K102947 · K-Jump Health Co., Ltd. · FLL · Feb 25, 2011 · General Hospital
Device Facts
Record ID
K102947
Device Name
NON-CONTACT INFRARED THERMOMETER
Applicant
K-Jump Health Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
Feb 25, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The non-contact infrared thermometer, Model KI-8280, can measure body temperature for infants and adults without contact to human body. It can be used by consumers in household environment and doctor in clinic as reference.
Device Story
Non-contact infrared thermometer (Model KI-8280) uses thermopile sensor to detect radiated infrared energy from human body surface (3-8cm distance). Device processes electronic signals to calculate and display oral-equivalent temperature. Operated by consumers or clinicians in home or clinic settings. Provides non-invasive temperature monitoring to assist in clinical decision-making.
Clinical Evidence
Clinical performance testing conducted per ASTM E1965 requirements. Study compared temperature readings between a digital thermometer and the KI-8280. Results demonstrated that temperature differences were within acceptable ranges, supporting the device's clinical accuracy for human body temperature measurement.
Technological Characteristics
Infrared thermopile sensor; 3-8cm measurement distance; LCD display; 10-set memory; powered by two 1.5V AAA batteries. Complies with ASTM E1965, EN 12470-5, IEC 60601-1 (safety), and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for measuring body temperature in infants and adults without physical contact. Suitable for use by consumers in household settings and by clinicians in medical facilities.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{0}------------------------------------------------
### EXHIBIT 1
### 510(k) SUMMARY
FEB 2 5 2011
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
K102947 The assigned 510(k) number is:
### 1. Submitter's Identification:
K-jump Health Co., Ltd. No. 56 Wu Kung 5th Road Wu Ku Industrial Park Taipei Hsien, 248, Taiwan Tel: +886-2-22991378 Fax: +886-2-22331386
Contact: Mr. Jason Cheng Date Summary Prepared: September 30, 2010
## 2. Name of Device:
Non-Contact Infrared Thermometer, Model KI-8280
Regulation Number: 21 CFR 880.2910 Regulation Name: Thermometer, Electronic, Clinical Regulatory Class: II Product Code: FLL
### 3. Predicate Device Information:
- K080759, ThermoFlash LX-26, JXB Co., Ltd., Guangzhou, China -
### 4. Device Description:
The Non-Contact Infrared Thermometer, Model KI-8280, uses infrared sensor
Page 1 of 4
{1}------------------------------------------------
### EXHIBIT 1
(thermopile) to detect the radiated infrared energy emitted by the objects, solid, liquid or gas. The intensity of the emitted energy depends on the temperature of the object and the infrared sensor can recognize it to transfer to the proper electronic signal. The electronic signal can be processed in the subject device to convert to the temperature reading. Therefore, the subject device is able to measure the temperature of a person by the energy the person emits. The predicate device, JXB Thermoflash, use the same detection principle to measure the patient's temperature.
The subject device also use the focusing design to collect the infrared emitted from nearby area of object surface. This mechanism make the subject device have the ability to detect the object surface temperature in the distance of 3-8cm. The subject device intended to detect the temperature of patients. The temperature is converted to the oral temperature. The temperature reading shown in the device display after measuring is the patient's oral temperature.
The compact, small and light-weight design, the K-jump Health Co, Ltd. Non-Contact Infrared Thermometer, Model KI-8280, enables to provide safe and reliable results and offers a very good clinical accuracy for human body temperature measurement.
### 5. Intend Use:
The non-contact infrared thermometer, Model KI-8280, can measure body temperature for infants and adults without contact to human body. It can be used by consumers in household environment and doctor in clinic as reference.
#### 6. Comparison to Predicate Devices:
The subject device is substantially equivalent to the predicate devices, K080759, ThermoFlash LX-26. The substantial equivalence chart is provided as follows:
| Characteristics | K-jump Device<br>(Subject Device) | ThermoFlash LX-26<br>JXB Co., Ltd.<br>K#080759 |
|-----------------------|--------------------------------------------------------------------|----------------------------------------------------------------------------------|
| Measurement<br>method | Infrared radiation detection | Infrared radiation detection |
| Measuring Range | Forehead Mode:<br>$32.2°C~43.3°C(90.0°F~109.9°F)$<br>Surface Mode: | In body mode: $32°C - 42.9°C$<br>(90°F - 109°F)<br>In Surface Temp mode: $0°C ~$ |
{2}------------------------------------------------
| | $0.0℃~100.0℃ (32°F~212°F)$ | $60°C (32°F to 140°F)$ |
|---------------------------|---------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Display Resolution | $0.1°C (0.1°F)$ | $0.1°C (0.1°F)$ |
| C/F Switchable | Yes | Yes |
| Measuring<br>accuracy | Forehead Mode:<br>$±0.2℃ (0.4°F)$<br>Surface Mode:<br>$±1°C(1.8°F)$ | $±0.2℃ (0.4°F)$ range from 36<br>$-39.0℃ (93.2-102.2°F)$<br>$±0.3℃ (0.5°F)$ range from 34 ~<br>$35.9℃ (93.2~96.6°F) & 39 ~$<br>$42.5℃ (102.2~108.5°F)$ |
| Display | LCD display | LCD display |
| Measurement<br>Distance | 3 - 8 cm | 5 - 8 cm |
| Key | Two button (On/Off, Memory) | Four button (measuring, Menu,<br>Setting*2) |
| Memory | 10 sets | 32 sets |
| Power source | Two 1.5V AAA batteries | Two 1.5V AA batteries |
| Low battery<br>Indication | Replace the battery if the low<br>battery indication appears | Replace the battery if the low<br>battery indication appears |
| Waterproof | No | No |
| Dimension | 138X90X45 mm | 196x150x50 mm |
| Weight | 125g (including batteries) | 220g |
| Operating<br>condition | $10℃~40℃ (50.0 ~ 104°F),$<br>< 95% RH, non-condensing | $10℃ ~ 40℃$<br>< 85% RH |
## 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
The following performance testing was conducted:
- · The Non-Contact Infrared Thermometer, KI-8280, is complied with voluntary standards includes ASTM E1965 and EN 12470-5.
- The recognized consensus standards for safety of medical electrical equipment: IEC 60601-1 with amendments for safety and IEC 60601-1-2 for eiectromagnetic compatibility are complied.
- Guidance documents included the "FDA Guidance on the Content of Premarket Notification 510(k) submissions for Clinical Electronic Thermometers".
## 8. Discussion of Clinical Tests Performed:
{3}------------------------------------------------
The clinical performance test protocol and data analysis is followed the requirements the ASTM E 1965. The test report shows the patient's temperature readings difference between digital thermometer and the subject device, KI-8280, are within acceptable range. It can conclude that the Non-Contact Infrared Thermometer, KI-8280, is acceptable to measure patient's temperature
### 9. Conclusions:
Non-Contact Infrared Thermometer, Model KI-8280, has the same intended use and similar characteristics as the predicate device. Moreover, the subject device demonstrates product safety by successful completion of testing to the IEC60601-1 standard and electromagnetic standard, IEC 60601-1-2. The performance test demonstrates the KI-8280 meets the ASTM E1965 and EN 12470-5 standards and concludes that any differences in their characteristics do not rise any safety and effectivness issues.
From the above information we conclude the the subject device, KI-8280, is substantially equivalent to the predicate devices, ThermoFalsh LX-26.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the United States Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol featuring three abstract human figures, with the figures appearing to be connected or overlapping.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Jason Cheng Regulatory Affairs K-jump Health Company, Limited No. 56 Wu Kung 5th Road Wu Ku Industrial Park Taipei Hsien, 248, Taiwan
# FEB 2 5 2011
Re: K102947
Trade/Device Name: Non-Contact Infrared Thermometer, Model KI-8280 Regulation Number: 21 CFR 880.2910 Regulation Name: Thermometer, Electronic, Clinical Regulatory Class: II Product Code: FLL Dated: February 16, 2011 Received: February 18, 2011
Dear Mr. Cheng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
### Page 2- Mr. Cheng
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance,
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital,
Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### Indications for Use Statement
Page of 1
#### 510(k) Number (if known): K102947
Device Name: Non-Contact Infrared Thermometer, Model KI-8280
Indications for Use:
Non-Contact Infrared Thermometer, Model KI-8280, can measure body temperature for infants and adults without contact to human body. It can be used by consumers in household environment and doctor in clinic as reference.
Prescription Use __ OR (Per 21 CFR 801 Subpart D)
Over-The Counter Use × (21 CFT 807 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (DE)
- Device Evaluation (100) 2/25/11
(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital
Division of Anesthesion Devices Division of Anaobal - Infection Control, Devices
510(k) Number: K102947
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.