The Reliance Endoscope Processing System is intended for washing and high level disinfection of up to two manually pre-cleaned, immersible, reusable, heat-sensitive, semi-critical devices such as GI flexible endoscopes, bronchoscopes and their accessories. High level disinfection is achieved within the 50 - 57°C HLD Phase of the Endoscope Processing Cycle (4-minute generation sequence followed by a 6-minute exposure sequence).
Device Story
Electromechanical washer/high-level disinfector; processes up to two manually pre-cleaned, immersible, heat-sensitive endoscopes. Input: endoscopes, water, and Reliance DG Dry Germicide (single-use package). Operation: microprocessor-controlled cycles; automated dilution of dry germicide to generate peracetic acid in situ; circulation of solution through device lumens and exterior surfaces at 50-57°C; automated rinse with 0.2-micron filtered water; HEPA-filtered air purge. Used in clinical settings by healthcare personnel. Output: processed, high-level disinfected endoscopes; printed cycle summary. Benefits: automated, standardized reprocessing; reduced user exposure to chemicals; validated microbial efficacy; filter integrity monitoring ensures disinfection reliability.
Clinical Evidence
In-use study conducted in a US hospital. Three flexible endoscope types processed according to instructions. Triplicate evaluations showed no organisms recovered after processing. Bioburden levels on clinically used endoscopes prior to disinfection were as high as 10^8 CFU/device.
Technological Characteristics
Electromechanical washer-disinfector; microprocessor-controlled. Internal components: stainless steel and silicone. Energy source: electrical. Connectivity: printer output for cycle documentation; optional barcode reader. Filtration: 0.2-micron water filter; HEPA air filter. Sterilization/Disinfection: High-level disinfection via in situ generated peracetic acid. Cycle parameters: 50-57°C HLD phase.
Indications for Use
Indicated for washing and high-level disinfection of up to two manually pre-cleaned, immersible, reusable, heat-sensitive, semi-critical medical devices, including GI flexible endoscopes and bronchoscopes. Not for use on non-immersible or non-heat-sensitive devices.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Predicate Devices
Custom Ultrasonics System 83 Plus
STERIS 20 Sterilant Concentrate
Sterilox Liquid Chemical HLD System
Related Devices
K203223 — Reliance EPS Endoscope Processing System and Reliance DG Dry Germicide · Steris · Jan 28, 2021
K110453 — RELIANCE ENDOSCOPE PROCESSING SYSTEM · STERIS Corporation · Mar 16, 2011
K102244 — RELIANCE ENDOSCOPE PROCESSING SYSTEM · STERIS Corporation · Dec 3, 2010
K220361 — Reliance EPS Endoscope Processing System and Reliance DG Dry Germicide · STERIS Corporation · Mar 10, 2022
K200989 — Reliance Endoscope Processing System · STERIS Corporation · May 14, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
#### Certification to electromagnetic standards, continued: Non-clinical Tests: Limits and Methods of Processor CISPR 22 (1997) Measurements of Radio Performance, (equivalent to EN 55022:1998) Disturbance Characteristics of continued Information Technology Equipment The Reliance Endoscope Processing System was evaluated in an inuse study in a US hospital. Three flexible endoscopes representing the range of types indicated in the product labeling were used in clinical procedures and processed according to instructions for use. In triplicate Clinical Tests evaluations of each endoscope, no orqanisms were recovered after processing. Bioburden levels on the clinically used endoscopes after manual cleaning and before high level disinfection were determined to be as high as 108 CFU/device. The above data document that the Reliance Endoscope Processing System meets the FDA requirements of the quidance documents: • Content and Format of Premarket Notification [510(k)] Submissions for Liquid Chemical Sterilants/High Level Disinfectants (2000), and · Guidance on Premarket Notification [510(k)] Submissions for Conclusion Automated Endoscope Washers (1993). Conformity to the requirements of these quidances demonstrates the safety and effectiveness of the Reliance Endoscope Processing System. Based on the results of all testing conducted, this system is as safe and effective as the predicate devices.
L-69
{1}------------------------------------------------
| Date | July 5, 2006 |
|----------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter | STERIS Corporation<br>5960 Heisley Road<br>Mentor Ohio 44060 USA<br><br>Telephone: 440-354-2600 |
| Contact | Nancy A. Robinson<br>Director, Advanced Sterilization<br>STERIS Corporation<br>5960 Heisley Road<br>Mentor, OH 44060 USA<br><br>Tel.: 440-392-7742<br>Fax: 440-392-8955<br>E-mail: nancy_robinson@steris.com |
| | Trade name: The Reliance® Endoscope Processing System |
| | Common name: Automated endoscope reprocessing system |
| Device Name | The System includes the following<br>Classification names:<br>Processor: Medical Washer-Disinfector for High Level Disinfection of Medical Devices Germicide: Liquid Chemical Germicide Cleaners: Accessories to the Medical Washer-Disinfector |
| Legally Marketed<br>Devices to which<br>Substantial<br>Equivalence is<br>Claimed | The Reliance Endoscope Processing System is substantially<br>equivalent to the following legally marketed Class II medical devices:<br>Reliance Processor:<br>Custom Ultrasonic System 83 Plus, based on similar indications for use, designs, and features Reliance™ DG Dry germicide:<br>STERIS 20™ Sterilant Concentrate, with respect to device design, active ingredient, conditions of use, and chemical monitoring and<br>Sterilox Liquid Chemical HLD System, with respect to indications for use (high level disinfectant) and <i>in situ</i> generation of the active ingredient |
--
:
L-70
.
:
-
{2}------------------------------------------------
| | Property ▼ | STERIS<br>Reliance Endoscope<br>Processor | | Custom Ultrasonics<br>System 83 Plus | |
|-----------------------------------------------------------------------------------------------------------------|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|
| | Intended Use | Wash and high level<br>disinfect up to two<br>manually pre-cleaned,<br>immersible, reusable,<br>heat-sensitive, semi-<br>critical devices such as GI<br>flexible endoscopes,<br>bronchoscopes and their<br>accessories.<br>High level disinfection is<br>achieved within the<br>50 - 57°C HLD Phase of<br>the Endoscope<br>Processing Cycle<br>(4-minute generation<br>sequence followed by a<br>6-minute exposure<br>sequence). | | Clean and high level<br>disinfect one or two<br>submersible flexible<br>endoscopes intended<br>used to treat and<br>diagnose disorders of the<br>gastrointestinal and/or<br>pulmonary tracts.<br>High level disinfection<br>contact conditions vary<br>with the germicide used<br>in the processor. | |
| Summary of<br>technological<br>characteristics for<br>comparison to<br>predicate:<br>a) Processor | Operational<br>Principles | Pressure in control<br>handle boot drives<br>washing/disinfecting/rinse<br>solutions through lumens.<br>Circulation pump sprays<br>exterior surfaces. | | Fluid connectors/<br>attachments flow<br>cleaning/disinfecting/rinse<br>solution through lumens.<br>Immersion system with<br>ultrasonic cavitations<br>during all phases except<br>disinfection. | |
| | | Flow units are required<br>only for endoscope<br>channels that do not open<br>in the endoscope control<br>handle and when<br>mechanical action is<br>required for valve<br>operation | | Adaptors are required to<br>flow solutions through<br>lumens. | |
| | Cycle Description | Phase | Time | Phase | Time |
| | | | Optional:<br>Washing 1<br>(Wash with<br>rinse) | Wash time<br>can be<br>adjusted to<br>be between<br>5 and 10<br>minutes | Wash |
| | | Rinse | 40 seconds | Rinse | 1.5 minutes |
| | Property ▼ | STERIS<br>Reliance Endoscope<br>Processor | | Custom Ultrasonics<br>System 83 Plus | |
| | Cycle Description,<br>continued | Phase | Time | Phase | Time |
| | | Optional:<br>Washing 2<br>(Wash with<br>rinse) | Wash time<br>can be<br>adjusted to<br>be between<br>5 and 10<br>minutes | Not applicable | |
| | | Rinse | 40 seconds | | |
| | | Dry Germicide<br>Generation | 4 minutes | | |
| | | Dry Germicide<br>Exposure | 6 minutes | Chemical<br>Immersion | Varies with<br>germicide<br>used |
| | | Rinse 1 | 40 seconds | Rinse 1 | 1 minute |
| Summary of<br>technological<br>characteristics for<br>comparison to<br>predicate:<br>a) Processor,<br>continued | | Rinse 2 | 40 seconds | Rinse 2 | 1.5 minutes |
| | | Air purge | Air purge<br>time can be<br>adjusted to<br>be between<br>4 and 30<br>minutes | Air purge | 40 seconds |
| | | Alcohol Flush: Product<br>labeling instructs the user to<br>follow departmental<br>procedures regarding<br>flushing the endoscope<br>channels with alcohol and<br>storage of the endoscope<br>before use. | | Alcohol Flush: Product<br>labeling indicates that an<br>alcohol rinse can be<br>performed manually by the<br>user while the device is<br>attached to the processor. | |
| | Critical Design<br>Features | Intended for use with<br>Reliance DG Dry Germicide<br>only | | Intended for use with<br>different types of liquid<br>chemical germicides<br>(aldehydes and most<br>oxidizing germicides) | |
| | | Microprocessor controlled | | Microprocessor controlled | |
| | | Internal components<br>constructed of stainless steel<br>and silicone | | Internal components<br>constructed of stainless<br>steel | |
:
.--
--
{3}------------------------------------------------
.
:
.
-----
-
L-72
…………
{4}------------------------------------------------
## RELIANCE® ENDOSCOPE PROCESSING SYSTEM
| | Property ▼ | STERIS<br>Reliance Endoscope<br>Processor | Custom Ultrasonics<br>System 83 Plus |
|-----------------------------------------------------------------------------------------------------------------|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Critical Design<br>Features,<br>continued | Processor provides<br>0.2 micron filtered water<br>for washing, disinfection<br>and rinsing | Water filtration system<br>external to device<br>(customer supplied) |
| | | Automated injection of<br>washing solution | Manual injection of<br>cleaning solution |
| | | Automated generation<br>and delivery of high level<br>disinfection solution | High level disinfectant<br>selected and added by<br>operator |
| | | Air intake for Air Purge is<br>HEPA filtered | No air cleaning option |
| Summary of<br>technological<br>characteristics for<br>comparison to<br>predicate:<br>a) Processor,<br>continued | Process<br>Parameters | Can select the following<br>endoscope processing<br>cycles:<br>• Disinfect only<br>• Wash and disinfect<br>Processor also provides<br>the following self-<br>decontamination cycles:<br>• D-SHORT (must be<br>run every 54 hours)<br>• D-LONG (must be<br>run when D-SHORT<br>has not been run in<br>the past 54 hours) | Can select the following<br>endoscope processing<br>cycles:<br>• Wash only<br>• Wash and disinfect<br>• Disinfect only |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | Disinfection time not<br>adjustable | Adjustable disinfection<br>time to accommodate<br>different HLD solutions. |
| | | Washing and air purge<br>times adjustable | Cleaning and air purge<br>times adjustable |
| | | Temperature not<br>adjustable | Temperature is<br>adjustable |
| | Property ▼ | STERIS<br>Reliance Endoscope<br>Processor | Custom Ultrasonics<br>System 83 Plus |
| Summary of<br>technological<br>characteristics for<br>comparison to<br>predicate:<br>a) Processor,<br>continued | Process<br>Parameters,<br>continued | Washing solution<br>concentration not adjustable<br>Reliance DG Dry Germicide<br>solution concentration not<br>adjustable | Cleaning solution is added by<br>operator (3 oz. Tergal 800)<br>Liquid chemical germicide<br>solution is adjustable |
| | | Water quality is not<br>adjustable; filtration systems<br>are provided with processor | Water quality external to<br>processor and provided by<br>customer |
| | Process Monitors | Control Handle Boot pressure<br>alarms if pressure too low to<br>process, or if too high and<br>could potentially damage<br>scopes | Pressure monitor cut off at<br>5 psi<br>Flow sensor for air/water<br>channel to detect<br>backpressure<br>Air restrictor allowing for<br>visual confirmation of flow (air<br>bubbles) through the air/water<br>channel. |
| | | Detection of a fresh Reliance<br>DG Dry Germicide container<br>in every processing cycle | No on-board monitoring for<br>presence of germicide or<br>germicide components |
| | | Washing solution level<br>monitored; alarm indicates<br>when container does not<br>have sufficient amount to<br>complete cycle | Cleaner solution level not<br>monitored |
| | | Temperature alarms if out of<br>range | No temperature monitoring<br>other than heater<br>thermocouple |
| | | Water filter integrity test | No water filter integrity test |
| | Maintenance<br>Requirements | D-SHORT decontamination<br>cycle required every 54 hours<br>D-LONG decontamination<br>cycle required if D-SHORT<br>not performed within past 54<br>hours. | Daily decontamination of<br>internal plumbing using<br>disinfectant solution |
| | | Periodic cleaning of debris<br>screen and spray arms | Daily cleaning with "Comet"<br>or "Ajax"<br>Cleaning when changing<br>disinfectant |
| | | Periodic replacement of<br>water and air filters | Filters provided by customer |
.
{5}------------------------------------------------
---
----
-
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{6}------------------------------------------------
| | Device▶ | Reliance DG Dry<br>Germicide | STERIS 20 | Sterilox Liquid<br>Chemical HLD<br>System |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| | General Features: The general features of Reliance DG Dry Germicide and the<br>predicate germicides are similar in that they are intended to process pre-<br>cleaned, immersible, reusable medical devices. The germicides are supplied in a<br>ready to use, single use dose. | | | |
| | Intended Use: | Dry germicide for<br>use only in the<br>Reliance<br>Endoscope<br>Processor. | Sterilant for use<br>only in the<br>SYSTEM 1<br>Processor. | High level<br>disinfectant for use<br>only with the<br>Sterilox Generator. |
| Summary of<br>technological<br>characteristics<br>for<br>comparison to<br>predicate:<br>b) Reliance<br>DG Dry<br>Germicide | | Provides high level<br>disinfection of up to<br>two manually pre-<br>cleaned,<br>immersible,<br>reusable, heat-<br>sensitive, semi-<br>critical medical<br>devices such as<br>endoscopes and<br>their accessories. | Provides sterile<br>processing of<br>manually pre-<br>cleaned<br>immersible,<br>reusable, semi-<br>critical and critical<br>devices. | Provides for high<br>level disinfection of<br>pre-cleaned,<br>immersible,<br>reusable, heat-<br>sensitive, semi-<br>critical medical<br>devices. |
| | Labeling: | Dry Germicide | Liquid Chemical<br>Sterilant | High Level<br>Disinfectant |
| | Germicide<br>Exposure Time<br>(min) for<br>intended use<br>(High Level<br>Disinfection or<br>Sterile<br>Processing) | 10<br>(4-minute<br>generation<br>sequence followed<br>by a 6-minute<br>exposure<br>sequence) | 12 | 10 |
| | Use<br>Temperature | 50 - 57°C | 50 - 56°C | 25°C |
| | Reuse | Single use | Single use | Single use |
| | Device▶ | Reliance DG Dry<br>Germicide | STERIS 20 | Sterilox Liquid<br>Chemical HLD<br>System |
| Human<br>Factors | | Dispensed ready to<br>use.<br>Container opened<br>automatically by<br>the processor,<br>limiting user<br>exposure to the<br>germicide. | Dispensed ready<br>to use.<br>Container opened<br>automatically by<br>the processor,<br>limiting user<br>exposure to the<br>germicide | Dispensed ready to<br>use. |
| Physical and Chemical Properties: The physical and chemical properties of<br>Reliance DG Dry Germicide and the predicate germicides are similar in that they<br>are supplied in a single use, ready to use dose in a form that is readily soluble in<br>water and results in a ~ pH neutral solution | | | | |
| Summary of<br>technological<br>characteristics<br>for<br>comparison to<br>predicate:<br>b) Reliance<br>DG Dry<br>Germicide,<br>continued | Container | Single-use,<br>2 Compartment<br>Container | Single-use,<br>2 Compartment<br>Container | Single use. |
| | Composition,<br>Packaged<br>Product | Dry/Dry | Dry/Liquid | Liquid |
| | Appearance | Translucent liquid | Lime green liquid | Colorless liquid |
| | Odor | Slightly acidic | Slightly acidic | odorless |
| | pH | 7.0 - 8.5 | 6 - 8 | 5 – 7 |
| | Solubility in<br>Water…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.