MICROLIFE NON-CONTACT INFRARED FOREHEAD THERMOMETER, MODEL FR1DZ1

K100953 · Microlife Intellectual Property GmbH · FLL · May 17, 2010 · General Hospital

Device Facts

Record IDK100953
Device NameMICROLIFE NON-CONTACT INFRARED FOREHEAD THERMOMETER, MODEL FR1DZ1
ApplicantMicrolife Intellectual Property GmbH
Product CodeFLL · General Hospital
Decision DateMay 17, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2910
Device ClassClass 2
AttributesPediatric

Indications for Use

The Microlife Non-Contact Infrared Forehead Thermometer, Model FR1DZ1 is intended for the intermittent measurement and monitoring of human body temperature. The device is indicated for use by people of all ages in the home.

Device Story

Non-contact infrared forehead thermometer; uses thermopile sensor to detect infrared radiation from forehead; ASIC and E2PROM IC process signals; provides temperature readings comparable to rectal, oral, or axillary sites; intended for home use by patients of all ages; output displayed on LCD with backlight; buzzer provides alerts; battery-powered (2x AAA); assists users in monitoring body temperature for clinical decision-making.

Clinical Evidence

Controlled human clinical studies conducted per ASTM E1965. Data evaluated clinical bias, clinical uncertainty, and clinical repeatability to demonstrate performance accuracy.

Technological Characteristics

Infrared thermopile sensor; ASIC and E2PROM IC processing; LCD with backlight; 3-key interface; buzzer; 2x AAA battery power. Compliance with ASTM E1965, IEC60601-1, and IEC60601-1-2.

Indications for Use

Indicated for intermittent measurement and monitoring of human body temperature for all ages in home settings.

Regulatory Classification

Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; (2) Device is not a clinical thermometer with continuous temperature measurement functions; and (3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ MAY 1 7 2010 # EXHIBIT #1a K100953 # 510(K) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92. The assigned 5l0(k) number is: ### Submitter's Identification: 1. Microlife Intellectual Property GmbH, Switzerland Espenstrasse 139 9443 Widnau / Switzerland Date Summary Prepared: May 3, 2010 Mr. Gerhard Frick Contact: Vice President of Technical and Service Microlife Intellectual Property GmbH, Switzerland Tel: +41 79 216 0070 E-Mail: gerhard frick@microlife.ch ### 2. Name of the Device: Microlife Non-Contact Infrared Forehead Thermometer, Model FR1DZ1 Requlation Number: 21 CFR 880.2910 Regulation Name: Thermometer, Electronic, Clinical. Requlatory Class: II Product Code: FLL ### 3. Predicate Device Information: Microlife Digital Infrared Forehead Thermometer, Model FR1DM1, K#033820 Thermofocus Non-contact Infrared Forehead Thermometer, Model 01500, K#072108. ### 4. Device Description: The Microlife Non-contact Infrared Forehead Thermometer, Model FR1DZ1 is an electronic thermometer using an infrared sensor (thermopile) {1}------------------------------------------------ to detect body temperature from the forehead. Microlife FR1DZ1 specially enables you to take measurements and judge the readings according to vour local habits. It can measure for three types of readings comparable to readings measured at such three conventional measuring sites as rectal, oral, and axillary with an ordinary pen-type thermometer. This Infrared Forehead Thermometer enables very safe and reliable measurements and with its technology the thermometer offers a very high clinical accuracy and has been designed to provide a maximum of userfriendliness. The Microlife Non-Contact Infrared Forehead Thermometer consists mainly of seven parts: - a) Thermopile Sensor - b) ASIC - c) E2PROM IC - d) Lens - e) LCD and Backlight - f) 3 Keys (Offset 2 keys: optional), 1 Buzzer - g) 2 batteries AAA (LR03) ### ട. Intended Use: The Microlife Non-Contact Infrared Forehead Thermometer, Model FR1DZ1 is intended for the intermittent measurement and monitoring of human body temperature. The device is indicated for use by people of all ages in the home. ### 6. Comparison to Predicate Devices: The Microlife Non-contact Infrared Forehead Thermometer, Model FR1DZ1 is substantially equivalent to Microlife Digital Infrared Forehead Thermometer, Model FR1DM1, K#033820, and Thermofocus Non-contact Infrared Forehead Thermometer, Model 01500, K#072108, which has the same intended use and is similar in design to the predicate devices. The Microlife Non-contact Infrared Forehead Thermometer FR1DZ1 and the predicate devices are identical in the temperature measurements algorithm and fundamental scientific technology, Microlife FR1DZ1 and FR1DM1 mainly differ by "non-contact", reference site, power supply, memory sets, backlight, lens etc., Microlife FR1DZ1 and Thermofocus 01500 mainly differ by the measuring range, power supply, memory sets, backlight, etc. ### Discussion of Non-Clinical Tests Performed for Determination of 7. Substantial Equivalence are as follows: {2}------------------------------------------------ Compliance to applicable voluntary standards includes ASTM E1965, as well as IEC60601-1and IEC60601-1-2 requirements. Guidance documents included the "FDA Guidance on the Content of Premarket Notification 510(K) Submissions for Clinical Electronic Thermometers. ### Discussion of Clinical Tests Performed: 8. Controlled human clinical studies were conducted in accordance with ASTM E1965 using the Microlife Non-Contact Infrared Forehead Thermometer Model FR1DZ1. Clinical data was presented evaluating clinical bias, clinical uncertainty and clinical repeatability per clinical validation for Microlife FR1DZ1. ### 9. Conclusions: The Microlife Non-Contact Infrared Forehead Thermometer, Model FR1DZ1 has the same intended use and similar technological characteristics as the Microlife Digital Infrared Forehead Thermometer Model FR1DM1 and Thermofocus Non-Contact Infrared Forehead Thermometer Model 01500. Moreover, bench testing contained in this submission supplied demonstrates that any differences in their characteristics do not raise any new questions of safety or effectiveness. Thus, the Microlife Non-Contact Infrared Forehead Thermometer FR1DZ1 is substantially equivalent to the predicate devices FR1DM1 and Thermofocus 01500. ## EXHIBIT #1a-3 {3}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized image of an eagle with three curved lines representing its wings and body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MD 20993-0002 Microlife Intellectual Property GmbH C/O Ms. Susan D. Goldstein-Falk MDI Consultants, Incorporated 55 Northern Boulevard, Suite 200 Great Neck, New York 11021 MAY 1 7 2010 Re: K100953 Trade/Device Name: Microlife Non-Contact Infrared Forehead Thermometer, Model FRIDZ1 Regulation Number: 21 CFR 880.2910 Regulation Name: Thermometer, Electronic, Clinical Regulatory Class: II Product Code: FLL Dated: April 1, 2010 Received: April 6, 2010 Dear Ms. Goldstein-Falk: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. Nh for Anthony Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health ## Enclosure {5}------------------------------------------------ ## EXHIBIT B # Indications for Use 510 (k) Number (if known): Device Name: Microlife Non-Contact Infrared Forehead Thermometer, Model FR1DZ1 Indications For Use: The Microlife Non-Contact Infrared Forehead Thermometer, Model FR1DZ1 is intended for the intermittent measurement and monitoring of human body temperature. The device is indicated for use by people of all ages in the home. | Prescription Use<br>(Part21 CFR 801 Subpart D) | | |------------------------------------------------|-----------------------------| | AND/OR Over-The-Counter Use | X<br>(21 CFR 801 Subpart C) | (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Charles (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K100953
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...