in which temperature is higher than absolute zero degree will transmit some infrared radiation energy according to its own temperature. The radiation energy and its distribution per wavelength are closely associated with its surface temperature. Based on the principle, it is possible to measure the forehead or ear temperature and then adjust the offset between forehead or ear temperature and actual body temperature, which will result in the correct display of body temperature.
### 6. Device Description
EFT-16 series (Model: EFT-161, EFT-162, EFT-164, EFT-165, EFT-166) measure the body temperature through receiving the infrared energy radiation from the surface of objects. It is with forehead temperature mode, and ear temperature mode, directly show on the LCD screen.
- . The device is widely used for home healthcare, medical institutes and many other occasions.
- . The product is mainly composed with infrared temperature sensors, signal receiving processor, buttons, buzzer, LCD display, battery, etc.
- It focuses the infrared from the human's forehead or ear by the Fresnel lens.
### 7. Performance Data
Performance data includes "Non-Clinical Data" and "Clinical Data", brief description of which are shown as below.
### Non-Clinical Data:
The following performance data were provided in support of the substantial equivalence determination.
### Biocompatibility testing
The biocompatibility evaluation for the EFT-16 series Infrared thermometer were conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. Since the worst case of the whole system is considered tissue contacting for duration of less than 24 hours. And the necessary testing necessary for this contacting category material include Cytotoxicity, Skin Sensitization and Irritation per the Annex A Biological evaluation tests of ISO 10993-1, which have been conducted on the EFT-16 series Infrared thermometer device.
{6}------------------------------------------------
### Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the EFT-16 series Infrared thermometer device, consisting of all the modules and accessories in the system. The system complies with the ANSI/AAMI ES60601-1:2005/(R)2012and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 Medical electrical equipment Part 1: General requirements for basic safety and essential performance for safety and the IEC 60601-1-2: 2014 Medical electrical equipment –Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – Requirements and tests standard for EMC.
#### Bench Testing
Bench testing was conducted on the EFT-16 series Infrared thermometer device, consisting of all the accessories in the system. The system complies with the IEC 60601-1-11: 2010 MEDICAL ELECTRICAL EQUIPMENT –Part 1-11: General requirements for basic safety and essential performance –Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment, ISO 80601-2-56: 2009 Medical electrical equipment — Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement standards for performance effectiveness.
#### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff. "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Moderate" level of concern, since a failure or latent flaw in the software could directly result in serious injury or death to the patient or operator.
#### Clinical data:
Clinical testing is conducted per ASTM E 1965-98 Standard Specification For Infrared Thermometers For Intermittent Determination Of Patient Temperature.
#### Summary
Based on the non-clinical and clinical performance as documented in the device development, the subject devices were found to have a performance that is substantially equivalent to the predicate device.
{7}------------------------------------------------
# 8. Comparison to predicate device
| ITEM | Proposed Device | Predicate Device<br>K113761 | Comparison<br>Result |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Manufacture | Shenzhen Brav Electronic<br>Technologies Co., Ltd. | Famidoc Technology Co.,<br>Ltd. | --- |
| Model | EFT-161, EFT-162,<br>EFT-163, EFT-164,<br>EFT-165, EFT-166 | FDIR-V2, FDIR-V4,<br>FIDIR-V5, FDIR-V6,<br>FDIR-V7,FIDIR-V9,<br>FDIR-V9-3, FDIR-V10,<br>FDIR-V15 | --- |
| Indications for Use | The infrared thermometer is<br>intended for the<br>measurement and<br>monitoring of human body<br>temperature by doctors or<br>customers in the hospital or<br>at home. | Famidoc Infrared<br>Thermometers are intended<br>for intermittent measurement<br>and monitoring of human<br>body temperature by<br>consumers in the home. | Different1 |
| Measuring<br>position | Forehead and ear | Forehead and ear | Same |
| Measuring<br>distance | 0cm, Contact | 0cm, Contact | Same |
| Measuring mode | Forehead temperature<br>mode and ear temperature<br>mode | Forehead temperature mode,<br>and ear temperature mode | Same |
| Detection method | Temperature is detected by<br>thermistor and calculated. | Temperature is detected by<br>thermistor and calculated. | Same |
| Material contact to<br>patient | ABS and PE lens filters | ABS | Different2 |
| Design principle | Based on Infrared Sensor<br>technology | Based on Infrared Sensor<br>technology | Same |
| Measurement<br>Range | 32.0°C~42.9°C<br>(89.6°F~109.2°F) | 32.0°C~42.9°C<br>(89.6°F~109.2°F) | Same |
| Accuracy for body<br>temperature<br>measurement | Body Mode: ±0.2°C(0.4°F) | ±0.2°C / 0.4°F @<br>35.5-42.0°C(95.9-107.6°F )<br>±0.3°C / 0.5°F<br>@ 32.0-35.4°C(89.6-95.7°F )<br>+0.3°C / 0.5°F | Different3 |
| | | @ | |
| | | 42.1-42.9°C(107.8-109.2°F) | |
| Resolution of<br>Display | 0.1°C/°F | 0.1°C/°F | Same |
| Operating<br>Environment | 15°C ~ 35°C ( 59°F to<br>95 °F) ≤85% moisture<br>condensation | 10° C ~ 40 ° C (60.8°F to<br>95 °F) , RH 80% or below | Different4 |
| Storage<br>Environment | -20°C~55°C (-4°F~131°F)<br>≤90% moisture<br>condensation | - 25 ° ~ 55 ° C (-13° F to<br>131° F) RH 80% or below | Different5 |
| Power supply | 3Vdc battery | 3Vdc battery | Same |
| Applicable<br>standards | AAMI ANSI ES60601-1,<br>IEC 60601-1-2, IEC<br>60601-1-11, and ASTM<br>E1965-98 | ASTM E1965-98, IEC<br>60601-1 and IEC 60601-1-2 | Different6 |
| Biocompatibility<br>information | ISO 10993-5, ISO<br>10993-10 | ISO 10993-5, ISO 10993-10 | Same |
| Memory Data<br>Limit | memorize 20<br>measurements<br>automatically | memorize 19 measurements<br>automatically | Different7 |
| Product<br>configuration | It is mainly composed with<br>infrared sensor, signal<br>receiving processor,<br>buttons, buzzer, LCD<br>display, battery and etc. | It is mainly composed with<br>infrared sensor, signal<br>receiving processor, buttons,<br>buzzer, LCD display, battery<br>and etc. | Same |
| Temperature unit<br>and conversion | Dual temperature units "°C"<br>and "°F" optional, and the<br>two units can convert by the<br>conversion key<br>automatically | Dual temperature units "°C"<br>and "°F" optional and the two<br>units can convert by the<br>conversion key automatically | Same |
{8}------------------------------------------------
#### Justification for the differences:
- 1) Different Indications for Use
As indicated in the comparison table, the subject device can be used both in hospital and home, which the predicate can be used at home only. But the scenario where the device is intended to use require the same performance and clinical requirement considering the usage of such general public device. Moreover, the EFT-16 series Infrared thermometer has been verified and the results show PASS for all test, so the different extra hospital scenario will not introduce any new questions of safety and effectiveness.
{9}------------------------------------------------
- 2) Different patient-contacting
The patient-contacting materials of the subject include ABS and PE, and both materials show biocompatible after tested per the international standards ISO10993-5 and ISO 10993-10.
- 3) Different accuracy
The difference accuracy is that the subject shows the accuracy is ±0.2°C(0.4°F) for the whole measurement range, while the accuracy of the predicate is the same when the temperature is 35.5-42.0°C(95.9-107.6°F ) , and minor different accuracy when lower than 35.5°C(95.9°F ) or higher than 42.0°C(107.6°F ). The critical feature meets the requirements of the international standards ISO 80601-2-56:2019 and the FDA recognized specific standard ASTM E 1965-98.
- 4) Different Operating Environment
The subject can operate in a narrower environment condition range than the predicate, while can store in wider environment condition range, but both are minor. More important is that such testing has been verified during the design and development process for the environmental testing, and the results show PASS, so such different will not cause any new safety and effectiveness questions.
- 5) Different Storage Environment
Same explanation as the justification for 4# difference.
- 6) Different Applicable Standards
The difference is that the subject device and the predicate refer to different standard, the former uses the US standard AAMI ANSI ES60601-1, while the predicate uses the international standard IEC60601-1.
- 7) Different Memory Data Limit
The subject can store one more data than the predicate, which is verified during all the testing, and the function shows stable.
As seen in the comparison tables, the subject and predicate devices have similar design features and performance specifications. The main technological differences between the subject and predicate devices are minor differences, including only the appearance, control keys, and operating/storage environment which do not raise different questions of safety or effectiveness. Moreover, as demonstrated in the non-clinical testing, the different technological characteristics do not affect the performance of the EFT-16 series Infrared thermometer system.
{10}------------------------------------------------
## 9 Conclusion
Based on the above considerations table, the Proposed Device, the EFT-16 series Infrared thermometer is substantially equivalent to the predicate device Famidoc Infrared thermometer (K113761).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.