DIACAP POLYSULFONE HIFLO 23 HEMODILAYZER , MODEL 72036811
K100334 · B. Braun Avitum AG · KDI · Apr 22, 2010 · Gastroenterology, Urology
Device Facts
Record ID
K100334
Device Name
DIACAP POLYSULFONE HIFLO 23 HEMODILAYZER , MODEL 72036811
Applicant
B. Braun Avitum AG
Product Code
KDI · Gastroenterology, Urology
Decision Date
Apr 22, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Diacap® Polysulfone HiFlo 23 Hemodialyzer is designed for single use in acute and chronic hemodialysis.
Device Story
Diacap Polysulfone HiFlo 23 Hemodialyzer is a high-permeability dialyzer for acute and chronic hemodialysis. Device consists of polysulfone hollow fiber capillary membranes housed in polycarbonate with polyurethane potting resin and silicone O-ring. It functions as an extracorporeal blood filter to remove waste products and excess fluid from patient blood during dialysis. Intended for single-use in clinical settings; operated by trained healthcare professionals. Output is the filtered blood returned to the patient. Benefits include effective solute clearance and fluid management for patients with renal failure. Design is an extension of existing B. Braun polysulfone dialyzer line, featuring an increased surface area of 2.3m².
Indicated for patients requiring acute or chronic hemodialysis treatment.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”
Predicate Devices
Diacap® LO PS (10, 12, 15) and Diacap® HI PS (10, 12, 15, 18, 20) Hemodialyzers (K071518)
Submission Summary (Full Text)
{0}------------------------------------------------
B. Braun Avitum AG Special 510(k) Premarket Notification Diacap Polysulfone HiFlo 23 Hemodialyzer
#### 510(k) SUMMARY ഗ്
# APR 9 8 2010
B. Braun Avitum AG APPLICANT: Schwarzenberger Weg 73-79 34212 Melsungen, Germany Establishment Registration Number: 3002879653
SUBMITTER:
B. Braun Medical Inc. 901 Marcon Boulevard Allentown, PA 18109-9341 610-266-0500
#### CONTACT:
Angela J. Caravella Sr. Regulatory Affairs Analyst Phone: 610-596-2966 Fax: 610-596-2502 E-mail: Angela.Caravella@bbraun.com
DEVICE NAME:
COMMON OR USUAL NAME:
DEVICE
High Permeability Hemodialysis System
Diacap® Polysulfone HiFlo 23 Hemodialyzer
Class II per 21 CFR § 876.5860 Product Code KDI
## CURRENTLY MARKETED DEVICE (PREDICATE):
CLASSIFICATION:
DESCRIPTION:
Diacap® LO PS (10, 12, 15) and Diacap® HI PS (10, 12, 15, 18, 20) Hemodialyzers, (K071518)
The Diacap® Polysulfone HiFlo 23 Hemodialyzer is a high permeability dialyzer intended to be used in acute and chronic hemodialysis. It is composed of polysulfone hollow fiber capillary membranes in a polycarbonate housing with a polyurethane potting resin and a silicone O-ring. The proposed Diacap HiFlo 23 hemodialyzer will be available in one model, the 23, which represents the effective surface area of 2.3m². The dialyzer is intended for single use and is provided sterile.
## INTENDED USE:
The Diacap® Polysulfone HiFlo 23 Hemodialyzer is designed for single use in acute and chronic hemodialysis.
18
{1}------------------------------------------------
B. Braun Avitum AG Special 510(k) Premarket Notification Diacap Polysulfone HiFlo 23 Hemodialyzer
### SUBSTANTIAL EQUIVALENCE:
Diacap® Polysulfone HiFlo 23 Hemodialyzer has the same intended use and utilizes the same fundamental technology as the predicate device, the currently marketed Diacap LO PS (10, 12, 15) and Diacap HI PS (10, 12, 15, 18, 20) Hemodialyzers, (K071518). The Diacap HiFlo 23 Hemodialyzer is a modified version of the existing Diacap LO PS and HI PS dialyzers and can be considered an extension to the line of B. Braun Polysulfone Dialyzers.
10033.
The technological characteristics of the Diacap Polysulfone HiFlo 23 Hemodialyzer are equivalent to those of the existing Diacap polysulfone dialyzers. The Diacap Polysulfone HiFlo 23 Hemodialyzer is similar to the predicate device in material composition. Changes in the design of the proposed dialyzer include an increased surface area and two alternate materials have been added: polyurethane (PUR) potting resin and an alternate silicone material qualified for the dialyzer O-ring.
These changes do not pose a significant impact upon the fundamental technology of the Diacap polysulfone capillary dialyzers, as identified within the Risk Analysis. Verification and validation testing for the proposed dialyzer have been completed and all acceptance criteria have been met. The testing demonstrated that there are no differences between the predicate and the proposed device that raise new issues of safety or effectiveness.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" around the perimeter. The text is written in all capital letters.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
JUN -3 2010
B.Braun Avitum AG % Ms. Angela J. Caravella Sr. Regulatory Affairs Analyst B. Braun Medical, Inc. 901 Marcon Boulevard ALLENTOWN PA 18109-9341
Re: K100334
Diacap® Polysulfone HiFlo 23 Hemodialyzer Regulation Number: 21 CFR §876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: II Product Code: KDI Dated: March 30, 2010 Received: April 1, 2010
Dear Ms. Caravella:
This letter corrects our substantially equivalent letter of April 22, 2010. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth
{3}------------------------------------------------
Page 2 - Ms. Angela J. Caravella
in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Herbert Kemen
Janine M. Morris Acting Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
B. Braun Avitum AG Special SIO(k) Premarket Notification Diacap Polysulfone HiFlo 23 Hemodialyzer
| 4. INDICATIONS FOR USE STATEMENT | |
|-------------------------------------------------------------------------------------------------------------|-------------------------------------------|
| | Page 1 of 1 |
| 510(k) Number (if known): | K100334 |
| Device Name: | Diacap® Polysulfone HiFlo 23 Hemodialyzer |
| Indications For Use: | |
| The Diacap® Polysulfone HiFlo 23 Hemodialyzer is designed for single use in acute and chronic hemodialysis. | |
| Prescription Use X (Per 21 CFR 801.109) | OR Over-The-Counter Use |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Signature
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 00334 510(k) Number.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.