COULTER LH 750 HEMATOLOGY ANALYZER, MODEL 6605632

K061574 · Beckman Coulter, Inc. · GKZ · Jul 21, 2006 · Hematology

Device Facts

Record IDK061574
Device NameCOULTER LH 750 HEMATOLOGY ANALYZER, MODEL 6605632
ApplicantBeckman Coulter, Inc.
Product CodeGKZ · Hematology
Decision DateJul 21, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.5220
Device ClassClass 2

Intended Use

The COULTER LH 750 Hematology Analyzer is a quantitative, automated hematology analyzer and leukocyte differential counter for in vitro diagnostic use in clinical laboratories. The COULTER LH 750 Hematology Analyzer provides automated reticulocyte analysis and enumeration of nucleated red blood cells (NRBCs) as well as an automated method for enumeration of RBCs and WBCs in body fluids.

Device Story

Automated hematology analyzer; performs complete blood cell analysis, leukocyte differential, reticulocyte analysis, NRBC enumeration, and body fluid cell counts. Inputs: blood or body fluid samples. Principle: Coulter Principle (electrical impedance) for cell counting/sizing; VCS technology (volume, conductivity, light scatter) for WBC differential/reticulocyte analysis; single beam photometer for hemoglobinometry. Operation: automated diluting/mixing; manual mode for body fluids. Output: quantitative hematology results. Used in clinical laboratories by trained personnel. Output used by clinicians for diagnostic decision-making regarding hematological status.

Clinical Evidence

No clinical data provided. Substantial equivalence supported by analytical performance data regarding MCV accuracy specifications.

Technological Characteristics

Automated hematology analyzer utilizing Coulter Principle (impedance) and VCS (Volume, Conductivity, Light Scatter) technology. Includes single beam photometer for hemoglobinometry. Reagent-based system (diluents, lytic reagents, cleaners). Software version 2B or higher. Connectivity/integration via standard laboratory workflows.

Indications for Use

Indicated for in vitro diagnostic use in clinical laboratories for quantitative automated hematology analysis, leukocyte differential counting, reticulocyte analysis, nucleated red blood cell (NRBC) enumeration, and enumeration of RBCs and WBCs in body fluids.

Regulatory Classification

Identification

An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K061574 # Summary of Safety and Effectiveness for COULTER® LH 750 Hematology Analyzer JUL 2 1 2006 #### 1.0 General Information | Device Generic Name(s): | Automated differential cell counter | |-----------------------------|----------------------------------------------------------------------------------------------------| | Device Trade Name(s): | COULTER®LH 750 Hematology Analyzer | | Device Classification: | The COULTER® LH 750 Hematology Analyzer is a Class II<br>medical device. | | Applicant Name and Address: | Beckman Coulter, Inc.<br>Cellular Analysis Division<br>11800 SW 147 Avenue<br>Miami, FL 33196-2500 | | Date: | June 6 2006 | #### 2.0 Legally Marketed Device(s) The modified COULTER® LH 750 Hematology Analyzer with claims substantial equivalence to the previously cleared COULTER® LH 750 Hematology Analyzer and COULTER® LH 750 Body Fluids Application. FDA 510(k) Number(s): K011342 and K030606 #### 3.0 Device Description The product is an automated hematology analyzer capable of supplying a complete blood cell analysis and includes a differential leukocyte cell count. The product also provides automated reticulocyte analysis and enumeration of nucleated red blood cells (NRBCs) as well as an automated method for enumeration of RBCs and WBCs in body fluids. The following reagents, with 510(k) numbers indicated where applicable, are qualified for use on the COULTER LH 750 Hematology Analyzer with Software Version 2B (or higher): - · 5C Cell Control (K912133) and COULTER RETIC-C™ Cell control (K930119) hematology quality control materials used to monitor the instrument performance. COULTER® LIN-C® linearity control (K955334) verifies reportable range of the CBC parameters. - · COULTER® LH Series Diluent and ISOTON® 4 Diluent. Intended for use as a diluent for counting and sizing blood cells. - · COULTER Lyse S® III diff lytic reagent and COULTER Lyse S® 4 lytic reagent. Intended for the simultaneous quantitative determination of hemoglobin and for leukocyte counting and sizing. - · COULTER CLENZ® cleaning agent to prevent protein buildup on surfaces. - · COULTER Latron™ Primer and Latron Control (K885028) to monitor VCS performance. LH 750 Revised MCV Accuracy Specification- Special 510k Beckman Coulter, Inc. {1}------------------------------------------------ - · COULTER LH Series Pak, containing Erythrolyse™ lytic reagent and Stabilyse " to preserve leukocytes in near-native state to allow differentiation into subpopulations. - · COULTER LH Series RETIC Pak for preparing samples for reticulocyte analysis. - · COULTER S-CAL® Hematology Calibrator (K840794), alternative to whole blood reference method of calibration. Intended for use in ensuring accurate instrument measurements. #### Principle of Method: 4.0 The COULTER LH 750 Hematology Analyzer utilizes the Coulter Principle for enumerating and sizing blood cells, automatic diluting and mixing for sample processing and a single beam photometer for hemoglobinometry. It uses COULTER VCS (yolume, conductivity, light scatter) technology for WBC Differential analysis and classification and reticulocyte analysis. The analyzer uses a reagent system consisting of an isotonic diluent, lytic reagents to lyse the red cells without significantly affecting the white cells and an instrument cleaner. Additionally, all systems include reagents used for reticulocyte staining and analysis. For body fluids analysis, the specimen is aspirated into the LH 750 Analyzer via the manual mode and is diluted in separate WBC and RBC baths. The Coulter Method of counting cells is used to detect and measure changes in electrical resistance when a cell, suspended in a conductive diluent, passes through a small aperture. Each suspended cell acts as an insulator. As the cell passes through the aperture, it momentarily increases the resistance of the electrical path between two submerged electrodes, one located on each side of the aperture. The resistance generates an electrical pulse. The accumulation of electrical pulses are channelized, processed for coincidence correction, and multiplied by a calibration factor, yielding the WBC and RBC counts. The WBC result represents the TNC (total nucleated cell count) in the analysis of Body Fluids. #### Indications for Use: 5.0 The COULTER LH 750 Hematology Analyzer is a quantitative, automated hematology analyzer and leukocyte differential counter For In Vitro Diagnostic Use in clinical laboratories. The COULTER® LH 750 Hematology Analyzer also provides automated reticulocyte analysis and nucleated red blood cell (NRBC) enumeration as well as an automated method for enumeration of RBCs and WBCs in body fluids. #### 6.0 Description of the modification: The labeling for the marketed COULTER LH 750 hematology analyzer will be modified relative to the draft labeling submitted in the original Premarket notification (K011342) to reflect a revised MCV accuracy specification. {2}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, which is a symbol of medicine, with three lines representing the three branches of government. The caduceus is surrounded by a circle of text that reads "U.S. Department of Health and Human Services". Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Ms. Nancy Nadler Staff Regulatory Affairs Specialist Beckman Coulter, Inc. 11800 S.W. 147 Avenue M/S 31-B06 Miami, Florida 33196-2500 JUL 21 2006 Re: k061574 > Trade/Device Name: COULTER® LH 750 Hematology Analyzer Regulation Number: 21 CFR § 864.5220 Regulation Name: Automated differential cell counter Regulatory Class: II Product Code: GKZ Dated: June 6, 2006 Received: June 7, 2006 Dear Ms. Nadler: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus. permits your device to proceed to the market. {3}------------------------------------------------ Page 2 -- If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll from the (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html Sincerely yours, lobetz Beckerh Robert L. Becker, Jr., MD, PK Director Division of Immunology and Hematology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## INDICATIONS FOR USE 510(k) Number (if known): K061574 Device Name: COULTER® LH 750 Hematology Analyzer ### Indications For Use: The COULTER LH 750 Hematology Analyzer is a quantitative, automated hematology analyzer and leukocyte differential counter For In Vitro Diagnostic Use in clinical laboratories. The COULTER LH 750 Hematology Analyzer provides automated Reticulocyte analysis and enumeration of nucleated red blood cells (NRBCs) as well as an automated method for enumeration of RBCs and WBCs in body fluids. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Page 1 of 1 Robert Hooper Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safe 310(k) K061574 LH 750 Revised MCV Accuracy Specification- Special 510k Beckman Coulter, Inc.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...