FRESH CELLS VERO

K973209 · Diagnostic Hybrids, Inc. · KIR · Sep 25, 1997 · Hematology

Device Facts

Record IDK973209
Device NameFRESH CELLS VERO
ApplicantDiagnostic Hybrids, Inc.
Product CodeKIR · Hematology
Decision DateSep 25, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.2280
Device ClassClass 1

Indications for Use

Cell cultures to be used as hosts for the isolation and identification of specific viruses. The subject of this 510(k) Notification, the cell lines HEL, HFF, LLC-MK2, Mv1Lu, NCI H292, Vero and WI-38 are susceptible to and can be used in the isolation and confirmation of the following viruses from clinical samples: [Table of cell lines and corresponding viruses provided in source]. FreshCells™ are indicated for use in the isolation of various viruses and Chlamydia from clinical specimens. Viruses and Chlamydia are intracellular parasites which can be cultured or grown only in specific cellular hosts. They cause a variety of diseases in man with some virus infections being lethal, particularly if the individual is immunocompromised. With the introduction of several new and specific antiviral drugs over the last several years, the need to determine the identity of the viral agent has become even more important. Culture of viruses in specific cell lines has become the standard for the identification of these viruses.

Device Story

FreshCells™ are cultured animal and human cell lines (HEL, HFF, LLC-MK2, Mv1Lu, NCI H292, Vero, WI-38) provided as nearly confluent monolayers in multiwell plates, shell vials, or tubes. Used in clinical laboratories by technicians/microbiologists to isolate and identify specific viruses and Chlamydia from patient clinical samples. Cells act as biological hosts for intracellular parasites; viral/Chlamydia growth in these specific lines confirms presence of pathogen. Results assist clinicians in identifying viral agents to guide antiviral therapy. Benefits include standardized, ready-to-use host cells for accurate diagnostic confirmation of infections.

Clinical Evidence

Bench testing only. Product characterized via appearance, growth characteristics, sterility, isoenzyme analysis, and virus susceptibility testing. No clinical trial data presented.

Technological Characteristics

Cultured human and animal cell lines (HEL, HFF, LLC-MK2, Mv1Lu, NCI H292, Vero, WI-38). Provided as nearly confluent monolayers in multiwell plates, shell vials, or tubes. Biological substrate for viral/Chlamydia isolation. No electronic or software components.

Indications for Use

Indicated for the isolation and identification of viruses and Chlamydia from clinical specimens in patients, including immunocompromised individuals, where viral or Chlamydia infection is suspected.

Regulatory Classification

Identification

Cultured animal and human cells are in vitro cultivated cell lines from the tissue of humans or other animals which are used in various diagnostic procedures, particularly diagnostic virology and cytogenetic studies.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ SEP 25 397 ## 510 (k) Summary - a.1. Diagnostic Hybrids, Inc. 1 President Street Athens, OH 45701 (614) 593-1784 FAX: (614) 593-0980 Attn: J.L. Brown Date of Preparation: June 10, 1997 - Trade Name: FreshCells™ a.2. Cells, Animal and Human, Cultured. Classification Name: - A predicate device is that of BioWhittaker, marketed as cell cultures, a.3. Appendix II, pp. A-11 to A-13 of this 510(k) Notification. - The subject device provides HEL, HFF, LLC-MK2, MVILu, NCI H292, Vero/and a.4. WI-38 cells (in addition to MRC-5, McCOY, BGMK, and CV-1, under S10(k) K936271 and A549 and HEp-2, under 510(k) K962306 which have previously been cleared for marketing under the same name, Fresh Cella™) as nearly confluent monolayers ready for use upon receipt after a short pre-incubation period. - Intended Use: Cell cultures to be used as hosts for the isolation and a.5. identification of specific viruses. The subject of this 510(k) Notification, the cell lines HEL, HFF, LLC-MK2, Mv1Lu, NCI H292, Vero and WI-38 are susceptible to and can be used in the isolation and confirmation of the following viruses from clinical samples: | CELL LINE/ORIGIN | SPECIFIC VIRUSES | |----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | HEL/Human Embryonic Lung | Adenovirus, CMV, Echovirus, HSV,<br>Poliovirus, Rhinovirus, Vesicular stomatitis<br>(Indiana Strain) virus and VZV. | | HFF/Human Foreskin Fibroblasts | Adenovirus, CMV, Echovirus, HSV, Mumps,<br>Poliovirus, Rhinovirus, VZV. | | LLC-MK2, Original/Rhesus Monkey Kidney | Poliovirus type 1, Enterovirus, Rhinovirus,<br>Myxovirus and Poxvirus groups. | | Mv1Lu/Mink Lung | HSV, CMV. | | NCI-H292/Human, Pulmonary muco-<br>epidermoid carcinoma. | Vaccinia virus, HSV, Adenovirus, BK<br>polyomavirus, Reoviruses, Measles virus,<br>RSV, some strains of Influenza type A, most<br>Enteroviruses and Rhinoviruses, Parainfluenza<br>and Mumps. | | Vero/African Green Monkey | Adenovirus, Coxsackie B, HSV, Measles,<br>Mumps, Poliovirus type 3, Rotavirus, Rubella | | WI-38/Human Lung | Adenovirus, CMV, Echovirus, HSV.,<br>Influenza, Mumps, Poliovirus, Rhinovirus,<br>RSV, VZV. | {1}------------------------------------------------ ## a.6. A comparison of Technological Characteristics: | Characteristics | Predicate Device | Subject Device | |-----------------------------------------------|--------------------------------------------------------------------|---------------------------| | Source of Cell Line. | ATCC or another approved supplier. | Same as predicate device. | | Provided as nearly con-<br>fluent monolayers. | Cells are provided<br>routinely as nearly<br>confluent monolayers. | Same as predicate device. | | Intended Use. | Isolation & Confirmation<br>of specific viruses. | Same as predicate device. | - b.1. The non-clinical tests consist of those used to characterize the product such as appearance, growth characteristics, sterility, isoenzyme analysis and virus susceptibility. - b.2. Not applicable. - b.3. Not applicable. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines forming its body and head. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 SEP 2 5 1997 James L. Brown Vice President Product Development and · Regulatory Affairs Diagnostic Hybrids, Inc. One President Street Athens, Ohio 45701 Re: K973209 Trade Name: Fresh Cells™ Vero Regulatory Class: I Product Code: KIR Dated: August 22, 1997 Received: August 27, 1997 Dear Mr. Brown: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ Page 2 This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in-vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Steven Butman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use Statement 510 (k) Number (if known) : FreahCella™ in Multiwell Plates, Shell Vials and Tubes. Device Name: FreshCells™ are indicated for use in the isolation of Indications for Use: various viruses and Chlamydia from clinical specimens. Viruses and Chlamydia are intracellular parasites which can be cultured or grown only in specific cellular hosts. They cause a variety of diseases in man with some virus infections being lethal, particularly if the individual is immunocompromised. With the introduction of several new and specific antiviral drugs over the last several years, the need to determine the identity of the viral agent has become even more important. Culture of viruses in specific cell lines has become the standard for the identification of these viruses. Ar Pah DIVISION Sign .. : of Childed Laboratory Devices 510(k) Number PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED). Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) ーディ ﭼ ﮯ۔ OR Over-The-Counter Use
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...