FRESH CELLS WI-38

K973212 · Diagnostic Hybrids, Inc. · KIR · Sep 25, 1997 · Hematology

Device Facts

Record IDK973212
Device NameFRESH CELLS WI-38
ApplicantDiagnostic Hybrids, Inc.
Product CodeKIR · Hematology
Decision DateSep 25, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.2280
Device ClassClass 1

Indications for Use

FreshCells™ are indicated for use in the isolation of various viruses and Chlamydia from clinical specimens. Viruses and Chlamydia are intracellular parasites which can be cultured or grown only in specific cellular hosts. They cause a variety of diseases in man with some virus infections being lethal, particularly if the individual is immunocompromised. With the introduction of several new and specific antiviral drugs over the last several years, the need to determine the identity of the viral agent has become even more important. Culture of viruses in specific cell lines has become the standard for the identification of these viruses.

Device Story

FreshCells™ consist of nearly confluent monolayers of specific cell lines (HEL, HFF, LLC-MK2, MvlLu, NCI H292, Vero, WI-38) provided in multiwell plates, shell vials, or tubes. These cell cultures serve as host systems for the isolation and identification of viruses and Chlamydia from clinical samples. Used in clinical laboratory settings by trained personnel; the device facilitates the growth of intracellular parasites. Healthcare providers use the resulting viral identification to guide clinical decision-making, particularly for selecting appropriate antiviral therapies. The device benefits patients by enabling accurate diagnosis of viral infections, which is critical for managing potentially lethal conditions.

Clinical Evidence

Bench testing only. Product characterization included appearance, growth characteristics, sterility, isoenzyme analysis, and virus susceptibility testing to confirm the ability of the cell lines to support the growth of specific viruses.

Technological Characteristics

Cell lines (HEL, HFF, LLC-MK2, MvlLu, NCI H292, Vero, WI-38) provided as nearly confluent monolayers in multiwell plates, shell vials, or tubes. Biological material sourced from ATCC or approved suppliers. No electronic or software components.

Indications for Use

Indicated for the isolation and identification of viruses and Chlamydia from clinical specimens in patients, including immunocompromised individuals, where viral identification is necessary for clinical management.

Regulatory Classification

Identification

Cultured animal and human cells are in vitro cultivated cell lines from the tissue of humans or other animals which are used in various diagnostic procedures, particularly diagnostic virology and cytogenetic studies.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 510 (k) Summary SEP 2 5 ga- - Diagnostic Hybrids, Inc. a.1. 1 President Street Athens, OH 45701 (614)593-1784 FAX: (614) 593-0980 Attn: J.L. Brown Date of Preparation: June 10, 1997 - Trade Name: FreshCells™ a.2. Classification Name: Cells, Animal and Human, Cultured. - A predicate device is that of BioWhittaker, marketed as cell cultures, a.3. Appendix II, pp. A-11 to A-13 of this 510(k) Notification. - The subject device provides HEL, HFF, LLC-MK2, MVILu, NCI H292, Vero and a.4. WI-38 cells (in addition to MRC-5, McCOY, BGMK, and CV-1, under 510 (k) K936271 and A549 and HEp-2, under 510(k) K962306 which have previously been cleared for marketing under the same name, Fresh Cella™) as nearly confluent monolayers ready for use upon receipt a short pre-incubation period. after - Cell cultures to be used as hosts for the isolation and a.5. Intended Use: identification of specific viruses. The subject of this 510(k) Notification, the cell lines HEL, HFF, LLC-MK2, MvlLu, NCI H292, Vero and WI-38 are susceptible to and can be used in the isolation and confirmation of the following viruses from clinical samples: | CELL LINE/ORIGIN | SPECIFIC VIRUSES | |----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | HEL/Human Embryonic Lung | Adenovirus, CMV, Echovirus, HSV,<br>Poliovirus, Rhinovirus, Vesicular stomatitis<br>(Indiana Strain) virus and VZV. | | HFF/Human Foreskin Fibroblasts | Adenovirus, CMV, Echovirus, HSV, Mumps,<br>Poliovirus, Rhinovirus, VZV. | | LLC-MK2, Original/Rhesus Monkey Kidney | Poliovirus type 1, Enterovirus, Rhinovirus,<br>Myxovirus and Poxvirus groups. | | MvlLu/Mink Lung | HSV, CMV. | | NCI-H292/Human, Pulmonary muco-<br>epidermoid carcinoma. | Vaccinia virus, HSV, Adenovirus, BK<br>polyomavirus, Reoviruses, Measles virus,<br>RSV, some strains of Influenza type A, most<br>Enteroviruses and Rhinoviruses, Parainfluenza<br>and Mumps. | | Vero/African Green Monkey | Adenovirus, Coxsackie B, HSV, Measles,<br>Mumps, Poliovirus type 3, Rotavirus, Rubella | | WI-38/Human Lung | Adenovirus, CMV, Echovirus, HSV,<br>Influenza, Mumps, Poliovirus, Rhinovirus,<br>RSV, VZV. | {1}------------------------------------------------ a.6. A comparison of Technological Characteristics: | Characteristics | Predicate Device | Subject Device | |-----------------------------------------------|--------------------------------------------------------------------|---------------------------| | Source of Cell Line. | ATCC or another approved supplier. | Same as predicate device. | | Provided as nearly con-<br>fluent monolayers. | Cells are provided<br>routinely as nearly<br>confluent monolayers. | Same as predicate device. | | Intended Use. | Isolation & Confirmation<br>of specific viruses. | Same as predicate device. | - The non-clinical tests consist of those used to characterize the product b.l. rne non as appearance, growth characteristics, sterility, isoenzyme analysis and virus susceptibility. Not applicable. b.2. b.3. Not applicable. {2}------------------------------------------------ Food and Drug Administration 2098 Gaither Road Rockville MD 20850 James L. Brown Vice President Product Development and · Regulatory Affairs Diagnostic Hybrids, Inc. One President Street Athens, Ohio 45701 SEP 2 5 1997 Re: K973212 Trade Name: Fresh Cells™ WI-38 Regulatory Class: I Product Code: KIR Dated: August 22, 1997 Received: August 27, 1997 Dear Mr. Brown: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ Page 2 This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in-vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Steven Butman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use Statement 510(k) Number (if known): FreshCella™ in Multiwell Plates, Shell Vials and Tubes. Device Name: FreshCells™ are indicated for use in the isolation of Indications for Use: various viruses and Chlamydia from clinical specimens. Viruses and Chlamydia are intracellular parasites which can be cultured or grown only in specific cellular hosts. They cause a variety of diseases in man with some virus infections being lethal, particularly if the individual is immunocompromised. With the introduction of several new and specific antiviral drugs over the last several years, the need to determine the identity of the viral agent has Culture of viruses in specific cell lines has become even more important. become the standard for the identification of these viruses. Pa (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED). OR Concurrence of CDRH, Office of Device Evaluation (ODE) 5 Prescription Use I (Per 21 CFR 801.109) Over-The-Counter Use_ Steatures in to and any to 彩票 ప్రా :
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...