COULTER LH 750 HEMATOLOGY ANALYZER
Device Facts
| Record ID | K032342 |
|---|---|
| Device Name | COULTER LH 750 HEMATOLOGY ANALYZER |
| Applicant | Beckman Coulter, Inc. |
| Product Code | GKZ · Hematology |
| Decision Date | Sep 26, 2003 |
| Decision | SESE |
| Submission Type | Abbreviated |
| Regulation | 21 CFR 864.5220 |
| Device Class | Class 2 |
Indications for Use
The Coulter LH 750 Hematology Analyzer with Version 2B Software is designed for in vitro diagnostic use in clinical laboratories. The LH 750 provides automated complete blood count, leukocyte differential, NRBC enumeration and reticulocyte analysis. The purpose of the LH 750 Hematology Analyzer with Version 2B Software is to separate the normal patient, with all normal system-generated parameters, from the patient who needs additional studies of any of these parameters. These studies might include further measurements of cell size, and platelet distribution, manual WBC differential or any other definitive test that helps diagnose the patient's condition.
Device Story
Automated hematology analyzer; processes whole blood samples via open or closed tube sampling; utilizes barcode identification; performs complete blood count, leukocyte differential, NRBC enumeration, and reticulocyte analysis; identifies normal vs. abnormal patient samples to flag need for manual review or further diagnostic testing; used in clinical laboratory settings by trained laboratory personnel; output provided to healthcare providers to assist in clinical decision-making and patient diagnosis.
Clinical Evidence
Bench testing only. Performance validation focused on improved NRBC enumeration and WBC correction fail-safes compared to the predicate device.
Technological Characteristics
Automated hematology analyzer; utilizes Coulter Principle (impedance) for cell counting/sizing, single beam photometry for hemoglobin, and VCS (volume, conductivity, light scatter) technology for WBC/reticulocyte analysis. Reagent-based system (diluent, lytic agents, cleaner). Software-controlled; supports cyanide-free reagents.
Indications for Use
Indicated for use in clinical laboratories for quantitative, automated hematology analysis, including complete blood count, leukocyte differential, NRBC enumeration, and reticulocyte analysis.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
Predicate Devices
- Coulter LH 750 Hematology Analyzer with Version 2A Software (K011342)
Related Devices
- K022161 — MODIFICATION TO COULTER LH 750 HEMATOLOGY ANALYZER, MODEL 6605632 · Beckman Coulter, Inc. · Jul 29, 2002
- K052905 — MODIFICATION TO COULTER LH 750 HEMATOLOGY ANALYZER · Beckman Coulter, Inc. · Nov 18, 2005
- K061574 — COULTER LH 750 HEMATOLOGY ANALYZER, MODEL 6605632 · Beckman Coulter, Inc. · Jul 21, 2006
- K050057 — COULTER LH 750 HEMATOLOGY ANALYZER · Beckman Coulter, Inc. · Feb 10, 2005