K120130 · Medigroup, Inc. · FJS · May 21, 2012 · Gastroenterology, Urology
Device Facts
Record ID
K120130
Device Name
FALLER TROCAR
Applicant
Medigroup, Inc.
Product Code
FJS · Gastroenterology, Urology
Decision Date
May 21, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Faller Trocar is used to create an ante-grade subcutaneous tunnel for a peritoneal dialysis catheter from the initial implantation site to the preferred exit-site. It separates the tissue as it leads the catheter through to the skin exit-site.
Device Story
Faller Trocar is a surgical tool used by physicians to create subcutaneous tunnels for peritoneal dialysis catheters. Device features a sharp tip for tissue separation and a barbed end for catheter attachment. Available in stainless steel or approved plastic; curved design facilitates ante-grade tunneling from implantation site to skin exit-site. Operates as a manual surgical instrument; no electronic or automated components. Used in clinical settings during catheter implantation procedures. Benefits include facilitating precise catheter placement, reducing procedural complexity, and leveraging established design principles proven by decades of clinical use.
Clinical Evidence
Bench testing only. In-house evaluation compared barbed ends and cutting tips of the subject device against predicate devices. No new clinical trials were conducted; safety and effectiveness are supported by historical data and 20+ years of clinical use of the predicate devices.
Technological Characteristics
Manual surgical instrument. Materials: stainless steel or approved plastic. Design: curved shaft with sharp distal tip and proximal barbed end for catheter attachment. No energy source, connectivity, or software. Sterilization method not specified.
Indications for Use
Indicated for creating an ante-grade subcutaneous tunnel for a peritoneal dialysis catheter from the initial implantation site to the preferred exit-site in patients requiring peritoneal dialysis.
Regulatory Classification
Identification
(1) A peritoneal dialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions, and that consists of a peritoneal access device, an administration set for peritoneal dialysis, a source of dialysate, and, in some cases, a water purification mechanism. After the dialysate is instilled into the patient's peritoneal cavity, it is allowed to dwell there so that undesirable substances from the patient's blood pass through the lining membrane of the peritoneal cavity into this dialysate. These substances are then removed when the dialysate is drained from the patient. The peritoneal dialysis system may regulate and monitor the dialysate temperature, volume, and delivery rate together with the time course of each cycle of filling, dwell time, and draining of the peritoneal cavity or manual controls may be used. This generic device includes the semiautomatic and the automatic peritoneal delivery system.(2) The peritoneal access device is a flexible tube that is implanted through the abdominal wall into the peritoneal cavity and that may have attached cuffs to provide anchoring and a skin seal. The device is either a single use peritioneal catheter, intended to remain in the peritoneal cavity for less than 30 days, or a long term peritoneal catheter. Accessories include stylets and trocars to aid in the insertion of the catheter and an obturator to maintain the patency of the surgical fistula in the abdominal wall between treatments. (3) The disposable administration set for peritoneal dialysis consists of tubing, an optional reservoir bag, and appropriate connectors. It may include a peritoneal dialysate filter to trap and remove contaminating particles. (4) The source of dialysate may be sterile prepackaged dialysate (for semiautomatic peritoneal dialysate delivery systems or “cycler systems”) or dialysate prepared from dialysate concentrate and sterile purified water (for automatic peritoneal dialysate delivery systems or “reverse osmosis” systems). Prepackaged dialysate intended for use with either of the peritoneal dialysate delivery systems is regulated by FDA as a drug.
Special Controls
*Classification.* Class II (special controls). The following accessories are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9: A catheter finger grip that is non-patient contacting and intended for single use with a peritoneal catheter; a continuous ambulatory peritoneal dialysis (CAPD) belt; and a catheter stand that does not include weigh scales.
{0}------------------------------------------------
K120130 R
MAY 2 1 2012
Image /page/0/Picture/2 description: The image shows the logo for MEDIGROUP, Inc. The logo consists of a stylized heart shape above the text "MEDIGROUP, Inc." The heart shape is made up of two interlocking shapes, with diagonal lines filling the space between them.
# 510(k) Summary
# Basic Information
| Submitter: | Medigroup Inc |
|------------|---------------|
|------------|---------------|
14A Stonehill Rd
Oswego IL 60543
Establishment Registration Number: #1450420 Contact: Irene K. Navis Vice-President, Finance & Regulatory Telephone: 630-254-2533 Fax: 630-554-2535 E-mail: inavis@medigroupinc.com Date of Submission: January 13, 2012
# Device Information
Trade Name: Common Name: Product Code: Class:
Faller Trocar Catheter Tunneling Device FJS, Accessory =
#### Predicate Devices
K910786 Trocar (Faller Stylet) K823331 Tunnelor Tool
### Product Description
The Faller Trocar is a surgical tool, made of either stainless steel or approved plastic, with a barbed section on one end and a sharp tip on the other end. The device is curved so that the physician using it can follow a tunnel path to lead the catheter from the initial catheter implantation site to the exit-site in an ante-grade manner.
#### Intended Use
The Faller Trocar is used to create an ante-grade subcutaneous tunnel for a peritoneal dialysis catheter from the initial implantation site to the preferred exit-site. It separates the tissue as it leads the catheter through to the skin exit-site.
MEDIGROUP, Inc
14 A Stonehill Road, Oswego, IL 60543-9400 USA 1 (630) 524-5533 | (800) 323-5389 | E (630) 254-5235 www.medigroupinc.com
Image /page/0/Picture/21 description: The image contains two logos. The logo on the left is a circular seal with the letters 'R' in the center. The text around the circle reads 'LLOYD'S REGISTER QUALITY ASSURANCE ISO 14001'. The logo on the right is a square with the letters 'UL' inside.
Image /page/0/Picture/22 description: The image shows a logo with a crown on top of a check mark. Below the check mark is the text "UKAS" in bold letters. Underneath "UKAS" are two lines of smaller text that are difficult to read. At the very bottom of the logo is the number "001".
{1}------------------------------------------------
K12015
Image /page/1/Picture/1 description: The image shows the logo for MEDIGROUP, Inc. The logo consists of a stylized heart shape above the text "MEDIGROUP, Inc." The heart shape is black and has a striped pattern at the bottom.
#### Substantial Equivalence
The new Faller Trocars have the same Indications for Use as the two predicate devices. They are all used to create a subcutaneous tunnel through which to lead the catheter from the initial implantation site through the exit-site. The predicate device Tunnelor® Tool does this in a retro-grade manor, as opposed to an ante-grade manner, but the end result is the same - getting the distal end of the catheter out through the exit-site in the skin.
The new Faller Trocars also have a barbed end to which the catheter is attached. This barbed end is exactly the same as the predicate, Tunnelor® Tool. The new Faller Trocars both have a sharp tip, just as the predicate Covidien's Trocar (Faller Stylet), which is inserted into the initial implantation site to open up the tunnel path through the subcutaneous tissue.
The new Faller Trocars are made of the same materials as the predicate devices, stainless steel like Covidien's Trocar (Faller stylet) and an approved plastic like Medigroup's Tunnelor® Tool. The shape and bend of the new stainless steel Faller Trocar is similar to the predicate Covidien's Trocar (Faller Stylet). The curvature of the new plastic Faller Trocar is similar to the predicate Medigroup's Tunnelor® Tool. The two different shapes give an option to the implanting surgeon to use which one he prefers.
#### Testing
The barbed ends and the cutting tip were tested in-house with acceptable comparisons between the new Faller Trocars and the two predicate devices. In addition, both of the predicate devices have been used by physicians for twenty years or more. Evaluation of Journal articles and historical data of the predicate devices indicate their safety and effectiveness. Because of the equivalency of the new Faller Trocars to those predicates, the new devices will function as safely and effectively.
#### Conclusions
The Faller Trocars, either made of stainless steel or an approved plastic, will function as designed and intended. Because of their substantial equivalence to the two predicate devices, they will be as safe and effective as the predicates which have a proven record of more than twenty years of successful use.
Image /page/1/Picture/12 description: The image contains two logos. The logo on the left is a circular logo with the text "LLOYD'S REGISTER QUALITY ASSURANCE" surrounding a large letter "R". Below the "R" is the text "ISO 1485". The logo on the right is a square logo with a crown above a check mark. Below the check mark is the text "UKAS QUALITY MANAGEMENT 001".
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three wavy lines extending from its body, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the upper portion of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Irene Navis Vice-President, Finance & Regulatory Medigroup, Inc. 14 A Stonehill Road OSWEGO IL 60543
MAY 2 1 2012
Re: K120130
> Trade/Device Name: Faller Trocar Regulation Number: 21 CFR§ 876.5630 Regulation Name: Peritoneal dialysis system and accessories Regulatory Class: II Product Code: FJS Dated: May 14, 2012 Received: May 16, 2012
Dear Ms. Navis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
Page 2-
device-related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Sincerery, yours,
Benjamin K. Evans
Benjamin R. Fisher, Ph.D. Director Division of Reproductive. Gastro-Renal. and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): · K120130
Device Name: Faller Trocar
Indications for Use: The Faller Trocar is used to create an ante-grade subcutaneous tunnel for a peritoneal dialysis catheter from the initial implantation site to the preferred exit-site. It separates the tissue as it leads the catheter through to the skin exit-site.
Prescription Use _ V (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Helene Leun
luctive, Gastro-Renal, and
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.