K120014 · Diagnostica Stago, Inc. · GGN · Mar 14, 2013 · Hematology
Device Facts
Record ID
K120014
Device Name
STA- COAG CONTROL (N+ABN) PLUS
Applicant
Diagnostica Stago, Inc.
Product Code
GGN · Hematology
Decision Date
Mar 14, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5425
Device Class
Class 2
Indications for Use
The STA® - Coag Control (N +ABN) PLUS is a kit containing assayed normal and abnormal plasmas intended for the quality control of the following quantitative tests on STA-R® and STA Compact® analyzers: prothrombin time (PT), activated partial thromboplastin time (APTT), fibrinogen, thrombin time (TT), and antithrombin (AT).
Device Story
STA-Coag Control (N + ABN) PLUS is a kit of two levels of lyophilized, citrated human plasma (normal and abnormal). Used as quality control material for coagulation assays (PT, APTT, fibrinogen, TT, AT) on STA-R and STA Compact analyzers. Analyzers use chronometric (viscosity-based) detection for clotting assays and photometric (absorbance) detection for chromogenic assays. Reconstituted with distilled water; stable for 24 hours on-board. Healthcare providers use control results to verify analyzer performance and ensure accuracy of patient test results.
Clinical Evidence
No clinical data. Bench testing only. Precision studies performed per CLSI EP5-A2 across three sites using three lots on STA-R and STA Compact analyzers. Stability studies (open-vial and closed-vial) performed per CLSI EP25-A to support 24-hour on-board and 24-month shelf-life claims. All results met pre-determined acceptance criteria.
Technological Characteristics
Lyophilized citrated human plasma. Reconstituted with distilled water. Analyzers utilize chronometric (viscosity-based) and photometric (absorbance) detection principles. 24-month shelf-life at 2-8°C. 24-hour on-board stability.
Indications for Use
Indicated for quality control of PT, APTT, fibrinogen, TT, and AT assays on STA-R and STA Compact analyzers using normal and abnormal citrated human plasma controls.
Regulatory Classification
Identification
A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.
Special Controls
*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
Predicate Devices
STA® - System Control N + P (k943518)
Submission Summary (Full Text)
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k120014
B. Purpose for Submission:
Clearance of a new device
C. Measurand:
PT, APTT, Fibrinogen, TT, and AT
D. Type of Test:
Quality Control Material, Assayed
E. Applicant:
Diagnostica Stago, Inc.
F. Proprietary and Established Names:
STA® - Coag Control (N + ABN) PLUS
G. Regulatory Information:
1. Regulation section:
21 CFR § 864.5425, Multipurpose system for in vitro coagulation studies
2. Classification:
Class II
3. Product code:
GGN, Plasma, Coagulation Control
4. Panel:
Hematology (81)
{2}
H. Intended Use:
1. Intended use(s):
The STA® - Coag Control (N +ABN) PLUS is a kit containing assayed normal and abnormal plasmas intended for the quality control of the following quantitative tests on STA-R® and STA Compact® analyzers: prothrombin time (PT), activated partial thromboplastin time (APTT), fibrinogen, thrombin time (TT), and antithrombin (AT).
2. Indication(s) for use:
Same as intended use
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
STA-R® and STA Compact® analyzers
I. Device Description:
The STA® - Coag Control (N + ABN) PLUS kit is a set of two control levels. Each kit provides:
- 12 x 2-mL vials of Reagent 1 (STA® - Coag Control N PLUS), citrated normal human plasma, lyophilized.
- 12 x 2-mL vials of Reagent 2 (STA® - Coag Control ABN PLUS), citrated abnormal human plasma, lyophilized.
J. Substantial Equivalence Information:
1. Predicate device name(s):
STA® - System Control N + P
2. Predicate 510(k) number(s):
k943518
{3}
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device: STA® - Coag Control (N + ABN) PLUS | Predicate: STA® - System Control N + P |
| Intended Use | For the quality control of the following coagulation assays: prothrombin time (PT), activated partial thromboplastin time (APTT), fibrinogen, thrombin time (TT) and antithrombin (AT) in the normal and abnormal ranges. | Same, except: - additional analytes claimed for the intended use in the normal and abnormal ranges - thrombin time determined in the normal range only. |
| Number of levels per set | 2 (one normal and one abnormal) | Same |
| Matrix | Citrated Human plasma | Same |
| Form | Lyophilized | Same |
| Reagent preparation | Reconstitution with distilled water | Same |
| Analyzers | STA-R® and STA Compact® | Same in addition to STA Satellite® |
| Shelf-life | 24 months at 2-8°C in intact vials | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device: STA® - Coag Control (N + ABN) PLUS | Predicate: STA® - System Control N + P |
| Reconstituted reagent stability | 24 hours on-board analyzers | 8 hours on-board analyzers |
K. Standard/Guidance Document Referenced (if applicable):
CLSI EP5-A2, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline-Second Edition
CLSI EP25-A, Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline-First Edition
L. Test Principle:
STA® - Coag Control (N + ABN) PLUS are used as controls for clotting assays (PT, APTT, fibrinogen, and TT) and chromogenic assays (AT) performed on STA-R® and STA Compact® analyzers. These analyzers utilize the chronometric principle (viscosity based detection system) for clotting assays while the chromogenic assays are based on the photometric method (measurement of absorbance of monochromatic light).
{4}
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Precision studies were carried out according to the CLSI document EP5-A2. Three lots of STA® - Coag Control (N + ABN) PLUS and one lot of assay reagents were analyzed in duplicate, two runs per day for 20 operating days at an internal site and two external sites. Each control level was tested as an unknown sample and assayed on a STA-R® analyzer for: PT (STA® - Neoplastine® CI and STA® - Neoplastine® CI Plus), APTT (STA® - PTT A, STA® - C.K. Prest® and STA® - Cephascreen®), fibrinogen (STA® - Fibrinogen), TT (STA® - Thrombin), and AT (STA® - Stachrom® AT III).
In addition, each control level was tested as an unknown sample and assayed on a STA Compact® analyzer at the internal site using the same assay reagents as above.
For each site, the coefficient of variation (CV) and standard deviation (SD) for within-run, between-run, within-day, between-day, within-laboratory, between-lot, between-site (laboratory) and total precision for each lot and control level were estimated.
For pooled data on the STA-R® analyzer from the three sites, the CV and SD for within-run, between-run, between-lot, between-site and total precision were estimated by using a mixed model fitted to the pooled data (i.e., a fully nested ANOVA model).
All results met the acceptance criteria.
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Value assignment:
Prothrombin time (PT): For each control level, the target clotting time corresponds to the mean of at least three determinations, each using at least 3 different lots of STA® - Neoplastine® CI or STA® - Neoplastine® CI Plus. The determinations are carried out on a minimum of 2 analyzers (STA-R® and STA Compact®) and performed by a minimum of 2 laboratory technicians. The control ranges for PT (values in sec and %) are then assigned by calculating the lower and upper limits.
Activated Partial Thromboplastin Time (APTT): For each control level, the target clotting time corresponds to the mean of at least three determinations, each using at least 3 different lots of STA® - PTT A, STA® - C.K. Prest® and 5 different lots of
4
{5}
STA® - Cephascreen®. The determinations are carried out on a minimum of 2 analyzers (STA-R® and STA Compact®) and by a minimum of 2 laboratory technicians. The control ranges for APTT (sec) are then assigned by calculating the lower and upper limits.
Fibrinogen Assay: For each control level, the target fibrinogen level corresponds to the mean of at least four determinations using a total of at least 2 different lots of STA® - Fibrinogen and STA® - Fib. The determinations are carried out on a minimum of 2 analyzers (STA-R® and STA Compact®) and by a minimum of 2 laboratory technicians. The control ranges for fibrinogen (g/L) are then assigned by calculating the lower and upper limits.
Thrombin Time (TT): For each control level, the target clotting time corresponds to the mean of at least three determinations, each using at least 3 different lots of STA® - Thrombin. The determinations are carried out on a minimum of 2 analyzers (STA-R® and STA Compact®) and by a minimum of 2 laboratory technicians. The control ranges for TT (sec) are then assigned by calculating the lower and upper limits.
Antithrombin Assay (AT): For each control level, the target antithrombin level corresponds to the mean of at least three determinations, each made on two vials. The determinations are carried out using a total of at least 3 different lots of STA® - Stachrom® AT III, on a minimum of 2 analyzers (STA-R® and STA Compact®) and by a minimum of 2 laboratory technicians. The control ranges for AT (%) are then assigned by calculating the lower and upper limits.
Open vial stability: The 24-hour open vial stability claim was verified using one lot of STA® - Coag Control (N + ABN) PLUS Reagents. Three vials of each control level were reconstituted, prepared and stored onboard the STA Compact® and STA-R® analyzers in their original vials and tested at 4, 8, 17, 24 and 25 hour time points in parallel with freshly reconstituted reagents from the same lot that were used immediately. At each time interval, each vial was tested in triplicate for PT (with STA® - Neoplastine® CI and STA® Neoplastine® CI Plus), APTT (with STA® - PTT A, STA® - C.K. Prest® and STA® - Cephascreen®), fibrinogen (with STA® - Fibrinogen), TT (with STA® - Thrombin), and AT (with STA® - Stachrom® AT III). The median values were compared to their established baseline range (using the median value at zero hour as a baseline).
Results obtained were analyzed using linear regression (median results (y-axis) versus test hour (x-axis)) with the stated allowable drift acceptance criteria. In addition, the 95% confidence interval limits were computed and plotted.
In each case, the upper or lower limit of the 95% confidence interval of the regression line did not intersect with the acceptance criteria. The stability duration was taken as the maximum time point tested minus one hour time.
5
{6}
All results met the acceptance criteria supporting an on-board stability claim of 24 hours.
**Closed vial stability:** A 24-month closed vial stability (shelf-life) claim was verified using three lots of STA® - Coag Control N PLUS and STA® - Coag Control ABN PLUS reagents as patient samples. Each control reagent was tested on the STA-R® analyzer to perform PT (with STA® - Neoplastine® CI and STA® - Neoplastine® CI Plus), APTT (with STA® - PTT A, STA® - C.K. Prest® and STA® - Cephascreen®), fibrinogen (with STA® - Fibrinogen), TT (with STA® - Thrombin), and AT (with STA® - Stachrom® AT III) assays.
Control reagents stored at 2-8°C in their original vials were tested in 10 replicates at the following time points: T0, T0 + 6 months, T0 + 12 months, T0 + 18 months, T0 + 24 months and T0 + 25 months. All test results obtained were compared to the median value at T0 as a baseline.
The results met pre-determined acceptance criteria (i.e. deviation in absolute value from the median value at T0) and support the 24-month shelf-life claim.
**d. Detection limit:**
Not applicable
**e. Analytical specificity:**
Not applicable
**f. Assay cut-off:**
Not applicable
**2. Comparison studies:**
**a. Method comparison with predicate device:**
Not applicable
**b. Matrix comparison:**
Not applicable
**3. Clinical studies:**
**a. Clinical Sensitivity:**
Not applicable
{7}
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
The end-user is instructed to refer to the product assay sheet accompanying the product information sheet.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.