MODIFICATION TO:FRESENIUS 2008T HEMODIALYSIS MACHINE, MODEL 190573
K093902 · Fresenius Medical Care North America · KDI · May 27, 2010 · Gastroenterology, Urology
Device Facts
Record ID
K093902
Device Name
MODIFICATION TO:FRESENIUS 2008T HEMODIALYSIS MACHINE, MODEL 190573
Applicant
Fresenius Medical Care North America
Product Code
KDI · Gastroenterology, Urology
Decision Date
May 27, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Fresenius 2008T is indicated for acute and chronic hemodialysis
Device Story
Fresenius 2008T Hemodialysis Machine modified with CDX option; incorporates standard Windows/Linux PC board for data management. Device takes treatment data as input; transfers data directly to in-center medical information systems via machine's alphanumeric keyboard. Added support for Citrasate and DRYalysate acid concentrates; user interface updated to display concentrate constituents. Used in clinical settings by trained healthcare professionals (nurses/technicians) under physician direction. Provides hemodialysis therapy; facilitates electronic data transfer to improve clinical documentation and workflow. Benefits include streamlined data management and expanded compatibility with specific acid concentrates.
Clinical Evidence
Bench testing only. No clinical data. Validation included software regression testing, EMC testing per IEC 60601-1-2 (2007), and electrical safety testing per UL 60601-1 and CAN/CSA-C22.2 No. 601.1-M90. System performance testing confirmed functionality of Citrasate acid concentrate integration.
Technological Characteristics
Hemodialysis delivery system. Modifications: integrated PC board (Windows/Linux capable) for data management. Connectivity: interface to in-center medical information systems. Electrical safety: UL 60601-1, CAN/CSA-C22.2 No. 601.1-M90. EMC: IEC 60601-1-2 Class A. Software: data management integration.
Indications for Use
Indicated for acute and chronic dialysis therapy in patients requiring hemodialysis. Must be administered by trained operators at the direction of a physician.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”
{0}------------------------------------------------
This 510(k) Summary is in accordance with the requirements of 21CFR Part 807.92
# A. Submitter's Information:
MAY 2 7 2010
| Name: | Fresenius Medical Care North America |
|----------------------|---------------------------------------------------|
| Address: | 920 Winter Street<br>Waltham, MA 02451 |
| Phone: | 1-781-699-4475 |
| Fax: | (781) 699-9635 |
| Contact Person: | Janet Kay, Director of Regulatory Affairs-Devices |
| Date of Preparation: | December 21, 2009 |
- B. Device Name: Fresenius 2008T Hemodialysis Machine
Common Name: Hemodialysis Machine
Product Code/Classification Panel: 78 KDI / Gastroenterology/Urology
Classification: Class II per § 876.5860
- C. Legally Marketed Predicate Device (unmodified devices); Fresenius 2008T Hemodialysis Machine
# D. Device Description:
- 1. The optional CDX modification incorporates a standard Windows/Linux capable computer (PC board) into the 2008T giving clinics the option to load data management software into the machine. With data management software on the machine the user may then interface with the program, using the machines preexisting alphanumeric keyboard, then transfer treatment data directly from the 2008T machine to their In-center medical information system.
- 2. Additionally, the selection of Citrasate® (K000792) and DRYalvsate® (K980659) acid concentrates have been added to the selection of acid concentrates available to our customers on the 2008T hemodialysis machine. The user interface has been changed to allow the user to
{1}------------------------------------------------
select Citrasate/DRYalysate from the dialysis screen of the 2008T hemodialysis machine and to display the proper constituents of the concentrate.
#### E. Indications for Use:
The Fresenius 2008T is indicated for acute and chronic dialysis therapy.
### Substantial Equivalence Decision Making Process
#### 1. Is the product a device?
YES - The modified Fresenius 2008T Hemodialysis Machine is a device pursuant to 21 CFR §201 [321] (h).
### 2. Does the new device have the same intended use?
YES - The intended use for the modified 2008T Hemodialysis Machine is identical to that for the unmodified 2008T (K080964) as follows:
#### Intended Use
The Fresenius 2008T is indicated for acute and chronic hemodialysis
# 3. Does the device have technological characteristics that raise new types of safety or effectiveness questions?
NO - The Fresenius 2008T with CDX option is a modified version of the Fresenius 2008T Dialysate Delivery System. The performance and technological characteristics of the modified Fresenius 2008T are equivalent to those of the Fresenius 2008T (K080964) Dialysate Delivery System and raise no new types of safety or effectiveness questions.
### 4. Does descriptive or performance information demonstrate equivalence?
YES - Fresenius Medical Care North America believes the information provided in this submission clearly describes the modified Fresenius 2008T and demonstrates that it is substantially equivalent to the original Fresenius 2008T Dialysate Delivery system.
# F. Safety Summary
The Fresenius 2008T hemodialysis machine with CDX option incorporates a standard Windows/Linux capable computer (PC board) into the machine. Additionally, Citrasate/DRYalysate was added to the 2008T's selectable acid concentrates. All water requirements, module options, functional options, performance limits, control parameters, compatible bloodlines.
{2}------------------------------------------------
and language options remain unchanged from the predicate device, the 2008T (K080964). A Risk Analysis has been completed and potential hazards associated with the modifications have been identified, mitigated and where applicable mitigations have been verified. All potential risks were deemed acceptable after mitigation.
# G. Summary of non-clinical tests submitted with the premarket notification of the device
- 1. Full system validation and software regression testing was performed to ensure that the modifications to the 2008T hemodialysis functions as intended and that the modifications did not negatively impact the overall 2008T hemodialysis machine system. Testing included:
- . Software validation and regression testing
- · Electromagnetic compatibility (EMC) testing
- · Electrical safety testing
- · System performance testing using Citrasate® acid concentrate
The results from the testing demonstrated that all modifications functioned as intended and met pre-determined acceptance criteria.
- 2. Electromagnetic compatibility testing (EMC) was conducted according to the IEC 60601-1-2 (2007) Class A Testing at CKC Laboratories, Inc. The modified 2008T hemodialysis machine, outlined in this submission, met the requirements for IEC 60601-1-2 devices.
- 3. Electrical safety testing was conducted according to the following standards:
- UL 60601-1, 1st Edition, 2006-04-26, (Medical Electrical Equipment, . Part 1:General Requirements for Safety)
CAN/CSA-C22.2 No. 601.1-M90, 2005, (Medical Equipment Electrical . Equipment, Part 1: General Requirements for Safety) The modified 2008T hemodialysis machine with was found to comply with the above standards.
# H. General Safety and Effectiveness Concerns
Operators of the 2008T Hemodialysis machine with CDX option must be trained to administer hemodialysis at the direction of a physician. In addition, the operator should be:
- . Knowledgeable of hemodialysis methodology and relevant physiology.
- . Proficient in healthcare procedures regarding aseptic techniques.
{3}------------------------------------------------
- Thoroughly familiar with the contents of the Operator's manual.
Fully trained and qualified to operate this machine, and able to distinguish between normal and abnormal operation.
.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal for the Department of Health & Human Services USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center . WO66-G6( Silver Spring, MD 20993-0002
Ms. Janet C. Kay Director of Regulatory Affairs Renal Therapies Group Fresenius Medical Care North America 920 Winter Street WALTHAM MA 02451
MAY 2 7 2010
Re: K093902
Trade/Device Name: Fresenius 2008T Hemodialysis Machine Regulation Number: 21 CFR §876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: II Product Code: KDI Dated: April 27, 2010 Received: April 28, 2010
Dear Ms. Kay:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{5}------------------------------------------------
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Janine M. Morris
Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{6}------------------------------------------------
Indications for Use
510(k) Number (if known): K093902
Device Name: Fresenius 2008T Hemodialysis Machine
Indications for Use:
Fresenius 2008T is indicated for acute and chronic dialysis therapy
Prescription Use _ √ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Huld Reum
(Division Sign-Off)
Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
Page_1_of_1
127
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.