ConFirm 20 Minute Incubator

K233681 · STERIS Corporation · FRC · Dec 15, 2023 · General Hospital

Device Facts

Record IDK233681
Device NameConFirm 20 Minute Incubator
ApplicantSTERIS Corporation
Product CodeFRC · General Hospital
Decision DateDec 15, 2023
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 880.2800
Device ClassClass 2

Indications for Use

Use the ConFirm 20 Minute Incubate and automatically read ConFirm Rapid 20 Minute Biological Indicators for Steam and Vaporized Hydrogen Peroxide sterilization at 55°C-60°C for a fluorescent result.

Device Story

ConFirm 20 Minute Incubator is a benchtop device for processing biological indicators (BIs) used in steam and vaporized hydrogen peroxide sterilization. The device features 8 wells arranged in two banks of four. The operator scans the BI barcode, which informs the system of the appropriate algorithm, and places the activated BI into a well. The incubator maintains a temperature of 59°C (-4°C/+1°C). During incubation, UV LEDs excite fluorescent molecules produced by enzymatic activity (a-glucosidase) within the BI; a photodiode detects the emitted light. The system monitors for fluorescence to determine sterilization success or failure. Results are displayed via LCD screen, LED indicators, and audible alarms. A positive result indicates potential sterilization failure, requiring operator acknowledgment. The device is intended for use in clinical or healthcare settings to verify sterilization efficacy, providing rapid feedback to ensure patient safety by confirming the sterility of processed instruments.

Clinical Evidence

Bench testing only. Regression testing was performed to evaluate performance and demonstrate substantial equivalence to the predicate device. No clinical data was required or provided.

Technological Characteristics

Benchtop incubator with 8 wells; 100-240 VAC power with 12 VDC conversion; 55-60°C operating temperature. Sensing principle: UV LED excitation of fluorogenic substrate and photodiode detection of emitted fluorescence. Barcode scanner for input. User interface: LCD screen, LED status lights, audible alarm. Factory calibrated.

Indications for Use

Indicated for use in incubating and automatically reading ConFirm Rapid 20 Minute Biological Indicators for Steam and Vaporized Hydrogen Peroxide sterilization at 55°C-60°C to detect fluorescent results.

Regulatory Classification

Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below. December 15, 2023 Steris Corporation Logan Persons Regulatory Affairs Specialist 5960 Heisley Rd Mentor, Ohio 44060 Re: K233681 Trade/Device Name: ConFirm 20 Minute Incubator Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: FRC Dated: November 16, 2023 Received: November 16, 2023 Dear Logan Persons: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Christopher K. Dugard -S Christopher K. Dugard, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical {2}------------------------------------------------ and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K233681 Device Name ConFirm 20 Minute Incubator Indications for Use (Describe) Use the ConFirm 20 Minute Incubate and automatically read ConFirm Rapid 20 Minute Biological Indicators for Steam and Vaporized Hydrogen Peroxide sterilization at 55°C-60°C for a fluorescent result. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the STERIS logo. The word STERIS is in all caps and in a serif font. Below the word STERIS is a graphic of several horizontal wavy lines in blue. # 510(k) Summary For ConFirm 20 Minute Incubator K233681 STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198 Contact: Mr. Logan Persons Regulatory Affairs Specialist Tel: 440-392-7325 Fax: 440-357-9198 Logan persons@steris.com Summary Date: December 15, 2023 {5}------------------------------------------------ #### 1. Device Name | Trade Name: | ConFirm 20 Minute Incubator | |------------------------|-------------------------------------------------| | Device Classification: | II | | Common Name: | Incubator | | Classification Name: | Indicator, Biological Sterilization Process | | Classification Panel: | General Hospital and Personal Use Devices Panel | | Classification Number: | 21 CFR 880.2800 | | Product Code | FRC | #### 2. Predicate Device K223715 - Celerity Incubator Reference Device K173634 - ConFirm Biological Indicator (cleared as CELERITY 20 Steam Biological Indicator) #### 3. Device Description The ConFirm 20 Minute Incubator incubates and reads fluorescent Vaporized Hydrogen Peroxide (VHP) and Steam Biological Indicators (BIs). The Bl is incubated at 59°C with an acceptable tolerance of -4°C/+1°C. During incubation, the BI is monitored for a potential fluorescence signal generated as a result of the production of a-glucosidase. When a growth response has been detected or when the required incubation time has elapsed, the incubator indicates the results to the user. #### Indications for Use 4. Use the ConFirm 20 Minute Incubator to incubate and automatically read ConFirm Rapid 20 Minute Biological Indicators for Steam and Vaporized Hydrogen Peroxide sterilization at 55°C-60°C for a fluorescent result. #### ડ. Technological Characteristics Comparison Table A comparison of technical characteristics between the proposed and predicate devices is summarized in Table 1. | Feature | Modified Device<br>ConFirm 20 Minute<br>Incubator | Predicate Device<br>Celerity Indicator<br>(K223715) | Comparison | |---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------| | Indications<br>for Use | Use the ConFirm 20 | Use the Celerity Incubator | Same, only | | Feature | Modified Device<br>ConFirm 20 Minute<br>Incubator | Predicate Device<br>Celerity Indicator<br>(K223715) | Comparison | | | Minute Incubator to<br>incubate and automatically<br>read ConFirm Rapid 20<br>Minute Biological<br> | to incubate and<br>automatically read STERIS<br>Celerity Biological<br>Indicators for Steam and<br>Vaporized Hydrogen<br>Peroxide sterilization at<br>55°C - 60°C for a<br>fluorescent result. | change is<br>branding of<br>incubator and<br>BI with which<br>it is used | | Basis of<br>Readout | Photodiode detects<br>fluorescence produced by<br>enzymatic activity that<br>results from viable<br>biological indicator<br>organisms. | Photodiode detects<br>fluorescence produced by<br>enzymatic activity that<br>results from viable<br>biological indicator<br>organisms. | Same | | Incubation<br>Temperature Range | 55 - 60 °C | 55 - 60 °C | Same | | Voltage<br>Range | 100-240 VAC with 12<br>VDC conversion | 100-240 VAC with 12<br>VDC conversion | Same | | Test<br>Capacity | 8 wells | 8 wells | Same | | Calibration | Factory calibration – no<br>calibration by customer | Factory calibration – no<br>calibration by customer | Same | | Incubation<br>Time | 20 minutes | 5 minutes or 20 minutes<br>depending on Biological<br>Indicator | Similar, the 5-<br>minute BI is<br>not being<br>rebranded. | | Fluorescence<br>Detection | UV LEDs are used to<br>excite the fluorescent<br>molecule produced by<br>enzyme cleavage of the<br>fluorogenic substrate<br>contained in the BI's<br>media. The emitted light is<br>detected by a photodiode. | UV LEDs are used to<br>excite the fluorescent<br>molecule produced by<br>enzyme cleavage of the<br>fluorogenic substrate<br>contained in the BI's<br>media. The emitted light is<br>detected by a photodiode. | Same | | Indication of<br>Results | Positive – audible alarm,<br>visual LED lights and | Positive – audible alarm,<br>visual LED lights and | Same | | Feature | Modified Device<br>ConFirm 20 Minute<br>Incubator | Predicate Device<br>Celerity Indicator<br>(K223715) | Comparison | | | screen<br>Negative - optional alarm,<br>visual indication with LED<br>lights and LCD screen<br>User must acknowledge<br>results | screen<br>Negative - optional alarm,<br>visual indication with LED<br>lights and LCD screen<br>User must acknowledge<br>results | | | System<br>Operation | The reader/incubator wells<br>are arranged in 2 banks of<br>4 wells and preset to 59°C.<br>The user scans the barcode<br>on the label of an activated<br>BI using the system's<br>barcode scanner and places<br>it into an open well. The<br>system detects the well the<br>BI was placed into and<br>begins measurement of<br>fluorescence; a blinking<br>yellow light indicates the<br>incubation is in process<br>and the read initiated. The<br>System uses information<br>from the barcode to apply<br>the appropriate algorithm<br>to the well. A "positive"<br>reading is interpreted as an<br>indication of a potential<br>sterilization cycle failure.<br>A “positive” finding is<br>indicated to the user by red<br>light on the front panel<br>adjacent to the well, by an<br>audible alarm, and by text<br>displayed on the LCD<br>screen. The alarm must be<br>muted by the operator<br>when a positive result is<br>obtained. The LCD screen<br>provides instructions for<br>the user to turn off the<br>alarm for that specific BI. | The reader/incubator wells<br>are arranged in 2 banks of<br>4 wells and preset to 59°C.<br>The user scans the barcode<br>on the label of an activated<br>BI using the system's<br>barcode scanner and places<br>it into an open well. The<br>system detects the well the<br>BI was placed into and<br>begins measurement of<br>fluorescence; a blinking<br>yellow light indicates the<br>incubation is in process<br>and the read initiated. The<br>System uses information<br>from the barcode to apply<br>the appropriate algorithm<br>to the well. A "positive"<br>reading is interpreted as an<br>indication of a potential<br>sterilization cycle failure.<br>A “positive” finding is<br>indicated to the user by red<br>light on the front panel<br>adjacent to the well, by an<br>audible alarm, and by text<br>displayed on the LCD<br>screen. The alarm must be<br>muted by the operator<br>when a positive result is<br>obtained. The LCD screen<br>provides instructions for<br>the user to turn off the<br>alarm for that specific BI. | Same | | Feature | Modified Device<br>ConFirm 20 Minute<br>Incubator | Predicate Device<br>Celerity Indicator<br>(K223715) | Comparison | | | Should another BI become<br>"positive", the alarm will<br>sound again and the above<br>actions are repeated.<br>If the result is not positive<br>at the end of the incubation<br>time, the result is negative.<br>Negative results are<br>identified by a green light<br>on the front panel adjacent<br>to the well with the<br>"negative" BI and by text<br>on the LCD. In addition, an<br>optional alarm is available<br>for negative results. | Should another BI become<br>"positive", the alarm will<br>sound again and the above<br>actions are repeated.<br>If the result is not positive<br>at the end of the incubation<br>time, the result is negative.<br>Negative results are<br>identified by a green light<br>on the front panel adjacent<br>to the well with the<br>"negative" BI and by text<br>on the LCD. In addition, an<br>optional alarm is available<br>for negative results. | | ## Table 1. Technological Characteristics Comparison Table to predicate device {6}------------------------------------------------ {7}------------------------------------------------ {8}------------------------------------------------ #### 6. Summary of Non-Clinical Testing Testing was performed to evaluate performance and demonstrate substantial equivalence to the predicate as summarized in Table 2. # Table 2. Performance Testing | Test | Acceptance Criteria | Result | |-----------------------|-------------------------|--------| | Regression<br>Testing | No unresolved anomalies | PASS | #### 7. Conclusion The ConFirm 20 Minute Incubator has met the established performance criteria. Based on the intended use, technological characteristics and non-clinical performance data, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device (K223715), Class II (21 CFR 880.2800), product code FRC.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...