CELL-DYN 3500 SYSTEM WITH IMMATURE RETICULOCYTE FRACTION (IRF), CELL-DYN 3700 SYSTEM WITH IMMATURE RETICULOCYTE FRACTION

K991605 · Abbott Laboratories · GKZ · Jul 9, 1999 · Hematology

Device Facts

Record IDK991605
Device NameCELL-DYN 3500 SYSTEM WITH IMMATURE RETICULOCYTE FRACTION (IRF), CELL-DYN 3700 SYSTEM WITH IMMATURE RETICULOCYTE FRACTION
ApplicantAbbott Laboratories
Product CodeGKZ · Hematology
Decision DateJul 9, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.5220
Device ClassClass 2

Indications for Use

The CELL-DYN 3500 and the CELL-DYN 3700 Systems are automated, multiparameter hematology analyzers intended to classify the following formed elements of EDTA anticoagulated blood: WBC -- White Blood Cell or Leukocyte Count; NEU -- Neutrophil Absolute Count; %N -- Neutrophil Percent; LYM -- Lymphocyte Absolute Count; %L -- Lymphocyte Percent; MONO -- Monocyte Absolute Count; %M -- Monocyte Percent; EOS -- Eosinophil Absolute Count; %E -- Eosinophil Percent; BASO -- Basophil Absolute Count; %B -- Basophil Percent; PLT -- Platelet Count; MPV -- Mean Platelet Volume; RBC -- Red Blood Cell or Erythrocyte Count; HCT - Hematocrit; MCV -- Mean Corpuscular Volume; RDW -- Red Cell Distribution Width; HGB -- Hemoglobin Concentration; MCH -- Mean Corpuscular Hemoglobin; MCHC -- Mean Corpuscular Hemoglobin Concentration; RETC -- Reticulocyte Concentration; % RETC -- Reticulocyte Concentration; IRF -- Immature Reticulocyte Fraction; *PDW -- Platelet Distribution Width; *PCT -- Plateletcrit. * These parameters are provided for laboratory use only and are not reportable in the US.

Device Story

Automated hematology analyzers (CELL-DYN 3500/3700) process EDTA-anticoagulated whole blood. System modules include analyzer, sample loader, data station, and printer. Input: whole blood specimen; reticulocyte samples require off-line staining with New Methylene Blue. Operation: laser optical light scatter, impedance, and absorption spectrophotometry. Helium-Neon laser source detects scatter from cell surfaces/internal structures. Data station controls processing and operator interface. Output: quantitative hematology parameters including WBC, RBC, platelet, hemoglobin, and reticulocyte counts/fractions. Used in clinical laboratories by trained personnel. Results assist clinicians in hematological assessment and patient diagnosis.

Clinical Evidence

Bench testing only. Performance data including accuracy, precision, linearity, and carryover were evaluated at two clinical sites (Scottsdale Health Care, VAMC Phoenix) and one internal Abbott site. Results demonstrated performance consistent with manufacturer specifications and substantial equivalence to the predicate device.

Technological Characteristics

Multi-parameter hematology analyzer. Sensing: Laser optical light scatter (He-Ne laser), impedance, and absorption spectrophotometry. Sample handling: Automated sample loader. Data processing: Integrated data station. Connectivity: Color printer output. Software: Automated analysis of cell counts and fractions. Materials: Standard laboratory instrumentation components.

Indications for Use

Indicated for the classification of formed elements in EDTA-anticoagulated whole blood samples using automated hematology analysis. Intended for clinical laboratory use.

Regulatory Classification

Identification

An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JUL - 9 1999 ## 16. 510(k) Summary # CELL-DYN 3500 System and the CELL-DYN 3700 System with Immature Reticulocyte Fraction (IRF) | Submitted by: | Abbott Laboratories<br>5440 Patrick Henry Drive<br>Santa Clara, CA 95054 | |----------------------|----------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Michelle B. Roeding<br>Phone: 408 567-3781<br>Fax:<br>408 982-4863 | | Date Prepared: | May 7, 1999 | | Proprietary Name: | CELL-DYN 3500 System with Immature<br>Reticulocyte Fraction (IRF)<br>CELL-DYN 3700 System with Immature<br>Reticulocyte Fraction (IRF) | | Common Name: | Automated Hematology Analyzers | | Classification Name: | Automated Differential Cell Counter (21 CFR 864.5220) | | Predicate Device: | CELL-DYN 4000 System, 510(k) #971152/S1 | K991605 510(k) Notification {1}------------------------------------------------ ### 16. 510(k) Summary (cont'd) ### CELL-DYN 3500 System and the CELL-DYN 3700 System with Immature Reticulocyte Fraction (IRF) ### Intended Use The CELL-DYN 3500 and the CELL-DYN 3700 Systems are automated, multi-parameter hematology analyzers intended to classify the following formed elements of EDTA anticoagulated blood including: | White Blood Cell Parameters: | Red Blood Cell Parameters: | |--------------------------------------------|--------------------------------------------| | WBC -- White Blood Cell or Leukocyte Count | RBC -- Red Blood Cell or Erythrocyte Count | | NEU -- Neutrophil Absolute Count | HCT -- Hematocrit | | %N -- Neutrophil Percent | MCV -- Mean Corpuscular Volume | | LYM -- Lymphocyte Absolute Count | RDW -- Red Cell Distribution Width | | %L -- Lymphocyte Percent | | | MONO -- Monocyte Absolute Count | Hemoglobin Parameters: | | %M -- Monocyte Percent | HGB -- Hemoglobin Concentration | | EOS -- Eosinophil Absolute Count | MCH -- Mean Corpuscular Hemoglobin | | %E -- Eosinophil Percent | MCHC -- Mean Corpuscular Hemoglobin | | BASO -- Basophil Absolute Count | Concentration | | %B -- Basophil Percent | | | Platelet Parameters: | Reticulocyte Parameters: | | PLT -- Platelet Count | RETC -- Reticulocyte Concentration | | MPV -- Mean Platelet Volume | %R -- Reticulocyte Percent of RBC Count | | *PDW -- Platelet Distribution Width | IRF -- Immature Reticulocyte Fraction | | *PCT -- Plateletcrit | | * These parameters are provided for laboratory use only and are not reportable in the US. ### Device Description: The CELL-DYN 3500 and the CELL-DYN 3700 Systems have four main modules: 1) the Analyzer, which aspirates, dilutes and analyzes each whole blood specimen 2) the Sample Loader, which automatically identifies, mixes, and presents specimens for processing 3) the Data Station, which controls all system processing and provides the primary operator interface with the system and; 4) the Color Printer, which generates reports automatically or on demand. {2}------------------------------------------------ ### 16. 510(k) Summary (cont'd) ### CELL-DYN 3500 System and the CELL-DYN 3700 System with Immature Reticulocyte Fraction (IRF) ### Device Description (cont'd): The analyzer counts, sizes and classifies blood cells by the combination of methods: Laser Optical Light Scatter, Impedance, and Absorption Spectrophotometry. The IRF is derived from the data measured for the Reticulocyte parameters. The CELL-DYN 3500 and the CELL-DYN 3700 Systems use a Helium-Neon laser as the optical light source. The Optical Bench detects light in the form of scatter from blood cell surfaces and internal structures. For the reticulocyte parameters, an off-line dilution of blood and Reticulocyte Reagent is prepared and stained for 15 minutes. The dilution is aspirated and the reticulocytes are counted in the WOC channel. Data are collected for scatter (0°, 10°, and 90°) as each cell passes through the laser beam. The CELL-DYN 3500 and the CELL-DYN 3700 Systems report Reticulocyte % and Absolute Count. The IRF is a fraction of the total reticulocyte absolute count. The CELL-DYN 3500 and the CELL-DYN 3700 Systems are designed to analyze EDTA-anticoagulated whole blood specimens and report the additional Immature Reticulocyte Parameter. ### Similarities and Differences: The CELL-DYN 3500 and the CELL-DYN 3700 Systems with IRF, and the CELL-DYN 4000 System are similar in that the systems use impedance for counting and sizing RBCs. The Reticulocyte Reagent enables both the CELL-DYN 3500 and the CELL-DYN 3700 Systems with IRF and the CELL-DYN 4000 System to count and classify mature and immature reticulocytes. The CELL-DYN 3500 and the CELL-DYN 3700 Systems with IRF and the CELL-DYN 4000 System with IRF, enumerate reticulocytes in EDTA-anticoagulated whole blood using optical laser scatter. They are different in that CELL-DYN 3500 and the CELL-DYN 3700 Systems with IRF specimens are externally stained in an off-line preparation step with New Methylene Blue while the CELL-DYN 4000 System with IRF automatically dilutes the specimen with a fast acting proprietary dye that requires no incubation prior to measurement. {3}------------------------------------------------ #### 16. 510(k) Summary (cont'd) ### CELL-DYN 3500 System and the CELL-DYN 3700 System with Immature Reticulocyte Fraction (IRF) #### Equivalency Data: The data compiled to support the claim that the CELL-DYN 3500 and the CELL-DYN 3700 Systems IRF are substantially equivalent to the Abbott CELL-DYN 4000 System include background, accuracy, precision, linearity, and carryover. The data supports the claim that the CELL-DYN 3500 and the CELL-DYN 3700 Systems IRF parameters are substantially equivalent to the Abbott CELL-DYN 4000 System IRF parameter. The accuracy, precision, and linearity data shows performance to manufacturer's specifications. Data was collected at Scottsdale Health Care - Shea, Scottsdale, AZ; VAMC, Phoenix, AZ; and an internal Abbott Diagnostics Division site. #### Conclusion: The CELL-DYN 3500 and the CELL-DYN 3700 Systems IRF demonstrate substantial equivalence to the predicate device. {4}------------------------------------------------ #### 17. Financial Disclosure Data was collected from two field clinical sites and one internal site. The two clinical sites were Scottsdale Health Care - Shea, Scottsdale, Arizona and Veteran's Affairs Medical Center (VAMC), Phoenix, Arizona. Scottsdale Health Care was paid \$14,000 on 12/11/98, VAMC was paid \$5,984.78 on 12/9/98. Neither site's reimbursement exceeded \$25,000.00 for this clinical study. The internal Abbott Diagnostics Division site in Santa Clara, CA also collected data for this clinical study. Financial disclosure is not applicable as there was no payment involved. {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features the department's name encircling a symbol. The symbol consists of three stylized human profiles facing right, stacked on top of each other. The profiles are rendered in a flowing, wave-like design. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 JUL - 9 1999 Ms. Michelle B. Roeding Regulatory Affairs Supervisor Abbott Laboratories Diagnostic Division 5440 Patrick Henry Drive Santa Clara, California 95054 - K991605 Re: Trade Name: CELL-DYN 3500 System with Immature Reticulocyte Fraction CELL-DYN 3700 System with Immature Reticulocyte Fraction Regulatory Class: III Product Code: GKZ Dated: May 7, 1999 Received: May 10, 1999 Dear Ms. Roeding: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {6}------------------------------------------------ #### Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Steven Butman Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ 19. #### Indications For Use Statement #### K980614/A1 510(k): CELL-DYN 3500 System with Immature Reticulocyte Fraction Device Name: CELL-DYN 3700 System with Immature Reticulocyte Fraction Indications for Use: The CELL-DYN 3500 and the CELL-DYN 3700 Systems are automated, multiparameter hematology analyzers intended to classify the following formed elements of EDTA anticoagulated blood: | WBC -- White Blood Cell or Leukocyte Count | RBC -- Red Blood Cell or Erythrocyte Count | |--------------------------------------------|--------------------------------------------| | NEU -- Neutrophil Absolute Count | HCT - Hematocrit | | %N -- Neutrophil Percent | MCV -- Mean Corpuscular Volume | | LYM -- Lymphocyte Absolute Count | RDW -- Red Cell Distribution Width | | %L -- Lymphocyte Percent | HGB -- Hemoglobin Concentration | | MONO -- Monocyte Absolute Count | MCH -- Mean Corpuscular Hemoglobin | | %M -- Monocyte Percent | MCHC -- Mean Corpuscular Hemoglobin | | EOS -- Eosinophil Absolute Count | Concentration | | %E -- Eosinophil Percent | RETC -- Reticulocyte Concentration | | BASO -- Basophil Absolute Count | % RETC -- Reticulocyte Concentration | | %B -- Basophil Percent | IRF -- Immature Reticulocyte Fraction | | PLT -- Platelet Count | *PDW -- Platelet Distribution Width | | MPV -- Mean Platelet Volume | *PCT -- Plateletcrit | * These parameters are provided for laboratory use only and are not reportable in the US. PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE) Alkem (Division Sign-Off)/ Sign-on aboratory Devices Nag 1605 Clinical (Division of Clinical 510(k) Number Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...