KC 1A COAGULATION ANALYZER/SET
Device Facts
| Record ID | K955724 |
|---|---|
| Device Name | KC 1A COAGULATION ANALYZER/SET |
| Applicant | Sigma Diagnostics, Inc. |
| Product Code | JPA · Hematology |
| Decision Date | Mar 12, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 864.5425 |
| Device Class | Class 2 |
Indications for Use
The Amelung KC 1 A Micro Coagulation Analyzer is a semi-automated mechanical clot detection system designed for the determination of prothrombin times (PT), activated partial thromboplastin times (APTT), fibrinogen concentrations and other clotting tests. Any clotting time test that has fibrin formation as its endpoint may be performed on the KC 1 A.
Device Story
Semi-automated mechanical clot detection system; determines PT, APTT, fibrinogen, and other clotting tests with fibrin formation endpoints. Used in physician office laboratories (POL) and clinical settings; operated by laboratory personnel. Device measures clotting time via mechanical detection; results used by clinicians to assess coagulation status and guide therapeutic decisions. Benefits include standardized, reliable coagulation monitoring.
Clinical Evidence
Performance studies conducted at three physician office laboratories (n > 45 samples) and in-house (n > 100 samples). Compared KC 1 A against reference laboratory results and predicate devices (Fibrometer, CA-5000). PT correlation (r) ranged 0.981-0.992; APTT correlation (r) ranged 0.867-0.935; Fibrinogen correlation (r) = 0.932. Precision studies showed CV < 6%.
Technological Characteristics
Semi-automated mechanical clot detection system. Benchtop form factor. No specific materials, connectivity, or software architecture details provided.
Indications for Use
Indicated for patients requiring coagulation testing, specifically PT, APTT, and fibrinogen concentration determination, where fibrin formation is the endpoint.
Regulatory Classification
Identification
A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.
Special Controls
*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
Predicate Devices
- Fibrometer
- CA-5000
Related Devices
- K955735 — KC 4A MICRO COAGULATION ANALYZER SET/SYSTEM · Sigma Diagnostics, Inc. · Mar 11, 1996
- K960875 — KC 40 MICRO COAGULATION ANALYZER SYSTEM (K3632) · Sigma Diagnostics, Inc. · Sep 10, 1996
- K955720 — AMAX CS-190 COAGULATION ANALYZER SYSTEM · Sigma Diagnostics, Inc. · Mar 13, 1996
- K021162 — AMAX DENSITY COAGULATION ANALYZER, AMAX DENSTINY COAGULATION ANALYZER-EU, AMAX DESTINY COAGULATION ANALYZER-REFURBISHED · Sigma Diagnostics, Inc. · Aug 30, 2002
- K972260 — CS-400 COAGULATION ANALYZER SYSTEM (A1208) · Sigma Diagnostics, Inc. · Oct 20, 1997