SERIM BLOOD LEAK TEST STRIP
Device Facts
| Record ID | K990206 |
|---|---|
| Device Name | SERIM BLOOD LEAK TEST STRIP |
| Applicant | Serim Research Corp. |
| Product Code | FJD · Gastroenterology, Urology |
| Decision Date | Apr 21, 1999 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.5820 |
| Device Class | Class 2 |
Indications for Use
SERIM Blood Leak Test Strips provide a rapid and convenient means for testing spent dialysate for blood in hemodialysis clinics. The presence of blood in dialysate indicates a leak in the dialysis membrane.
Device Story
Device consists of reagent pad on polystyrene handle; used to detect occult blood in hemodialysis fluids. Operator immerses pad in dialysate sample; reacts for 60 seconds; compares color to bottle label blocks (Negative/Positive). Principle: peroxidase-like activity of hemoglobin catalyzes oxidation of chromogen to green color. Positive result indicates hemoglobin concentration ≥1.5 mg/dL (approx. 0.1 mL/min blood loss). Used by dialysis technicians to confirm photometric sensor alarms, reducing unnecessary session interruptions. Benefits: quick, convenient confirmation of membrane leaks; prevents premature termination of dialysis when false alarms occur.
Clinical Evidence
Bench testing and clinical comparison performed. 12 participants tested blood standards in blind studies (n=100 negative, n=100 positive) achieving 100% sensitivity and 100% specificity. Comparison study with 101 clinical dialysate samples showed equivalent performance to predicate Chemistrip occult blood strips; 33 spiked samples (1.5 mg/dL hemoglobin) correctly identified as positive by both methods.
Technological Characteristics
Reagent pad containing organic peroxide and chromogen attached to polystyrene handle. Sensing principle: peroxidase-like catalytic oxidation of chromogen by hemoglobin. Dimensions: 0.2 x 0.2 inch pad on 0.2 x 3.25 inch handle. Visual readout via colorimetric comparison. Non-electronic, manual test strip.
Indications for Use
Indicated for use in hemodialysis clinics to test spent dialysate for the presence of blood, which indicates a leak in the dialysis membrane.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
Predicate Devices
- Chemistrip 10 urine test strip (K896454)
Reference Devices
- cyanomethemoglobin assay