KC 40 MICRO COAGULATION ANALYZER SYSTEM (K3632)
Device Facts
| Record ID | K960875 |
|---|---|
| Device Name | KC 40 MICRO COAGULATION ANALYZER SYSTEM (K3632) |
| Applicant | Sigma Diagnostics, Inc. |
| Product Code | JPA · Hematology |
| Decision Date | Sep 10, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 864.5425 |
| Device Class | Class 2 |
Indications for Use
The Amelung KC 40 Micro Coagulation Analyzer is an automated random access multipurpose analyzer. The KC 40 can be used for the detection of fibrin formation utilizing mechanical principles (ball method) to perform clot based tests such as prothrombin time (PT), activated partial thromboplastin time (APTT), fibrinogen, factor assays, and other clotting tests.
Device Story
Automated random access coagulation analyzer; utilizes mechanical ball method to detect fibrin formation; performs PT, APTT, fibrinogen, and factor assays; used in clinical laboratory settings by trained personnel; provides quantitative clotting test results; assists clinicians in assessing patient coagulation status and managing anticoagulant therapy.
Clinical Evidence
Bench testing only. Comparison studies performed against TOA Medical Electronics CA-5000 and BBL Fibrometer. PT, APTT, Fibrinogen, Factor IX, and Factor X assays evaluated. Regression analysis showed high correlation (r=0.915 to 0.980). Precision studies demonstrated coefficient of variation <5% for PT/APTT, <6% for fibrinogen, and <10% for factor assays.
Technological Characteristics
Automated mechanical coagulation analyzer; utilizes 'ball method' for fibrin detection; random access architecture; designed for clinical laboratory environment.
Indications for Use
Indicated for clinical laboratory use for the detection of fibrin formation to perform clot-based coagulation assays including PT, APTT, fibrinogen, and factor assays.
Regulatory Classification
Identification
A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.
Special Controls
*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
Predicate Devices
- TOA Medical Electronics CA-5000
- BBL Fibrometer
Related Devices
- K955720 — AMAX CS-190 COAGULATION ANALYZER SYSTEM · Sigma Diagnostics, Inc. · Mar 13, 1996
- K021162 — AMAX DENSITY COAGULATION ANALYZER, AMAX DENSTINY COAGULATION ANALYZER-EU, AMAX DESTINY COAGULATION ANALYZER-REFURBISHED · Sigma Diagnostics, Inc. · Aug 30, 2002
- K955735 — KC 4A MICRO COAGULATION ANALYZER SET/SYSTEM · Sigma Diagnostics, Inc. · Mar 11, 1996
- K970431 — BEHRING COAGULATION SYSTEM · Behring Diagnostics, Inc. · May 5, 1997
- K962664 — ELECTRA 1800C AUTOMATIC COAGULATION ANALYZER · Medical Laboratory Automation Systems, Inc. · Sep 10, 1996