DyeVert Plus Contrast Reduction System, DyeVert NG Contrast Reduction System

K171217 · Osprey Medical, Inc. · DXT · May 26, 2017 · Cardiovascular

Device Facts

Record IDK171217
Device NameDyeVert Plus Contrast Reduction System, DyeVert NG Contrast Reduction System
ApplicantOsprey Medical, Inc.
Product CodeDXT · Cardiovascular
Decision DateMay 26, 2017
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1650
Device ClassClass 2

Intended Use

The DyeVert™ NG Contrast Reduction System is to be used for the controlled infusion and contrast volume reduction of radiopaque contrast media for angiographic procedures with the following agents: Iodixanol 270 or 320 mg/mL, Iohexol 300 or 350 mgI/mL and Iopamidol 370 mgI/mL. The DyeVert™ Plus Contrast Reduction System consists of a Display and DyeVert Plus Disposable Kit. The system is to be used for contrast volume reduction and for the monitoring of radiopaque contrast media during angiographic or CT procedures with the following agents: Iodixanol 270 or 320 mg/mL, Iohexol 300 or 350 mg/mL and Iopamidol 370 mg/ mL.

Device Story

DyeVert NG and DyeVert Plus systems modulate fluid pathway resistance during manual contrast injections; minimizes excess 'refluxed' contrast volume in patient vasculature; maintains image quality while reducing total contrast agent exposure. DyeVert Plus includes a display for monitoring contrast media. Used in angiographic or CT procedures; operated by clinicians. Devices facilitate reduced contrast load, potentially benefiting patients by lowering risk of contrast-associated complications. Modification from machined to molded components; fundamental technology and principle of operation remain unchanged.

Clinical Evidence

No clinical testing performed. Bench testing only: included flow rate, peak pressure, contrast diversion, mechanical testing (high pressure, leak, tensile), durability cycling, sterilization validation (ISO 11135-1:2014), simulated use/usability, packaging/distribution (ASTM D4169:2016), and biocompatibility (ISO 10993-1:2009). All tests met acceptance criteria.

Technological Characteristics

Angiographic injector/syringe system. Components transitioned from machined to molded. Sterilization via Ethylene Oxide (ISO 11135-1:2014). Biocompatibility per ISO 10993-1:2009. Packaging per ISO 11607. Connectivity: DyeVert Plus includes a display unit for monitoring. No active electronic sensing/actuation described beyond fluid pathway resistance modulation.

Indications for Use

Indicated for patients undergoing angiographic or CT procedures requiring radiopaque contrast media (Iodixanol, Iohexol, or Iopamidol) where contrast volume reduction and/or monitoring is desired.

Regulatory Classification

Special Controls

*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Public Health Service May 26, 2017 Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Osprey Medical, Inc. Melanie Hess Vice President Regulatory Affairs 5600 Rowland Road Suite 250 Minnetonka, Minnesota 55343 Re: K171217 Trade/Device Name: DyeVert Plus Contrast Reduction System, DyeVert NG Contrast Reduction System Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic Injector And Syringe Regulatory Class: Class II Product Code: DXT Dated: April 25, 2017 Received: April 26, 2017 Dear Melanie Hess: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply {1}------------------------------------------------ with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, Fernando Aquel-S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K171217 Device Name DyeVert™ NG Contrast Reduction System #### Indications for Use (Describe) The DyeVert™ NG Contrast Reduction System is to be used for the controlled infusion and contrast volume reduction of radiopaque contrast media for angiographic procedures with the following agents: Iodixanol 270 or 320 mg/mL, Iohexol 300 or 350 mgI/mL and Iopamidol 370 mgI/mL. Type of Use (Select one or both, as applicable) | <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | |----------------------------------------------------------------------------------| | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) | # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K171217 Device Name DyeVert™ Plus Contrast Reduction System #### Indications for Use (Describe) The DyeVert™ Plus Contrast Reduction System consists of a Display and DyeVert Plus Disposable Kit. The system is to be used for contrast volume reduction and for the monitoring of radiopaque contrast media during angiographic or CT procedures with the following agents: Iodixanol 270 or 320 mg/mL, Iohexol 300 or 350 mg/mL and Iopamidol 370 mg/ mL. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------| | <div> <span> <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ## 510(k) Summary # K171217 510(k) Summary As required by 21CFR 807.92(c) 510(k) Number: Submission Type: Date Prepared: Submitter's Name/Address: Contact Person: K171217 Special pre-market notification (510(k)), bundled April 25, 2017 Osprey Medical 5600 Rowland Road Suite 250 Minnetonka, MN 55343 Melanie Hess Vice President, Regulatory Affairs Tel: 952-955-8252 Fax: 952-955-8171 Mhess@ospreymed.com #### Device Information: | Trade Name/Proprietary Name: | DyeVert™ NG Contrast Reduction System | |------------------------------|------------------------------------------------------| | Manufacturer: | Osprey Medical, Inc. | | Common Name: | Injector and Syringe, Angiographic | | Classification Registration: | 21 CFR § 870.1650 | | Product Code: | DXT | | FDA Center/Branch: | CDRH/Interventional Cardiology Devices Branch (ICDB) | | Trade Name/Proprietary Name: | DyeVert™ Plus Contrast Reduction System | | Manufacturer: | Osprey Medical, Inc. | | Common Name: | Injector and Syringe, Angiographic | | Classification Registration: | 21 CFR § 870.1650 | | Product Code: | DXT | | FDA Center/Branch: | CDRH/Interventional Cardiology Devices Branch (ICDB) | ## Device Description: The Osprey Medical DyeVert" NG and DyeVert" Plus Contrast Reduction Systems are compatible to manual contrast injections and provide fluid pathway resistance modulation such that excess contrast volume (i.e. contrast that is not needed for diagnostic or therapeutic purposes also referred to as "refluxed contrast') is minimized in the patient's vasculature. This allows for a reduction in the total contrast agent volume during coronary or peripheral imaging to the patient; while maintaining adequate image quality. #### Indications for Use: The DyeVert™ NG Contrast Reduction System is to be used for the controlled infusion and contrast volume reduction of radiopaque contrast media for angiographic procedures with the following agents: Iodixanol 270 or 320 mgI/mL, Iohexol 300 or 350 mgI/mL and Iopamidol 370 mgI/mL. The DyeVert™ Plus Contrast Reduction System consists of a Display and DyeVert Plus Disposable Kit. The system is to be used for contrast volume reduction and for the monitoring of radiopaque contrast media during angiographic or CT procedures with the following agents: Iodixanol 270 or 320 mg/mL, Iohexol 300 or 350 mgI/mL and Iopamidol 370 mgI/mL. ## Predicate Device: Trade Name/Proprietary Name: Manufacturer: DyeVert™ NG Contrast Reduction System Osprey Medical, Inc. {5}------------------------------------------------ | Common Name: | Injector and Syringe, Angiographic | |------------------|------------------------------------| | Classification: | II | | CFR Reference: | 21 CFR § 870.1650 | | Product Code: | DXT | | 510(k) number(s) | K161505 | #### Predicate Device: | Trade Name/Proprietary Name: | DyeVert™ PLUS Contrast Reduction System | |------------------------------|-----------------------------------------| | Manufacturer: | Osprey Medical, Inc. | | Common Name: | Injector and Syringe, Angiographic | | Classification: | II | | CFR Reference: | 21 CFR § 870.1650 | | Product Code: | DXT | | 510(k) number(s): | K163054 | # Comparison to the Predicate Device: The proposed devices are substantially equivalent to the previously cleared predicates, in that they are both designed for use during the controlled infusion of radiopaque contrast media for angiographic procedures. The proposed devices have identical product performance specifications, sterilization processes, shelf life, packaging and benefit risk profiles. The subject 510(k) is to allow for the modification from machined to molded components. The fundamental scientific technology, principle of operation and primary mechanism of action remains unchanged. No new intended use, intended user or different questions of safety or effectiveness are raised with the proposed modification. # Summary of Non-Clinical Testing: Bench testing was performed and/or leveraged to support this submission and results demonstrate the materials, design considerations and manufacturing processes continue to meet product specifications and performance requirements. The following testing was successfully completed and/or leveraged within this submission: - . Confirmatory device performance testing (design verification) included flow rate, peak pressure, contrast diversion, mechanical testing (high pressure use, leak and tensile strength) and visual verifications to design specifications for specific contrasts and catheter configurations. Durability testing (cycling) was leveraged. All testing passed and demonstrated product performance met all prior established acceptance criteria. - Sterilization validation was leveraged - Sterilization conditions have been validated in accordance with ISO 11135-1:2014, Sterilization of health care products – Ethylene Oxide Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices to provide a Sterility Assurance Level of 10°. All testing passed. - Simulated Use (Bench) design validation was leveraged and included the assessment of the ease ● of use, system set up and device priming ability (usability). In addition, design validation was leveraged for injection pressure, contrast diversion and image analysis testing. All testing passed. - Packaging, shelf life and distribution testing was leveraged. Distribution testing was conducted per ASTM D4169:2016 Standard Practice for Performance Testing of Shipping Containers and Systems. Testing included visual inspection, cycle testing, dye leak/penetration test, seal strength test and functional testing. All testing passed and demonstrated product performance met all prior established acceptance criteria. Packaging was deemed to be in compliance with ISO 11607 part 1 and 2:2006 Packaging for terminally sterilized medical devices. - . Confirmatory biocompatibility testing was performed in accordance with ISO 10993-1:2009 Biological Evaluation of Medical Devices – Part I: Evaluation and Testing. Testing included cytotoxicity, sensitization and acute systemic toxicity, and hemocompatibility. All testing passed and met prior established acceptance criteria. {6}------------------------------------------------ No performance standards have been established under Section 514 of the Food, Drug and Cosmetic Act for angiographic injectors and syringes. # Clinical Testing: No clinical testing was performed to support this 510(k) Premarket Notification. ## Statement of Equivalence: The proposed subject devices are substantially equivalent in intended use, indications for use statement and fundamental scientific technology as the predicate devices. Based on this and data analyzed in accordance with Osprey Medical Quality System Procedure in compliance with EN ISO 13485:2012 Medical Devices – Quality management systems - regulatory purposes and EN ISO 14791:2012 Risk management for medical devices, the proposed subject devices have been shown to be substantially equivalent under 21 CFR Part 807 subpart E.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...