INFRARED THERMOMETER

K113761 · Famidoc Technology Co., Ltd. · FLL · Sep 7, 2012 · General Hospital

Device Facts

Record IDK113761
Device NameINFRARED THERMOMETER
ApplicantFamidoc Technology Co., Ltd.
Product CodeFLL · General Hospital
Decision DateSep 7, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2910
Device ClassClass 2

Indications for Use

Famidoc Infrared Thermometers, Models: FDIR-V2, FDIR-V4, FDIR-V6, FDIR-V6, FDIR-V7, FDIR-V9, FDIR-V9-3, FDIR-V10, FDIR-V15, are intended for intermittent measurement and monitoring of human body temperature by consumers in the home.

Device Story

Hand-held, battery-operated infrared thermometers; measure human body temperature via forehead skin or tympanic (ear) canal. Infrared sensor detects thermal radiation; ASIC converts sensor signals to digital values; LCD displays temperature. Used by consumers in home settings. Provides intermittent temperature monitoring to assist in health assessment.

Clinical Evidence

Bench testing only. Conformance to ASTM E1965-98, IEC 60601-1, and IEC 60601-1-2 standards.

Technological Characteristics

Hand-held, battery-operated infrared sensor technology. ASIC-based signal processing. LCD display. Conforms to ASTM E1965-98, IEC 60601-1, and IEC 60601-1-2.

Indications for Use

Indicated for intermittent measurement and monitoring of human body temperature by consumers in the home setting.

Regulatory Classification

Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; (2) Device is not a clinical thermometer with continuous temperature measurement functions; and (3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K113761. Sponsor: Famidoc Technology Co, Ltd. Infrared Thermometer, models: FDIR-V2, FDIR-V4, FDIR-V5, FDIR-V6, FDIR-V7, Subject Device: FDIR-V9, FDIR-V9-3, FDIR-V10, FDIR-V10, FDIR-V15 File No.: 510(k) submission file (V1.1) - Chapter 7 # 510(k) Summary SEP 7 2012 This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR §807.92. ### 1.0 Submitter's Information #### Establishment Registration: Famidoc Technology Co., Ltd. Address 1: East 2/F Zhixiang Building, 71 Block Baoan District, Shenzhen 518101, Guangdong, P.R. China Phone: +86-755-27864342 Fax: +86-755-27864151 Address 2: No. 212 Yilong Road, Hexi Industrial Zone, Jinxia, Changan Town, Dongguan , Guangdong Province, P.R.China. Contact Person: Mr. Cao Liang E-mail: leon@famidoc.com ## 2.0 Device Information | Type of 510(k) submission: | Traditional | |----------------------------|----------------------------------------------| | Trade Name: | Famidoc Infrared Thermometer | | Model: | FDIR-V2, FDIR-V4, FDIR-V5, FDIR-V6, FDIR-V7, | | | FDIR-V9, FDIR-V9-3, FDIR-V10, FDIR-V15 | | Classification name: | thermometer, electronic, clinical | | Review Panel: | General Hospital | | Product Code: | FLL | | Device Class: | 2 | | Regulation Number: | 880.2910 | #### 3.0 Predicate Device Information | Sponsor: | Famidoc Technology Co., Ltd. | |----------------|-------------------------------------| | Device: | Infrared Thermometer, model FDIR-V1 | | 510(K) Number: | K052849 | Chapter 7, Page 1 of 3 {1}------------------------------------------------ Famidoc Technology Co, Ltd. Sponsor: Infrared Thermometer, models: FDIR-V2, FDIR-V4, FDIR-V5, FDIR-V6, FDIR-V7, Subject Device: FDIR-V9, FDIR-V9-3, FDIR-V10, FDIR-V15 510(k) submission file (V1.1) - Chapter 7 File No.: ### 4.0 Device Description Famidoc Infrared FDIR series Thermometers are hand-held, reusable, battery operated devices, which can measure human body temperature in two ways. - On forehead, the skin temperature on one's forehead. - In ear, the tympanic temperature in one's ear. The operation principle is based on Infrared Sensor technology. The IR sensor can output different signal when measuring different object temperature or in different ambient temperature, and the ASIC can turn the signal from IR Sensor to a digital value and display it on the LCD. The subject devices (Models: FDIR-V2, FDIR-V4, FDIR-V6, FDIR-V6, FDIR-V7, FDIR-V9, FDIR-V9-3, FDIR-V10, FDIR-V15) are the same series in Famidoc. Their design principle, intended use, indications for use, material, control keys, software and the applicable standards are the same. The differences between them are appearance, some parameters (e.g. LCD size), and some minor functions (e.g. Voice function). ### 5.0 Intended Use Famidoc Infrared Thermometers, Models: FDIR-V2, FDIR-V4, FDIR-V6, FDIR-V6, FDIR-V7, FDIR-V9, FDIR-V9-3, FDIR-V10, FDIR-V15, are intended for intermittent measurement and monitoring of human body temperature by consumers in the home. #### 6.0 Performance Summary The devices conform to applicable standards including ASTM E1965-98, IEC 60601-1 and IEC 60601-1-2 requirements. ### 7.0 Comparison to predicate device and conclusion The subject device "Infrared Thermometers, models: FDIR-V4, FDIR-V4, FDIR-V5, FDIR-V6, FDIR-V7, FDIR-V9, FDIR-V9-3, FDIR-V10, FDIR-V15", and the predicate device "Infrared Thermometer, model: FDIR-V1" are the same series infrared thermometer in Famidoc. {2}------------------------------------------------ Sponsor: Famidoc Technology Co, Ltd. Infrared Thermometer, models: FDIR-V2, FDIR-V4, FDIR-V5, FDIR-V6, FDIR-V7, Subject Device: FDIR-V9, FDIR-V9-3, FDIR-V10, FDIR-V15 File No.: ' 510(k) submission file (V1.1) - Chapter 7 The subject devices and the predicate device are the same in design principle, intended use, functions, material and the applicable standards. Only their appearance and some characteristics (e.g. control keys, software, and voice function) are different. But the tests in this submission provide demonstrate these differences do not raise new issue of safety and effectiveness. ### 8.0 Conclusions The Infrared Thermometer, Models: FDIR-V2, FDIR-V4, FDIR-V5, FDIR-V6, FDIR-V7, FDIR-V9, FDIR-V9-3, FDIR-V10, and FDIR-V15 are substantially equivalent to the predicate device. #### 9.0 Summary prepared date September 6, 2012 {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three overlapping wing segments, symbolizing health and human services. The eagle is positioned to the right of the department's name, which is arranged in a circular fashion around the logo. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 SEP Mr. Cao Liang General Manager Famidoc Technology Company, LIMITED No. 212 Yilong Road , Hexi Industrial Zone Jinxia, Changan Town, Dongguan Guangdong Province, P R China Re: K113761 Trade/Device Name: Famidoc Infrared Thermometer, Models: FDIR-V2, FDIR-V4, FDIR-V5, FDIR-V6, FDIR-V7, FDIR-V9, FDIR-V9-3, FDIR-V10, FDIR-V15 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: July 23, 2012 Received: July 23, 2012 Dear Mr. Liang We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. 7 2012 {4}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. For Robert Smith, M.D. Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {5}------------------------------------------------ Famidoc Technology Co, Ltd. Sponsor: Infrared Thermometer, models: FDIR-V2, FDIR-V4, FDIR-V5, FDIR-V6, FDIR-V7, Subject Device: FDIR-V9. FDIR-V9-3. FDIR-V10, FDIR-V10, FDIR-V15 510(k) submission file (V1.0) - Chapter 6 File No.: # Indications for Use 510(k) Number (if known): Device Name: Famidoc Infrared Thermometer Models: FDIR-V2, FDIR-V4, FDIR-V5, FDIR-V6, FDIR-V7, FDIR-V9, FDIR-V9-3, FDIR-V10, FDIR-V15 Indications for Use: Famidoc Infrared Thermometers, models: FDIR-V2, FDIR-V4, FDIR-V5, FDIR-V6, FDIR-V7, FDIR-V9, FDIR-V9-3, FDIR-V10, FDIR-V15, are intended for intermittent measurement and monitoring of human body temperature by consumers in the home. Prescription Use Over-The-Counter Use X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) RH, Office of Device Evaluation (ODE) Page 1 of (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: Chapter 6, Page 1 of 1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...