Infrared Thermometer

K181215 · Shenzhen Finicare Co., Ltd. · FLL · Jan 2, 2019 · General Hospital

Device Facts

Record IDK181215
Device NameInfrared Thermometer
ApplicantShenzhen Finicare Co., Ltd.
Product CodeFLL · General Hospital
Decision DateJan 2, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2910
Device ClassClass 2
AttributesPediatric

Indications for Use

The Infrared Thermometer is intended for the measurement of human body temperatures. The forehead mode is indicated for people of all ages and the tympanic mode is indicated for people above three months old. The device can be used in the home setting and medical institutes.

Device Story

Infrared thermometer (models FC-IR100, FC-IR101, FC-IR200, FC-IR300, FC-IR400) measures body temperature via infrared energy radiation from forehead or ear; uses Fresnel lens to focus energy onto infrared sensor. Signal receiving processor converts input to temperature reading; displayed on LCD. Operated by patients or clinicians in home or medical settings. FC-IR101 includes Bluetooth for wireless data transmission to app. Provides immediate temperature feedback to assist in clinical decision-making; aids in monitoring patient health status.

Clinical Evidence

Clinical study conducted per ASTM E 1965-98 (2009). Randomized, simple blind homologous control study with minimum 144 subjects across age groups referenced in ISO 80601-2-56. Data supports performance effectiveness for body temperature measurement.

Technological Characteristics

Infrared radiation detection; Fresnel lens focusing. Materials: ABS plastic (surface-contacting, <24h). Power: 2x1.5V AAA batteries. Connectivity: Bluetooth (FC-IR101 only). Standards: AAMI ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 80601-2-56, ASTM E1965-98, ISO 10993-5/10. Non-sterile.

Indications for Use

Indicated for human body temperature measurement. Forehead mode for all ages; tympanic mode for individuals >3 months old. Suitable for home and clinical settings.

Regulatory Classification

Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; (2) Device is not a clinical thermometer with continuous temperature measurement functions; and (3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the agency's name on the right. The symbol on the left is a stylized representation of human services. To the right of the symbol is a blue square containing the acronym "FDA" in white letters. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue. January 2, 2019 Shenzhen Finicare Co., Ltd. Chao Li General Manager Room 201, A Building NO. 1 Qianwan Road Shengang Cooperative Zone, Qianhai Shenzhen, 518052 China Re: K181215 Trade/Device Name: Infrared Thermometer Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: July 17, 2018 Received: August 13, 2018 Dear Chao Li: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Geeta K. Pamidimukkala -S for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K181215 Device Name Infrared thermometer, models FC-IR100, FC-IR101, FC-IR200, FC-IR300, FC-IR400 Indications for Use (Describe) The Infrared Thermometer is intended for the measurement of human body temperatures. The forehead mode is indicated for people of all ages and the tympanic mode is indicated for people above three months old. The device can be used in the home setting and medical institutes. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| Prescription Use (Part 21 CFR 801 Subpart D) |X | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## K181215 # 510(k) Summary ## Prepared in accordance with the requirements of 21 CFR Part 807.92 | 1. Submitter: | Shenzhen Finicare Co., Ltd. | |------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Room 201, A Building, NO.1 Qianwan Road, Shengang Cooperative Zone, Qianhai, Shenzhen, | | | China, 518052, Shenzhen, P.R. China | | | Tel.: +86 -755-23013503 | | | Fax: +86 -755-23574425 | | Contact<br>Person: | Chao Li | | Prepare<br>date: | 2018-03-31 | | 2. Device<br>name and<br>classification: | Device Name: Infrared Thermometer | | | Models: FC-IR100, FC-IR101, FC-IR200, FC-IR300, FC-IR400 | | | Classification Name: | | | 21 CFR 880.2910 | | | Clinical Electronic Thermometers-Temperature Monitor with Probe | | | Product code: FLL | | | Regulatory Class: Class II | | 3. Reason for<br>Submission | New Application. | | 4. Predicate<br>Device(s): | Shenzhen Brav Electronic Technologies Co., Ltd., EFT-165 Infrared Thermometer /K171214 | | 5. Device<br>Description: | FC-IR series infrared thermometers (Models: FC-IR100, FC-IR101, FC-IR200, FC-IR300, FC-IR400) measure the body temperature through receiving infrared energy radiation via the ear or forehead. These thermometers have the capability to measure temperature via forehead temperature mode or ear temperature mode, and the temperature is directly shown on the LCD display. | | | <ul><li>The device is intended to be used in the home setting and medical institutes.</li><li>The product is mainly composed of infrared temperature sensors, signal receiving processor, buttons, buzzer, LCD display. It is powered by 2 X 1.5 AAA batteries.</li><li>It focuses the infrared energy received from the human's forehead or ear by using the Fresnel lens of the thermometer.</li></ul> | In addition, when measuring the ear temperature, the ear cover must be used by all thermometer models except for the FC-IR200. FC-IR101 model incorporates Bluetooth technology to wirelessly transmit temperature reading to an APP. {4}------------------------------------------------ | Model(s)<br>Configuration | FC-IR100 | FC-IR101 | FC-IR200 | FC-IR300 | FC-IR400 | |---------------------------|----------|----------|----------|----------|----------| | Internal Battery | √ | √ | √ | √ | √ | | °C/°F Transaction | √ | √ | √ | √ | √ | | Forehead Measurement | √ | √ | √ | × | × | | Tympanic<br>Measurement | √ | √ | × | √ | √ | | Ear Cover | √ | √ | × | √ | √ | | Bluetooth | × | √ | × | × | × | | High Temperature Hint | √ | √ | √ | √ | √ | | Backlight | √ | √ | √ | √ | √ | ## 6. Indications for Use: The Infrared Thermometer is intended for the measurement of human body temperatures. The forehead mode is indicated for people of all ages and the tympanic mode is indicated for people above three months old. The device can be used in the home setting and medical institutes. ## 7. Predicate Device Comparison In comparison to the predicate device, the subject device has the same intended use, similar product design, same performance effectiveness, and performance safety as the predicate device. Please refer to the following table that notes differences between the subject device and predicate device. All the differences do not affect the basic design principle, usage, effectiveness and safety of the subject device. No new questions are raised regarding effectiveness and safety. | ITEM | Proposed Device<br>FC-IR series | Predicate Device<br>EFT-165/K171214 | Comparison<br>Result | |--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------| | Manufacture | Shenzhen Finicare Co., Ltd. | Brav Electronic Technologies<br>Co., Ltd. | --- | | Indications for<br>Use | The Infrared Thermometer is<br>intended for the measurement of<br>human body temperatures. The<br>forehead mode is indicated for<br>people of all ages and the<br>tympanic mode is indicated for<br>people above three months old.<br>The device can be used in the<br>home setting and medical<br>institutes. | The infrared thermometer is<br>intended for the measurement<br>and monitoring of human body<br>temperature by doctor or<br>customers in the hospital or<br>home. | Different | | Operational Specifications | | | | | Operational<br>Principle | Infrared radiation detection | Infrared radiation detection | Same | | Measuring Mode | Forehead and ear | Forehead and ear | Same | | | | | | | Measurement<br>Range | $32.0°C~42.9°C$<br>$(89.6°F~109.2°F)$ | $32.0°C~42.9°C$<br>$(89.6°F~109.2°F)$ | Same | | Measurement<br>Distance | 0 cm | 0 cm | Same | | Accuracy | Body Mode: ±0.2°C(0.4°F) | Body Mode: ±0.2°C(0.4°F) | Same | | Memory Data<br>Limit | memorize 20 measurements<br>automatically | memorize 20 measurements<br>automatically | Same | | Product<br>configuration | It is mainly composed with<br>infrared sensor, signal receiving<br>processor, buttons, buzzer, LCD<br>display, battery and etc. | It is mainly composed with<br>infrared sensor, signal receiving<br>processor, buttons, buzzer, LCD<br>display, battery and etc. | Same | | Temperature unit<br>and conversion | Dual temperature units “°C” and<br>“°F” optional, and the two units<br>can convert by the conversion<br>key automatically | Dual temperature units “°C” and<br>“°F” optional, and the two units<br>can convert by the conversion<br>key automatically | Same | | Bluetooth<br>Compatible | YES (FC-IR 101 only) | YES | Same | | Applicable<br>Standards | AAMI ANSI ES60601-1, IEC<br>60601-1-2, IEC 60601-1-11, and<br>ASTM E1965-98, ISO 10993-5,<br>ISO 10993-10 | AAMI ANSI ES60601-1, IEC<br>60601-1-2, IEC 60601-1-11, and<br>ASTM E1965-98, ISO 10993-5,<br>ISO 10993-10 | Same | | Display | $0.1°C/°F$ , LCD | $0.1°C/°F$ , LCD | Same | | Operating<br>Environment | 15°C ~ 35°C (59°F to 95 °F)<br><85% moisture condensation | 15°C ~ 35°C (59°F to 95 °F)<br><85% moisture condensation | Same | | Storage<br>Environment | -20°C~55°C(-4°F~131°F)≤90%<br>moisture condensation | -20°C~55°C(-4°F~131°F)≤90%<br>moisture condensation | Same | | Power supply | 2 X 1.5V AAA | 2 X 1.5V AAA | Same | | Physical Specifications | | | | | Weight | 70g (without battery) | 72 g (without battery) | Different | | Dimensions<br>(length x width x<br>height) | FC-IR100:<br>163 mm *39 mm *39mm<br>FC-IR101:<br>163 mm *39 mm *39mm<br>FC-IR200:<br>166 mm *53 mm *41 mm<br>FC-IR300:<br>122 mm *59 mm *41mm<br>FC-IR400:<br>161 mm *68 mm *43mm | 152mm * 44 mm *30mm | | | Biological Specifications | | | | | Patient<br>Contacting<br>Materials | ABS | ABS | Same | | Patient<br>Contacting | Surface-contacting,<br>Less than 24 h | Surface-contacting,<br>Less than 24 h | Same | Table 1 Comparison between the main predicate device EFT-165 and the subject device {5}------------------------------------------------ As seen in the comparison tables, the subject and predicate devices have almost the same design features and performance specifications. The only differences between the subject and predicate devices are the indications for use (which link specific age groups with specific modes of temperature reading) and the appearance and operating/storage environment, which do not raise different questions of safety or effectiveness. Moreover, as demonstrated in the non-clinical testing, the different indications and {6}------------------------------------------------ technological characteristics do not affect the safety and effectiveness of the FC-IR series Infrared thermometer system. #### 8. Performance Testing: Performance data includes "Non-Clinical Data" and "Clinical Data". Summary of the data provided is below: #### Non-Clinical Data: The following performance data were provided in support of the substantial equivalence determination. #### Biocompatibility testing The biocompatibility evaluation for the FC-IR series Infrared thermometer was conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The worst case of the whole system is considered tissue-contacting for a duration of less than 24 hours. The following tests were conducted: |Cytotoxicity ■Skin Sensitization Skin Irritation #### Reprocessing: Cleaning and Disinfection The sterilization and shelf life information provided are acceptable because the subject devices are non-sterile devices. The validation testing results for cleaning and disinfection of the subject devices are adequate and acceptable. #### Electrical safety and electromagnetic compatibility (EMC) Electrical safety and EMC testing were conducted on the FC-IR series Infrared thermometer device. consisting of all the modules and accessories in the system complies with ANSI/AAMI ES60601-1:2005/(R)2012and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 Medical electrical equipment Part 1: General requirements for basic safety and essential performance for safety, IEC 60601-1-11: 2010 MEDICAL ELECTRICAL EQUIPMENT –Part 1-11: General requirements for basic safety and essential performance –Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment, and the IEC 60601-1-2: 2014 Medical electrical equipment –Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances - Requirements and tests standard for EMC. #### Bench Testing Bench testing was conducted on the FC-IR series Infrared thermometer device, consisting of all the accessories in the system. The system complies with ISO 80601-2-56: 2009 Medical electrical equipment -Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement for performance effectiveness. #### Software Verification and Validation Testing Software documentation including verification was provided in accordance with FDA Guidance: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices for software with a moderate level of concern. #### Risk Analysis A risk analysis was conducted in accordance with ISO 14971: 2007 - Medical devices - Application of risk management to medical devices. {7}------------------------------------------------ #### Clinical data: Clinical testing was conducted per ASTM E 1965-98 (2009) Standard Specification for Infrared Thermometers For Intermittent Determination Of Patient Temperature. This clinical study was randomized, with a simple blind homologous control, which consisted of a minimum of 144 subjects with the various age groups as referenced in Section 201.102 of ISO 80601-2-56. ### 9. Substantial Equivalence Conclusion: Differences between the indications for use and technological characteristics of the subject device compared to the predicate do not raise different questions of safety and effectiveness. The performance of the device is supported by non-clinical testing, clinical testing, and risk management activities. The Shenzhen Finicare Co., Ltd. FC-IR Series Infrared Thermometer is Substantially Equivalent (SE) to the Shenzhen Brav Electronic Technologies Co., Ltd., EFT-165 Infrared Thermometer, cleared under K171214.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...