RELIANCE ENDOSCOPE PROCESSING SYSTEM

K110453 · STERIS Corporation · NZA · Mar 16, 2011 · Gastroenterology, Urology

Device Facts

Record IDK110453
Device NameRELIANCE ENDOSCOPE PROCESSING SYSTEM
ApplicantSTERIS Corporation
Product CodeNZA · Gastroenterology, Urology
Decision DateMar 16, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K110453 · Mar 16, 2011RELIANCE ENDOSCOPE PROCESSING SYSTEMSTERIS CorporationUS hospital clinical settingAn in-use study was conducted in a US hospital to evaluate the performance of the Reliance Endoscope Processing System on flexible endoscopes used in actual clinical procedures.in-use study; clinical setting; endoscope reprocessing

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
In-use study in a US hospital; In-use clinical evaluationFlexible endoscopes used in clinical procedures; Sample Size: 3 flexible endoscopes; Number of Sites: 1Not applicable for this studyRecovery of organisms after processing; bioburden levels after manual cleaning

Indications for Use

The Reliance Endoscope Processing System is intended for washing and high level disinfection of up to two manually pre-cleaned, immersible, reusable, heat-sensitive, semi-critical devices such as GI flexible endoscopes, bronchoscopes and their accessories. High level disinfection is achieved within the 50 - 57℃ HLD Phase of the Endoscope Processing Cycle (4 minute generation sequence followed by a 6-minute exposure sequence).

Device Story

Electromechanical washer/high-level disinfector; processes up to two manually cleaned, immersible, heat-sensitive, semi-critical devices (e.g., GI flexible endoscopes, bronchoscopes). System utilizes Reliance DG Dry Germicide (two-part, dry, single-use peracetic acid-generating chemistry) and optional Klenzyme Enzymatic Presoak. Processor features microprocessor-based controller for automated cycles: optional wash phase (5-10 mins), high-level disinfection phase (4-min generation at 50-57°C, 6-min exposure), rinse phase (0.2-micron bacterial-retentive filtered water), and HEPA-filtered air purge. Includes two self-decontamination cycles (D-SHORT for biofilm prevention; D-LONG with CIP 200 cleaner). Output includes processed devices and printed cycle summary (optional barcode reader for tracking). Used in clinical settings by staff; ensures patient safety by reducing bioburden on endoscopes. System includes internal filter integrity check and alarm system.

Clinical Evidence

Bench testing and simulated-use studies performed. Germicidal efficacy validated via AOAC sporicidal, tuberculocidal, virucidal, bactericidal, and fungicidal tests. Simulated-use achieved >6 log10 reduction of Mycobacterium terrae. In-use study in a US hospital on three flexible endoscope types showed no organisms recovered post-processing (pre-disinfection bioburden up to 10^5 CFU/device). Material compatibility confirmed after 300 cycles. Processor parameters validated under worst-case conditions.

Technological Characteristics

Electromechanical washer/disinfector; microprocessor-controlled. Uses peracetic acid-generating oxidative chemistry (Reliance DG). Features 0.2-micron bacterial-retentive water filtration and HEPA air filtration. Complies with ANSI/UL 61010-1, CAN/CSA C22.2 61010-1, IEC 61010-1, IEC 61010-2-40, and IEC 61326-1. Standalone unit.

Indications for Use

Indicated for washing and high-level disinfection of up to two manually pre-cleaned, immersible, reusable, heat-sensitive, semi-critical devices, including GI flexible endoscopes, bronchoscopes, and their accessories.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K110453 STERIS® p1 of 7 MAR 1 6 2011 510(k) Summary For Reliance® Endoscope Processing System STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198 Contact: Robert F. Sullivan Senior Director, Regulatory Affairs Telephone: (440) 392-7695 Fax No: Summary Date: February 15, 2011 STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600 {2}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the word "STERIS" in large, bold letters. Above the word "STERIS" is the text "K110453 p.2 of 7" in a handwritten font. Below the word "STERIS" is a graphic of several horizontal wavy lines stacked on top of each other. #### 1. Device Name | Trade Name: | Reliance Endoscope Processing System | |----------------------|----------------------------------------------| | Common/usual Name: | Automated Endoscope Reprocessor | | Classification Name: | 21 CFR 876.1500<br>Endoscope and Accessories | ### 2. Predicate Device Reliance Endoscope Processing System K102244 ### 3. Description of Device - The Reliance Endoscope Processing System is an economical, easy-to-use high level disinfection system intended to wash and high level disinfect up to two manually cleaned, immersible, reusable, heat-sensitive, semi-critical devices such as GI flexible endoscopes and related accessories. The system utilizes Reliance™ DG Dry Germicide, a proprietary, safe, and dry peracetic acid generating oxidative chemistry. The Reliance Endoscope Processing System was designed to be versatile in meeting the growing demands of the modern flexible endoscope processing department, while offering the highest level of patient and staff safety. The Reliance Endoscope Processing System is a combination of products that are used to wash and high level disinfect flexible endoscopes and their accessories. - The Reliance Endoscope Processor is an electromechanical washer/high level . disinfector with a microprocessor-based controller that provides for automated endoscope processing cycles and processor self-decontamination cycles. - Reliance DG Dry Germicide is a proprietary, two-part, dry, single-use . oxidative chemistry, designed to generate the high level disinfection solution upon automatic dilution in water within the Reliance Endoscope Processor. - Optional washing is provided through the automated delivery of Klenzyme . Enzymatic Presoak and Cleaner during the wash phase of the cycle. - CIP 200 Acid-Based Process and Research Cleaner, a general cleaning agent, . is used in one of the two self-decontamination cycles provided by the processor. - Various accessories are available to accommodate the processing needs of . specific endoscopes and endoscopic accessories. {3}------------------------------------------------ - - . Verify Reliance CI Process Indicator is available to monitor for the presence of the Reliance DG active ingredient, peracetic acid. The Reliance Endoscope Processing Cycle has the following features: - > The first part of this cycle is an optional programmable washing phase. This phase consists of a wash that uses Klenzyme, followed by a rinse. The washing phase can be programmed on or off. In the "on" mode, the user can choose either one or two washing phases per processing cycle, and the wash time can be adjusted to be between 5 and 10 minutes. The Reliance cleaning phase does not replace manual cleaning by the user. - > The second part is a high level disinfection phase that is non-optional and the parameters cannot be changed by the user. In this phase, the proprietary Reliance DG components, provided in a single use container, are dissolved with water at ~50°C for four minutes of generation time and circulated throughout the processor and through device lumens for 6 minutes of high level disinfection solution exposure time. - > Following the high level disinfection phase, the Reliance Endoscope Processor removes the high level disinfection solution through a rinse phase which is non-optional and the parameters cannot be changed by the user. The processor filters the rinse water (as well as all of the water used throughout the cycle) through a 0.2 micron bacterial-retentive filter. It also incorporates an automatic internal integrity check of this filter at the end of each processing cycle. If the integrity check fails, an alarm alerts the user, and the processor does not complete the cycle. - > The last step in the processing cycle is an air purge phase using HEPAfiltered air. The air purge helps to remove excess rinse water from the processed devices. The final air purge is preset to run for 4 minutes; additional air purge time may be selected by the operator. - >> The processor will print a detailed cycle summary at the end of each cycle that includes information such as processor number, cycle date, start and stop times, as well as phase parameters. With an optional bar code reader, the printouts can also include identification numbers for the operator, patient, device, doctor and procedure. The processor features two decontamination cycles that are to be used without endoscopes in the processor: - > The first, called D-SHORT, consists of hot water circulating through the processor for 10 minutes, followed by a 10-minute hot air purge. This cycle is to be run every 54 hours. D-SHORT is intended to prevent biofilm from forming. - > The second, called D-LONG, consists of a cycle in which CIP 200 Acid-Based Process and Research Cleaner is added to hot water. The cleaning solution is then circulated through the processor for 20 minutes; this is followed by three rinses to remove the solution from the processor and a 10- {4}------------------------------------------------ minute hot air purge. D-LONG is to be used on those occasions when the D-SHORT cycle has not been run within the past 54 hours. #### 4. Intended Use The Reliance Endoscope Processing System is intended for washing and high level disinfection of up to two manually cleaned, immersible, reusable, heat-sensitive, semi-critical devices such as GI flexible endoscopes, bronchoscopes and their accessories. High level disinfection is achieved within the 50 - 57°C HLD Phase of the Endoscope Processing Cycle (4 minute generation sequence followed by a 6minute exposure sequence). ### 5. Description of Safety and Substantial Equivalence The Reliance Endoscope Processing System has the same technological characteristics as the predicate device. The redesigned Reliance DG Dry Germicide secondary packaging configuration has been tested through the labeled shelf life of the product under defined storage conditions that represent a range of environmental stresses. After removal from the packaging throughout its shelf life, as well as after an additional two-week use life outside the packaging, Reliance DG containers met established acceptance criteria for visual physical integrity of the package and its contents, component concentrations, pH, and chelation capacity. At end of shelf life the product met its performance criteria in the Reliance EPS processor. The proposed modification does not change the integrity of the Reliance DG Dry Germicide over its labeled shelf life. The device is substantially equivalent to its predicate. A summary of performance testing applicable to both the proposed Reliance Endoscope Processing System and its predicate Reliance Endoscope Processing System is provided in the Table attached. {5}------------------------------------------------ K110453 p547 Non-clinical Tests: Germicide Efficacy The Reliance Endoscope Processing System was developed and validated in accord with two primary FDA Guidance documents: - Content and Format of Premarket Notification [510(k)] Submissions for Liquid Chemical Sterilants/High Level Disinfectants (2000), and - Guidance on Premarket Notification [510(k)] Submissions for Automated Endoscope Washers (1993). A summary of this testing follows: Reliance DG Dry Germicide Efficacy: Reliance DG Dry Germicide was tested and shown to generate an effective high level disinfection solution using the standard array of microbiological tests for germicidal efficacy. The following testing was performed at conditions of use that were worst case with respect to contact time, circulation, water hardness, temperature, artificial soiling, and germicide concentration (at or below the minimum effective dose of 9000 mg/L PAA minutes): ⇒ **Sporicidal:** Reliance DG Dry Germicide was proven to be sporicidal as defined by AOAC Sporicidal Activity Test with an *in situ* exposure time of 6 minutes. Confirmatory testing was completed successfully and supplemental confirmatory testing was completed in the Reliance Endoscope Processor. Potency was subsequently confirmed in the processor using Reliance DG Dry Germicide containers that were aged beyond the end of its shelf life. ⇒ **Tuberculocidal:** Reliance DG Dry Germicide was proven to be tuberculocidal as defined by the AOAC Tuberculocidal Activity Test with an exposure time of 6 minutes. Potency was subsequently confirmed using Reliance DG Dry Germicide aged beyond the end of its shelf life. ⇒ **Virucidal:** The Reliance Process was proven to reduce the viable population of poliovirus Type 1, adenovirus Type 5, and herpes simplex virus Type 1 by > 4 log₁₀. ⇒ **Bactericidal:** Reliance DG Dry Germicide was proven to be bactericidal as defined by the AOAC Bactericidal Activity Test with an exposure time of 6 minutes at worst case conditions, whether performed *in situ* or *in vitro*. ⇒ **Fungicidal:** Reliance DG Dry Germicide was proven to be fungicidal as defined by the AOAC Fungicidal Activity Test with an exposure time of 6 minutes, whether performed *in situ* or *in vitro*. **Simulated-Use:** Reliance DG Dry Germicide, at the minimum recommended dose, reproducibly achieved greater than a 6 log₁₀ reduction of *Mycobacterium tarrae* in triplicate trials within the Reliance Endoscope Processor for selected clinically relevant flexible endoscopes and their accessories. The test articles represented the range of most challenging devices, accessories, and processing situations. **In-Use:** Reliance Endoscope Processing System was evaluated in an in use study in a US hospital. Three flexible endoscopes representing the range of types indicated in the product labeling were used in clinical procedures and processed according to instructions for use. In triplicate evaluations of each endoscope, no organisms were recovered after processing. Bioburden levels on the clinically used endoscopes after manual cleaning and before high level disinfection were determined to be as high as 10⁵ CFU/device. {6}------------------------------------------------ 10453 ## Biocompatibility: The Reliance chemical formulations, as supplied in packaging as well as in use dilutions, can be safely handled and used by customers. Residues that may remain on medical endoscopes and accessories are below established residue limits and do not pose a risk to patients. Safety statements in product labeling are appropriate to the potential risk. - Reliance DG Dry Germicide, its components, reaction products, fi and residuals remaining on medical devices were evaluated for biocompatibility and possible risks to users. Testing included acute oral and ocular toxicity tests, dermal irritation studies, in vitro bacterial mutation genotoxicity studies, sensitization tests, and in vitro cytotoxicity evaluations; literature reviews of raw material toxicity data were also performed. Certain components in the single-use container, which under normal use conditions never contact the user, have the potential for irritation or skin sensitization: therefore appropriate warnings and instructions are displayed on labeling for the unusual event of a spill or container breakage. - > Use dilution reaches non-cytotoxic levels with minimal dilution. - > Biocompatibility testing of extracts from processed medical devices demonstrated that no toxic residuals remain on devices under worst case circumstances. The test data indicate that the worst case residue levels for the components of maximum potential risk are far below the allowable limits. Furthermore, the processor final rinse water was found to be non-cytotoxic. ## Reliance Endoscope Processing System Material Compatibility: The Reliance System was evaluated for its effect on intact medical devices, including flexible endoscopes and/or common materials of device construction. After 300 processing cycles, no deleterious effects were observed other than minor cosmetic changes. No functional changes in flexible endoscopes were observed. ## Reliance DG Dry Germicide Stability: Reliance DG was tested and found stable for 18 months in the unopened moisture-resistant packaging at the stated conditions for storage. Once opened, the individual cups within each tray are to be used within 2 weeks, or by the expiration date on the Reliance DG cup, whichever comes first. {7}------------------------------------------------ K110453 p7f7 | <b>Reliance Endoscope Processor Performance</b><br>The critical process parameters for the processor (water temperature and volume, fresh Reliance DG Dry Germicide container detection, boot pressure, delivery of washing solutions and high level disinfection solution and water filter integrity testing) were each evaluated in replicate under worst case conditions and found to be within required specifications.<br>Each processor phase or cycle was separately evaluated and documented to be effective under worst case conditions:<br>⇒ Validation of the washing phase of the Endoscope Processing Cycle documented that after the shortest possible washing phase, devices pre-soiled with a combination of eggs, blood, mucin and serum in saline: 1) were visually clean, and 2) achieved greatly reduced yield of extractable protein per cm² device area (assayed to be reduced from >173 to < 5 μg/cm²).<br>⇒ Validation of the high level disinfection phase of the Endoscope Processing Cycle was performed through a simulated-use study in challenging clinically relevant endoscopes, as well as in an in-use study.<br>⇒ The rinse phase of the endoscope processing cycle was shown to be effective. Evaluations of extracts of devices exposed to worst case conditions in the Endoscope Processing Cycle documented that levels of residuals remaining on devices were far below allowable limits and were not cytotoxic.<br>⇒ The air purge phase was validated to confirm the ability to remove rinse water from processed medical devices.<br>⇒ The filter integrity test system of the processor was documented to reliably detect filter failure.<br>⇒ The two self-decontamination cycles were shown to be effective as follows: D-LONG cycle - can disinfect the Reliance Endoscope Processor after a high level challenge with <i>Pseudomonas aeruginosa</i> followed by a 5 day inactive period; D-SHORT cycle - can kill bacteria that have potential to form biofilm. | | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | ANSI/UL -61010-1, 2nd Ed., | Electrical equipment for measurement, control and laboratory use Part 1, 2nd Edition | | CAN/CSA C22.2 61010-1, 2nd Ed. | Electrical equipment for measurement, control and laboratory use Part 1, 2nd Edition | | IEC 61010-1:2001 | Safety requirements for electrical equipment for measurement, control and laboratory use, Part 1: General requirements, | | IEC 61010-2-40, 1st Ed. | Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-040: Particular requirements for sterilizers and washer-disinfectors used to treat medical materials | | IEC 61326-1:2005, 1st Ed. | Electrical equipment for measurement, control and laboratory use, EMC requirements Part 1: General requirements | : . - - । - {8}------------------------------------------------ Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Mr. Robert Sullivan Senior Director, Regulatory Affairs STERIS Corporation 5960 Heisley Road Mentor, Ohio 44060 MAR 1 6 231 Re: K110453 Trade/Device Name: Reliance® Endoscope Processing System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: NZA Dated: February 15, 2011 Received: February 16, 2011 Dear Mr. Sullivan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {9}------------------------------------------------ Page 2- Mr. Sullivan Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Juns J. Koscielniak for R Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health ## Enclosure {10}------------------------------------------------ # Indications for Use 510(k) Number (if known): Device Name: Reliance® Endoscope Processing System Indications for Use: The Reliance Endoscope Processing System is intended for washing and high level disinfection of up to two manually pre-cleaned, immersible, reusable, heat-sensitive, semi-critical devices such as GI flexible endoscopes, bronchoscopes and their accessories. High level disinfection is achieved within the 50 - 57℃ HLD Phase of the Endoscope Processing Cycle (4 minute generation sequence followed by a 6-minute exposure sequence). Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ X (21 CFR 801 Subpart C) # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Elizabeth F. Clausing Williams (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices KII 0453 510(k) Number: Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...