IR EAR/FOREHEAD THERMOMETER MODE DX6635, DX6637, DTH1081
Applicant
Shenzhen Dongdixin Technology Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
Jul 22, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
3rd-Party Reviewed
Intended Use
The device is intended for the intermittent measurement and monitoring of human body temperature by consumers in the home.
Device Story
Hand-held, battery-operated infrared thermometer (models DX6635, DX6637, TH1081, TH1091); measures body temperature via forehead skin or ear tympanic membrane. Infrared sensor receives thermal energy; converts to voltage; MCU uses A/D circuit to convert voltage to digital data; internal algorithms calculate final temperature; displayed on LCD. Used by consumers in home environment for intermittent monitoring. Provides immediate temperature readout to assist in health status assessment.
Clinical Evidence
Bench testing only. Conformance to ASTM E1965-98 (2003), IEC 60601-1, and IEC 60601-1-2 standards demonstrated.
Technological Characteristics
Infrared sensor technology; 3V DC power; LCD display; handheld form factor. Measurement range 32.0-42.9°C (89.6-109.2°F). Accuracy ±0.2°C (35.5-42.0°C) and ±0.3°C otherwise. Standards: ASTM E1965-98 (2003), IEC 60601-1, IEC 60601-1-2.
Indications for Use
Indicated for intermittent measurement and monitoring of human body temperature in home settings by consumers.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{0}------------------------------------------------
K100488q.lot3
# Caretalk
File No: WMI-04-JR-T-FDA-05 Version: 1.I
## JUL 2.2 2010
# 510(k) SUMMARY
.
IR Ear/Forehead thermometer, K (
Date of Submission: 01/20/2009 Submitter's Name: Shenzhen Dongdixin Technology Co., Ltd. Address: No.3bldვ. xiliyangguang industrial estae xilixiaobaimang Nanshan district Shenzhen, China 518108 Telephone: +86(755) 27652471 Fax: +86(755) 27652674 info@dundex.com E-mail: microcong@gmail.com Contact: Zhigang Zhao
{1}------------------------------------------------
K100486p. 2 of 3
lined
피
ビ
ਦਿੱਤਾ ਸ
ビ
는 그
F
는 글
res
### હિંદ Caretalk
1. Proposed Device:
File No: WMI-04-IR-T-FDA-05 Version: 1.1
Trade Name: IR Ear/Forehead thermometer Classification Name: Clinical electronic thermometer Regulation Number: 21 CFR 880.2910 Product Code: FLL Device Class: II
- 2. Predicate Device:
| Legally Marketed Device: | Infrared Thermometer Model:FDIR-VI |
|--------------------------|------------------------------------|
| 510(k) Number: | K052849 |
| Manufacturer: | FAMIDOC TECHNOLOGY CO., LTD |
#### 3. Description of Proposed Device:
IR Ear/Forehead thermometer, which includes models DX6635, DX6637, TH1081 and TH1091, are a hand-held, reusable, battery operated device that can measures human body temperature by two ways:
- On forehead, the skin temperature on people's forehead �
- In ear, the tympanic temperature in people's ear �
The working principle based on infrared sensor technology. It utilizes infrared technology to receive infrared energy emitted from the eardrum or the surface of the object and convert to a voltage. The voltage value is responsible with the different temperature between the target object and infrared sensor. Then MCU will convert this value to digital data by an analog to digital (A/D) circuit. This data will be changed to the finalized temperature data by algorithms and displayed on LCD.
#### 4. Proposed Device Intended for Use Statement:
### Device Name:
IR Ear/Forehead thermometer, Model: DX6635, DX6637, TH1081, TH1091
### Indications for Use:
The device is intended for the intermittent measurement and monitoring of human body temperature by consumers in the home.
#### Technological Characteristics and Substantial Equivalence ડ.
Section 05-Page 2 of 3
{2}------------------------------------------------
K100488 p. 3 of 3
## Caretalk®
File No: WMI-04-IR-T-FDA-05 Version: 1.1
Both the IR Ear/Forehead thermometer and the Predicate device thermometer have the same intended use and fundamental technology. A side-by-side comparison of the IR Ear/Forehead thermometer and the cited predicate devices is included in the 510(k) submission. The IR Ear/Forehead thermometer is substantially equivalent to the technological features as the predicate devices.
| | 1 20 - 2 8 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 | New device 1 8 8 1 1 1 1 1 1 1 1 1 1<br>r : | Predicate device ========== |
|----|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|----------------------------------|
| | 510K# | K | K052849 |
| 2 | Device Name | IR Ear/Forehead thermometer Mode: | Infrared<br>Thermometer<br>Mode: |
| | | DX6635, DX6636, TH1081, TH1091 | FDIR-V1 |
| 3 | Manufacturer | Shenzhen Dongdixin | FAMIDOC TECHNOLOGY Co., Ltd |
| | | Technology Co., Ltd. | |
| 4 | Measurement temp range | 32.0~42.9℃ | 32.0~42.9℃ |
| | | (89.6~109.2°F) | (89.6~109.2°F) |
| S | Accuracy | 35.5℃ (95.9ºF ) ~42.0℃ (107.6ºF ) | 35.5℃(95.9°F)~42℃(107.6°F) |
| | | ±0.2℃ (0.4℃ ) , Other ±0.3℃ (0.5℃ ) | ±0.2℃(0.4℃),Other ±0.3℃(0.5℃ |
| હ | Ambient range | 10℃~~ 40℃ (50.0°F~104.0°F) for Ear | 10℃~ 40℃ (50.0°F~104.0°F) |
| | | 15℃~ 40℃ (59.0℃~104.0℃)<br>for | |
| | | Forehead | |
| 7 | Storage range | -20℃~ 55℃ (-47~~ 131°F) | -25℃~ 55℃ (-137~~131°Г) |
| 8 | Display type | LCD | LCD |
| ਰੇ | Activation | Scan button | Scan button |
| 10 | Power requirements | 3V DC | 3V DC |
| 11 | Classification | Clinical Electronic thermometer | Clinical Electronic thermometer |
| | | Class II 21 CFR 880.2910 | Class II 21 CFR 880.2910 |
Basic technological characteristics, new device vs. Predicate device
#### 6. Performance Summary:
The device conforms to applicable standards included ASTM E1965-98 (2003), IEC 60601-1, and IEC 60601-1-2 requirements.
#### 7. Conclusions:
IR Ear/Forehead thermometer, which includes models DX6635,DX6637,TH1081 and TH1091, has the same intended use and similar technological characteristics as the predicate device. Moreover, bench testing, safety report and Risk Analysis Report documentation supplied in this submission demonstrates that the difference in their technological characteristics do not raise any new question of safety or effectiveness. Thus, the IR Ear/Forehead thermometer is substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in bold, black font. The text is centered and appears to be a title or heading. The words are all capitalized and evenly spaced.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Shenzhen Dongdixin Technology Company, Limited C/O Mr. Ned Devine Responsible Third Party Official Underwriters Laboratories, Incorporated 333 Pfingsten Road Northbrook, Illinois 60062
## JUL 2-2 2010
Re: K100488
Trade/Device Name: IR Ear/Forehead Thermometer, Model: DX6635, DX6637, TH1081, TH1091 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: June 30, 2010 Received: July 15, 2010
Dear Mr. Devine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/3/Picture/10 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized human figure with three arms reaching upwards, representing health, services, and people. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure.
{4}------------------------------------------------
Page 2- Mr. Devine
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOTfices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Antony V. Watt
Anthony D. Watson, B.S., M.S., M.B.A. Diréctor Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{5}------------------------------------------------
## છે Caretalk
File No: WMI-04-LT3000-FDA-04 Version: 1.1
## Indications for Use
510(k) Number (
):
Device Name:
IR Ear/Forehead thermometer, Model: DX6635, DX6637, TH1081, TH1091
## Indications for Use:
The device is intended for the intermittent measurement and monitoring of human body temperature by consumers in the home.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ ✓ _ (21 CFR 801 Subpart C)
> :::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::: .. 1 - 5
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Light Music for RZC July 22, 2010
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K100488
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.