The Edwards Aquarius System is indicated for continuous solute and/or fluid removal in patients with acute renal failure or fluid overload. The Edwards Aquarius system may also be used in Therapeutic Plasma Exchange (TPE) therapies. The Edwards Aquarius system is indicated for use in a clinical setting and not for home use.
Device Story
Portable automated fluid balance monitor for renal replacement or plasma therapies. System comprises three circuits: extracorporeal (blood), substitution/dialysate, and filtrate. Uses two scales for precise filtration/substitution volume balancing. Includes heparin pump for anticoagulation; 2-channel protective system for patient safety; manual hand-crank for blood pump override. Operated by clinicians in clinical settings. Provides continuous solute/fluid removal or TPE. Benefits patients by managing fluid overload and renal failure. Software version 6.02.00 manages system functions.
Clinical Evidence
No clinical data presented. Submission is a special 510(k) for software and labeling modifications; evidence consists of design control activities and risk/hazard analysis.
Technological Characteristics
Portable hemodialysis system; 170cm x 50cm x 60cm; 75kg. Power: 104-126V, 47-63Hz. Features dual-scale fluid management, pressure monitoring, degassing unit, and fluid warmer. Software v6.02.00. Complies with UL and EMC safety standards.
Indications for Use
Indicated for patients with acute renal failure or fluid overload requiring continuous solute and/or fluid removal, or Therapeutic Plasma Exchange (TPE). For clinical use only; not for home use.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”
{0}------------------------------------------------
NOV 1 9 2009
Image /page/0/Picture/2 description: The image shows a logo with the letter 'E' in a stylized font, set against a textured, square background. Below the letter, the word 'Edwards' is printed in a simple, sans-serif font. The overall design is clean and professional, suggesting a corporate or business identity. The letter 'E' is white, contrasting with the darker background.
Edwards Lifesciences Services GmbH Edisonstrasse 6 85716 Unterschleissheim Germany
Phone: 011 49 89 95475 203 Fax: 011 49 89 95475 305
## "510(k) Summary"
as required by section 807.92(c)
Summary:
This summary was prepared on June 9 ", 2009
Submitter:
Robert Madino Director of Regulatory Affairs Edwards Lifesciences Services GmbH Edisonstrasse 6 85716 Unterschleissheim Germany Phone: 011 49 89 95475 203 Fax: 011 49 89 95475 305 e-mail: robert madino@edwards.com
Owner:
Edwards Lifesciences Services GmbH Edisonstrasse 6 85716 Unterschleissheim Germany Establishment Registration Number: 3006550990
Contact Person:
Neil R. Armstrong Consultant and RA Advisor to Edwards MeddiQuest Limited BioPark Broadwater Road Welwyn Garden City Herts. AL7 3AX
{1}------------------------------------------------
·
.
:
.
| United Kingdom | |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Phone (UK): | 011 44 1763 222505 |
| Phone (USA): | (240) 949 2477 |
| Fax: | 011 44 1763 222049 |
| e-Mail: | neil.armstrong@meddiquest.com |
| or | |
| | |
| | Robert Madjno |
| | Director of Regulatory Affairs |
| | Edwards Lifesciences Services GmbH |
| | Edisonstrasse 6 |
| | 85716 Unterschleissheim |
| | Germany |
| Phone: | 011 49 89 95475 203 |
| Fax: | 011 49 89 95475 305 |
| e-mail: | robert_madjno@edwards.com |
| Device Trade Name: | Aquarius system |
| Classification Name: | High permeability hemodialysis system |
| Device Classification: | The FDA has classified: 21 CFR 876.5860 High Permeability Hemodialysis System (78 KDI) as a Class II. |
| Common Name: | Hemofiltration System |
| Predicate Device: | Edwards Aquarius system cleared under premarket notification K070320 on June 7, 2007 |
| Brief Description: | The general description of the device has not changed from the predicate version, K070320, except that the system has been validated to use specific legally market filter as opposed to a special Edwards Filter. |
The Edwards Aquarius System needs to be used in conjunction with a tubing set and a filter to provide Hemofiltration treatment to the patient.
The tubing set has already been the subject of a PreMarket Notification, K063293 Edwards Aqualine Sterile Tubing Sets and suitable legally marketed filters have been identified.
The Aquarius system is an Automated Fluid Balance Monitor, designed to be used with various extracorporeal treatments in the field of renal replacement therapies or plasma therapies. All therapies must be prescribed by a physician.
. . .
·
{2}------------------------------------------------
The Aquarius system is divided into three circuits: the extracorporeal (blood) circuit, the substitution/dialysate circuit and the filtrate circuit.
Toxic substances are removed by filters and clean blood is returned to the patient.
The Aquarius system allows the patient to be positioned left or right of the instrument.
The Aquarius system uses two scales to accurately measure and precisely balance filtration and substitution volumes.
Heparin may be supplied to the extracorporeal circuit via an anticoagulant pump (Heparin pump). The prescribing physician may select continuous or intermittent options.
The Aquarius protective system is designed as a 2-channel system to protect the patient from foreseeable danger.
At the back of the scale system a removable hand-crank is mounted. This can be used to manually turn the blood pump in case a pump stops.
The Aquarius system is portable. It has a wheeled base connected with a handle to move or carry the Aquarius.
Intended Use:
Continuous Renal Replacement Therapy: SCUF. CVVH. CVVHD. CVVHDF Plasma exchange (TPE)
Indications for Use:
"The Aquarius System is indicated for continuous solute and/or fluid removal in patients with acute renal failure or fluid overload. The Aquarius System may also be used in Therapeutic Plasma Exchange (TPE) therapies.
"The Aquarius System is indicated for use in a clinical setting and not for home use."
These are the same indications for use as the predicate Aquarius System cleared under K070320.
Substantial Equivalence: The Aquarius System is substantially equivalent to the predicate Aquarius System as follows:
- same intended use .
- same indications for use .
- same labels .
{3}------------------------------------------------
- revised operating manual based on the draft labeling supplied in the . predicate submission
- same operating principle ●
- . same fundamental scientific technology with the exception of the algorithm improvements
- . same hardware
- updated software (6.02.00) performing the same functions as the . predicate software (version 6.00.04)
- . same performance specifications
- same accessories .
- . continues to meet the safety requirements of UL and EMC standards
In summary, the Aquarius System is substantially equivalent to the predicate Aquarius System (K070320), with the same intended use and fundamental scientific technology.
| Technological | |
|------------------|--|
| | |
| Characteristics: | |
| Characteristic | Aquarius System<br>K070320<br>(Predicate Device) | Modified Aquarius System<br>(Proposed Device) |
|-----------------------------------|--------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| Weight | Approx. 75 Kg | Same as predicate. |
| Dimensions (H x W x<br>D) | 170 cm x 50 cm x 60 cm | No physical change,<br>but quoted dimensions<br>corrected in Operating<br>Manual |
| Power Supply | 115 v (a.c.) ± 10% at<br>60Hz (U.S.) | 104 - 126 v at 47 - 63<br>Hz |
| Power Consumption | 500 VA (theoretical<br>value) | Maxium measured<br>value 350w |
| Fluids Management | All parameters*, except:<br>Heparin syringe size 30<br>ml or 50 ml | Same as predicate,<br>except<br>50 ml only |
| Scales | All parameters* | Same as predicate. |
| Fluid Warmer | All parameters* | Same as predicate. |
| Pressure Monitoring | All parameters* | Same as predicate. |
| Degassing Unit | All parameters* | Same as predicate. |
| Monitor / Detection<br>Parameters | All parameters* | Same as predicate. |
| Software Revision | 6.00.04 | 6.02.00 (see above for<br>description of<br>differences between<br>software revisions) |
{4}------------------------------------------------
* refers to the parameters listed in the 510(k) summary for K070320 bttp://www.fda.gov/cdrh/pdf7/K070320.pdf
Performance Testing:
The submission is a special 510(k) for changes to software and labeling. No clinical or non-clinical data is presented, however a summary of design control activities and risk/hazard analysis for software changes is presented.
Conclusion:
The proposed modifications, included in this submission, update the software and labeling for the device. Edwards Lifesciences Services GmbH have demonstrated that these modifications do not alter the fundamental scientific technology or the intended use.
Edwards Lifesciences believes that these modifications are eligible for review through the Special 510(k) process and that the proposed device is substantially equivalent to the predicate device.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the seal is a stylized image of what appears to be an abstract representation of a human figure or symbol. The image is in black and white.
## EPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Edwards Lifesciences Services GMBH c/o Mr. Neil R. Armstrong Consultant and RA Advisor to Edwards Meddiquest Limited Biopark, Broadwater Road Welwyn Garden City, Herts. UNITED KINGDOM AL7 3AX
NOV 1 9 2009
Re: K090682
Trade/Device Name: Edwards' Aquarius system Regulation Number: 21 CFR §876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: II Product Code: KDI Dated: October 13, 2009 Received: October 20, 2009
## Dear Mr. Armstrong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. .
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{6}------------------------------------------------
Page 2 -
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Indications for Use
510(k) Number (if known): K090682
Device Name:
Edwards' Aquarius system
Indications for Use:
The Edwards Aquarius System is indicated for continuous solute and/or fluid removal in patients with acute renal failure or fluid overload. The Edwards Aquarius system may also be used in Therapeutic Plasma Exchange (TPE) therapies.
The Edwards Aquarius system is indicated for use in a clinical setting and not for home use.
| Prescription Use | X | AND/OR | Over-The-Counter Use | |
|-----------------------------|---|--------|------------------------|--|
| (Part 21 CFR 801 Subpart D) | | | (21 CFR 801 Subpart C) | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|---------------------------------------------------------------|---------|
| Division of Reproductive, Abdominal, and Radiological Devices | |
| 510(k) Number | K690682 |
| (Posted November 13, 2003) | Page 1 of 1 |
|----------------------------|-------------|
|----------------------------|-------------|
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.