K962968 · Behring Diagnostics, Inc. · GGP · Dec 13, 1996 · Hematology
Device Facts
Record ID
K962968
Device Name
PROC APC ASSAY, PROC CONTROL PLASMA
Applicant
Behring Diagnostics, Inc.
Product Code
GGP · Hematology
Decision Date
Dec 13, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.7290
Device Class
Class 2
Indications for Use
ProC APC is intended for the determination of the activated Protein C sensitivity of the activated partial thromboplastin time (aPTT) in citrated human plasma.
Device Story
ProC APC assay; in vitro diagnostic reagent kit. Inputs: citrated human plasma samples. Principle: measures activated partial thromboplastin time (aPTT) in presence and absence of activated Protein C. Components: liquid aPTT reagent (phospholipids, silica activator), lyophilized activated Protein C reagent, lyophilized control plasma. Used in clinical laboratory settings by trained technicians. Output: quantitative/qualitative assessment of APC sensitivity. Clinical utility: aids in identifying Factor V Leiden carriers or abnormal APC sensitivity states; supports coagulation disorder diagnosis.
Clinical Evidence
Comparative study of 116 plasma samples (22 Factor V Leiden carriers, 94 normal donors) against predicate. ProC APC correctly identified all 116 samples. Precision studies per NCCLS EP5: within-run precision 1.7-2.9%; total precision 2.5-7.3%.
Technological Characteristics
In vitro diagnostic reagent kit. Components: liquid aPTT reagent (phospholipids, silica-based activator), lyophilized activated Protein C reagent, lyophilized control plasma. Manual or automated coagulation analyzer compatible. No software or electronic components.
Indications for Use
Indicated for the determination of activated Protein C sensitivity in citrated human plasma to assess for Factor V Leiden or other causes of abnormal APC sensitivity.
Predicate Devices
Chromogenix Coatest APC Resistance-C
Related Devices
K260551 — HemosIL Factor V Leiden (APC Resistance V) · Instrumentation Laboratory (IL) Co. · Mar 20, 2026
K983898 — BIOCLOT APC SENSITIVITY · Biopool Intl., Inc. · Jan 6, 1999
K083729 — HEMOCLOT QUANTI- V-L, FACTOR V-L CALIBRATOR, BIOPHEN V-L CAL (UNDILUTED), BIOPHEN ACT. PC-R CONTROL PLASMA AND BIOPHEN · Aniara Diagnostica, LLC · May 8, 2009
K981709 — COAGULATION FACTOR V DEFICIENT PLASMA · Dade Behring, Inc. · Oct 13, 1998
Submission Summary (Full Text)
{0}
KAC29G8
DEC 13 1996
BEHRING
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS For ProC APC
## 1. Manufacturer and Contact Information:
**Manufacturer:** Behringwerke AG
Postfach 1140
35001 Marburg, Germany
**Contact Information:**
Paul Rogers
Behring Diagnostics Inc.
3403 Yerba Buena Road
P.O. Box 49013
San Jose, CA 95161-9013
Tel: 408-239-2000
## 2. Device Classification Name:
The ProC APC is a Class II device and has a classification name of partial thromboplastin time test (21 CFR §864.7925).
## 3. Intended Use:
ProC APC is intended for the determination of the activated Protein C sensitivity of the activated partial thromboplastin time (aPTT) in citrated human plasma.
## 4. Device Description and Characteristics:
The ProC APC assay consists of a liquid aPTT Reagent (containing phospholipids, silica-based activator), a lyophilized activated Protein C time (APCT) Reagent (containing activated human Protein C), and lyophilized Control Plasma (in the pathological range). This is similar to the Chromogenix Coatest APC Resistance-C (Coatest APC) assay, the predicate device, which consists a liquid aPTT reagent (phospholipids, silica-based activator), a lyophilized APC/CaCl2 reagent (activated human Protein C), and two lyophilized Control Plasmas (for normal and pathological ranges).
**Comparative Analysis:** A total of 116 plasma samples (22 Factor V Leiden carriers, 94 normal blood donors) were tested by the ProC APC and the Coatest APC Resistance-C. The ProC APC correctly identified the 116 samples as either having normal or abnormal (Factor V Leiden) APC sensitivities. The Coatest APC identified 20 of the 22 Factor V Leiden samples as having abnormal APC sensitivities and 92 of the 94 normal blood donor samples as having normal APC sensitivities.
**Precision:** Precision studies were performed following the NCCLS EP5 guideline using two plasma samples which were in the normal and abnormal APC sensitivity ranges. The within-run precision ranged from 1.7 to 2.9%. The total precision ranged from 2.5 to 7.3%.
## 5. Substantial Equivalence:
Behring Diagnostics Inc. considers the ProC APC to be substantially equivalent to the Chromogenix Coatest APC Resistance-C in terms of intended use, reagent composition, and overall performance characteristics.
Syva® and Behring
diagnostic products
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.