The IL HemosIL vWF Activity Kit is an in vitro diagnostic automated latex enhanced immunoturbidometric assay for the quantitative determination of vWF activity in human citrated plasma on IL Coagulation Systems.
Device Story
The HemosIL vWF Activity Kit is an in vitro diagnostic assay used on IL Coagulation Systems (ACL 8000, 9000, 10000) to quantify vWF activity in human citrated plasma. The device utilizes polystyrene latex particles coated with anti-vWF monoclonal antibodies directed against the glycoprotein Ib receptor binding site. When mixed with patient plasma, the vWF in the sample causes latex particle agglutination. The degree of agglutination is proportional to vWF activity and is measured by the decrease in transmitted light (immunoturbidometry). The assay is performed in clinical laboratory settings by trained technicians. The resulting quantitative output allows clinicians to assess vWF activity levels, aiding in the diagnosis or monitoring of vWF-related deficiencies. The system provides automated analysis, replacing manual or ELISA-based methods with a faster, automated turbidimetric approach.
Clinical Evidence
Bench testing only. Precision evaluated per NCCLS EP5-A (N=84, CVs 1.7–9.0%). Linearity confirmed (R2 0.9845–0.9941) across 21–200% range. Method comparison (N=108) against Shield Diagnostics ELISA showed r=0.9723. Comparison (N=114) against Dade Behring BC vWF:RCo reagent showed r=0.9497. Reference ranges established using 248 healthy donors stratified by blood type.
Technological Characteristics
Latex particle enhanced immunoturbidometric assay. Components: lyophilized polystyrene latex particles coated with anti-vWF monoclonal antibody (hybridoma RFF-VIII:R/1) and Tris buffer. Energy source: optical detection (transmitted light) via IL Coagulation Systems. Calibration traceable to NIBSC WHO/ECBS 4th International Standard (Code: 97/586).
Indications for Use
Indicated for the quantitative determination of von Willebrand Factor Activity in human citrated plasma for patients requiring assessment of VWF activity levels. For professional in vitro diagnostic use on IL Coagulation Systems.
Predicate Devices
Shield Von Willebrand Factor Activity ELISA (K000398)
Related Devices
K992704 — IL TEST VON WILLEBRAND FACTOR · Instrumentation Laboratory CO · Nov 1, 1999
K133005 — HEMOSIL VON WILLEBRAND ACTIVITY ASSAY · Instrumentation Laboratory CO · Mar 13, 2014
K200033 — HemosIL von Willebrand Factor Antigen · Instrumentation Laboratory CO · Aug 19, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
APR 30 2004
K040843
## Section 3 HemosIL von Willebrand Factor Activity 510(k) Summary (Summary of Safety and Effectiveness)
### Submitted by:
Instrumentation Laboratory Co. 113 Hartwell Avenue Lexington, MA 02421 Phone: 781-861-4467 781-861-4207 Fax:
### Contact Person:
Carol Marble, Regulatory Affairs Director Phone: 781-861-4467 / Fax: 781-861-4207
### Summary Prepared:
March 31, 2004
### Name of the Device:
HemosIL von Willebrand Factor Activity
### Regulatory Information:
| Regulation Section: | Factor Deficiency Test (864.7290) |
|---------------------|-----------------------------------|
| Classification: | Class II |
| Product Code: | GGP |
| Panel: | Hematology |
### Identification of predicate device(s):
K000398 Shield Von Willebrand Factor Activity ELISA
### Description of the device/intended use(s):
HemosIL von Willebrand Factor Activity is an automated latex enhanced immunoassay intended for the in vitro diagnostic quantitative determination of von Willebrand Factor Activity (VWF Activity) in human citrated plasma on IL Coagulation Systems.
The VWF Activity kit is a latex particle enhanced immunoturbidimetric assay to quantify VWF Activity in plasma. The activity of VWF is determined by measuring the increase of turbidity produced by the agglutination of the latex reagent. A specific anti-VWF monoclonal antibody adsorbed onto the latex reagent, directed against the platelet binding site of VWF (Glycoprotein Ib receptor), reacts with the VWF of patient plasma. The degree of agglutination is directly proportional to the activity of VWF in the sample and is determined by measuring the decrease of transmitted light caused by the aggregates.
### Statement of Technological Characteristics of the Device Compared to Predicate Device:
HemosIL von Willebrand Factor Activity is substantially equivalent to the commercially available predicate device (Shield Von Willebrand Factor Activity ELISA) in performance and intended use.
{1}------------------------------------------------
## Section 3 (Cont.) HemosIL von Willebrand Factor Activity 510(k) Summary (Summary of Safety and Effectiveness)
### Summary of Performance Data:
### Method Comparison
In separate method comparison studies on an ACL 9000 (n=108) using citrated plasma samples ranging from 0 to 192% VWF activity, the correlation statistics for HemosIL von Willebrand Factor Activity versus the predicate device are shown below:
| IL System | % VWF Activity<br>Slope | Intercept | r |
|-----------|-------------------------|-----------|-------|
| ACL 9000 | 0.832 | 4.782 | 0.972 |
#### Precision
Within run and total precision assessed over multiple runs using both three levels of control plasma gave the following results:
| ACL 9000: | Mean (% VWF Activity) | CV% (Within run) | CV% (Total) |
|-------------------------------|-----------------------|------------------|-------------|
| Normal Control | 79.6 | 2.7 | 4.9 |
| Special Test Controls Level 1 | 49.8 | 4.6 | 9.0 |
| Special Test Controls Level 2 | 27.7 | 7.5 | 8.7 |
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the three levels of government: federal, state, and local. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
APR 3 0 2004
Ms. Carol Marble Director, Regulatory Affairs Instrumentation Labortory Co. 113 Hartwell Avenue Lexington, MA 02421
Re: k040843
> Trade/Device Name: HemosIL von Willebrand Factor Activity Regulation Name: Factor deficiency test Regulation Number: 21 CFR 864.7290 Regulatory Class: Class II Product Code: GGP Dated: March 31, 2004 Received: April 1, 2004
Dear Ms. Marble:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{3}------------------------------------------------
Page 2
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Joseph L. Arblett
Joseph L. Hackett, Ph.D. Acting Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use Statement
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: HemosIL von Willebrand Factor Activity
### Indications for Use:
HemosIL von Willebrand Factor Activity is an in vitro diagnostic automated latex enhanced immunoassay intended for the quantitative determination of von Willebrand Factor Activity (VWF Activity) in human citrated plasma on IL Coagulation Systems by immunoturbidity.
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
| 510(k) | K040843 |
|---------------------------------------|-------------------------|
| Prescription Use (Per 21 CFR 801.019) | OR Over-The-Counter Use |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.