COATEST APC (TM) RESISTANCE V/COATEST APC (TM) RESISTANCE V S

K963111 · Chromogenix AB · GGW · Dec 23, 1996 · Hematology

Device Facts

Record IDK963111
Device NameCOATEST APC (TM) RESISTANCE V/COATEST APC (TM) RESISTANCE V S
ApplicantChromogenix AB
Product CodeGGW · Hematology
Decision DateDec 23, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.7925
Device ClassClass 2

Indications for Use

The COATEST® APC™ RESISTANCE V / COATEST® APC™ RESISTANCE V S is intended for the diagnosis of the APC resistance phenotype.

Device Story

COATEST® APC™ RESISTANCE V/V S is an in vitro diagnostic assay for detecting APC resistance phenotype, which is associated with factor V mutation (FV:Q506). The device uses plasma samples from patients. The principle of operation is an APTT-based clotting assay utilizing a supplementary component, V-DEF Plasma, to increase sensitivity to the factor V mutation. The assay is performed in clinical laboratory settings using automated coagulation analyzers (e.g., Thrombolyzer, MLA/Electra 900C, ACL 300, ST-4). The device produces an APC ratio based on clotting times. Healthcare providers use this ratio to discriminate between normal, heterozygous, and homozygous factor V genotypes. This aids in identifying patients at risk for deep venous thrombosis or pulmonary emboli, allowing for improved clinical management of coagulation disorders.

Clinical Evidence

Clinical validation included studies on 218 acutely ill patients and 399 plasmapheresis donors, demonstrating high discrimination for FV:Q506. Additional studies on patients with myocardial infarction (n=97), healthy controls (n=132), thrombotic patients (n=50), and patients on OAC/heparin therapy (n=147/61 and n=129/62) confirmed performance across various instruments. The device demonstrated >99% sensitivity for FV:Q506. Repeatability and reproducibility were comparable to the predicate. Testing confirmed correct classification of factor V genotypes in patients with phospholipid antibody syndrome and those receiving heparin up to 1 IU/mL.

Technological Characteristics

In vitro diagnostic reagent kit for APTT-based clotting assays. Includes V-DEF Plasma component. Operates on automated coagulation analyzers (e.g., Thrombolyzer, MLA/Electra 900C, ACL 300, ST-4) using various clot detection principles. Designed for plasma analysis. No specific materials of construction or software algorithm class (beyond standard instrument-based clot detection) are specified.

Indications for Use

Indicated for patients with suspected venous thrombosis or pulmonary emboli, including those on oral anticoagulant therapy, heparin (unfractionated or LMW), or with phospholipid antibody syndrome, to detect APC resistance phenotype and factor V mutation (FV:Q506).

Regulatory Classification

Identification

A partial thromboplastin time test is a device used for primary screening for coagulation abnormalities, for evaluation of the effect of therapy on procoagulant disorders, and as an assay for coagulation factor deficiencies of the intrinsic coagulation pathway.

Predicate Devices

Submission Summary (Full Text)

{0} CHROMOGENIX AB K9a3111 DEC 23 1996 # 510(k) SUMMARY This summary of 510(k) safety and effectiveness is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. APC resistance is a newly detected abnormality in hemostasis, which appears to be prevalent in 20 % to 50 % of patients with deep venous thrombosis. This abnormality appears to be at least 10 times more common than inherited deficiencies of any of the other anticoagulant proteins. APC resistance reflects an impairment of the expression of activated protein C (APC) activity in an individual's plasma and will result in a reduced ability to stop thrombin formation. The original device COATEST® APC™ RESISTANCE C / COATEST® APC™ RESISTANCE SC received 510(K) clearance for diagnosis of the APC resistance phenotype on August 8, 1995 (K945940). The new device COATEST® APC™ RESISTANCE V / COATEST® APC™ RESISTANCE V S contains a supplementary component, (factor) V-DEF Plasma. The present documentation intends to show the applicability of V-DEF Plasma in a modification of the original COATEST® APC™ Resistance kit and the sensitivity for the mutation in coagulation factor V is considerably increased. 90 to 95% of all cases with APC resistance are explained by a mutation in the factor V protein rendering the activated form of factor V less susceptible to degradation by APC. The documentation shows the applicability of V-DEF Plasma in a modification of the original COATEST® APC™ RESISTANCE kit with a considerably increased sensitivity for the mutation in factor V. COATEST® APC™ RESISTANCE V or V S test is based on the use of activated protein C in an APTT clotting assay, as was the original test kit. Two studies support the substantially equivalency of COATEST® APC™ RESISTANCE V or V S (modified method) with COATEST® APC™ RESISTANCE (original method, K945940). The first study shows a correlation (R= 0.74) between the two methods and their discrimination for FV:Q⁵⁰⁶ on analysis of plasma from 218 acutely ill patients with suspected venous thrombosis or pulmonary emboli and from family members. The second study showed a similar correlation and discrimination for FV:Q⁵⁰⁶, noted above, between the two methods on analysis of plasma from 399 plasmapheresis donors. Thus, the data supports COATEST® APC™ RESISTANCE V or V S Substantial Equivalency to the predicated device COATEST® APC RESISTANCE C (K945940) and there is a high discrimination for FV:Q⁵⁰⁶. The new method, COATEST® APC™ RESISTANCE V or V S, correlated well with regard to cases with thrombosis and with homo- or heterozygous factor V mutation. All showed pronounced APC resistance. The homozygous cases were properly discriminated from the heterozygous cases which were separated from the rest. The safety and efficacy of COATEST® APC™ RESISTANCE V or V S is further supported by studies which show that this kit can be utilized on instruments with different clot detection principles and that appropriate results are obtained on analysis of plasmas from patients receiving oral anticoagulants, normal or low molecular weight (LMW) heparin or patients with the phospholipid antibody syndrome. Thrombolyzer and MLA/Electra 900C, were utilized for samples of plasma from patients with myocardial infarction (97) and from healthy controls (132) in the former instrument; and on plasma from thrombotic patients or family members (50), in the latter instrument. The factor V gene nucleotide at XI {1} # CHROMOGENIX AB # 510(k) Application ## COATEST® APC™ RESISTANCE position 1691 was determined in almost all individuals. A clear and complete discrimination is obtained for FV:Q$^{506}$ by both instruments and an improvement is obtained as compared to the original kit method. This is illustrated for the patient group. Evaluation of the 50 family members by the MLA/Electra 900C showed that for normal Factor V genotype an APC ratio of 2.4 - 3.4 was obtained, for the Heterozygous genotype 1.6 - 1.9 and for the Homozygous 1.2 to 1.3. **COATEST® APC™ RESISTANCE V or V S** was used for analysis of plasma from 147 patients on OAC therapy and 61 healthy controls by the ACL 300 and of plasma from 129 patients on OAC therapy and 62 healthy controls by the ST-4 instrument. The factor V gene nucleotide at position 1691 was determined in all. Results show a clear discrimination obtained for the FV:Q$^{506}$ mutation on both ACL and ST-4. Similar ranges of APC ratios are obtained for healthy controls vs patients on OAC, for individuals with a normal factor V genotype, and with heterozygosity and homozygosity for FV:Q$^{506}$, respectively. Results from the ACL and ST-4 instruments for plasmas from patients on unfractionated and low molecular weight (LMW) heparin at concentrations ≤ 1 IU/mL show that in contrast to the original method **COATEST® APC™ RESISTANCE V or V S** provides normal APT times to at least 1 IU/mL of normal heparin. All samples from factor V genotyped patients receiving normal heparin were correctly classified as well as all of the genotyped LMW samples. The data also supports that a correct classification with regard to the factor V genotype is obtained for patients with the phospholipid antibody syndrome. **COATEST® APC™ RESISTANCE V or V S** provides a similar repeatability and reproducibility as Coatest APC Resistance. Similar results were obtained on the four instruments (Thrombolyzer, MLA/Electra 900, ST-4, ACL 300) included in the studies. Calibration and Determination of the Cut-Off Value for the APC Ratio was performed with the **COATEST® APC™ RESISTANCE V or V S** with three instruments in a population of normal factor V genotype males and females. The results illustrate that a significantly narrower range is obtained with **COATEST® APC™ RESISTANCE V or V S** compared to the original Coatest APC Resistance kit. Further, based on the data presented and analyzed in the same manner as the original Kit, the results support the fact that the **COATEST® APC™ RESISTANCE V or V S** has a greater than 99% sensitivity for FV:Q$^{506}$. Reagent V-DEF Plasma and the stability of other reconstituted reagent was determined through the effect of storage at different temperatures on the APT time and the APC ratio on the analysis of plasmas containing factor V of different genotype origin. The above noted data support the assigned stabilities in the package insert of reconstituted V-DEF plasma. Further, a similar stability is obtained for 3 mL and 4 mL vials. Further, the data supports that whole blood and plasma can be stored for at least 7 hours at room temperature. ![img-0.jpeg](img-0.jpeg) RONALD G. LEONARDI, Ph.D. August 9, 1996
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...