PBZ · Image Processing Device For Estimation Of External Blood Loss

General Hospital · 21 CFR 880.2750 · Class 2

Overview

Product CodePBZ
Device NameImage Processing Device For Estimation Of External Blood Loss
Regulation21 CFR 880.2750
Device ClassClass 2
Review PanelGeneral Hospital

Identification

An image processing device for estimation of external blood loss is a device to be used as an aid in estimation of patient external blood loss. The device may include software and/or hardware that is used to process images capturing externally lost blood to estimate the hemoglobin mass and/or the blood volume present in the images.

Classification Rationale

Class II (special controls). The special controls for this device are:

Special Controls

In combination with the general controls of the Food, Drug & Cosmetic Act, the Image processing device for estimation of external blood loss is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. Demonstration of the performance characteristics must include a comparison to a scientifically valid alternative method for measuring deposited hemoglobin mass. The following use conditions must be tested: (i) Lighting conditions; (ii) Range of expected hemoglobin concentrations; (iii) Range of expected blood volume absorption; and (iv) Presence of other non-sanguineous fluids ( *e.g.,* saline irrigation fluid).(2) Human factors testing and analysis must validate that the device design and labeling are sufficient for appropriate use by intended users of the device. (3) Appropriate analysis and non-clinical testing must validate the electromagnetic compatibility (EMC) and wireless performance of the device. (4) Appropriate software verification, validation, and hazard analysis must be performed. (5) Software display must include an estimate of the cumulative error associated with estimated blood loss values. (6) Labeling must include: (i) Warnings, cautions, and limitations needed for safe use of the device; (ii) A detailed summary of the performance testing pertinent to use of the device, including a description of the bias and variance the device exhibited during testing; (iii) The validated surgical materials, range of hemoglobin mass, software, hardware, and accessories that the device is intended to be used with; and (iv) EMC and wireless technology instructions and information.

Recent Cleared Devices (6 of 6)

RecordDevice NameApplicantDecision DateDecision
K252282SurgiCount+ SystemStryker InstrumentsMar 16, 2026SESE
K232250SurgiCount+ SystemStryker InstrumentsJan 11, 2024SESE
K163507Triton Sponge SystemGauss Surgical, Inc.,Apr 25, 2017SESE
K160338Triton SystemGauss Surgical, Inc.,Aug 5, 2016SESE
K142801Triton Canister SystemGauss Surgical, Inc.,Mar 12, 2015SESE
DEN130015PIXEL 3 SYSTEMGauss Surgical, Inc.,May 9, 2014DENG

Top Applicants

Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...