K223037 · Shen Zhen Rong Feng Technology Co., Ltd. · FLL · Jul 21, 2023 · General Hospital
Device Facts
Record ID
K223037
Device Name
Infrared Thermometer, Model: IR8807
Applicant
Shen Zhen Rong Feng Technology Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
Jul 21, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Infrared Thermometer (Model: IR8807) is intended for the intermittent measurement of human body temperature from the forehead. It is a non-contact clinical thermometer and intended for use on people of all ages except for neonates/newborns.
Device Story
Non-contact infrared thermometer; measures infrared energy radiated from forehead via thermopile sensor; converts energy to temperature value using internal microcontroller; features forehead and object measurement modes. Operated by users (OTC) in home or clinical settings. Device includes LCD display, two buttons (scan/setting), and 32-group memory storage. Provides temperature readings in Celsius or Fahrenheit; includes low battery indicator, buzzer, and adjustable alarm. Output assists healthcare providers or users in monitoring body temperature for clinical decision-making.
Clinical Evidence
Clinical study conducted on 169 subjects across three age groups: infants (<1 year), children (1-5 years), and adults (>5 years). At least 50 subjects per group; 30% had temperatures ≥37.5°C. Study excluded subjects with inflammation at measurement sites or using temperature-altering medications. Results confirmed compliance with ASTM E1965-98 (2016) performance requirements.
Technological Characteristics
Infrared thermopile sensor; ABS plastic shell; 2x1.5V AAA battery power. Dimensions: 182x54x47mm; Weight: ~125g. Conforms to ASTM E1965-98, IEC 60601-1, IEC 60601-1-2, and IEC 60601-1-11. Biocompatibility per ISO 10993-5 and ISO 10993-10. Software Class B per IEC 62304.
Indications for Use
Indicated for intermittent forehead temperature measurement in patients of all ages, excluding neonates/newborns.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in a larger, bolder font, and "ADMINISTRATION" in a smaller font below.
July 21, 2023
Shen Zhen Rong Feng Technology Co., Ltd Iris Fung Regulation Manager 3/F R Building ShaSi Industrial Park, Shajing Town BaoAn district Shenzhen, Guangdong China
Re: K223037
Trade/Device Name: Infrared Thermometer, Model: IR8807 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: June 18, 2023 Received: June 20, 2023
Dear Iris Fung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Porsche Bennett
Porsche Bennett For David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223037
Device Name Infrared Thermometer, Model: IR8807
Indications for Use (Describe)
The Infrared Thermometer (Model: IR8807) is intended for the intermittent measurement of human body temperature from the forehead. It is a non-contact clinical thermometer and intended for use on people of all ages except for neonates/newborns.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary-K223037
This summary of 510(k) substantial equivalence information is being submitted in accordance with the requirements of 21 CFR 807.92.
#### 1. Submitter Information
# Sponsor Company Name: SHEN ZHEN RONG FENG TECHNOLOGY CO., LTD
- Address: 3/F R building ShaSi Industrial Park, ShajingTown, BaoAndistrict, ShenZhen � city, GuangDong Province, China
- Phone: +86-13715151187 � Fax: +86-020-82591255
- Contact Person (including title): Hong Xiao Lan (Engineer) �
- E-mail: 844352259@qq.com �
# Application Correspondent: SHEN ZHEN RONG FENG TECHNOLOGY CO., LTD
Address: 3/F R building ShaSi Industrial Park, Shajing Town, BaoAn district, ShenZhen city,
GuangDong Province, China
- � Contact Person: Iris Fung
- Tile: Requlation Manager �
- Tel: +86-13211147965 �
- � Email: mdc-fs@foxmail.com; med-jl@foxmail.com
- Summary Prepared Date � July 21, 2023
- Subject Device Information: 2.
- � Type of 510(k) submission: Traditional
- � Common Name: Infrared Thermometer
- Trade Name: Infrared Thermometer, Model: IR8807 �
- Classification Name: Clinical Electronic Thermometer �
- Review Panel: General Hospital �
- � Product Code: FLL
- Regulation Number: 21 CFR 880.2910 �
- Regulation Class: 2 �
#### Predicate Device Information: 3.
- 510(k) number: K173048 �
- � Sponsor: Dongguan SIMZO Electronic Technology Co., Ltd.
{4}------------------------------------------------
- Trade Name: Non-contact Forehead Thermometer, Model: HW-2/HW-3/HW-3/HW-4 /HW-� 4S/HW-302/HW-303
- Common Name: Infrared Thermometer �
- Review Panel: General Hospital �
- � Product Code: FLL
- Regulation Number: 21 CFR 880.2910 �
- Requlation Class: 2 �
#### Device Description 4.
The Infrared Thermometer (Model: IR8807) is an electronic thermometer using an infrared sensor to measure infrared energy radiated from the forehead. This energy is collected through the lens and converted to a temperature value. The Infrared Thermometer (Model: IR8807) has a forehead measurement mode and an object measurement mode.
The Infrared Thermometer (Model: IR8807), consists of the following parts:
- a) Thermopile Sensor
- b) Application-Specific Integrated Circuitry
- c) Erasable Programmable Read-Only Memory Integrated Circuit
- d) Capacitance-touch Integrated Circuit
- e) LCD display
- f) 2 buttons (Scan button, Setting button,)
- q) 2×1.5V AAA dry batteries
The operation principle is based on infrared sensor technology. The sensor in the thermometer detects the infrared energy, and the device processes the data to body temperature result with its internal microcontroller unit. It operates in an adjusted mode.
The main functions of Infrared Thermometer (Model: IR8807) are as followings:
- Forehead temperature measure functions �
- Wide range of temperature readings: from 32.0ºCto 42.9ºC (89.6ºF ~ 109.2ºF) �
- The measurement results can be displayed in either °Cor °F units �
- The LCD display showing clear measurement result. �
- Memory data reading/removal function: The device can store a total of 32 groups of � memorable values. All memory values can be removed after the battery is removed
- Low battery indicator �
- Buzzer reminder function �
- Users can set temperature alarm value �
- Environment or object temperature measurement function � The reference body site of the output temperature is Oral.
{5}------------------------------------------------
# 5. Indications for use
The Infrared Thermometer (Model: IR8807) is intended for the intermittent measurement of human body temperature from the forehead. It is a non-contact clinical thermometer and intended for use on people of all ages except for neonates/newborns.
| Elements of<br>Comparison | Subject Device | Predicate Device | Comparison |
|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Manufacturer | SHEN ZHEN RONG FENG<br>TECHNOLOGY CO., LTD. | Dongguan SIMZO Electronic<br>Technology Co.Ltd. | – |
| 510 (k) Number | K223037 | K173048 | – |
| Product Name | Infrared Thermometer | Non-contact Forehead<br>Thermometer | – |
| Models | Model: IR8807 | HW-2/HW-2S/HW-3/HW-4<br>/HW-4S/HW-302/HW-<br>303 | Note 1 |
| Intended Use | The Infrared Thermometer (Model:<br>IR8807) is intended for the<br>intermittent measurement of human<br>body temperature from the forehead.<br>It is a non-contact clinical<br>thermometer and intended for use<br>on people of all ages except for<br>neonates/newborns. | The Non-contact Forehead<br>Thermometer is an infrared<br>thermometer intended for the<br>intermittent measurement of<br>human body temperature in<br>people of all ages. | Note 1 |
| Sensor | Infrared Sensor | Infrared Sensor | Same |
| Measurement<br>Mode | Forehead measurement Mode | Forehead measurement Mode | Same |
| Measurement<br>contact type | Non-contact type | Non-contact type | Same |
| Measurement<br>type | Adjusted | Adjusted | Same |
| Measurement<br>method | Infrared radiation detection | Infrared radiation detection | Same |
| Measurement<br>time | ≤3S | 1S | Note 2 |
| Measurement<br>Distance | 2-5cm | HW-2/HW-2S/HW-3/HW-302/<br>HW-303:5-8cm HW-4/HW-<br>4S:1-2cm | Note 3 |
| Measuring<br>Range | 32°C ~ 42.9°C | Forehead mode: 35.5°C<br>~42.9°C (95.9°F<br>~109.2°F) | Note 4 |
| Accuracy | Forehead mode:±0.2°C , for<br>range 35.0°C ~ 42.0°C<br>±0.3°C, outside this temperature<br>range | Forehead mode:<br>±0.2°C (0.4°F) within 35.5°C<br>~42.9°C<br>(95.9°F~109.2°F),<br>±0.3°C(0.5°F) other range | Note 5 |
| Display<br>Resolution | 0.1°C/0.1°F | 0.1°C/0.1°F | Same |
| Sensor type | Thermopile | Thermopile | Same |
| Elements of<br>Comparison | Subject Device | Predicate Device | Comparison |
| High<br>temperature<br>indication | Yes | Yes | Same |
| Buzzer | Yes | Yes | Same |
| Display screen | LCD | LCD | Same |
| Auto power-off<br>while no<br>operation | Yes | Yes | Same |
| °C/°Fswitchable | Yes | Yes | Same |
| Memory | 32 sets | HW-2/2S:32<br>sets HW-<br>3/4/4S: 1 set<br>HW-302/303:64 Sets | Note 6 |
| Power Supply | 2 x AAA batteries | 2 x AAA | Same |
| Operating<br>Conditions | Temperature: 10°C ~ 40°C Relative<br>humidity: ≤85% RH | Temperature: 10~40°C<br>Relative humidity: ≤85%<br>RH | Same |
| Storage<br>Conditions | Temperature: -20°C ~55°C Relative<br>humidity: ≤93%RH, non-<br>condensing | Temperature: -25°C- +55°C<br>Relative humidity: ≤90% | Note 7 |
| Dimension | 182(L) x 54(W) x 47(H) mm | 93*153*41mm | Note 8 |
| Weight | Approximately 125 g | 90-125 g | Note 8 |
| Conformance<br>standard | EN60601-1(Safety),<br>IEC60601-1-2(EMC)<br>ASTME1965-98(performance) | EN60601-<br>1(Safety),<br>IEC60601-1-<br>2(EMC)<br>ASTME1965-98(performance) | Same |
| Biocompatibility | ISO 10993-10 Skin Irritation test ISO<br>10993-10Skin Sensitization test<br>ISO 10993-5 Vitro Cytotoxicity test | ISO 10993-10 Skin Irritation<br>test ISO 10993-10Skin<br>Sensitization test<br>ISO 10993-5 Vitro Cytotoxicity<br>test | Same |
| Sensor and<br>patient contact<br>materials | Infrared Sensor ABS<br>plastic shell | Infrared Sensor<br>ABS plastic shell | Same |
#### Comparison to Predicate Device 6.
{6}------------------------------------------------
# Note 1
The Intended Use of subject device and predicate device is non-contact measurement of human body temperature from the forehead. The target population of the subject device is a subset of the predicate. This difference does not raise new or different questions of safety and effectiveness.
# Note 2
Measurement time of the subject device is ≤3 seconds, the predicate device is 1 second. Both the subject and predict devices comply with ASTM E 1965-98. It is demonstrated that Page 4 of 7
{7}------------------------------------------------
the measure time does not affect the measurement accuracy. This difference does not raise new or different questions of safety and effectiveness.
### Note 3
Measurement distance of the subject devices is 2-5cm, the predicate device measurement distance is 5- 8cm (models HW-2/HW-2S/HW-3/HW-302/ HW-303) or 1-2cm (models HW-4/HW-4S). Performance testing of the of the subject device demonstrates the accuracy meets the requirements within the distance range. The difference does not raise new or different questions of safety and effectiveness.
### Note 4
Both devices have different measurement ranges; however, the measurement range of the subject device meets the requirements of ASTME1965-98. Therefore, the difference does not raise new or different questions of safety and effectiveness.
### Note 5
Both devices have different accuracy requirements ; however, the measurement accuracy of the subject device and the predicate device meet the requirements of ASTME1965-98. Performance testing was conducted on the subject device according to ASTM E1965-98. Therefore, the difference does not raise new or different questions of safety and effectiveness.
## Note 6
The memory capacity of the subject device is the same as the predicate device model: HW-2/2S. The memory capacity is only an auxiliary function. Software verification and validation testing conducted on the subject device demonstrates that the difference does not raise any new or different safety and effectiveness questions.
### Note 7
There is a difference between the storage conditions of the predicate device and the subject device; however, both devices comply with IEC60601-1-11 and ASTM E1965-98. Therefore, the difference does not raise new or different questions of safety and effectiveness.
### Note 8
There is a difference between the dimensions and weight of the predicate device and the subject device, however, the differences do not affect the safety and effectiveness as the subject device demonstrated adequate performance to requirements.
{8}------------------------------------------------
# 7. Test Summary - Non-Clinical Test
The subject device conforms to applicable standards that include:
- � ASTM E 1965-98:2016 Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature
- � IEC 60601-1:2012, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance
- � IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility
- � IEC 62304:2006+AMD1:2015 Medical Device Software - Software Life Cycle Processes
- � IEC60601-1-11:2015 medical electrical equipment - part 1-11: general requirements for basic safety and essential performance - collateral standard: requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- � ISO 10993-5:2009, Biological evaluation of medical devices --Part 5: Testsfor In Vitro cytotoxicity,
- � ISO 10993-10:2010, Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization
#### Summary of Clinical Test 8.
Clinical tests were conducted on Model: IR8807. The study excluded subjects with medical conditions such as inflammation at the measuring sites and subjects using medications known to affect body temperature. The clinical tests evaluated 169 subjects which were divided into three age groups: infant (less than 1 year), children (1 t o 5 years old) and adult (greater than 5 years old). No less than 50 subjects in each group and 30% of them had temperature equaling or exceeding 37.5 °C.
Based on the result, the clinical performance of the subject device complied with the requirement of ASTM E1965-98 (2016).
#### 9. Summary of software verification and validation Test
According to Software Safety Classification of IEC 62304, the infrared forehead thermometer IR8807 software shall be Class B. The Infrared Thermometer (model: IR8807) display data about the patient's body temperature supported by the software. According to the software unitlevel test, integration test and system level test, we hereby certify that all the functions those support the infrared forehead thermometer system specifications are proper. The Infrared Thermometer (model: IR8807) can be used safely and effectively to measure body temperature.
{9}------------------------------------------------
## 10. Conclusion
After analyzing the intended use, technological characteristics, verification and validation testing, it can be concluded that the differences between the subject and predicate device do not raise any new or different questions of safety and effectiveness. Therefore, subject device, Infrared Thermometer, Model: IR8807 is substantially equivalent to the predicate device, Noncontact Forehead Thermometer, Model: HW-2/HW-2S/HW-3/HW-4 /HW-4S/HW-302/HW-303, cleared under K173048.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.