VERIFY All-In-One STEAM Reusable Test Pack, VERIFY All-In-One STEAM Reusable Test Pack with Step Counter, VERIFY All-In-One STEAM Reusable Test Pack with Tracker Tag
K182931 · Steris Corporations · FRC · Nov 20, 2018 · General Hospital
Device Facts
Record ID
K182931
Device Name
VERIFY All-In-One STEAM Reusable Test Pack, VERIFY All-In-One STEAM Reusable Test Pack with Step Counter, VERIFY All-In-One STEAM Reusable Test Pack with Tracker Tag
Applicant
Steris Corporations
Product Code
FRC · General Hospital
Decision Date
Nov 20, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
The VERIFY All-In-One STEAM Reusable Test Pack, when used with a VERIFY Bowie Dick Indicator Strip, is designed for testing the air removal efficiency of pre-vacuum steam sterilizers. The indicator will demonstrate complete dye migration when proper sterilization conditions are met and proper vacuum is achieved. If there is inadequate vacuum such that would result in a 2°C (+1°C/-0°C) temperature difference between the center of the ANSI/AAMI/ISO 11140-5:2007 Standard Test Pack and the drain temperature at the beginning of the final minute of a three and a half minute cycle, the indicator will demonstrate incomplete color change in the indicator window). The test parameters used for validating this device were 134℃ for 3.5 minutes with a 2 minute dry time. VERIFY All-In-One STEAM Reusable Test Pack is designed to challenge steam sterilization process in healthcare facilities. It is intended to be used for routine monitoring of pre-vacuum steam sterilization cycles. It is validated to be used in 4-minute 270°F/132°C pre-vacuum steam sterilization a biological indicator, integrating indicator or both a biological and integrating indicator from the following list. - 3M ATTEST Rapid Read Biological Indicators catalog number 1292 manufactured by 3M Company - 3M ATTEST Super Rapid Read Biological Indicators catalog number 1492V manufactured by 3M Company - Smart-Read EZTest Self Contained Biological Indicators for steam manufactured by MesaLabs, Inc. - Celerity 20 Steam Biological Indicator - VERIFY STEAM Integrating Indicator - VERIFY SixCess MF4 STEAM Emulating Indicator VERIFY All-In-One STEAM Reusable Test Pack is designed to challenge the steam sterilization process in healthcare facilities. It is intended to be used for routine monitoring of pre-vacuum steam sterilization cycles. It is validated to be used in 4-minute 270°F/132°C pre-vacuum steam sterilization cycles with the VERIFY STEAM Integrating Indicator. The VERIFY All-In-One STEAM Resuable Test Pack employing the All-in-One PV10 Assembly (a disposable polypropylene sleeve assembly) with a 3M ATTEST Rapid Readout Biological Indicator catalog number 1292 can be used for routine monitoring of 10-Minute 270°P/132°C pre-vacuum steam sterilization cycles. The polypropylene sleeve assembly is intended to be used only with the VERIFY All-In-One STEAM Resuable Test Pack. The VERIFY All-In-One STEAM Reusable Test Pack employing the Celerity Xtend 10 Assembly with a Celerity 20 STEAM Biological Indicator and a Celerity 10 STEAM Chemical Indicator can be used for routine monitoring of 10minute 270°F/132°C dynamic air removal steam sterilization cycles. The sleeve assembly is intended to be used only with the VERIFY All-In-One STEAM Reusable Test Pack.
Device Story
Reusable aluminum test pack; contains tortuous path for air removal/steam penetration; used to challenge steam sterilization processes. User places Celerity Xtend 10 Assembly (polypropylene tube/silicone cap containing Celerity 20 Biological Indicator and Celerity 10 Chemical Indicator) inside pack; places pack in sterilizer chamber. After 10-minute 270°F cycle, user removes assembly; evaluates Chemical Indicator (yellow to blue/purple color change). If pass, Biological Indicator activated by twisting cap to release media; incubated in Celerity Steam Incubator for 20-minute viability determination. Provides healthcare providers with confirmation of sterilization cycle efficacy; ensures patient safety by verifying sterilization conditions.
Clinical Evidence
Bench testing only. Simulated use testing demonstrated that the performance of the biological indicator within the reusable test pack is equivalent to the predicate. Testing confirmed the pack provides a greater challenge to the sterilization process than the Celerity Xtend 10 Assembly alone.
Technological Characteristics
Reusable aluminum chamber with removable aluminum plug and permanently pressed aluminum cap; precision-machined spiral provides tortuous path for air removal. Dimensions accommodate biological and chemical indicators. Compatible with Celerity Xtend 10 Assembly (polypropylene tube, silicone rubber cap).
Indications for Use
Indicated for routine monitoring of pre-vacuum steam sterilization cycles in healthcare facilities. Used for testing air removal efficiency of pre-vacuum steam sterilizers and monitoring 4-minute or 10-minute 270°F/132°C cycles using specified biological and chemical indicators.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
Predicate Devices
VERIFY All-in-One STEAM Reusable Test Pack (K173633)
K173633 — VERIFY All-In-One STEAM Reusable Test Pack, VERIFY All-In-One STEAM Reusable Test Pack with Step Counter, VERIFY All-In-One STEAM Reusable Test Pack with Tracker Tag · STERIS Corporation · Feb 28, 2018
K133549 — STERITEC BIOLOGICAL TEST PACK WITH INSTANT READOUT INTEGRATOR · Steritec Products Mfg Co, Inc. · Aug 27, 2014
K110253 — REUSABLE TEST PACK FOR DANA EMULATING INDICATORS · Dana Products, Inc. · Apr 29, 2011
K140620 — SPORVIEW PLUS BI TEST PACK · Crosstex International · Aug 7, 2014
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 20, 2018
Steris Corporations Jennifer Nalepka Senior Regulatory Affairs Specialist 5960 Heisley Road Mentor, Ohio 44060
Re: K182931
Trade/Device Name: VERIFY All-In-One STEAM Reusable Test Pack, VERIFY All-In-One STEAM Reusable Test Pack with Step Counter, VERIFY All-In-One STEAM Reusable Test Pack with Tracker Tag
Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: FRC Dated: October 19, 2018 Received: October 22, 2018
Dear Jennifer Nalepka:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Elizabeth F. Claverie -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
### Device Name
VERIFY All-In-One STEAM Reusable Test Pack
### Indications for Use (Describe)
The VERIFY All-In-One STEAM Reusable Test Pack, when used with a VERIFY Bowie Dick Indicator Strip, is designed for testing the air removal efficiency of pre-vacuum steam sterilizers. The indicator will demonstrate complete dye migration when proper sterilization conditions are met and proper vacuum is achieved. If there is inadequate vacuum such that would result in a 2°C (+1°C/-0°C) temperature difference between the center of the ANSI/AAMI/ISO 11140-5:2007 Standard Test Pack and the drain temperature at the beginning of the final minute of a three and a half minute cycle, the indicator will demonstrate incomplete color change in the indicator window). The test parameters used for validating this device were 134℃ for 3.5 minutes with a 2 minute dry time.
VERIFY All-In-One STEAM Reusable Test Pack is designed to challenge steam sterilization process in healthcare facilities. It is intended to be used for routine monitoring of pre-vacuum steam sterilization cycles. It is validated to be used in 4-minute 270°F/132°C pre-vacuum steam sterilization a biological indicator, integrating indicator or both a biological and integrating indicator from the following list.
- · 3M ATTEST Rapid Read Biological Indicators catalog number 1292 manufactured by 3M Company
- · 3M ATTEST Super Rapid Read Biological Indicators catalog number 1492V manufactured by 3M Company
- Smart-Read EZTest Self Contained Biological Indicators for steam manufactured by MesaLabs, Inc.
- · Celerity 20 Steam Biological Indicator
- · VERIFY STEAM Integrating Indicator
- · VERIFY SixCess MF4 STEAM Emulating Indicator
VERIFY All-In-One STEAM Reusable Test Pack is designed to challenge the steam sterilization process in healthcare facilities. It is intended to be used for routine monitoring of pre-vacuum steam sterilization cycles. It is validated to be used in 4-minute 270°F/132°C pre-vacuum steam sterilization cycles with the VERIFY STEAM Integrating Indicator.
The VERIFY All-In-One STEAM Resuable Test Pack employing the All-in-One PV10 Assembly (a disposable polypropylene sleeve assembly) with a 3M ATTEST Rapid Readout Biological Indicator catalog number 1292 can be used for routine monitoring of 10-Minute 270°P/132°C pre-vacuum steam sterilization cycles. The polypropylene sleeve assembly is intended to be used only with the VERIFY All-In-One STEAM Resuable Test Pack.
The VERIFY All-In-One STEAM Reusable Test Pack employing the Celerity Xtend 10 Assembly with a Celerity 20 STEAM Biological Indicator and a Celerity 10 STEAM Chemical Indicator can be used for routine monitoring of 10minute 270°F/132°C dynamic air removal steam sterilization cycles. The sleeve assembly is intended to be used only with the VERIFY All-In-One STEAM Reusable Test Pack.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary For VERIFY™ All-in-One STEAM Reusable Test Pack
### Sponsor Facility
STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198
### Manufacturing Facility
STERIS Corporation Franklin Park Facility 11457 Melrose Ave. Franklin Park, IL 60131 Phone: (847) 455-2881
Jennifer Nalepka Contact: Senior Regulatory Affairs Specialist
Telephone: (440) 392-7458 Fax No: (440) 357-9189 e-mail: jennifer_nalepka@steris.com
Submission Date: October 19, 2018
STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600
{5}------------------------------------------------
#### 1. Device Name
| Trade Name: | VERIFY™ All-in-One STEAM Reusable Test Pack |
|------------------------|-----------------------------------------------------------------------|
| Common/usual Name: | Biological Indicator Challenge Pack |
| Device Classification: | Class II |
| Classification Name: | Indicator, Biological Sterilization Process<br>(21 CFR 880.2800, FRC) |
#### 2. Predicate Device
VERIFY All-in-One STEAM Reusable Test Pack K173633
#### 3. Description of Device
The VERIFY All-in-One STEAM Reusable Test Pack consists of a reusable aluminum tube with a tortuous path challenge for air removal and steam penetration. The VERIFY All-in-One STEAM Reusable Test Pack will be used with the Celerity Xtend 10 Assembly containing the Celerity 20 STEAM Biological Indicator and the Celerity 10 STEAM Chemical Indicator.
The Celerity Xtend 10 Assembly consists of a tube and cap design, comprised of one (1) polypropylene tube and one (1) silicone rubber cap containing one (1) Celerity 20 STEAM Biological Indicator and one (1) Celerity 10 STEAM Chemical Indicator and can only be used with the VERIFY A11-in-One STEAM Reusable Test Pack. The Celerity Xtend 10 Assembly has a tamper evident seal that will be broken once the assembly has been opened after use. A broken tamper evident seal will indicate to the user that it cannot be reused. The Celerity Xtend 10 Assembly is a single use device used for routine monitoring of 10-minute 270°F pre-vacuum steam sterilization cycles. The polypropylene sleeve assembly is intended to be used only with the VERIFY All-In-One STEAM Reusable Test Pack.
The Celerity 20 STEAM Biological Indicator monitors the critical parameters of steam sterilization cycles by producing an optical change (signal) that is detected by the STERIS proprietary reader, Celerity STEAM Incubator, in 20 minutes to confirm viability of the biological indicator at the end of a steam sterilization process. Both the Celerity 20 STEAM Biological Indicator has been cleared under K173634 and K181429 and the Incubator has been cleared under K173670.
The Celerity 10 STEAM Chemical Indicator is a single-use device used to monitor a 10-minute 270°F/132°C dynamic air removal steam sterilization cycle and is subject of a concurrent premarket notification.
{6}------------------------------------------------
The user will place the Celerity Xtend 10 Assembly inside the VERIFY All-in-One STEAM Reusable Test Pack and screw on the lid. The user will place the VERIFY All-in-One STEAM Reusable Test Pack into the sterilizer chamber and perform a 10-minute 270°F dynamic air removal steam sterilization cycle.
At the end of the cycle, the user will remove the Celerity Xtend 10 Assembly from the pack. The Celerity 10 STEAM Chemical Indicator can be evaluated for a pass or fail prior to opening the assembly. The Celerity 10 STEAM Chemical Indicator indicates a pass by a color change of yellow to blue/purple. If there is no color change or an incomplete color change, the Celerity 10 STEAM Chemical Indicator indicates a fail.
If the CI indicates a pass, the Celerity 20 STEAM Biological Indicator can either be immediately activated or held at room temperature for a maximum of 72 hours (3 days) prior to activation by twisting the cap and thus puncturing the cap to release the contained media. The activated BI is incubated at 55-60°C in the Celerity Steam Incubator for a final determination of viability within 20 minutes of incubation.
#### 4. Intended Use/ Indications for Use
The VERIFY All-In-One STEAM Reusable Test Pack, when used with a VERIFY Bowie Dick Indicator Strip, is designed for testing the air removal efficiency of prevacuum steam sterilizers. The indicator will demonstrate complete dye migration when proper sterilization conditions are met and proper vacuum is achieved. If there is inadequate vacuum such that would result in a 2°C (+1°C/-0°C) temperature difference between the center of the ANSI/AAMI/ISO 1140-5:2007 Standard Test Pack and the drain temperature at the beginning of the final minute of a three and a half minute cycle, the indicator will demonstrate incomplete dye migration (incomplete color change in the indicator window). The test parameters used for validating this device were 134°C for 3.5 minutes with a 2 minute dry time.
VERIFY All-In-One STEAM Reusable Test Pack is designed to challenge steam sterilization process in healthcare facilities. It is intended to be used for routine monitoring of pre-vacuum steam sterilization cycles. It is validated to be used in 4minute 270°F/132°C pre-vacuum steam sterilization cycles utilizing a biological indicator, integrating indicator or both a biological and integrating indicator from the following list.
- . 3M ATTEST Rapid Read Biological Indicators catalog number 1292 manufactured by 3M Company
- 3M ATTEST Super Rapid Read Biological Indicators catalog number ● 1492V manufactured by 3M Company
- Smart-Read EZTest Self Contained Biological Indicators for steam ● manufactured by MesaLabs, Inc.
{7}------------------------------------------------
- Celerity 20 STEAM Biological Indicator ●
- VERIFY STEAM Integrating Indicator ●
- . VERIFY SixCess MF4 |STEAM Emulating Indicator
VERIFY All-In-One STEAM Reusable Test Pack is designed to challenge the steam sterilization process in healthcare facilities. It is intended to be used for routine monitoring of pre-vacuum steam sterilization cycles. It is validated to be used in 4-minute 270°F/132°C pre-vacuum steam sterilization cycles with the VERIFY |STEAM Integrating Indicator.
The VERIFY All-In-One STEAM Reusable Test Pack employing the All-in-One PV10 Assembly (a disposable polypropylene sleeve assembly) with a 3M ATTEST Rapid Readout Biological Indicator catalog number 1292 can be used for routine monitoring of 10-Minute 270°F/132°C pre-vacuum steam sterilization cycles. The polypropylene sleeve assembly is intended to be used only with the VERIFY All-In-One STEAM Reusable Test Pack.
The VERIFY All-in-One STEAM Reusable Test Pack employing the Celerity Xtend 10 Assembly with a Celerity 20 STEAM Biological Indicator and a Celerity 10 STEAM Chemical Indicator can be used for routine monitoring of 10-Minute 270°F/132°C dynamic air removal steam sterilization cycles. The sleeve assembly is intended to be used only with the VERIFY All-In-One STEAM Reusable Test Pack.
#### 5. Technical Characteristics Comparison Table
A comparison of technical characteristics are summarized in Table 6-1.
| Feature | VERIFY All-In-One<br><span>STEAM</span> Reusable Test<br>Pack<br>(proposed) | VERIFY All-In-One<br><span>STEAM</span> Reusable Test<br>Pack<br>(K173633) | Comparison |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use /<br>Indication for<br>Use | The VERIFY All-In-One<br><span>STEAM</span> Reusable Test<br>Pack, when used with a<br>VERIFY Bowie Dick<br>Indicator Strip, is designed<br>for testing the air removal<br>efficiency of pre-vacuum<br>steam sterilizers. The<br>indicator will demonstrate<br>complete dye migration<br>when proper sterilization<br>conditions are met and<br>proper vacuum is<br>achieved. If there is | The VERIFY All-In-One<br><span>STEAM</span> Reusable Test<br>Pack, when used with a<br>VERIFY Bowie Dick<br>Indicator Strip, is designed<br>for testing the air removal<br>efficiency of pre-vacuum<br>steam sterilizers. The<br>indicator will demonstrate<br>complete dye migration<br>when proper sterilization<br>conditions are met and<br>proper vacuum is<br>achieved. If there is | The proposed and predicate<br>devices are identical. The<br>current premarket<br>notification is to add the use<br>of the Celerity 10 Xtend<br>Assembly for routine<br>monitoring and qualification<br>of 10-minute 270°F/132°C<br>prevacuum steam<br>sterilization cycles in the<br>VERIFY All-In-One<br><span>STEAM</span> Reusable Test<br>Pack. |
| Feature | VERIFY All-In-One<br>STEAM Reusable Test<br>Pack<br>(proposed) | VERIFY All-In-One<br>STEAM Reusable Test<br>Pack<br>(K173633) | Comparison |
| | inadequate vacuum such<br>that would result in a 2°C<br>(+1°C/-0°C) temperature<br>difference between the<br>center of the<br>ANSI/AAMI/ISO 1140-<br>5:2007 Standard Test Pack<br>and the drain temperature<br>at the beginning of the<br>final minute of a three and<br>a half minute cycle, the<br>indicator will demonstrate<br>incomplete dye migration<br>(incomplete color change<br>in the indicator window).<br>The test parameters used<br>for validating this device<br>were 134°C for 3.5<br>minutes with a 2 minute<br>dry time. | inadequate vacuum such<br>that would result in a 2°C<br>(+1°C/-0°C) temperature<br>difference between the<br>center of the<br>ANSI/AAMI/ISO 1140-<br>5:2007 Standard Test Pack<br>and the drain temperature<br>at the beginning of the<br>final minute of a three and<br>a half minute cycle, the<br>indicator will demonstrate<br>incomplete dye migration<br>(incomplete color change<br>in the indicator window).<br>The test parameters used<br>for validating this device<br>were 134°C for 3.5<br>minutes with a 2 minute<br>dry time. | The following claims for use<br>of the VERIFY All-In-One<br>Reusable Test Pack have<br>been previously cleared:<br><br>Bowie Dick Indicator when<br>used with the VERIFY<br>Bowie Dick Indicator Strip<br>(K120592)<br><br>Routine Monitoring of a 4-<br>minute 270°F/132°C pre-<br>vacuum steam sterilization<br>cycle when used with:<br>• 3M ATTEST Rapid Read<br>Biological Indicators<br>catalog number 1292<br>manufactured by 3M<br>Company (K092944)<br>• Smart-Read EZTest Self<br>Contained Biological<br>Indicators for steam<br>manufactured by<br>MesaLabs, Inc.<br>(K130135)<br>• VERIFY STEAM<br>Integrating Indicator<br>(K102761)<br>• Celerity 20 STEAM<br>Biological Indicator<br>(K173633)<br>• VERIFY SixCess MF4<br>STEAM Emulating<br>Indicator (K110253)<br>o The VERIFY SixCess<br>MF4 STEAM<br>Emulating Indicator<br>was inadvertently left<br>out of the Indications<br>for Use in the previous<br>clearance for the<br>Reusable Test Pack<br>due to a typographical<br>error.<br><br>Routine Monitoring of a 10-<br>minute 270°F/132°C pre-<br>vacuum steam sterilization<br>cycle when used with the<br>additional polypropylene |
| | VERIFY All-In-One<br>STEAM Reusable Test<br>Pack is designed to<br>challenge steam<br>sterilization process in<br>healthcare facilities. It is<br>intended to be used for<br>routine monitoring of pre-<br>vacuum steam sterilization<br>cycles. It is validated to<br>be used in 4-minute<br>270°F/132°C pre-vacuum<br>steam sterilization cycles<br>utilizing a biological<br>indicator, integrating<br>indicator or both a<br>biological and integrating<br>indicator from the<br>following list.<br>• 3M ATTEST Rapid<br>Read Biological<br>Indicators catalog<br>number 1292<br>manufactured by 3M<br>Company<br>• 3M ATTEST Super<br>Rapid Read Biological<br>Indicators catalog<br>number 1492V | VERIFY All-In-One<br>STEAM Reusable Test<br>Pack is designed to<br>challenge steam<br>sterilization process in<br>healthcare facilities. It is<br>intended to be used for<br>routine monitoring of pre-<br>vacuum steam sterilization<br>cycles. It is validated to<br>be used in 4-minute<br>270°F/132°C pre-vacuum<br>steam sterilization cycles<br>utilizing a biological<br>indicator, integrating<br>indicator or both a<br>biological and integrating<br>indicator from the<br>following list.<br>• 3M ATTEST Rapid<br>Read Biological<br>Indicators catalog<br>number 1292<br>manufactured by 3M<br>Company<br>• 3M ATTEST Super<br>Rapid Read Biological<br>Indicators catalog<br>number 1492V | |
| Feature | VERIFY All-In-One<br>STEAM Reusable Test<br>Pack<br>(proposed) | VERIFY All-In-One<br>STEAM Reusable Test<br>Pack<br>(K173633) | Comparison |
| | manufactured by 3M<br>Company<br>• Smart-Read EZTest<br>Self Contained<br>Biological Indicators<br>for steam manufactured<br>by MesaLabs, Inc.<br>• Celerity 20 Steam<br>Biological Indicator<br>• VERIFY STEAM<br>Integrating Indicator<br>• VERIFY SixCess MF4<br>STEAM Emulating<br>Indicator | manufactured by 3M<br>Company<br>• Smart-Read EZTest<br>Self Contained<br>Biological Indicators<br>for steam manufactured<br>by MesaLabs, Inc.<br>• Celerity 20 Steam<br>Biological Indicator<br>• VERIFY STEAM<br>Integrating Indicator | ATTEST Rapid Readout<br>Biological Indicator catalog<br>number 1292 (K103321)<br>NOTE: In previous<br>submissions, the VERIFY<br>All-In-One STEAM<br>Reusable Pack was referred<br>to as the Dana Reusable Test<br>Pack, the VERIFY STEAM<br>Integrating Integrator was<br>called the SteriScan<br>Integrator and the VERIFY<br>SixCess MF4 STEAM<br>Emulating Indicator was<br>called the Dana Emulating<br>Indicator. |
| | VERIFY All-In-One<br>STEAM Reusable Test<br>Pack is designed to<br>challenge the steam<br>sterilization process in<br>healthcare facilities. It is<br>intended to be used for<br>routine monitoring of pre-<br>vacuum steam sterilization<br>cycles. It is validated to<br>be used in 4-minute<br>270°F/132°C pre-vacuum<br>steam sterilization cycles<br>with the VERIFY<br>STEAM Integrating<br>Indicator. | VERIFY All-In-One<br>STEAM Reusable Test<br>Pack is designed to<br>challenge the steam<br>sterilization process in<br>healthcare facilities. It is<br>intended to be used for<br>routine monitoring of pre-<br>vacuum steam sterilization<br>cycles. It is validated to<br>be used in 4-minute<br>270°F/132°C pre-vacuum<br>steam sterilization cycles<br>with the VERIFY STEAM<br>Integrating Indicator. | |
| | The VERIFY All-In-One<br>STEAM Reusable Test<br>Pack employing the All-<br>in-One PV10 Assembly (a<br>disposable polypropylene<br>sleeve assembly) with a<br>3M ATTEST Rapid<br>Readout Biological<br>Indicator catalog number<br>1292 can be used for<br>routine monitoring of 10-<br>Minute 270°F/132°C pre-<br>vacuum steam sterilization<br>cycles. The polypropylene<br>sleeve assembly is<br>intended to be used only<br>with the VERIFY All-In-<br>One STEAM Reusable | The VERIFY All-In-One<br>STEAM Reusable Test<br>Pack employing the All-<br>in-One PV10 Assembly (a<br>disposable polypropylene<br>sleeve assembly) with a<br>3M ATTEST Rapid<br>Readout Biological<br>Indicator catalog number<br>1292 can be used for<br>routine monitoring of 10-<br>Minute 270°F/132°C pre-<br>vacuum steam sterilization<br>cycles. The polypropylene<br>sleeve assembly is<br>intended to be used only<br>with the VERIFY All-In-<br>One STEAM Reusable | |
| Feature | VERIFY All-In-One<br><b>STEAM</b> Reusable Test<br>Pack<br>(proposed) | VERIFY All-In-One<br><b>STEAM</b> Reusable Test<br>Pack<br>(K173633) | Comparison |
| | The VERIFY All-in-One<br><b>STEAM</b> Reusable Test<br>Pack employing the<br>Celerity Xtend 10<br>Assembly with a Celerity<br>20 <b>STEAM</b> Biological<br>Indicator and a Celerity 10<br><b>STEAM</b> Chemical<br>Indicator can be used for<br>routine monitoring of 10-<br>Minute 270°F/132°C pre-<br>vacuum steam<br>sterilization cycles. The<br>sleeve assembly is<br>intended to be used only<br>with the VERIFY All-In-<br>One STEAM Reusable<br>Test Pack. | | |
| General<br>Design | Aluminum chamber that<br>can be sealed at one end<br>using a removable<br>aluminum plug. The other<br>end has a permanently<br>pressed in aluminum cap<br>with a precision machined<br>spiral that provides a<br>tortuous path for air<br>removal. The aluminum<br>chamber is large enough to<br>contain a biological<br>indicator and chemical<br>indicator. | Aluminum chamber that<br>can be sealed at one end<br>using a removable<br>aluminum plug. The other<br>end has a permanently<br>pressed in aluminum cap<br>with a precision machined<br>spiral that provides a<br>tortuous path for air<br>removal. The aluminum<br>chamber is large enough to<br>contain a biological<br>indicator and chemical<br>indicator. | Identical configuration |
Table 6-1 Summary of pack Physical Description and Technological Properties
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
| | VERIFY All-In-One<br>STEAM Reusable Test<br>Pack<br>(proposed) | VERIFY All-In-One<br>STEAM Reusable Test<br>Pack<br>(K173633) | |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | | | Comparison |
| Biological<br>Indicator | 3M ATTEST Rapid Read Biological Indicators catalog number 1292 manufactured by 3M Company 3M ATTEST Super Rapid Read Biological Indicators catalog number 1492V manufactured by 3M Company Smart-Read EZTest Self Contained Biological Indicators for steam manufactured by MesaLabs, Inc. Celerity 20 STEAM Biological Indicator | 3M ATTEST Rapid Read Biological Indicators catalog number 1292 manufactured by 3M Company 3M ATTEST Super Rapid Read Biological Indicators catalog number 1492V manufactured by 3M Company Smart-Read EZTest Self Contained Biological Indicators for steam manufactured by MesaLabs, Inc. Celerity 20 STEAM Biological Indicator | Identical |
| Chemical<br>Indicator | VERIFY STEAM Integrating Indicator | VERIFY STEAM Integrating Indicator | Identical |
| Extended<br>Cycle<br>Assemblies | PV10 Assembly with a 3M ATTEST Rapid Readout Biological Indicator catalog number 1292 Celerity Xtend 10 Assembly containing a Celerity 20 STEAM Biological Indicator and Celerity 10 STEAM Chemical Indicator | PV10 Assembly with a 3M ATTEST Rapid Readout Biological Indicator catalog number 1292 | The proposed and predicate devices are identical. The current premarket notification is to add the use of the Celerity 10 Xtend Assembly for routine monitoring and qualification of 10-minute 270°F/132°C prevacuum steam sterilization cycles in the VERIFY All-In-One STEAM Reusable Test Pack. |
#### 6. Summary of Nonclinical Tests
Performance testing to demonstrate substantial equivalence to the predicate has been completed and is summarized in Table 6-2 below.
| Test | Acceptance Criteria | Conclusion |
|----------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Simulated Use | Performance of the BI in the Reusable Test Pack is<br>equivalent to the performance of the predicate | PASS |
| Celerity Xtend 10<br>Assembly in pack vs<br>Celerity Xtend 10<br>Assembly outside of<br>pack | Reusable Test Pack containing the Celerity Xtend 10<br>Assembly provides a greater challenge to the process<br>than the Celerity Xtend 10 Assembly itself. | PASS |
Table 6-2. Summary of Non-clinical Testing
{12}------------------------------------------------
### 7. Conclusion
The VERIFY All-in-One STEAM Reusable Test Pack has met the established performance criteria. Based on the conclusions drawn from the intended use, technological characteristics and nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device (21 CFR 880.2800, Product code FRC).
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.