K252282 · Stryker Instruments · PBZ · Mar 16, 2026 · General Hospital
Device Facts
Record ID
K252282
Device Name
SurgiCount+ System
Applicant
Stryker Instruments
Product Code
PBZ · General Hospital
Decision Date
Mar 16, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2750
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
Indications for Use
The SurgiCount+ System Software is a multi-functional software application intended to be used as an adjunct in the estimation of blood loss and management of surgical sponges and other absorbent items in operating and recovery rooms, procedure rooms, and surgical centers. The system incorporates three distinct software configurations: Triton AI, SC+ Sponge Counting, and Triton QBL. Additionally, combined workflows (SC+AI and SC+QBL) are provided for use in clinical environments that require both the tracking of absorbent surgical items and estimation of patient blood loss. The Triton AI configuration is intended to be used with surgical sponges, software, hardware, and accessory devices that have been validated for use with the application to estimate the hemoglobin (Hb) mass contained on used surgical sponges. The configuration is also intended to calculate an estimate of blood volume on used surgical sponges from the estimated Hb mass and a user-entered patient Hb value. The validated surgical sponges, hardware, software, accessory devices and Hb mass ranges are listed in the Instructions for Use. The SC+ Sponge Counting configuration is intended for use as an adjunctive technology for augmenting the manual process of counting, displaying, and recording the number of RFID-tagged absorbent articles used during surgical procedures, and providing a non-invasive means of locating RFID-tagged absorbent articles within an operating room and surgical sites. The Triton QBL configuration is intended to be used to record the weight of used surgical sponges and other absorbent items in order to calculate the quantity of fluid volume on the sponges/absorbent items.
Device Story
SurgiCount+ System is a multi-functional software application (SaMD) used in surgical environments to track absorbent items and estimate blood loss. Inputs include images of used surgical sponges (Triton AI), RFID signals from tagged sponges (SC+ Sponge Counting), and weight data from Bluetooth-enabled scales (Triton QBL). The system processes these inputs to calculate hemoglobin mass, estimate blood volume, count sponges, and locate missing items. Operated by clinical staff in ORs/procedure rooms, the system provides real-time data to assist in surgical management and blood loss monitoring. Output is displayed to the healthcare provider to support clinical decision-making regarding patient fluid status and surgical safety. Benefits include improved accuracy in blood loss estimation and reduced risk of retained surgical items.
Clinical Evidence
No clinical testing was provided to support this submission. Performance was supported by non-clinical design verification and validation testing leveraged from the predicate device.
Technological Characteristics
SaMD application; utilizes AI image analysis, RFID tracking, and weight-based fluid calculation. Hardware includes Apple iPad Pro, RFID-tagged sponges, RFID reader, and Bluetooth-enabled scale. System is networked/standalone; software configurations include Triton AI, SC+ Sponge Counting, and Triton QBL.
Indications for Use
Indicated for use in operating rooms, recovery rooms, procedure rooms, and surgical centers as an adjunct for estimating external blood loss and managing surgical sponges/absorbent items. Applicable to patients undergoing surgical procedures where tracking of absorbent items and blood loss estimation is required.
Regulatory Classification
Identification
An image processing device for estimation of external blood loss is a device to be used as an aid in estimation of patient external blood loss. The device may include software and/or hardware that is used to process images capturing externally lost blood to estimate the hemoglobin mass and/or the blood volume present in the images.
Special Controls
In combination with the general controls of the Food, Drug & Cosmetic Act, the Image processing device for estimation of external blood loss is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. Demonstration of the performance characteristics must include a comparison to a scientifically valid alternative method for measuring deposited hemoglobin mass. The following use conditions must be tested:
(i) Lighting conditions;
(ii) Range of expected hemoglobin concentrations;
(iii) Range of expected blood volume absorption; and
(iv) Presence of other non-sanguineous fluids (
*e.g.,* saline irrigation fluid).(2) Human factors testing and analysis must validate that the device design and labeling are sufficient for appropriate use by intended users of the device.
(3) Appropriate analysis and non-clinical testing must validate the electromagnetic compatibility (EMC) and wireless performance of the device.
(4) Appropriate software verification, validation, and hazard analysis must be performed.
(5) Software display must include an estimate of the cumulative error associated with estimated blood loss values.
(6) Labeling must include:
(i) Warnings, cautions, and limitations needed for safe use of the device;
(ii) A detailed summary of the performance testing pertinent to use of the device, including a description of the bias and variance the device exhibited during testing;
(iii) The validated surgical materials, range of hemoglobin mass, software, hardware, and accessories that the device is intended to be used with; and
(iv) EMC and wireless technology instructions and information.
K160338 — Triton System · Gauss Surgical, Inc., · Aug 5, 2016
K142801 — Triton Canister System · Gauss Surgical, Inc., · Mar 12, 2015
DEN130015 — PIXEL 3 SYSTEM · Gauss Surgical, Inc., · May 9, 2014
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 16, 2026
Stryker Instruments
Michael Mazelin
Senior Staff Regulatory Affairs Specialist
1941 Stryker Way
Portage, Michigan 49002
Re: K252282
Trade/Device Name: SurgiCount+ System
Regulation Number: 21 CFR 880.2750
Regulation Name: Image Processing Device For of External Blood Loss
Regulatory Class: II
Product Code: PBZ
Dated: February 12, 2026
Received: February 13, 2026
Dear Michael Mazelin:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252282 - Michael Mazelin
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K252282 - Michael Mazelin
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Monica D. Garcia -S
Monica D. Garcia, Ph.D.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252282 | |
| Device Name SurgiCount+ System | |
| Indications for Use (Describe) The SurgiCount+ System Software is a multi-functional software application intended to be used as an adjunct in the estimation of blood loss and management of surgical sponges and other absorbent items in operating and recovery rooms, procedure rooms, and surgical centers. The system incorporates three distinct software configurations: Triton AI, SC+ Sponge Counting, and Triton QBL. Additionally, combined workflows (SC+AI and SC+QBL) are provided for use in clinical environments that require both the tracking of absorbent surgical items and estimation of patient blood loss. The Triton AI configuration is intended to be used with surgical sponges, software, hardware, and accessory devices that have been validated for use with the application to estimate the hemoglobin (Hb) mass contained on used surgical sponges. The configuration is also intended to calculate an estimate of blood volume on used surgical sponges from the estimated Hb mass and a user-entered patient Hb value. The validated surgical sponges, hardware, software, accessory devices and Hb mass ranges are listed in the Instructions for Use. The SC+ Sponge Counting configuration is intended for use as an adjunctive technology for augmenting the manual process of counting, displaying, and recording the number of RFID-tagged absorbent articles used during surgical procedures, and providing a non-invasive means of locating RFID-tagged absorbent articles within an operating room and surgical sites. The Triton QBL configuration is intended to be used to record the weight of used surgical sponges and other absorbent items in order to calculate the quantity of fluid volume on the sponges/absorbent items. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K252282
stryker
510(k) Summary – K252282
| 807.92(a)(1) – Submitter Information | |
| --- | --- |
| 510(k) Submitter: | Stryker Instruments
1941 Stryker Way
Portage, MI 49002 USA |
| Contact Information: | Michael Mazelin
Sr. Staff Regulatory Affairs Specialist
Stryker Instruments
1941 Stryker Way
Portage, MI 49002
269-369-1909 |
| Date Summary Prepared: | March 13, 2026 |
| 807.92(a)(2) – Name of Device | |
| --- | --- |
| Trade Name(s): | SurgiCount+ System |
| Common Name: | External Blood Loss Estimation System |
| Regulation Name: | Image Processing Device for Estimation of External Blood Loss |
| Regulation Number: | 21 CFR §880.2750 |
| Device Class: | II |
| Product Code(s): | PBZ |
| 807.92(a)(3) – Predicate Device |
| --- |
| SurgiCount+ System – K232250
The predicate device has not been subject to a design-related recall. |
| 807.92(a)(4) – Device Description |
| The SurgiCount+ System Software is a multifunctional software application intended to be used as an adjunct in the estimation of blood loss and management of surgical sponges and other absorbent items in operating and recovery rooms, procedure rooms, and surgical centers.
The system incorporates three distinct software configurations: Triton AI, SC+ Sponge Counting, and Triton QBL. Additionally, combined workflows (SC+ AI and SC+QBL) are provided for use in clinical environments that require both the tracking of absorbent surgical items and estimation of patient blood loss.
The Triton AI software configuration estimates the hemoglobin (Hb) mass contained on used surgical sponges using an AI algorithm that analyzes the image of each sponge. It also calculates an estimate of blood volume on used surgical sponges from the estimated Hb mass and a user-entered patient Hb value. Workflow steps allow users to scan, identify, and count RFID-tagged surgical sponges and other absorbent items, and to locate missing surgical sponges inside the operating room and, noninvasively, in surgical sites.
The SaMD product includes the following nonmedical device and consumable and hardware accessories: a mobile device (Apple iPad Pro), RFID-tagged surgical sponges/absorbent items, an RFID reader, a Bluetooth-enabled |
Stryker Instruments
1941 Stryker, Portage, MI 49002 www.stryker.com
{5}
K252282
stryker
scale, and a stand (or optional wall mount) that houses the hardware accessories and connects the accessory devices to electrical power.
| 807.92(a)(5) – Intended Use/Indications for Use of the Device |
| --- |
| The SurgiCount+ System Software is a multi-functional software application intended to be used as an adjunct in the estimation of blood loss and management of surgical sponges and other absorbent items in operating and recovery rooms, procedure rooms, and surgical centers. The system incorporates three distinct software configurations: Triton AI, SC+ Sponge Counting, and Triton QBL. Additionally, combined workflows (SC+AI and SC+QBL) are provided for use in clinical environments that require both the tracking of absorbent surgical items and estimation of patient blood loss.
The Triton AI configuration is intended to be used with surgical sponges, software, hardware, and accessory devices that have been validated for use with the application to estimate the hemoglobin (Hb) mass contained on used surgical sponges. The configuration is also intended to calculate an estimate of blood volume on used surgical sponges from the estimated Hb mass and a user-entered patient Hb value. The validated surgical sponges, hardware, software, accessory devices and Hb mass ranges are listed in the Instructions for Use.
The SC+ Sponge Counting configuration is intended for use as an adjunctive technology for augmenting the manual process of counting, displaying, and recording the number of RFID-tagged absorbent articles used during surgical procedures, and providing a non-invasive means of locating RFID-tagged absorbent articles within an operating room and surgical sites.
The Triton QBL configuration is intended to be used to record the weight of used surgical sponges and other absorbent items in order to calculate the quantity of fluid volume on the sponges/absorbent items. |
| 807.92(a)(5) – Indications for Use Comparison |
| The Indications for Use Statement has been updated to specify the general environments of use, and the intended use of the subject device as compared to the predicate device remains the same (i.e., an adjunctive technology that is intended to be used to estimate external patient blood loss and in the management of surgical sponges). Key changes include the clarification of the general environments of use (i.e., operating and recovery rooms, procedure rooms, and surgical centers), which does not raise different questions of safety and effectiveness as compared to the predicate device. |
| 807.92(a)(6) Summary of the Technological Characteristics of the Device Compared to the Predicate Device |
| The subject and predicate device have identical technological characteristics, and there have been no changes to the design, software, and fundamental scientific technology. |
| 807.92(b)(1) – Nonclinical Testing to Support Submission |
| --- |
| The function and performance of the subject device was leveraged from the predicate device through non-clinical design verification and validation testing in conjunction with risk assessment. |
| The results demonstrate that the subject device successfully meets the requirements of its intended use. |
| 807.92(b)(2) – Clinical Testing |
| --- |
| No clinical testing was provided to support this submission. |
Stryker Instruments
1941 Stryker, Portage, MI 49002 www.stryker.com
{6}
Page 3 of 3
K252282
stryker
| 807.92(b)(3) – Conclusions Drawn from Testing Performed |
| --- |
| The subject device, in comparison with the legally marketed predicate device, has the same design, software, and fundamental scientific technology. Verification and Validation testing and Risk assessment leveraged from the predicate device demonstrates that the device is as safe and effective as the predicate device. |
| Conclusion: |
| --- |
| The information presented above demonstrates that the SurgiCount+ System is as safe and effective as the predicate device and supports a determination of substantial equivalence. |
Stryker Instruments
1941 Stryker, Portage, MI 49002 www.stryker.com
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.