DyeVert Plus Contrast Reduction System, DyeVert NG Contrast Reduction System
K171217 · Osprey Medical, Inc. · DXT · May 26, 2017 · Cardiovascular
Device Facts
Record ID
K171217
Device Name
DyeVert Plus Contrast Reduction System, DyeVert NG Contrast Reduction System
Applicant
Osprey Medical, Inc.
Product Code
DXT · Cardiovascular
Decision Date
May 26, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1650
Device Class
Class 2
Indications for Use
The DyeVert™ NG Contrast Reduction System is to be used for the controlled infusion and contrast volume reduction of radiopaque contrast media for angiographic procedures with the following agents: Iodixanol 270 or 320 mg/mL, Iohexol 300 or 350 mgI/mL and Iopamidol 370 mgI/mL. The DyeVert™ Plus Contrast Reduction System consists of a Display and DyeVert Plus Disposable Kit. The system is to be used for contrast volume reduction and for the monitoring of radiopaque contrast media during angiographic or CT procedures with the following agents: Iodixanol 270 or 320 mg/mL, Iohexol 300 or 350 mg/mL and Iopamidol 370 mg/ mL.
Device Story
DyeVert NG and DyeVert Plus systems modulate fluid pathway resistance during manual contrast injections; minimizes excess 'refluxed' contrast volume in patient vasculature; maintains image quality while reducing total contrast agent exposure. DyeVert Plus includes a display for monitoring contrast media. Used in angiographic or CT procedures; operated by clinicians. Devices facilitate reduced contrast load, potentially benefiting patients by lowering risk of contrast-associated complications. Modification from machined to molded components; fundamental technology and principle of operation remain unchanged.
Clinical Evidence
No clinical testing performed. Bench testing only: included flow rate, peak pressure, contrast diversion, mechanical testing (high pressure, leak, tensile), durability cycling, sterilization validation (ISO 11135-1:2014), simulated use/usability, packaging/distribution (ASTM D4169:2016), and biocompatibility (ISO 10993-1:2009). All tests met acceptance criteria.
Technological Characteristics
Angiographic injector/syringe system. Components transitioned from machined to molded. Sterilization via Ethylene Oxide (ISO 11135-1:2014). Biocompatibility per ISO 10993-1:2009. Packaging per ISO 11607. Connectivity: DyeVert Plus includes a display unit for monitoring. No active electronic sensing/actuation described beyond fluid pathway resistance modulation.
Indications for Use
Indicated for patients undergoing angiographic or CT procedures requiring radiopaque contrast media (Iodixanol, Iohexol, or Iopamidol) where contrast volume reduction and/or monitoring is desired.
Regulatory Classification
Special Controls
*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{1}------------------------------------------------
Public Health Service
May 26, 2017
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Osprey Medical, Inc. Melanie Hess Vice President Regulatory Affairs 5600 Rowland Road Suite 250 Minnetonka, Minnesota 55343
Re: K171217
Trade/Device Name: DyeVert Plus Contrast Reduction System, DyeVert NG Contrast Reduction System Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic Injector And Syringe Regulatory Class: Class II Product Code: DXT Dated: April 25, 2017 Received: April 26, 2017
Dear Melanie Hess:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{2}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Fernando
Aquel-S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171217
Device Name DyeVert™ NG Contrast Reduction System
#### Indications for Use (Describe)
The DyeVert™ NG Contrast Reduction System is to be used for the controlled infusion and contrast volume reduction of radiopaque contrast media for angiographic procedures with the following agents: Iodixanol 270 or 320 mg/mL, Iohexol 300 or 350 mgI/mL and Iopamidol 370 mgI/mL.
Type of Use (Select one or both, as applicable)
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171217
Device Name DyeVert™ Plus Contrast Reduction System
#### Indications for Use (Describe)
The DyeVert™ Plus Contrast Reduction System consists of a Display and DyeVert Plus Disposable Kit. The system is to be used for contrast volume reduction and for the monitoring of radiopaque contrast media during angiographic or CT procedures with the following agents: Iodixanol 270 or 320 mg/mL, Iohexol 300 or 350 mg/mL and Iopamidol 370 mg/ mL.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
## 510(k) Summary
# K171217
510(k) Summary As required by 21CFR 807.92(c)
510(k) Number: Submission Type: Date Prepared:
Submitter's Name/Address:
Contact Person:
K171217 Special pre-market notification (510(k)), bundled April 25, 2017
Osprey Medical 5600 Rowland Road Suite 250 Minnetonka, MN 55343
Melanie Hess Vice President, Regulatory Affairs Tel: 952-955-8252 Fax: 952-955-8171 Mhess@ospreymed.com
#### Device Information:
| Trade Name/Proprietary Name: | DyeVert™ NG Contrast Reduction System |
|------------------------------|------------------------------------------------------|
| Manufacturer: | Osprey Medical, Inc. |
| Common Name: | Injector and Syringe, Angiographic |
| Classification Registration: | 21 CFR § 870.1650 |
| Product Code: | DXT |
| FDA Center/Branch: | CDRH/Interventional Cardiology Devices Branch (ICDB) |
| Trade Name/Proprietary Name: | DyeVert™ Plus Contrast Reduction System |
| Manufacturer: | Osprey Medical, Inc. |
| Common Name: | Injector and Syringe, Angiographic |
| Classification Registration: | 21 CFR § 870.1650 |
| Product Code: | DXT |
| FDA Center/Branch: | CDRH/Interventional Cardiology Devices Branch (ICDB) |
## Device Description:
The Osprey Medical DyeVert" NG and DyeVert" Plus Contrast Reduction Systems are compatible to manual contrast injections and provide fluid pathway resistance modulation such that excess contrast volume (i.e. contrast that is not needed for diagnostic or therapeutic purposes also referred to as "refluxed contrast') is minimized in the patient's vasculature. This allows for a reduction in the total contrast agent volume during coronary or peripheral imaging to the patient; while maintaining adequate image quality.
#### Indications for Use:
The DyeVert™ NG Contrast Reduction System is to be used for the controlled infusion and contrast volume reduction of radiopaque contrast media for angiographic procedures with the following agents: Iodixanol 270 or 320 mgI/mL, Iohexol 300 or 350 mgI/mL and Iopamidol 370 mgI/mL.
The DyeVert™ Plus Contrast Reduction System consists of a Display and DyeVert Plus Disposable Kit. The system is to be used for contrast volume reduction and for the monitoring of radiopaque contrast media during angiographic or CT procedures with the following agents: Iodixanol 270 or 320 mg/mL, Iohexol 300 or 350 mgI/mL and Iopamidol 370 mgI/mL.
## Predicate Device:
Trade Name/Proprietary Name: Manufacturer:
DyeVert™ NG Contrast Reduction System Osprey Medical, Inc.
{6}------------------------------------------------
| Common Name: | Injector and Syringe, Angiographic |
|------------------|------------------------------------|
| Classification: | II |
| CFR Reference: | 21 CFR § 870.1650 |
| Product Code: | DXT |
| 510(k) number(s) | K161505 |
#### Predicate Device:
| Trade Name/Proprietary Name: | DyeVert™ PLUS Contrast Reduction System |
|------------------------------|-----------------------------------------|
| Manufacturer: | Osprey Medical, Inc. |
| Common Name: | Injector and Syringe, Angiographic |
| Classification: | II |
| CFR Reference: | 21 CFR § 870.1650 |
| Product Code: | DXT |
| 510(k) number(s): | K163054 |
# Comparison to the Predicate Device:
The proposed devices are substantially equivalent to the previously cleared predicates, in that they are both designed for use during the controlled infusion of radiopaque contrast media for angiographic procedures. The proposed devices have identical product performance specifications, sterilization processes, shelf life, packaging and benefit risk profiles. The subject 510(k) is to allow for the modification from machined to molded components. The fundamental scientific technology, principle of operation and primary mechanism of action remains unchanged. No new intended use, intended user or different questions of safety or effectiveness are raised with the proposed modification.
# Summary of Non-Clinical Testing:
Bench testing was performed and/or leveraged to support this submission and results demonstrate the materials, design considerations and manufacturing processes continue to meet product specifications and performance requirements. The following testing was successfully completed and/or leveraged within this submission:
- . Confirmatory device performance testing (design verification) included flow rate, peak pressure, contrast diversion, mechanical testing (high pressure use, leak and tensile strength) and visual verifications to design specifications for specific contrasts and catheter configurations. Durability testing (cycling) was leveraged. All testing passed and demonstrated product performance met all prior established acceptance criteria.
- Sterilization validation was leveraged - Sterilization conditions have been validated in accordance with ISO 11135-1:2014, Sterilization of health care products – Ethylene Oxide Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices to provide a Sterility Assurance Level of 10°. All testing passed.
- Simulated Use (Bench) design validation was leveraged and included the assessment of the ease ● of use, system set up and device priming ability (usability). In addition, design validation was leveraged for injection pressure, contrast diversion and image analysis testing. All testing passed.
- Packaging, shelf life and distribution testing was leveraged. Distribution testing was conducted per ASTM D4169:2016 Standard Practice for Performance Testing of Shipping Containers and Systems. Testing included visual inspection, cycle testing, dye leak/penetration test, seal strength test and functional testing. All testing passed and demonstrated product performance met all prior established acceptance criteria. Packaging was deemed to be in compliance with ISO 11607 part 1 and 2:2006 Packaging for terminally sterilized medical devices.
- . Confirmatory biocompatibility testing was performed in accordance with ISO 10993-1:2009 Biological Evaluation of Medical Devices – Part I: Evaluation and Testing. Testing included cytotoxicity, sensitization and acute systemic toxicity, and hemocompatibility. All testing passed and met prior established acceptance criteria.
{7}------------------------------------------------
No performance standards have been established under Section 514 of the Food, Drug and Cosmetic Act for angiographic injectors and syringes.
# Clinical Testing:
No clinical testing was performed to support this 510(k) Premarket Notification.
## Statement of Equivalence:
The proposed subject devices are substantially equivalent in intended use, indications for use statement and fundamental scientific technology as the predicate devices. Based on this and data analyzed in accordance with Osprey Medical Quality System Procedure in compliance with EN ISO 13485:2012 Medical Devices – Quality management systems - regulatory purposes and EN ISO 14791:2012 Risk management for medical devices, the proposed subject devices have been shown to be substantially equivalent under 21 CFR Part 807 subpart E.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.