HEMOCLOT QUANTI- V-L, FACTOR V-L CALIBRATOR, BIOPHEN V-L CAL (UNDILUTED), BIOPHEN ACT. PC-R CONTROL PLASMA AND BIOPHEN
Applicant
Aniara Diagnostica, LLC
Product Code
GGW · Hematology
Decision Date
May 8, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.7925
Device Class
Class 2
Indications for Use
HEMOCLOT Quanti VL is an in vitro diagnostic test for the quantitative determination of Factor V Leiden (FVL) activity in human citrated plasma, using automated or manual methods, by its resistance to the action of Activated Protein C ( APC). Patients with the Factor V-L (mutalion R506Q), are exposed to an increased thrombotic risk. Biophen V-L CAL (undilated), at defined Factor V-Leiden (FV-L) concentrations, for the callibration of Factor V-Leiden aclivity on human plasma, using the Hemoclot Quanti V-L kit. Factor V-L Cafibrator, pre diluted (1:20) human plasmas. at defined Factor V-Leiden (FV-L) concentrations, for the calibration of Factor V-L activity on human plasma using Hemoclot Quanti V-L Kit. Biophen Act PC-r Control Plasma contains human plasma, presenting an Activated Protein C Resistance (APC-R) is used as quality control plasma for the testing of Activated Protein C resistance (APC-R) on human plasma using Hemoclot Quarti V-L Kit. (Updated) BIOPHEN Normal Control Plasma is a set of 12 vials of normal citrated human plasma for the quality control of some coagulation factors. The following table shows parameters, which are measured using assays from HYPHEN BioMed or from other manufacturers, and according to the package inserts: Assays Reagents Manufacturers Reference ATIII Biophen ATIII Hyphen Biomed 221102/221105 Protein C Biophen Protein C Hyphen Biomed 221202/221205 aPC resistance (FV Leiden) Hemoclot Quanti V-L Hyphen Biomed CK065K Lupus Anticoagulant DVVtest®/DVVconfirm® American Diagnostica 810/815/815L DVVtest, DVVconfirm are registered trademarks from American Diagnostica, Inc. The BIOPHEN Normal Control Plasma is tested for the absence of Lupus Anticoagulant and can be used as a negative control for this investigation. This control plasma is also tested for the absence of Activated Protein C resistance (Act PC-7). When the APTT is performed with or without Activated Protein C (APC) the ratio obtained (APTT + APC/APTT) is ≥2.00.
Device Story
HEMOCLOT Quanti V-L is an in vitro diagnostic clotting assay measuring Factor V Leiden (FVL) activity in human citrated plasma. Input: patient citrated plasma. Process: plasma mixed with clotting factors (prothrombin, fibrinogen, protein S, APC) and heparin-neutralizing substance; purified Factor Xa and phospholipids added; clotting initiated by calcium. Principle: clotting time is inversely proportional to FVL concentration; resistance to APC indicates FVL mutation. Output: quantitative FVL activity level. Used in clinical laboratories; operated by technicians/lab personnel. Results interpreted by clinicians to identify APC resistance associated with Factor V R506Q mutation. Benefits: aids in diagnosis of thrombophilic conditions.
Clinical Evidence
Bench testing only. Four-site method comparison study (n=189) compared Hemoclot Quanti V-L to predicate (Coatest APCr). Results: 94.18% total agreement, 89.74% positive agreement, 97.30% negative agreement. Comparison with molecular biology diagnosis (n=63) showed 97% agreement. Precision studies (intra- and inter-assay) yielded CV < 6%. Linearity validated between 10% and 100% FVL (r² > 0.98). Specificity confirmed with factor-deficient plasmas and samples with heparin or AVK therapy.
Technological Characteristics
Lyophilized reagents (human fibrinogen, prothrombin, Protein S, APC, Factor Xa, rabbit brain phospholipids). Clotting-based assay. Quantitative measurement via calibration curves. Compatible with automated coagulation analyzers (e.g., STAR, KC10) or manual water bath. Reagents stable at 2-8°C.
Indications for Use
Indicated for the quantitative measurement of Factor V Leiden activity in human citrated plasma to assess resistance to Activated Protein C (APC) in patients suspected of having coagulation disorders. For prescription use only.
Regulatory Classification
Identification
A partial thromboplastin time test is a device used for primary screening for coagulation abnormalities, for evaluation of the effect of therapy on procoagulant disorders, and as an assay for coagulation factor deficiencies of the intrinsic coagulation pathway.
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k083729
B. Purpose for Submission:
Bundled submission for clearance of a new Factor V-Leiden (FVL) assay and its associated calibrators and control, and modification for a currently cleared coagulation control (k043451).
C. Measurand:
Factor V-Leiden
D. Type of Test:
Clotting Assay
E. Applicant:
ANIARA DIAGNOSTICA, LLC
F. Proprietary and Established Names:
Hemoclot Quanti V-L
Factor V-L Calibrator
Biophen V-L Cal (Undiluted)
Biophen ACT PC-r Control Plasma
Biophen Normal Control
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| GGW, Test Partial | Class II | 21 CFR 864.7925 | 81 Hematology |
| Thrombin Time | | 21 CFR 864.542521 | |
| GGN, Plasma, | | CFR 862.1150 | |
| Coagulation Control | | | |
| JIT, Calibrator, | | | |
| Secondary | | | |
H. Intended Use:
1. Intended use(s):
HEMOCLOT Quanti V-L is a clotting method for measuring the Factor V Leiden (FVL) activity in human citrated plasma, by its resistance to the action of Activated Protein C (APC).
Factor V-L Calibrator is lyophilized, pre-diluted (1:20) human plasmas, at defined Factor V-Leiden (FV-L) concentrations, for the calibration of Factor V-L activity quantitative clotting assay on human citrated plasma, using the HEMOCLOT Quanti. V-L Kit (ref CK065K).
BIOPHEN V-L CAL (Undiluted) is lyophilized, undiluted human plasmas, at defined Factor V-Leiden (FV-L) concentrations for the calibration curve of Factor
{2}
V-L activity quantitative clotting assay on human citrated plasma, using HEMOCLOT Quanti V-L Kit (ref CK065K).
**BIOPHEN Act PC-r Control Plasma** kit contains human plasma, presenting an activated Protein C Resistance (APC-R), usually correlated with the genetic mutation of Factor V R506Q. This plasma is used as quality control plasma for the testing of Activated Protein C Resistance (APC-R).
**BIOPHEN Normal Control Plasma** is a set of 12 vials of normal citrated human plasma for the quality control of some coagulation factors (ATIII, Protein C, aPC resistance FV Leiden, Lupus Anticoagulant).
2. **Indication(s) for use:**
Same as Intended Use.
3. **Special conditions for use statement(s):**
For Prescription use only.
4. **Special instrument requirements:**
Sigma Diagnostics KC 1A (k955724)
Sigma Diagnostics AMAX 400 (k972260)
American Bioproducts STA (k942117)
American Bioproducts STA-R (k082675)
Manual (Water Bath)
I. **Device Description:**
**HEMOCLOT Quanti V-L** is an in vitro diagnostic kit containing two reagents to measure the Factor V Leiden concentration in citrated plasma, by its resistance to the Activated Protein C. Reagent 1 is a clotting mixture containing human Fibrinogen, human Prothrombin, and Protein S at a constant concentration, optimized for the assay, and human Activated Protein C, lyophilized. It also contains a heparin neutralizing substance. Reagent 2 is Purified Human Factor Xa, containing rabbit brain phospholipids (cephalin), lyophilized. Calibrators and controls are required and sold separately.
**FVL Calibrator** kit: 12 vials (3 sets of 4 vials) of 1 mL of prediluted human plasma (1:20) at different concentrations of FV-L, ranging from about 10% to 100%.
**Biophen V-L Cal (Undiluted)** kit: 9 vials (3 sets of 3 vials) of 0.5 mL of undiluted human plasma at different concentrations of FV-L, to cover the assay range, from about 10% to 100%. (3 vials for each concentration, 4 points for the calibration curve).
**Act. PC-r Control Plasma** kit: 12 vials of 0.5 mL of human plasma, presenting an APC-R, citrated and lyophilized.
**BIOPHEN Normal Control Plasma**: 12 vials of 1 mL of normal citrated human plasma, lyophilized.
2
{3}
J. Substantial Equivalence Information:
| | New Device | Predicate Device |
| --- | --- | --- |
| | HEMOCLOT Quanti V-L, Factor V-L Calibrator, Biophen V-L CAL (undiluted), Biophen Act PC-r Control Plasma, Biophen Normal Control Plasma | COATEST APC Resistance V (K963111). Control Plasmas Level 1 and Level 2 are included in this device kit. |
| | Similarities | |
| Intended Use | Measurement of Factor V Leiden activity in human citrated plasma by its resistance to the action of Activated Protein C. | Same |
| Assay Type | Clotting Method | Same |
| Test Sample | Human Citrated Plasma | Same |
| Assay Principle | Measurement of Factor V Leiden coagulation activity based on insensitivity of Factor V Leiden to the action of Activated Protein C. | Same |
| Stability of Unopened Reagents | Stable at 2-8°C until expiry date | Same |
| Controls, Normal | Citrated human plasma, lyophilized, to provide quality control in the normal range. APC V Ratio > 2. | Same |
| Controls, Abnormal (APC Resistant) | Citrated human plasma, lyophilized, to provide quality control for APC resistant range of the assay. APC V Ratio < 1.80 | Same |
| | Differences | |
| Assay Principle | Assay is performed in the presence of Activated Protein C and Protein S (one single test for each patient). In the presence of APC, prolongation of clotting time is inversely related to the amount of Factor V Leiden. Normal Factor V is not measured. APC resistance due to Factor V Leiden is indicated when the %FVL value is above or equal to the cut-off value. | Assay is performed in the presence and absence of Activated Protein C (two tests for each patient). In the presence of APC, prolongation of clotting time is directly related to the concentration of Normal Factor V, and inversely related to the amount of Factor V Leiden. APC resistance due to Factor V Leiden is indicated when the APC-V ratio is below or equal to the cut-off value. |
| Assay Calibration | Assay of calibrator plasma at defined Factor V Leiden concentrations. Calibrator plasmas available in prediluted (FVL Calibrator) and undiluted (BIOPHEN V-L Cal) form, lyophilized. | Not Applicable. |
{4}
4
# K. Standard/Guidance Document Referenced (if applicable):
| STANDARDS |
| --- |
| Title and Reference Number |
| |
Other Standards
| GUIDANCE | | | |
| --- | --- | --- | --- |
| Document Title | Office | Division | Web Page |
| | | | |
# L. Test Principle:
The Hemoclot Quanti V-L kit is a clotting method triggered by purified Factor Xa, containing Phospholipids, and Calcium, and is performed on diluted test plasma, in presence of Activated Protein C and Protein S. In the first step, diluted plasma is mixed with purified clotting Factors [clotting mixture containing Prothrombin, Fibrinogen, Protein S and Activated Protein C (APC)], in a constant and optimized concentration. Then, the purified Factor Xa in the presence of Phospholipids (PLP) is added. Clotting is initiated by the addition of Calcium (Ca2+). The clotting time is then recorded. Clotting time measured is inversely proportional to the concentration of factor V-L. There is an inverse linear relationship, on a lin-log graph paper, between the factor V-L concentration and the corresponding clotting time.
FVL Calibrator and Biophen V-L Cal (Undiluted) are two calibrators that may be used with the HEMOCLOT Quanti.V-L kit. Biophen V-L CAL (Undiluted) is lyophilized undiluted human plasma containing different concentrations of FV-L, to cover the assay range, from about 10% to 100%. FVL Calibrator is lyophilized prediluted human plasma (1:20) at different concentrations of FV-L, to cover the assay range, from about 10% to 100%.
Biophen Normal Control Plasma and Biophen Act. PC-r Control Plasma are quality control plasmas which may be used with the kit.
# M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Two lots of Hemoclot Quanti V-L were analyzed for inter-assay precision. In one study the assay was performed with the waterbath, over four days using the internal reference calibrator at 4 different concentrations (10%, 25%, 50%, 100%) of FVL. Normal Control and the Biophen APCr control were used as controls. In another study using the waterbath, 9 vials of reagent was assayed at the 100% FVL level by several technicians. Inter-assay precision was also demonstrated using the KC10 analyzer by performing 10 assay runs over
{5}
several days using 2 technicians. The calibration curve was run, which represents the entire range of the assay, 10%-100% FVL.
Intra-assay precision was demonstrated on the waterbath by assaying 10 reagent vials from the same lot at the 100% FVL Level, in the single run.
All analyses yielded CV<6%.
Additional precision testing was performed at the medical decision points (10%, 25%) on the KC10 (10 runs), STA-R (10 runs), and waterbath (5 runs). Two samples were assayed over 3 days using 3 technicians. For each run a reference calibration curve and Normal Control and Biophen APCr was performed.
b. Linearity/assay reportable range:
The linearity of the Hemoclot Quanti V-L assay was evaluated by use of FVL Calibrator plasmas at 4 different concentrations of FVL, 10%, 25%, 50%, and 100%. The linearity of the assay was validated on the waterbath and several commercial analyzers (KC10, STA, STAR and Amax 400). The assay is linear between 10% and 100% FVL, r²>0.98. The linearity of the assay below 10% is not determined. Any value below 10% is considered normal. Values between 25% and 75% are expected for heterozygous plasmas. Values >75% are expected for homozygous plasmas. Values between 10 -25% are considered inconclusive.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The Internal Reference Standard used for assigning values to the calibrators and controls is traceable to pooled normal and APCr human plasmas, controlled for normal Factor V content against commercially available reference plasma pools, and verified against an NIBSC standard for normal Factor V (SSC/ISTH Secondary Coagulation Standard).
The calibration curve for the Hemoclot Quanti V-L kit is established with either Biophen V-L cal (undiluted) or Factor V-L Calibrator kits. By definition, normal plasma contains 0% FVL and heterozygous plasma, from patients carrying the R506Q mutation, contains 50% FVL. For the Biophen V-L cal (undiluted) kit, calibrators are made from mixtures of pooled heterozygous plasmas and pooled normal plasmas to yield calibration plasmas at concentrations of 10%, 25% and 50%. For the Factor V-L Calibrator kit, calibrators are made from mixtures of pooled heterozygous plasmas and pooled normal plasmas and diluted appropriately to yield calibration plasmas at concentrations of 10%, 25%, 50% and 100% FVL, ready to use with the Hemoclot Quanti V-L kit. The concentration of each calibrator level is accurately defined, against an internal reference standard, as the mean value from at least 10 vials tested in the same series by manual and/or automated methods, using the Hemoclot Quanti V-L Assay.
5
{6}
Biophen Act. PC-r Control Plasma and Biophen Normal Control Plasma (510k #043451) may be used for quality control with the Hemoclot Quanti V-L kit. Biophen Act PC-r Control Plasma is derived from pooled heterozygous plasma, from patients carrying the R506Q mutation, and contains about 50% FVL. Biophen Normal Control Plasma is derived from pooled normal plasmas and contains about 0% FVL. The FVL concentration of each control is accurately defined, against an internal reference standard, as the mean value of at least 3 vials tested in at least 3 independent series (i.e., at least 9 vials tested) by manual and/or automated methods, using the Hemoclot Quanti V-L Assay.
Stability: Reconstituted reagents are stable for 24 hrs at 2-8°C or 8 hrs at room temperature (18-25°C). Overheating studies support 30 month expiration dating for lyophilized reagents stored at 2-8°C.
d. Detection limit:
Not applicable
e. Analytical specificity:
The specificity of the assay was demonstrated by analyzing factor deficient plasmas. Results from Factor deficient plasmas (V, X, II, PS, PC) showed no interference. Since the assay can be performed in patients with heparin therapy (up to 1 IU/mL), plasma samples spiked with heparin were tested and showed no significant interference as the result of a heparin neutralizing substance in Reagent 1. To determine whether the assay can be performed in patients with Anti-Vitamin K (AVK) therapy, plasma samples from APCr patients with AVK therapy were tested and yielded expected values.
f. Assay cut-off:
The normal cut-off range was determined and validated by analyzing normal patient plasmas and aPCr patient plasmas, carrying R506Q mutation (FVL) identified with molecular technology as Heterozygous (HTZ) or Homozygous (HMZ). Analyses were conducted on water bath, KC10 and STAR analyzers. Normal plasma is expected to measure <10% FVL. Plasmas from patients with Factor V Leiden are expected to measure between 25% and 75% FVL (heterozygous) or >75% (homozygous).
Using the Hemoclot Quanti V-L method, the normal plasma range was determined by analyzing 20 normal plasma samples from a French blood bank on three different instruments with normal and APCr Controls and a selection of ActPCr plasma samples. To verify the cut-off, 42 plasma samples from FVL/HTZ patients out of therapy, 20 plasma samples from FVL/HTZ patients under Anti Vitamin K (AVK) therapy and 18 plasma samples from FVL/HMZ patients were analyzed. Results were acceptable with normal samples measuring <10% FVL, 25% and 75% FVL for FVL/HTZ and >75% FVL for FVL/HMZ samples.
2. Comparison studies:
a. Method comparison with predicate device:
A four site (two in the US and two in Europe) comparison study was
{7}
conducted. 189 plasma samples were analyzed using the test (Hemoclot Quanti V-L) and predicate (Coatest APCr) method. Samples were selected from patients who were previously tested for coagulation disorders and represented both normal and APC-Resistant populations (about 60% of the samples were from normal patients and 40% from APCr patients). Patient plasma samples were selected to cover the range of the assay. Overall, there was 94% agreement between the two methods for determining APC resistance due to the Factor V Leiden mutation. In one study, samples were identified as heterozygous (25), homozygous (11), or normal (27) with regard to FVL mutation using molecular biology methods. There was 97% agreement between the results of test method and the molecular biology diagnosis with regard to APCr resistance.
| All Sites | Predicate | | |
| --- | --- | --- | --- |
| | | Normal | Abnormal |
| Hemoclot Quanti V-L | Normal | 108 | 7 |
| | Abnormal | 1 | 70 |
| | Inconclusive* | 2 | 1 |
*Outside the range of the Hemoclot Quanti V-L
Percent Total Agreement 94.18% (178/189)
Percent Positive Agreement 89.74% (70/78)
Percent Negative Agreement 97.30% (108/111)
b. Matrix comparison:
Not applicable.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
Normal plasma is expected to measure <10% FVL. Plasmas from patients with Factor V Leiden are expected to measure between 25% and 75% FVL (heterozygous) or >75% (homozygous).
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.