CELERITY 20 Steam Biological Indicator

K181686 · STERIS Corporation · FRC · Jul 24, 2018 · General Hospital

Device Facts

Record IDK181686
Device NameCELERITY 20 Steam Biological Indicator
ApplicantSTERIS Corporation
Product CodeFRC · General Hospital
Decision DateJul 24, 2018
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 880.2800
Device ClassClass 2

Indications for Use

The Celerity 20 Steam Biological Indicator is used for routine monitoring, qualification testing, load monitoring and product testing of the following steam sterilization cycles: - Dynamic Air Removal 270°F (132°C) 4 minutes - Dynamic Air Removal 275°F (135°C) 3 minutes - Gravity 250°F (121°C) 30 minutes - Gravity 270°F (132°C) 15 minutes. When used in conjunction with the Celerity™ Steam Incubator, the Incubator provides a fluorescent result within 20 minutes.

Device Story

Biological indicator (BI) for steam sterilization monitoring; contains Geobacillus stearothermophilus spores and nutrient media. Device placed in steam sterilizer; after cycle, BI inserted into Celerity Steam Incubator. Incubator detects fluorescent signal produced by reporter enzyme reaction if spores remain viable. Provides result in 20 minutes. Used in clinical settings to confirm sterilization efficacy. Output informs healthcare providers whether sterilization cycle was successful; ensures sterile instruments for patient use.

Clinical Evidence

Bench testing only. Simulated use testing demonstrated no growth of Geobacillus stearothermophilus following exposure to full steam sterilization cycles, meeting established performance criteria.

Technological Characteristics

Self-contained biological indicator (SCBI) with Geobacillus stearothermophilus spores (1.0-4.0 x 10^6/SCBI). Mechanism: fluorogenic enzyme reaction. Dimensions: plastic vial with cap. Connectivity: requires Celerity Steam Incubator for automated reading. Shelf life: 10 months.

Indications for Use

Indicated for routine monitoring, qualification, load, and product testing of specified steam sterilization cycles (Dynamic Air Removal 270°F/275°F; Gravity 250°F/270°F).

Regulatory Classification

Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text. STERIS Corporation Tony Piotrkowski Senior Regulatory Affairs Manager 5960 Heisley Rd Mentor, Ohio 44060 Re: K181686 Trade/Device Name: Celerity 20 Steam Biological Indicator Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: FRC Dated: June 25, 2018 Received: June 26, 2018 Dear Tony Piotrkowski: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good {1}------------------------------------------------ manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Clarence W. Murray lii III -S For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K181686 Device Name Celerity 20 Steam Biological Indicator Indications for Use (Describe) The Celerity 20 Steam Biological Indicator is used for routine monitoring, qualification testing, load monitoring and product testing of the following steam sterilization cycles: - · Dynamic Air Removal 270°F (132°C) 4 minutes - · Dynamic Air Removal 275°F (135°C) 3 minutes - · Gravity 250°F (121°C) 30 minutes - · Gravity 270°F (132°C) 15 minutes. When used in conjunction with the Celerity™ Steam Incubator, the Incubator provides a fluorescent result within 20 minutes. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for STERIS. The word STERIS is in all caps and in a serif font. Below the word STERIS is a graphic of several horizontal wavy lines in blue. # 510(k) Summary For Celerity 20 Steam Biological Indicator ### Sponsor Facility STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198 ### Manufacturing Facility STERIS Corporation 9325 Pinecone Drive Mentor, OH 44060 Phone: (440) 392-7800 Fax No: (440) 392-7896 Contact: Tony Piotrkowski Senior Manager, Regulatory Affairs Telephone: (440) 392-7437 Fax No: (440) 357-9198 e-mail: tpiotrko@steris.com Submission Date: July 18, 2018 Premarket Notification Number: K181686 STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600 {4}------------------------------------------------ #### 1. Device Name | Trade Name: | Celerity 20 Steam Biological Indicator | |------------------------|-----------------------------------------------------------------------| | Common/usual Name: | Biological Indicator (BI, SCBI) | | Device Classification: | Class II | | Classification Name: | Indicator, Biological Sterilization Process<br>(21 CFR 880.2800, FRC) | #### 2. Predicate Device Celerity 20 Steam Biological Indicator, K173634 #### 3. Description of Device The product is intended to monitor the critical parameters of steam sterilization cycles described in the indications for use by producing an optical change (signal) that is detected by the STERIS proprietary reader, Celerity Steam Incubator in 20 minutes to confirm the viability of the biological indicator at the end of a steam sterilization process. The product consists of a biological organism known to be resistant to steam (Geobacillus stearothermophilus) and a defined nutrient media. A reporter enzyme, which is produced by the native organism, reacts with a fluorogenic substrate within the defined nutrient media to produce a fluorescent moiety. #### 4. Intended Use/ Indications for Use The Celerity 20 Steam Biological Indicator is used for routine monitoring, qualification testing, load monitoring and product testing of the following steam sterilization cycles: - · Dynamic Air Removal 270°F (132°C) 4 minutes - · Dynamic Air Removal 275°F (135°C) 3 minutes - Gravity 250°F (121°C) 30 minutes - · Gravity 270°F (132°C) 15 minutes. When used in conjunction with the Celerity™ Steam Incubator, the Incubator provides a fluorescent result within 20 minutes. {5}------------------------------------------------ #### 5. Summary of Technical Characteristics A comparison of technical characteristics are summarized in Table 5-1. | Feature | Celerity 20 Steam SCBI<br>(modified) | Celerity 20 Steam SCBI<br>Predicate (K173634) | Comparison | |------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------| | Intended<br>Use | The Celerity 20 Steam<br>Biological Indicator is used for<br>routine monitoring,<br>qualification testing, load<br>monitoring and product testing<br>of the following steam<br>sterilization cycles:<br>• Dynamic Air Removal<br>270°F (132°C) 4 minutes<br>• Dynamic Air Removal<br>275°F (135°C) 3 minutes<br>• Gravity 250°F (121°C) 30<br>minutes<br>• Gravity 270°F (132°C) 15<br>minutes.<br>When used in conjunction<br>with the Celerity™ Steam<br>Incubator, the Incubator<br>provides a fluorescent result<br>within 20 minutes. | The Celerity 20 Steam<br>Biological Indicator is used for<br>routine monitoring,<br>qualification testing, load<br>monitoring and product testing<br>of the following steam<br>sterilization cycles:<br>• Dynamic Air Removal<br>270°F (132°C) 4 minutes<br>• Dynamic Air Removal<br>275°F (135°C) 3 minutes<br>• Gravity 250°F (121°C) 30<br>minutes<br>• Gravity 270°F (132°C) 15<br>minutes.<br>When used in conjunction<br>with the Celerity™ Steam<br>Incubator, the Incubator<br>provides a fluorescent result<br>within 20 minutes. | Identical | | Indicator<br>organism | > 90% similarity to ATCC<br>7953 Geobacillus<br>stearothermophilus | > 90% similarity to ATCC<br>7953 Geobacillus<br>stearothermophilus | Identical | | Mechanism<br>of action | An enzyme, which is produced<br>by the organism, reacts with a<br>fluorogenic substrate within<br>the defined nutrient media to<br>produce a fluorescent moiety | An enzyme, which is produced<br>by the organism, reacts with a<br>fluorogenic substrate within<br>the defined nutrient media to<br>produce a fluorescent moiety | Identical | | Accessories | Automated incubator / reader | Automated incubator / reader | Identical | | Viable spore<br>population | 1.0 - 4.0 x 106 spore/SCBI | 1.0 - 4.0 x 106 spore/SCBI | Identical | | Resistance | D121 ≥ 1.5 min<br>D132 ≥ 10 s<br>D135 ≥ 8 s | D121 ≥ 1.5 min<br>D132 ≥ 10 s<br>D135 ≥ 8 s | Identical | | Survival<br>Time | Meets the longer of FDA and<br>ISO 11138-3 requirements | Meets the longer of FDA and<br>ISO 11138-3 requirements | Identical | | Kill Time | Meets the shorter of FDA and<br>ISO 11138-3 requirements | Meets the shorter of FDA and<br>ISO 11138-3 requirements | Identical | | Hold time | Up to 72 h | Up to 72 h | Identical | | Carrier<br>growth<br>inhibition /<br>media<br>growth | Positive growth of less than<br>100 spores after primary<br>packaging and media are<br>subject to worst case steam<br>exposure | Positive growth of less than<br>100 spores after primary<br>packaging and media are<br>subject to worst case steam<br>exposure | Identical | | Feature | Celerity 20 Steam SCBI<br>(modified) | Celerity 20 Steam SCBI<br>Predicate (K173634) | Comparison | | Culture<br>Conditions | 55- 59 °C, media included in<br>SCBI, 20-minute incubation<br>time. | 55- 59 °C, media included in<br>SCBI, 20-minute incubation<br>time. | Identical | | Primary<br>Packaging | Direct inoculum on plastic<br>vial, cap with recovery media. | Direct inoculum on plastic<br>vial, cap with recovery media. | Identical | | Shelf Life | 10 months | 10 months | Shelf-life testing is<br>ongoing | | Cap label | Single-ply label on cap edge<br>Process indicator on label | Two-ply label on cap top<br>Process indicator on label | Same adhesive,<br>process indicator ink<br>and substrate different<br>format and location | ## Table 5-1. Summary of SCBI Physical Description and Technological Properties {6}------------------------------------------------ ## Celerity 20 Steam Biological Indicator #### Summary of Nonclinical Tests 6. Performance testing to demonstrate substantial equivalence to the predicate has been completed and is summarized in Table 5-2 below. Table 5-2. Summary of Non-clinical Testing | Test | Acceptance Criteria | Conclusion | |---------------|-------------------------------------------------------|------------------------| | Simulated Use | Demonstrate no growth when exposed to a<br>full cycle | Full cycle - no growth | #### 7. Conclusion The Celerity 20 Steam Biological Indicator has met the established performance criteria. The conclusions drawn from the nonclinical tests performed demonstrate the subject device is as safe, as effective, and performs as well or better than the legally marketed predicate device, K173634 Class II (21 CFR 880.2800, Product code FRC).
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...