The EndoFLIP® EF-900 Gastric Tube is intended for use in bariatric surgical procedures to provide a sized support bougie, and to permit stomach decompression, gastric fluid drainage and removal. It is also intended for use to aid deployment of EndoFLIP® EF-620, EF-325, and BF-325 catheters.
Device Story
EndoFLIP® Gastric Tube EF-900 is a single-lumen, 43 Fr, 75 cm disposable tube used during bariatric surgery. Functions as a support bougie; facilitates stomach decompression, gastric fluid drainage, and removal via distal side-holes. Aids deployment of specific EndoFLIP® catheters (EF-620, EF-325, BF-325). Used in hospitals and surgery centers by clinicians. Supplied non-sterile with a suction connector and depth markings at 20 cm and 70 cm. Provides mechanical support and fluid management during surgical procedures.
Clinical Evidence
Bench testing only. Testing included age testing, kink testing, and biocompatibility testing per ISO 10993-1. No clinical data presented.
Technological Characteristics
43 Fr single-lumen tube, 75 cm length. Materials: PVC with Duraglide/PTFE coating. Features distal side-holes for suction and proximal connector. Biocompatibility per ISO 10993-1. Non-sterile, single-patient use.
Indications for Use
Indicated for patients undergoing gastric and bariatric surgery requiring a bougie or suctioning capabilities. No contraindications.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}------------------------------------------------
K110529
PAGE 1 OF 3
510(k) Summary Page 1 of 3 17-Jun-11
JUL 222 2011
| Company | Crospon Ltd.<br>Galway Business Park<br>Dangan<br>Galway, Ireland<br><br>Tel - 011 [353] (91) 519882<br>Fax - 011 [353] (91) 519873 |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact: | John O'Dea, Ph.D. |
| Proprietary or Trade Name: | EndoFLIP® Gastric Tube EF-900 |
| Common/Usual Name: | Gastrointestinal motility monitoring system |
| Classification / CFR: | KNT - CFR 876.1500 |
| Device: | EndoFLIP® Gastric Tube EF-900 |
| Predicate Devices: | K002838 - BioEnterics® Gastric Balloon Suction<br>Catheter |
#### Device Description:
The EndoFLIP® Gastric Tube is a simple single lumen tube which acts as a support bougie for a 43 Fr lumen and may be used to aid in the EndoFLIP® catheter, K092850 and K 102214, deployment. It may also be used for drainage, suction or irrigation as it is open at both ends and features a number of side-holes at the distal end. A connector is supplied to push-on the proximal end to aid attachment to a suction system. The tube is 75 cm long and markings are provided at 20 cm and 70 cm from the distal end. The EndoFLIP® Gastric Tube is supplied non-sterile and is single patient use, disposable. An alcohol swab is provided to wipe down the tube exterior prior to use.
#### Indications for Use:
The EndoFLIP® EF-900 Gastric Tube is intended for use in bariatric surgical procedures to provide a sized support bougie, and to permit stomach decompression, gastric fluid drainage and removal. It is also intended for use to aid deployment of EndoFLIP® EF-620, EF-325, and BF-325 catheters.
| Patient Population: | Patients undergoing gastric and bariatric surgery where a bougie<br>or suctioning capabilities are needed. |
|---------------------|------------------------------------------------------------------------------------------------------------|
| Environment of Use: | Hospitals and Surgery Centers |
| Contraindications: | None |
{1}------------------------------------------------
# KIIO529
PAGE 2 OF 3
.
## 510(k) Summary
| | EndoFLIP® Gastric Tube<br>EF-900 | BioEnterics® Gastric Balloon<br>Suction Catheter<br>K002838 |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Attributes | | |
| Indications for Use | The EndoFLIP® EF-900 Gastric<br>Tube is intended for use in bariatric<br>surgical procedures to provide a sized<br>support bougie, and to permit<br>stomach decompression, gastric fluid<br>drainage and removal. It is also<br>intended for use to aid deployment of<br>EndoFLIP® EF-620, EF-325, and<br>BF-325 catheters. | Indicated for use in gastric and<br>bariatric surgical procedures to<br>provide visible and tactile delineation<br>of the antrum of the stomach along<br>with the ability to decompress and<br>remove gastric fluid and size a gastric<br>pouch. |
| Typical Use | Use in gastric and bariatric<br>surgical procedures | Use in gastric and bariatric<br>surgical procedures |
| Environments of use | Hospital and surgery centers | Hospital and surgery centers |
| Patient Population | All individuals undergoing gastric<br>and bariatric procedures | All individuals undergoing gastric<br>and bariatric procedures |
| Contraindications | None | None |
| Prescription | Prescription use | Prescription use |
| Functions | Suction<br>Drainage | Suction<br>Drainage |
| Intraoperative use | Yes | Yes |
| French Size | 43 Fr | 40 - 56 Fr |
| Design | | |
| Tube | Length -750 mm<br>14.3 mm OD | Length - 685 mm<br>13.3 mm OD |
| Side holes at distal end<br>for suctioning | Yes | Yes |
| Depth markings | Yes | Yes |
| Connector for suction | Yes | Yes |
| Sterility | Supplied non-sterile, and are<br>single patient use, disposable | Supplied non-sterile, and are<br>single patient use, disposable |
| Performance | Age testing | N/A |
| Materials used | Tube -- PVC with<br>Duraglide/PTFE coating | Not identified |
| Biocompatibility | Tested according to ISO 10993-1 | ISO 10993-1 |
-
{2}------------------------------------------------
#### 510(k) Summary Page 3 of 3 17-Jun-11
#### Substantial Equivalence:
The EndoFLIP® Gastric Tube is viewed as substantially equivalent to the predicate device because:
#### Indications -
Similar to predicate - K002838 - BioEnterics® Gastric Balloon Suction Catheter - Indicated for use in bariatric surgical procedures to provide a sized support bougie, and to permit stomach decompression, gastric fluid drainage and removal. In both cases the intended use is to assist the clinician in gastric and bariatric procedures.
It is also intended for use to aid deployment of EndoFLIP® EF-620, EF-325, and BF-325 catheters.
#### Technology -
Similar to predicate - K002838 - BioEnterics® Gastric Balloon Suction Catheter except the proposed device does not have a balloon which is unnecessary for the intended use i.e. the balloon is specifically purposed for sizing the pouch created during a gastric band placement, whereas the EF-900 is not used for the creation of such pouches.
#### Materials -
The materials in contact with the patient have been tested per ISO 10993.
#### Environment of Use -
Identical to predicate -K002838 - BioEnterics® Gastric Balloon Suction Catheter.
#### Patient Population -
Identical to predicate - K002838 - BioEnterics® Gastric Balloon Suction Catheter.
#### Performance Testing -
We did perform age testing, ISO 10993 testing for material biocompatibility, and a kink test. The proposed device meets its performance specifications and does not raise any new safety or efficacy issues.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the top half of the circle. Inside the circle is an abstract symbol resembling an eagle or bird-like figure with three stylized wings or feathers.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Crospon, Ltd. % Mr. Paul Dryden President ProMedic, Inc. 24301 Woodsage Drive BONITA SPRINGS FL .34134-2958
JUL 2 2 2011
Re: K110529
Trade/Device Name: EndoFLIP® EF-900 Gastric Tube Regulation Number: 21 CFR §876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: KNT Dated: June 17, 2011 Received: June 20, 2011
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{4}------------------------------------------------
Page 2
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default:htm.
Sincerely yours,
Hubert Lenoir MD
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive. Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health.
Enclosure
{5}------------------------------------------------
### Indications for Use Statement
Page 1 of 1
510(k) Number:
Device Name:
K110529
EndoFLIP® EF-900 Gastric Tube
Indications for Use:
The EndoFLIP® EF-900 Gastric Tube is intended for use in bariatric surgical procedures to provide a sized support bougie, and to permit stomach decompression, gastric fluid drainage and removal. It is also intended for use to aid deployment of EndoFLIP® EF-620, EF-325, and BF-325 catheters.
Prescription Use (Part 21 CFR 801 Subpart D)
or
Over-the-counter use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
ision Sigh-Off of Reproductive, Gastro-Rena 10(k) Numbe
17
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.