ABBOTT CELL-DYN 3200 SYSTEM

K972354 · Abbott Laboratories · GKZ · Sep 16, 1997 · Hematology

Device Facts

Record IDK972354
Device NameABBOTT CELL-DYN 3200 SYSTEM
ApplicantAbbott Laboratories
Product CodeGKZ · Hematology
Decision DateSep 16, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.5220
Device ClassClass 2

Indications for Use

The CELL-DYN ® 3200 System is a fully automated hematology analyzer intended for in-vitro diagnostic use in the clinical laboratory.

Device Story

The CELL-DYN 3200 is a fully automated hematology analyzer for clinical laboratory use. It processes K2EDTA-anticoagulated whole blood specimens. The system consists of an analyzer module for specimen aspiration, dilution, and analysis; a data module for processing; and a display station for operator interface and reporting. The device utilizes flow cytometry, laser optical scatter (Helium-Neon laser), and colorimetric methods to count, size, and classify blood cells. WBC parameters are measured via four-angle light scatter (0°, 10°, 90°, and 90°D). RBC and PLT parameters are measured via optical scatter (0° and 10°). Hemoglobin is measured via absorbance at 540nm using a cyanide-free reagent. The system automatically generates reports for hematological parameters, aiding clinicians in hematological assessment and patient diagnosis.

Clinical Evidence

Bench testing only. Performance was validated through accuracy, precision, linearity, and carryover studies comparing the CELL-DYN 3200 to predicate systems (CELL-DYN 3500R, CELL-DYN 4000, Technicon H 3 RTC). Comparison to manual microhematocrit method (NCCLS H7-A2) and reference leukocyte differential count (NCCLS H20-A) was performed to arbitrate differential parameter differences. Results demonstrated performance consistent with manufacturer specifications.

Technological Characteristics

Automated hematology analyzer; Helium-Neon laser optical source; flow cytometry, laser optical scatter, and colorimetric (540nm) measurement principles. Analyzes K2EDTA-anticoagulated whole blood. Modular design: Analyzer, Data Module, Display Station. Cyanide-free reagents used for hemoglobin measurement.

Indications for Use

Indicated for in-vitro diagnostic use in clinical laboratories to analyze K2EDTA-anticoagulated whole blood specimens to provide hematological parameters including WBC, RBC, HGB, PLT, and WBC differential counts.

Regulatory Classification

Identification

An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ SEP 1 6 1997 510(k) Notification ## 510(k) Summary Of Safety And Effectiveness Information Supporting A Substantially Equivalent Determination The following information as presented in the Premarket Notification 510(k) for the CELL-DYN® 3200 System Hematology Analyzer constitutes data supporting a substantially-equivalent determination. Substantial equivalence is demonstrated between the CELL-DYN® 3200 System and the Abbott the CELL-DYN® 3500R Multi-Parameter Automated Hematology Analyzer, #K955715 System for the White Blood Cell (WBC), Red Blood Cell (RBC), Hemoglobin concentration (HGB), Platelet count (PLT), WBC differential parameters and the calculated parameter of Mean Corpuscular Hemoglobin (MCH). Substantial equivalence is also demonstrated between the CELL-DYN® 3200 System and the CELL-DYN® 4000 System for Mean Corpuscular Volume (MCV), Red Cell Distribution Width (RDW), and Mean Platelet Volume (MPV). Substantial equivalence is demonstrated between the CELL-DYN® 3200 System and the Technicon H 3 RTC #K930148 for the hemogram and white cell (WBC) differential parameters. Additional data was collected to demonstrate the comparison of the Mean Corpuscular Volume (MCV) of the CELL-DYN® 3200 System to the manual microhematocrit method, NCCLS H7-A2. The NCCLS H20-A, "Reference Leukocyte Differential Count (Proportional) and Evaluation of Instrumental Methods," Approved Voluntary Standard 1992 for the white cell differential parameters was used to arbitrate differences in the differential. ## Intended Use: The CELL-DYN ® 3200 System is a fully automated hematology analyzer intended for in-vitro diagnostic use in the clinical laboratory. ## Device Description: The CELL-DYN® 3200 System has three main modules: the Analyzer, which identifies, mixes, and presents specimens for processing, aspirates, dilutes and analyzes each whole blood specimen. The Data Module, which controls all system processing; and the Display Station, which provides the primary operator interface with the system and generates reports automatically or on demand. {1}------------------------------------------------ | specimens and report the following hematological parameters: | | |--------------------------------------------------------------|------------------------------------------------------| | White Blood Cell Parameters: | Red Blood Cell Parameters: | | WBC -- White Blood Cell | RBC -- Red Blood Cell or erythrocyte count | | NEU -- Neutrophil absolute count | HCT -- Hematocrit | | %N -- Neutrophil percent | MCV -- Mean Corpuscular Volume | | LYM -- Lymphocyte absolute count | RDW -- Red Cell Distribution Width | | %L -- Lymphocyte percent | Hemoglobin Parameters: | | MONO -- Monocyte absolute count | HGB -- Hemoglobin concentration | | %M -- Monocyte percent | MCH -- Mean Corpuscular Hemoglobin | | EOS -- Eosinophil absolute count | MCHC -- Mean Corpuscular Hemoglobin<br>Concentration | | %E -- Eosinophil percent | Platelet Parameters: | | BASO -- Basophil absolute count | PLT -- Platelet Count | | %B -- Basophil percent | MPV -- Mean Platelet Volume | | | *PDW -- Platelet Distribution Width | | | *PCT -- Plateletcrit | The CELL-DYN® 3200 is designed to analyze K2EDTA-anticoagulated whole blood C. 11 ------------------------------------------------------------------------------------------------------------------------------------------------------------------------ * These parameters are provided for laboratory use only and are not reportable in the US ## Principles of Operation: - The analyzer counts, sizes and classifies blood cells by the combination of a flow cytometry method, laser optical scatter, and colorimetric method. The CELL-DYN® 3200 System uses a Helium-Neon laser as the optical light source. The Optical Bench detects light in the form of scatter from blood cell surfaces and internal structures. For the WBC parameters, whole blood is diluted with a reagent to preserve cell integrity. Data are simultaneously collected for four angles (0°, 10°, 90°, and 90°D) of light scatter as each cell passes through the laser beam. For the RBC and the PLT parameters, whole blood is diluted with a reagent that prepares the cells for measurement. The dilution is split and measured by laser optical scatter at (0° and 10°). For the hemoglobin parameters, whole blood is diluted with a cyanide free reagent and the hemoglobin is measured optically by absorbance (540nm). {2}------------------------------------------------ 510(k) Notification #### Similarities and Differences: The CELL-DYN® 3200 System has similarities to one or more of the methods of determination for hemogram and automated WBCs differential parameters used by the following hematology analyzers: Abbott CELL-DYN® 3000 Series, Sysmex™ NE Series, Technicon H*1™ Series, and Abbott CELL-DYN® 4000. The CELL-DYN® 3200 is similar to the Abbott CELL-DYN® 3000 Series. Sysmex™ NE Series: Technicon H*1™ Series; and Abbott CELL-DYN® 4000 in that they use optical scatter to count and differentiate WBCs. WBCs are counted and classified by the Abbott CELL-DYN® 3000 Series Systems and the CELL-DYN® 4000 System in a very similar manner using four simultaneously collected angles of scattered laser light. The CELL-DYN® 3200 is different from the Coulter® Counters (Model ZBI, S-Plus Series and STKS), and the CELL-DYN® 4000 System in that they use impedance for counting and sizing RBCs and PLTs. The CELL-DYN® 3200 and the Sysmex™ NE series are different in that the Sysmex uses Focused Flow Impedance to count and size RBCs and PLTs. The CELL-DYN® 3200 is different from the Technicon H*1™ series and the CELL-DYN® 4000 System in that they both use Injection Metering to measure RBCs and PLTs optically. The CELL-DYN® 4000 System is different in that it counts RBCs and PLTs by both the optical and impedance methods and compares the data as an internal quality check. They are different in that the Abbott CELL-DYN® 3000 Series Systems use a helium neon laser, while the CELL-DYN® 4000 System uses an Argon-ion laser to collect optical scatter data. #### Equivalency Data: The data complied to support the claim that the CELL-DYN® 3200 System is substantially equivalent to the Abbott CELL-DYN® 3500R System includes background, accuracy, precision, linearity, comprehensive flagging and carryover. Substantial equivalence is demonstrated between the CELL-DYN® 3200 System and the Abbott the CELL-DYN® 3500R System Multi-Parameter Automated Hematology Analyzer, #K955715 for the White Blood Cell (WBC), Red Blood Cell (RBC), Hemoglobin concentration (HGB), Platelet count (PLT), WBC differential parameters and the calculated parameter of Mean Corpuscular Hemoglobin (MCH). Substantial equivalence is also demonstrated between the CELL-DYN® 3200 System and the CELL-DYN® 4000 System for Mean Corpuscular Volume (MCV), Red Cell Distribution Width (RDW), and Mean Platelet Volume (MPV). Substantial equivalence is demonstrated between the CELL-DYN® 3200 System and the Technicon H 3TM RTX #K930148 for the hemogram and white cell (WBC) differential parameters. {3}------------------------------------------------ The data supports the claim that the CELL-DYN® 3200 System is substantially equivalent to the Abbott CELL-DYN® 3500R System for the White Blood Cell (WBC) count, Red Blood Cell (RBC) count, Hemoglobin concentration (HGB), Platelet (PLT) count, WBC differential parameters and the calculated parameter of Mean Corpuscular Hemoglobin (MCH). The data supports the claim that the CELL-DYN® 3200 System is substantially equivalent to CELL-DYN® 4000 System for MCV, RDW and MPV. The accuracy, precision, and linearity data shows performance to manufacturer's specifications. Substantial equivalence is demonstrated between the CELL-DYN® 3200 System and the Technicon H 3 RTC #K930148 for the hemogram and WBC differential parameters. The accuracy, precision, and linearity data shows performance to manufacturer's specifications. #### Conclusion: The CELL-DYN® 3200 System shows an evolution of the technologies used on one or more of the currently available analyzers to count, size, and classify whole blood cells and their related parameters, and more specifically to the technologies used on the Abbott CELL-DYN® 3500R System and the CELL-DYN® 4000 System. The 510(k) Summary was prepared and submitted by: Sue E. Luptovic Sr. Regulatory Affairs Specialist Abbott Diagnostics 5440 Patrick Henry Drive Santa Clara, CA 95054 > Phone: 408 567-3389 Fax: 408 982-4863 {4}------------------------------------------------ Image /page/4/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a circular design with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, resembling a bird in flight. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Ms. Sue E. Luptovic Senior Requlatory Affairs Specialist Abbott Diagnostics Requlatory Affairs Abbott Laboratories SEP 1 6 1997 100 Abbott Park Road Abbott Park, Illinois 60064 Re : K972354 Trade Name: Abbott CELL-DYN 3200 System Regulatory Class: II Product Code: GKZ Dated: June 20, 1997 Received: June 24, 1997 Dear Ms. Luptovic: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labelinq, and prohibitions aqainst misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations. {5}------------------------------------------------ Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other qeneral information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html" Sincerely yours, Steven Autman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ Page__________________________________________________________________________________________________________________________________________________________________________ | 510(k) Number (if known): | K972354 | |---------------------------|---------| |---------------------------|---------| Device Name: # Indications For Use: CELL-DYN 3200 System 510(k) Notification Indications For Use: ﻧﺎ ﺑﮭﯿﻨﯿﮉﯾﭩﺮ ﮨﮯ ﺍﻭﺭ ﺍﺱ ﮐﯽ ﻣﺠﻤﻮﻋﯽ ﺁﺑﺎﺩﯼ ﺍﻭﺭ ﺍﺱ ﮐﮯ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ The CELL-DYN® 3200 System is a fully automated hematology analyzer intended for invitro diagnostic use in the clinical laboratory. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Cita E. Madim (Division Sign-Off) Division of Clinical Laboratory D 510(k) Number Prescription Use V (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...