The COULTER® LH 750 Hematology Analyzer is a quantitative, automated hematology Analyzer and leukocyte differential counter For In Vitro Diagnostic Use in clinical laboratories. The COULTER® LH 750 Hematology Analyzer also provides automated Reticulocyte analysis and enumeration of nucleated red blood cells (NRBCs). The system also provides an automated method for enumeration of RBCs and WBCs in body fluids.
Device Story
COULTER LH 750 Hematology Analyzer performs automated cell counting in body fluids (cerebrospinal, serous, synovial). Input: body fluid samples; reagent system (isotonic diluent, lytic reagents). Principle: Coulter Principle for cell sizing and enumeration; lytic reagents lyse red cells to isolate white cells. Output: quantitative WBC and RBC counts. Used in clinical laboratories by trained technicians. Replaces manual counting chamber methods. Benefits: increased efficiency and automation of cell enumeration compared to manual microscopy.
Clinical Evidence
No clinical data provided; substantial equivalence is based on comparison to the manual microscopic counting method.
Technological Characteristics
Automated hematology analyzer utilizing the Coulter Principle for cell sizing and enumeration. Reagent system includes isotonic diluent, lytic reagents, and instrument cleaner. System is designed for clinical laboratory use.
Indications for Use
Indicated for clinical laboratory use to provide quantitative, automated enumeration of WBCs and RBCs in cerebrospinal, serous, and synovial fluids, as well as automated hematology analysis, leukocyte differential counting, reticulocyte analysis, and NRBC enumeration in blood.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
Submission Summary (Full Text)
{0}------------------------------------------------
K030606
# APR 2 1 2003
### 510(k) Summary of Safety and Effectiveness for Section 1 D: COULTER® LH 750 Body Fluids Application
#### 1.0 General Information
| Applicant Name and Address: | Beckman Coulter, Inc. |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Cellular Analysis Division | |
| 11800 SW 147 Avenue | |
| Miami, FL 33196-2500 | |
| Primary Contact: | Stan Sugrue, Ph.D.<br>Senior Regulatory Affairs Specialist<br>Telephone: (305) 380-4552<br>FAX: (305) 380-3618<br>E-mail: stan.sugrue@coulter.com |
| Date: | February 24, 2003 |
| Device Trade Name(s): | COULTER® LH 750 Hematology Analyzer |
| Device Generic Name(s): | Automated differential cell counter |
| Device Classification: | The COULTER® LH 750 Hematology Analyzer is a Class II<br>medical device |
#### 2.0 Pre-amendment Predicate Method
The COULTER® LH 750 Body Fluids Application claims substantial equivalence to the preamendment predicate method for enumeration of WBCs and RBCs via manual cell count method in a counting chamber by a skilled competent technician.
FDA 510(k) Number(s): Not applicable
#### 3.0 Device Description
The COULTER® LH 750 Body Fluids Application is an automated method for enumeration of RBCs and WBCs in body fluids on the COULTER LH 750 Hematology Analyzer, an automated hematology analyzer capable of supplying a complete blood cell analysis and includes a differential leukocyte cell count. The LH 750 also provides automated reticulocyte analysis and enumeration of nucleated red blood cells (NRBCs).
#### 4.0 Principle of Method:
The COULTER LH 750 Body Fluids Application is an automated method for enumeration of RBCs and WBCs in body fluids on the COULTER LH 750 Hematology Analyzer. The LH 750 utilizes the Coulter Principle for automatically enumerating and sizing blood cells. The analyzer uses a reagent system consisting of an isotonic diluent, lytic reagents to lyse the red cells without significantly affecting the white cells and an instrument cleaner.
{1}------------------------------------------------
#### 5.0 Comparison to Predicate
| Similarities/<br>Differences | Characteristic | Manual method<br>(Predicate) | LH 750 Body Fluids<br>Application |
|------------------------------|-------------------------|-----------------------------------------------------------------------------------------------------------------------------|-----------------------------------|
| Similarities | Intended Use | To provide a quantitative<br>determination of blood cells<br>in cerebrospinal fluid,<br>serous fluid, and synovial<br>fluid | Same as manual method |
| | Specimen Analyzed | Body Fluids collected in a<br>container with or without<br>anti-coagulant | Same as manual method |
| Differences | WBC Count<br>(cells/µL) | Manual cell count<br>performed in a counting<br>chamber by a skilled<br>competent technician | Automated count |
| Differences | RBC count<br>(cells/µL) | Manual cell count<br>performed in a counting<br>chamber by a skilled<br>competent technician | Automated count |
#### 6.0 Indications for Use:
The COULTER LH 750 Hematology Analyzer is a quantitative, automated hematology analyzer and leukocyte differential counter For In Vitro Diagnostic Use in clinical laboratories.The LH 750 Body Fluids Application adds a quantitative, automated procedure for analyzing cerebrospinal fluid, serous fluid, and synovial fluid to the LH 750, providing enumeration of the WBCs and the RBCs.
#### 7.0 Conclusion:
The COULTER LH 750 Body Fluids Application is substantially equivalent to the manual microscopic predicate method for enumeration of RBCs and WBCs in body fluids.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, enclosed in a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the upper portion of the circle.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
## APR 2 1 2003
Stan Sugrue, Ph.D. Senior Regulatory Affairs Specialist Beckman Coulter, Inc. 11800 S.W. 147 Avenue P.O. Box 169015 Miami, FL 33116-9015
k030606 Re:
> Trade/Device Name: COULTER® LH 750 Hematology Analyzer Regulation Number: 21 CFR 864.5220 Regulation Name: Automated Differential Cell Counter Regulatory Class: Class II Product Code: GKZ Dated: April 2, 2003 Received: April 3, 2003
Dear Dr. Sugrue:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820).
{3}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Section 1C:
### INDICATIONS FOR USE
510(k) Number (if known):
Not assigned K030606
Device:
COULTER® LH 750 Hematology Analyzer
Indications For Use:
The COULTER® LH 750 Hematology Analyzer is a quantitative, automated hematology Analyzer and leukocyte differential counter For In Vitro Diagnostic Use in clinical laboratories. The COULTER® LH 750 Hematology Analyzer also provides automated Reticulocyte analysis and enumeration of nucleated red blood cells (NRBCs). The system also provides an automated method for enumeration of RBCs and WBCs in body fluids.
Future commercialization will add ISOTON® 4 diluent /Lyse S® 4 Lytic reagent to the indications for use.
### 21 CFR 864.5220 Automated differential cell counter
An automated differential cell counter is a device used to identify and classify one or more of the formed elements of blood.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use V Use (Per 21 CFR 801.109) OR
Over-The-Counter
Auchini Rautati
(Division Sigr Division of Clinical Laboratory De 510(k) Numbe
Beckman Coulter, Inc. COULTER® LH 750 Body Fluids Application