K992704 · Instrumentation Laboratory CO · GGP · Nov 1, 1999 · Hematology
Device Facts
Record ID
K992704
Device Name
IL TEST VON WILLEBRAND FACTOR
Applicant
Instrumentation Laboratory CO
Product Code
GGP · Hematology
Decision Date
Nov 1, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.7290
Device Class
Class 2
Indications for Use
IL Test™ von Willebrand Factor is an in vitro diagnostic automated latex enhanced immunoassay for the quantitative determination of von Willebrand Factor Antigen (vWF:Ag) in human citrated plasma on IL Coagulation Systems. When a plasma containing vWF:Ag is mixed with the Latex Reagent and the Reaction Buffer, the coated latex particles agglutinate. The degree of agglutination is directly proportional to the concentration of vWF:Ag in the sample and is determined by measuring the decrease of transmitted light caused by the aggregates (turbidimetric immunoassay).
Device Story
IL Test™ von Willebrand Factor is an in vitro diagnostic latex-enhanced immunoassay; used for quantitative determination of vWF:Ag in human citrated plasma. Operation involves mixing patient plasma with latex reagent and reaction buffer; coated latex particles agglutinate in presence of vWF:Ag. System measures decrease in transmitted light (turbidimetric immunoassay) to determine concentration. Device is used on automated IL Coagulation Systems (e.g., ACL 6000, ACL Futura) in clinical laboratory settings; operated by trained laboratory personnel. Output provides quantitative vWF:Ag levels; assists clinicians in assessing von Willebrand Factor status. Benefits include automated, standardized quantification of vWF:Ag to support clinical diagnosis.
Clinical Evidence
Bench testing only. Method comparison study evaluated 120 citrated plasma samples (range 0.3% to 634.3% vWF:Ag) on ACL 6000 and ACL Futura systems. Results showed slopes of 1.00-1.03 and correlation coefficients (r) of 0.996-0.997 compared to predicate. Precision studies (within-run) demonstrated %CVs ranging from 1.25% to 3.51% across normal and abnormal control levels.
Technological Characteristics
Automated latex-enhanced turbidimetric immunoassay. Components: latex reagent, reaction buffer. Principle: light transmission measurement via automated coagulation analyzer. Connectivity: integrated with IL Coagulation Systems (ACL 6000, ACL Futura).
Indications for Use
Indicated for the quantitative determination of von Willebrand Factor Antigen (vWF:Ag) in human citrated plasma using automated IL Coagulation Systems. For prescription use only.
K040843 — HEMOSIL VON WILLEBRAND ACTIVITY · Instrumentation Laboratory CO · Apr 30, 2004
K133005 — HEMOSIL VON WILLEBRAND ACTIVITY ASSAY · Instrumentation Laboratory CO · Mar 13, 2014
K200033 — HemosIL von Willebrand Factor Antigen · Instrumentation Laboratory CO · Aug 19, 2020
K223402 — HemosIL von Willebrand Factor Antigen · Instrumentation Laboratory CO · Dec 9, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
## Section 3 IL Test™ von Willebrand Factor - 510(k) Summary (Summary of Safety and Effectiveness)
#### Submitted by:
Instrumentation Laboratory Company 113 Hartwell Avenue Lexington, MA 02421 Phone: 781-861-4467 781-861-4464 Fax:
#### Contact Person:
Carol Marble, Regulatory Affairs Manager Phone: 781-861-4467 / Fax: 781-861-4464
#### Summary Prepared:
August 11, 1999
#### Name of the Device:
IL Test™ von Willebrand Factor
### Classification Name(s):
| 864.7290 | Factor Deficiency Test | Class II |
|----------|-----------------------------------------------------|----------|
| 81GGP | Test, Qualitative and Quantitative Factor Deficient | |
### Identification of predicate device(s):
K860371 Asserachrom® vWF
### Description of the device/intended use(s):
IL Test™ von Willebrand Factor is an in vitro diagnostic automated latex enhanced immunoassay for the quantitative determination of von Willebrand Factor Antigen (vWF.Ag) in human citrated plasma on IL Coagulation Systems. When a plasma containing vWF:Ag is mixed with the Latex Reagent and the Reaction Buffer, the coated latex particles agglutinate. The degree of agglutination is directly proportional to the concentration of vWF:Ag in the sample and is determined by measuring the decrease of transmitted light caused by the aggregates (turbidimetric immunoassay).
### Statement of Technological Characteristics of the Device Compared to Predicate Device:
IL Test™ von Willebrand Factor is substantially equivalent to the commercially available predicate device (Asserachrom® vWF) in performance and intended use.
{1}------------------------------------------------
# Section 3 IL Test™ von Willebrand Factor - 510(k) Summary (Cont.) (Summary of Safety and Effectiveness)
### Summary of Performance Data:
In method comparison studies evaluating 120 citrated plasma samples with vWF:Ag levels ranging from 0.3% to 634.3% on an ACL 6000 and an ACL Futura, the slopes and correlation coefficients (r) for IL Test™ von Willebrand Factor versus the predicate device are shown below:
| | %vWF:Ag | |
|------------|---------|-------|
| IL System | Slope | r |
| ACL Futura | 1.03 | 0.997 |
| ACL 6000 | 1.00 | 0.996 |
Within run precision assessed over multiple runs using both three levels of control plasma gave the following results:
| | | %vWF:Ag | | |
|------------------|-----------------|---------------------|----------------------|--|
| ACL Futura | Normal<br>Level | Abnormal<br>Level I | Abnormal<br>Level II | |
| Mean | 99.50 | 79.45 | 35.54 | |
| % CV | 3.51 | 2.48 | 3.16 | |
| | | %vWF:Ag | | |
| | Normal<br>Level | Abnormal<br>Level I | Abnormal<br>Level II | |
| ACL 6000<br>Mean | 101.52 | 80.83 | 33.67 | |
| % CV | 1.41 | 1.25 | 2.18 | |
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle with three human profiles incorporated into its design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.
### NOV - 1 1999
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Carol Marble Regulatory Affairs Manager Instrumentation Laboratory Company 113 Hartwell Avenue Lexington, Massachusetts 02421
Re: K992704 Trade Name: IL Test™ von Willebrand Fractor Regulatory Class: II Product Code: GGP Dated: August 11, 1999 Received: August 12, 1999
#### Dear Ms. Marble:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D. M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use Statement
510(k) Number (if known): K992704
Device Name: IL Test™ von Willebrand Factor
### Indications for Use:
IL Test™ von Willebrand Factor is an in vitro diagnostic automated latex enhanced immunoassay for the quantitative determination of von Willebrand Factor Antigen (vWF:Ag) in human citrated plasma on IL Coagulation Systems. When a plasma containing vWF:Ag is mixed with the Latex Reagent and the Reaction Buffer, the coated latex particles agglutinate. The degree of agglutination is directly proportional to the concentration of vWF:Ag in the sample and is determined by measuring the decrease of transmitted light caused by the aggregates (turbidimetric immunoassay).
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Pete-9. Madini
**Prescription Use**
(Per 21 CFR 801.019)
✓
OR Over-The-Counter Use
Section 2
IL Test™ von Willebrand Factor 510(k)
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.