STA LIATEST CONTROL [N]+[P] KIT

K964716 · American Bioproducts Co. · GGN · Jun 26, 1997 · Hematology

Device Facts

Record IDK964716
Device NameSTA LIATEST CONTROL [N]+[P] KIT
ApplicantAmerican Bioproducts Co.
Product CodeGGN · Hematology
Decision DateJun 26, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.5425
Device ClassClass 2

Indications for Use

The STA® Liatest® Control [N]+[P] kit is intended for use as control plasmas (Normal and Abnormal levels) for von Willebrand Factor (vWF) and D-dimer antigen assays by the immuno-turbidimetric method performed on STA® analyzers (Diagnostica Stago, France: STA® full-size model, K942117; STA® Compact model, K961579).

Device Story

STA®-Liatest® Control [N]+[P] kit consists of freeze-dried human plasma; provides normal and abnormal control levels for vWF and D-dimer antigen assays. Used in clinical laboratory settings on STA® analyzers (full-size and Compact models). Operates via immuno-turbidimetric method. Healthcare providers use control results to verify assay performance and ensure accuracy of patient testing. Benefits include standardized quality control for diagnostic coagulation assays.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Freeze-dried human plasma; citrated. Formulated at normal and abnormal levels for vWF and D-dimer. Stable for 24 months at 2-8°C; 8-hour stability post-reconstitution on-board STA® analyzers.

Indications for Use

Indicated for use as normal and abnormal quality control plasmas for vWF and D-dimer antigen assays performed on STA® analyzers. No specific patient population or contraindications described.

Regulatory Classification

Identification

A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.

Special Controls

*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} K964716 AMERICAN BIOPRODUCTS COMPANY Premarket 510(k) Notification STA®-Liatest® Control [N]+[P] Kit JUN 26 1997 ## VII. Safety and Effectiveness Summary The STA®-Liatest® Control [N]+[P] kit provides two citrated human plasmas intended for use as normal and abnormal controls for von Willebrand Factor (vWF) and D-dimer antigen assays by the immuno-turbidimetric method performed on STA® analyzers (Diagnostic Stago, France: STA® full-size model and STA® Compact model). Each STA®-Liatest® Control [N]+[P] kit provides 12 x 1-ml vials of freeze-dried human plasma containing vWF and D-dimer at normal levels (STA®-Liatest® Control [N]) and 12 x 1-ml vials of freeze-dried human plasma containing vWF and D-dimer at abnormal levels (STA®-Liatest® Control [P]). Both the normal and abnormal plasmas in the freeze-dried state are stable for 24 months after the date of manufacture, when stored at 2°-8°C; after reconstitution with water, both plasmas are stable for 8 hours on board STA® analyzers. {1} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 JUN 26 1997 Andrew Loc B. Le., Ph.D. Director, Regulatory Affairs and Quality Assurance American BioProducts Company Five Century Drive Parsippany, New Jersey 07054 Re: K964716 STA®-Liatest® Control [N]+[P] Kit Regulatory Class: II Product Code: GGN Dated: April 24, 1997 Received: April 28, 1997 Dear Dr. Le: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations. {2} Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3} Page 1 of 1 510(k) Number (if known): _________________________ Device Name: STA® Liatest® Control [N]+[P] Kit Indications for Use: The STA® Liatest® Control [N]+[P] kit is intended for use as control plasmas (Normal and Abnormal levels) for von Willebrand Factor (vWF) and D-dimer antigen assays by the immuno-turbidimetric method performed on STA® analyzers (Diagnostica Stago, France: STA® full-size model, K942117; STA® Compact model, K961579). ![img-0.jpeg](img-0.jpeg) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) ___________________________ (Concurrence of CDRH, Office of Device Evaluation (ODE)) Prescription Use ☑ (Per 21 CFR 801.109) OR Over-The-Counter Use ☐ (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...