NXSTAGE DIALYSATE PREPARATION MODULE, MODEL NX0269
K043436 · Nxstage Medical, Inc. · FKR · Mar 17, 2005 · Gastroenterology, Urology
Device Facts
Record ID
K043436
Device Name
NXSTAGE DIALYSATE PREPARATION MODULE, MODEL NX0269
Applicant
Nxstage Medical, Inc.
Product Code
FKR · Gastroenterology, Urology
Decision Date
Mar 17, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5820
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NxStage Dialysate Preparation Module is an optional accessory to the NxStage System One that prepares dialysate for use during hemodialysis, as prescribed by the physician.
Device Story
The NxStage Dialysate Preparation Module is an accessory to the NxStage System One hemodialysis system. It functions by mixing purified water (AAMI standard) with AAMI-compliant concentrate to produce dialysate meeting ANSI/AAMI RD52:2004 and RD62:2001 standards. The device incorporates safety features including conductivity measurement, an automatic safety clamp, a mix-to-use timer, and ultrafilters/I.V. sterilization filters to ensure dialysate purity. It is intended for use in hemodialysis therapies as prescribed by a physician. The module automates the preparation process, ensuring the dialysate meets required purity and composition specifications before use in the dialysis circuit, thereby supporting safe and effective hemodialysis treatment.
Clinical Evidence
No clinical data. Performance was demonstrated through bench testing and simulated use testing to characterize the module's performance and verify safety features against predicate device standards.
Technological Characteristics
Proportioning subsystem for hemodialysis. Mixes purified water and concentrate to produce AAMI-compliant dialysate. Features: conductivity measurement, automatic safety clamp, mix-to-use timer, ultrafilters, and I.V. sterilization filters. Designed for use with the NxStage System One.
Indications for Use
Indicated for use as an optional accessory to the NxStage System One to prepare dialysate for hemodialysis patients as prescribed by a physician. No known contraindications.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
{0}------------------------------------------------
#### NxStage Medical, Inc. NxStage Dialysate Prep Module MAR 1 7 2005 510(k) Premarket Notification
K043436
Page 1 of 2
# Section VIII: 510(k) Summary of Safety & Effectiveness
This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of SMDA 1990.
#### A. Submitter's Information:
| Name: | NxStage Medical, Inc. |
|----------------------|---------------------------------------------------------|
| Address: | 439 South Union Street, Suite 501<br>Lawrence, MA 01843 |
| Phone: | (978) 687-4700 |
| Fax: | (978) 687-4800 |
| Contact Person: | Norma LeMay<br>Manager, Regulatory Affairs |
| Date of Preparation: | December 10, 2004 |
#### B. Device Name:
| Trade Name: | NxStage Dialysate Preparation Module |
|----------------------|--------------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Subsystem, proportioning |
| Classification Name: | Hemodialysis System & Accessories (Class I<br>medical device under 21 CFR 875.5820, (Product<br>Code 78 FKR) |
### C. Substantial Equivalence/Predicate Devices:
The NxStage Dialysate Prep Module is substantially equivalent to the following legally marketed predicate devices previously cleared by FDA:
- Aksys PHD System, K010131 (cleared on 03/26/02) .
- Fresenius 2008K Hemodialysis System, K994267 (cleared on 03/16/00) .
- Gambro Bicarbonate Module BCM-10, K810127 (cleared on 02/25/81) .
{1}------------------------------------------------
# NxStage Medical, Inc. NxStage Dialysate Prep Module 510(k) Premarket Notification
K043436
Page 2 of 2
# Section VIII: 510(k) Summary of Safety & Effectiveness
# D. Device Description/Indications for Use:
Device Booonproniment and optional accessory to the NxStage System The NAGtage Dialyone in that mixes with concentrate to produce One that pro treats crequirements of the ANSI/AAMI Standards, RD52:2004 and The dialysate is to be used with the NxStage System One during RD62:2001. hemodialysis therapies.
### Indications for use:
The NxStage Dialysate Preparation Module is an optional accessory to the NxStage rne NAGtago Dialysate for use during hemodialysis, as prescribed by the physician.
There are no known contraindications with the use of the NxStage Dialysate Preparation Module.
## E. Technological Characteristics:
The proposed Prep Module has the same technological characteristics and is similar in design and configuration as compared to the predicate devices. It uses purified AAMI water that mixes with AAMI concentrate to produce AAMI dialysate. The proposed device is designed and assembled with components commonly found in the predicate devices. Importantly, the NxStage Prep Module incorporates the proper safety features, such as conductivity measurement, automatic safety clamp, mix to use timer, and ultrafilters and I.V. sterilization filters to insure dialysate purity.
## F. Summary of Non-Clinical Test/Performance Testing
NxStage believes that the information and data provided in this submission clearly describes the proposed device and demonstrates that the device is adequately designed for the labeled indication for use.
As there is no technological differences as compared to the predicate devices, performance testing was deemed appropriate and was conducted to characterize the performance of the proposed Prep Module to provide a basis of comparison to the predicate devices. The Prep Module utilizes FDA cleared devices with minor modifications, and additional components which are simple to operate using the Verification and Validation testing, which includes provided device labeling. simulated use testing, sufficiently demonstrates that the Prep Module can be used safely as intended. All testing performed is summarized in Section V of this 510(k) submission. Results of the testing presented in this 510(k) have demonstrated that the NxStage Prep Module is substantially equivalent to the predicate devices and is suitable for the labeled indication for use.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a circular seal or logo. The seal features an abstract image of a bird in flight, composed of three curved lines. Encircling the bird is text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES U.S.A."
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 7 2005
Ms. Norma LeMay Manager, Regulatory Affairs NxStage® Medical, Inc. 439 South Union Street, 5th Floor LAWRENCE MA 01843
Re: K043436
K043430
Trade/Device Name: NxStage® Dialysate Preparation Module Regulation Number: 21 CFR §876.5820 Regulation Name: Hemodialysis system and accessories Regulatory Class: II Product Code: 78 KPO and FKR Dated: March 3, 2005 Received: March 4, 2005
Dear Ms. LeMay:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced we have review of your better of substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to the cholosate) to regary mantowa of the Medical Device Amendments, or to devices that have been ready 20, 1776, the chance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that recrassition in accercation in premarket approval application (PMA). You may, therefore, market the do not require upproval er a ve visions of the Act. The general controls provisions of the Act de fice, subjoct to the general registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket It your devise is elective (t to such additional controls. Existing major regulations affecting your Applorary, the lia) 00 bely works of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that I reast of advised that 127 i our device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the I carral statues and regulations association and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) (21 CFR Part 801); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will allow you to begin marketing your actrice of your device of your device to a legally
premarket notification. The FDA finding of substantial equipand this pe premarket notification. The FDA Imaling of substantial equivalent of your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your device on our labering regarantists.
contact the Office of Compliance at one of the following numbers, based on the regulation number a the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Also, please note the regulation entitled, "onsonmants" of the sunder the Act from the 807.97). You may obtain other gelleral intornation on your respeared its toll-free number (800)
Division of Small Manufacturers, International and Consumer Assistance at its Division of Small 1143-6597 or at its Internet address 056-2041 or (2019/01/2019/11/industry/support/index.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
| 510(k) Number (if known): | K043436 |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | NxStage Dialysate Preparation Module |
| Indications for Use: | The NxStage Dialysate Preparation Module is an optional<br>accessory to the NxStage System One that prepares<br>dialysate for use during hemodialysis, as prescribed by the<br>physician. |
Prescription Use _____________________________________________________________________________________________________________________________________________________________ Use (Part 21 CFR 801 Subpart D)
Over-The-Counter AND/OR
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C. Brogdon
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number _
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.