K151971 · Advanced Sterilization Products · FRC · Oct 29, 2015 · General Hospital
Device Facts
Record ID
K151971
Device Name
STERRAD CYCLESURE 24 Biological Indicator
Applicant
Advanced Sterilization Products
Product Code
FRC · General Hospital
Decision Date
Oct 29, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
The STERRAD CYCLESURE 24 Biological Indicator is intended to be used as a standard method for frequent monitoring of the following STERRAD Sterilization Systems and has been validated for use in health care facilities: - STERRAD 100S - STERRAD 50 - STERRAD 200 - STERRAD NX® - STERRAD 100NX® (Standard, Flex and Express Cycles) - For STERRAD 100NX DUO Cycle in the United States, the STERRAD CYCLESURE 24 Biological Indicator should only be used in a test pack configuration (REF 20243).
Device Story
STERRAD CYCLESURE 24 Biological Indicator (BI) is a self-contained, stand-alone monitor for STERRAD sterilization processes. Device contains glass fiber disc with minimum 1x10^6 Geobacillus stearothermophilus spores and glass ampoule of nutrient growth medium. Cap features two small circular openings allowing hydrogen peroxide vapor diffusion into vial; openings restrict diffusion to simulate sterilization challenge. Used in healthcare facilities to verify sterilization efficacy. After sterilization cycle, BI is activated to release growth medium; incubation at 55-60°C for 24-72 hours allows spore growth if sterilization failed. Chemical indicator on cap changes color (red to golden yellow/bronze) to confirm exposure to hydrogen peroxide. Provides healthcare providers with objective verification of sterilization cycle success; ensures patient safety by confirming instruments are sterile before reuse.
Clinical Evidence
Bench testing only. Stability studies verified that spore population and resistance characteristics (D-value, survival time, kill time) meet specifications throughout the extended 9-month shelf life.
Technological Characteristics
Self-contained biological indicator. Materials: glass fiber disc carrier, glass ampoule with nutrient growth medium, cap with diffusion-restricting openings. Biological agent: Geobacillus stearothermophilus (ATCC 7953). Resistance: D-value ≥ ~1 second at 2.5 mg/L H2O2. Incubation: 55-60°C for 24-72 hours. Chemical indicator: color-changing (red to golden yellow/bronze) upon H2O2 exposure.
Indications for Use
Indicated for routine monitoring of STERRAD sterilization systems (100S, 50, 200, NX, 100NX) in healthcare facilities. For STERRAD 100NX DUO cycle, must be used in test pack configuration.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 29, 2015
Advanced Sterilization Products Ms. Sun Choi Regulatory Affairs Specialist IV 33 Technology Dr. Irvine, CA 92618
Re: K151971
Trade/Device Name: STERRAD® CYCLESURE® 24 Biological Indicator Regulation Number: 21 CFR 880.2800 Regulation Name: Biological Sterilization Process Indicator Regulatory Class: II Product Code: FRC Dated: October 13, 2015 Received: October 14, 2015
Dear Ms. Choi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Ms. Choi
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
**Tejashri Purohit-Sheth, M.D.**
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
## K151971
Device Name STERRAD® CYCLESURE® 24 Biological Indicator
Indications for Use (Describe)
The STERRAD CYCLESURE 24 Biological Indicator is intended to be used as a standard method for frequent montoring of the following STERRAD Sterlization Systems and has been validated for use in health care facilities:
- . STERRAD 100S
- o STERRAD 50
- . STERRAD 200
- o STERRAD NX®
- . STERRAD 100NX® (Standard, Flex and Express Cycles)
- For STERRAD 100NX DUO Cycle in the United States, the STERRAD CYCLESURE 24 Biological 0 Indicator should only be used in a test pack configuration (REF 20243).
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| <div style="display:inline-block"><input type="checkbox"/></div> Prescription Use (Part 21 CFR 801 Subpart D) | <div style="display:inline-block"><input checked="" type="checkbox"/></div> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
20
{3}------------------------------------------------
# 510(k) Summary
#### I. SUBMITTER
Advanced Sterilization Products 33 Technology Drive Irvine, CA 92618
Contact Person: Sun Choi Regulatory Affairs Specialist IV Telephone: (949) 453-6378 Fax: (949) 798-3900
Date Prepared: July 13, 2015
#### II. DEVICE
Name of Device: STERRAD® CYCLESURE® 24 Biological Indicator Common or Usual Name: Biological Indicator Classification Name: Biological Sterilization Process Indicator (21 CFR 880.2800) Regulatory Class: II Product Code: FRC
#### III. PREDICATE DEVICE
STERRAD CYCLESURE 24 Biological Indicator K123017
No reference devices were used in this submission.
#### IV. DEVICE DESCRIPTION
The STERRAD CYCLESURE 24 Biological Indicator (BI) is a self-contained stand- alone biological monitor intended for the routine monitoring of the STERRAD Sterilization Process. It consists of a glass fiber disc containing a minimum of 1x106 Geobacillus stearothermophilus spores, a glass ampoule containing nutrient growth medium, a cap and liner closing the vial and a chemical indicator on top of the cap. The cap contains two small circular openings that allow for diffusion of hydrogen peroxide vapor into the vial. The relatively small size of the circular openings serves as a restriction to this diffusion.
{4}------------------------------------------------
#### V. INDICATIONS FOR USE
The STERRAD CYCLESURE 24 Biological Indicator is intended to be used as a standard method for frequent monitoring of the following STERRAD Sterilization Systems and has been validated for use in health care facilities:
- STERRAD 100S
- STERRAD 50
- o STERRAD 200
- STERRAD NX® ●
- STERRAD 100NX® (Standard, Flex and Express Cycles) ●
- For STERRAD 100NX DUO Cycle in the United States, the STERRAD O CYCLESURE 24 Biological Indicator should only be used in a test pack configuration (REF 20243).
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The proposed modification is to extend the labeled shelf life of the STERRAD CYCLESURE 24 BI. The intended use/indications for use, technological characteristics, design, materials, and principles of operation have not changed. Refer to Table 1 for comparison between modified and predicate devices for the STERRAD CYCLESURE 24 BI.
| | Predicate Device:<br>STERRAD CYCLESURE 24 BI<br>(K123017) | Modified Device:<br>STERRAD<br>CYCLESURE 24 BI |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|
| Intended Use /<br>Indications for Use | The STERRAD CYCLESURE 24 Biological<br>Indicator is intended to be used as a standard<br>method for frequent monitoring of the following<br>STERRAD Sterilization Systems:<br>• STERRAD 100S<br>• STERRAD 50<br>• STERRAD 200<br>• STERRAD NX<br>• STERRAD 100NX (Standard, Flex and<br>Express Cycles)<br>o For STERRAD 100NX DUO Cycle in<br>the United States, the STERRAD<br>CYCLESURE 24 Biological Indicator<br>should only be used in a test pack<br>configuration (REF 20243). | Same as predicate |
Table 1: Comparison between Modified and Predicate Devices
{5}------------------------------------------------
| | Predicate Device:<br>STERRAD CYCLESURE 24 BI<br>(K123017) | Modified Device:<br>STERRAD<br>CYCLESURE 24 BI |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|
| Technological<br>Characteristics | The STERRAD CYCLESURE 24 BI is a self-<br>contained stand- alone biological monitor<br>intended for the routine monitoring of the<br>STERRAD Sterilization Process. It consists of<br>a glass fiber disc containing a minimum of<br>1x10⁶ Geobacillus stearothermophilus spores, a<br>glass ampoule containing nutrient growth<br>medium, a cap and liner closing the vial and a<br>chemical indicator on top of the cap. The cap<br>contains two small circular openings that allow<br>for diffusion of hydrogen peroxide vapor into<br>the vial. The relatively small size of the circular<br>openings serves as a restriction to this diffusion. | Same as predicate |
| Organism | Geobacillus stearothermophilus ATCC 7953 | Same as predicate |
| Viable Spore Population | Minimum of 1 x 10⁶ spores/BI | Same as predicate |
| Carrier Material | Glass Fiber | Same as predicate |
| Resistance Characteristics | D-value at 2.5 mg/L of hydrogen peroxide:<br>≥ ~ 1 second<br>Survival Time: calculated based on D-value<br>Kill Time: 60 seconds | Same as predicate |
| Incubation Temperature | 55 - 60 °C | Same as predicate |
| Incubation Time | 24 hours to 3 days (72 hours) | Same as predicate |
| Chemical Indicator | A throughput indicator, changes color from red<br>to golden yellow or bronze to indicate exposure<br>to hydrogen peroxide | Same as predicate |
| Shelf-life | 6 months | 9 months |
{6}------------------------------------------------
#### VII. PERFORMANCE DATA
The modified and predicate devices have the same intended use/indications for use, technological characteristics, design, materials, and principles of operation. The proposed modification is to extend the labeled shelf life of the STERRAD CYCLESURE 24 BI from 6 months to 9 months. Stability studies were conducted to demonstrate that the specifications (D-value, survival time, kill time, spore population, and positive BI color test) are maintained throughout the proposed labeled shelf life of the modified device. Refer to Table 2 for the summary of stability studies.
| Study Type | Description | Results |
|------------|---------------------------------------------------------------------------------------------------------------------------------|---------|
| Stability | Verification of spore population for 9 month labeled<br>product shelf life | Passed |
| Stability | Verification of resistance characteristics (D-value,<br>survival time, and kill time) for 9 month labeled<br>product shelf life | Passed |
Table 2: Summary of Performance Testing
# VIII. CONCLUSIONS
Stability study results demonstrate that the specifications (D-value, survival time, kill time, spore population, and positive BI color test) are maintained throughout the proposed labeled shelf life (9 months) of the STERRAD CYCLESURE 24 BI. Modified and predicate devices have the same intended use/indications for use, technological characteristics, design, materials, and principles of operation.
Therefore, Advanced Sterilization Products considers the modified device with the extended shelf life to be substantially equivalent to its predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.