The Infrared Forehead Thermometer (Model: T80、T81、T82、T83、T84、T85) is a non-contact thermometer used to measurement forehead temperature at home or hospital. The device is indicated for use on people of all ages except for neonates/newborns.
Device Story
Non-contact infrared thermometer; measures infrared energy radiated from forehead. Thermopile sensor collects energy; converts to analog signals; MCU with AD converter analyzes signals to calculate temperature. User operates via 4 buttons (Setting, Memory, ON/OFF, Measurement Trigger); LCD displays results. Used in home or hospital by patients or clinicians. Provides rapid temperature readings; aids in fever detection/monitoring. Benefits include non-invasive, hygienic temperature assessment.
Clinical Evidence
Clinical study evaluated 150 subjects across three age groups (infants <1yr, children 1-5yrs, adults >5yrs; n=50 per group). 30% of subjects had temperatures ≥37.5°C. Evaluated per ISO 80601-2-56 for clinical bias and repeatability. Results confirm accuracy within acceptable limits per ISO 80601-2-56 and ASTM E1965-98.
Technological Characteristics
Thermopile infrared sensor; MCU-based signal processing; LCD display; 3.0V DC (2xAA batteries). Materials evaluated for biocompatibility per ISO 10993-5/10. Standards: ASTM E1965-98, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 80601-2-56, IEC 62304.
Indications for Use
Indicated for non-contact forehead temperature measurement in home or hospital settings for all ages, excluding neonates/newborns.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
K210171 — Non-contact Infrared Thermometer (Models WL-301, WL-501, WL-601, WL-701) · Sichuan Youkede Medical Equipment Co.,Ltd · Mar 18, 2022
K181242 — Braun BNT400 No Touch + Forehead Thermometer · Kaz USA, Inc., A Helen of Troy Company · Jul 23, 2018
K202111 — Medical infrared forehead thermometer · Hunan Honggao Electronic Technolofy Co., Ltd. · Feb 26, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 4, 2021
Guandong Genial Technology CO., LIMITED % Jet Li Regulation Manager Guangzhou KINDA Biology Technology Co., Ltd. 6F, No.1 TianTai road, Science City, LuoGang District Guangzhou, Guangdong China
Re: K210014
Trade/Device Name: Infrared Forehead Thermometer (Model: T80, T81, T82, T83, T84, T85) Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: April 19, 2021 Received: May 4, 2021
Dear Jet Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Payal Patel Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210014
Device Name
Infrared Forehead Thermometer (Model: T80、T81、T82、T83、T84、T85)
Indications for Use (Describe)
The Infrared Forehead Thermometer (Model: T80、T81、T82、T83、T84、T85) is a non-contact thermometer used to measurement forehead temperature at home or hospital. The device is indicated for use on people of all ages except for neonates/newborns.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K210014
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
#### Submitter Information 1.
## Sponsor Company Name: GUANGDONG GENIAL TECHNOLOGY CO., LIMITED.
Establishment Registration Number: Applying
- � Address: 1-6-05-02, 11th Road, Area B, Guangfozhao Economic Cooperation Zone, Zhagang Town, Huaiji County, Zhaoging City, Guangdong, 526437, China
- Phone: +86-020-37236838 �
Fax: +86-020-37236838
- Contact Person (including title): Jun Yu (General Manager) �
- E-mail: admin@genial.cn �
## Application Correspondent: Guangzhou KINDA Biology Technology Co., Ltd.
- Address: 6F, No.1 TianTairoad, Science City, LuoGangDistrict, GuangZhouCity, China �
- Contact Person: Mr. JetLi �
- � Tile: Regulation Manager
- � Tel: +86-18588874857
- Email: med-jl@foxmail.com �
### 3 Subject Device Information:
- � Type of 510(k) submission: Traditional
- � Common Name: Infrared Forehead Thermometer
- � Trade Name: Infrared Forehead Thermometer(Model: T80、T81、T82、T83、T84、T85)
- � Classification Name: Clinical Electronic Thermometer
- � Review Panel: General Hospital
- � Product Code: FLL
- � Regulation Number: 21 CFR880.2910
- � Regulation Class: 2
### 4 Predicate Device Information:
- Sponsor: Dongguan SIMZO Electronic TechnologyCo.Ltd. �
{4}------------------------------------------------
- � Trade Name: Non-contact Forehead Thermometer
- � 510(k) number: K173048
- � Review Panel: General Hospital
- � Product Code: FLL
- � Regulation Number: 21 CFR880.2910
- � RequlationClass: 2
### 2. Device Description
The Infrared Forehead Thermometer (Model: T80、T81、T84、T84、T85) is an electronic thermometer using an infrared sensor to measure infrared energy radiated from the forehead. This energy is collected through the lens and converted to a temperature value.
The Infrared Forehead Thermometer (Model: T80、T81、T84、T85),consists of the following parts:
- a) Thermopile Sensor
- b) Application-Specific Integrated Circuitry
- c) Erasable Programmable Read-Only Memory Integrated Circuit
- d) Capacitance-touch Integrated Circuit
- e) LCD and Backlight
- f) 4 buttons (Setting button, Memory button, ON/OFF button, Measurement Trigger)
- g) 2×1.5V AA dry batteries
The operation principle is based on Infrared Sensor technology. The sensor can turn body's temperature to analog signals, and a MCU with an AD can get the result by intelligent analyze and treat the analog signals.
### 3. Intended Use
The Infrared Forehead Thermometer (Model: T80、T81、T82、T83、T84、T85) is a non-contact thermometer used to measurement forehead temperature at home or hospital. The device is indicated for use on people of all ages except for neonates/newborns.
### 4. Test Summary
Infrared Forehead Thermometer conforms to applicable standards that include:
- � ASTM E 1965-98 Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature
- � IEC 60601-1:2012, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance
{5}------------------------------------------------
- � IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility
- � IEC 62304:2006+AMD1:2015 Medical Device Software -Software Life Cycle Processes
- � IEC60601-1-11:2015 medical electrical equipment - part 1-11: general requirements for basic safety
- � and essential performance - collateral standard: requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- � ISO 80601-2-56:2017/AMD 1:2018 Medical electrical equipment - part 2-56: particular requirements f or basic safety and essential performance of clinical thermometers for body temperature measurement
- � ISO 10993-5:2009, Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity,
- � ISO 10993-10:2010, Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization
#### 5. Comparison to Predicate Device
Compare with predicate device, the subject device is very similar in design principle, intended use, sensor, measurement mode, measuring range, accuracy and the applicable standards. The differences between subject device and predicate device do not raise and new questions of safety or effectiveness.
| Elements of<br>Comparison | Subject Device | Predicate Device | Verdict |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Manufacturer | GUANGDONG GENIAL<br>TECHNOLOGY CO., LIMITED. | Dongguan SIMZO Electronic<br>Technology Co.Ltd. | -- |
| 510 (k) Number | K210014 | K173048 | -- |
| Product Name | Infrared Forehead Thermometer | Non-contact Forehead<br>Thermometer | -- |
| Models | Model: T80、T81、T82、T83、<br>T84、T85 | HW-2/HW-2S/HW-3/HW-4 /HW-<br>4S/HW-302/HW-303 | -- |
| Intended Use | The Infrared Forehead<br>Thermometer (Model: T80、T81、<br>T82、T83、T84、T85) is a non-<br>contact thermometer used to<br>measurement forehead temperature<br>at home or hospital. The device is<br>indicated for use on people of all<br>ages except for neonates/newborns | The Non-contact Forehead<br>Thermometer is an infrared<br>thermometer intended for the<br>intermittent measurement of human<br>body temperature in people of all<br>ages. | Note 1 |
| Sensor | Infrared Sensor | Infrared Sensor | SE |
| Measurement<br>Mode | Forehead measurement Mode | Forehead measurement Mode | SE |
| Measurement<br>type | Non-contact type | Non-contact type | SE |
| Measurement<br>method | Infrared radiation detection | Infrared radiation detection | SE |
| Elements of<br>Comparison | Subject Device | Predicate Device | Verdict |
| Measurement<br>time | <3S | 1S | Note 2 |
| Measurement<br>Distance | 1-5cm | HW-2/HW-2S/HW-3/HW-302/ HW-<br>303:5-8cm HW-4/HW-4S:1-2cm | Note 3 |
| Measuring<br>Range | 32°C ~ 42.9°C (89.6°F ~ 109.2°F) | Forehead mode: 35.5°C ~42.9°C<br>(95.9°F ~109.2°F) | Note 4 |
| Accuracy | Forehead mode: ±0.2°C , for<br>range 35.0°C ~ 42.0°C (95.9°F<br>~109.2°F);<br>±0.3°C (0.5°F),for the range<br>32°C - 34.9°C (89.6°F -94.8°F)<br>and 42.1-42.9°C (107.8°F -<br>109.2°F) | Forehead mode:<br>±0.2°C (0.4°F) within 35.5°C<br>~42.9°C<br>(95.9°F ~109.2°F),<br>±0.3°C(0.5°F) other range | Note 5 |
| Display<br>Resolution | 0.1℃/0.1°F | 0.1℃/0.1°F | SE |
| Sensor type | Thermopile | Thermopile | SE |
| High<br>temperature hint | Yes | Yes | SE |
| Buzzer | Yes | Yes | SE |
| Display screen | LCD | LCD | SE |
| Auto power-off<br>while no<br>operation | Yes | Yes | SE |
| °C/°F switchable | Yes | Yes | SE |
| Memory | 32 sets | HW-2/2S:32 sets<br>HW-3/4/4S: 1 set<br>HW-302/303:64 Sets | Note 6 |
| Power Supply | 3.0V DC, offered by two AA<br>batteries | 2 x AAA | Note 7 |
| Operating<br>Conditions | Temperature: 15°C ~ 40°C (60.8°F<br>~ 95°F)<br>Relative humidity: ≤85% RH | 10~40°C(50°F ~104 °F) RH < 85% | Note 8 |
| Storage<br>Conditions | Temperature: -20°C ~55°C (-4°F ~<br>131°F)<br>Relative humidity: ≤93%RH,non-<br>condensing | -25°C - +55°C (-13° F-+131° F)<br>RH≤90% | Note 8 |
| Dimension | 150(L) x 48(W) x 105(H) mm | 93*153*41mm | Similar |
| Weight | Approximately 125 g | 90-125 g | Similar |
| Conformance<br>standard | EN60601-1(Safety), IEC60601-1-<br>2(EMC) ASTME1965-<br>98(performance) | EN60601-1(Safety), IEC60601-1-<br>2(EMC) ASTME1965-<br>98(performance) | SE |
{6}------------------------------------------------
Note 1
{7}------------------------------------------------
Even there is minor difference subject group difference between subject device and predicate device, but the labelling had provided detail statement for applicable subject group. This difference does not affect the safety and effectiveness.
### Note 2
Measurement time of the subject devices is ≤ 3S , the predicate device's will be 1S. But both of them are complied with ASTM E 1965 and the measure time will not affect the measurement accuracy. This difference does not affect thesafety and effectiveness.
### Note 3
Measurement distance of the subject devices is 1-5cm, the predicate device's will be in the range of 5-8cm or 1-2cm. But the performance test result of subject device shows the accuracy meets the requirements within thedistance range.
### Note 4
Although there is little difference for the contact type of the subject device and predicate device, both of them are complied with ASTM E 1965 . This difference does not affect the safety and effectiveness.
### Note 5
Both devices have different measurement range, but they have the same accuracy and the measurement range of subject devices meet the requirements of ASTME 1965-98.
### Note 6
The memory capacity of predicate devices (HW-3/4/4S,HW-302/303)are difference from subject
devices ,but that does not impact the performance of subject devices.
### Note 7
Although the battery size between the predicate device and subject device is different, the power requirements are similar (voltage outputs are the same), and both of them are complied with IEC 60601-1. This difference does not affect thesafety and effectiveness.
## Note 8
There is minor difference on operation environment and storage condition between predicate device and subject device, but both device comply with IEC60601-1-11 and ASTM E1965, so This difference does not affect the safety and effectiveness.
#### Summary of Clinical Test 6.
Clinical tests were conducted on Model: T80、T81、T82、T85、The study excluded subjects with medical conditions such as inflammation at the measuring sites and subjects using medications known to affect body temperature. The clinical tests evaluated 150 subjects which were divided into three group age ranges-Infants (less than 1 year), children (1 to 5 years old)and adult (greater than 5 years old). No less than 50 subjects in each group and 30% of them got temperature equaling or exceeding 37.5℃.
{8}------------------------------------------------
The clinical accuracy of the proposed device was evaluated by ISO 80601-2-56 - clinical bias with stated uncertainty and clinical repeatability. The clinical test results showed that the accuracy of the proposed device is within acceptable scope specified in ISO 81061-2-56.
Based on the result, the clinical performance of the subject device complied with the requirement of ASTM E1965-98 (2016).
### 7. Conclusion
The subject device Infrared Forehead Thermometer has all features of the predicate device. The few differences do not affect the safety and effectiveness of the subject device. Thus, the subject device is substantially equivalent to the predicate device.
# 8. Summary Prepared Date
28 May 2021
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.