90XL INSTRUMENTATION SYSTEM
Device Facts
| Record ID | K050812 |
|---|---|
| Device Name | 90XL INSTRUMENTATION SYSTEM |
| Applicant | Mesa Laboratories, Inc. |
| Product Code | FIZ · Gastroenterology, Urology |
| Decision Date | May 5, 2005 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 876.5820 |
| Device Class | Class 2 |
Indications for Use
The 90XL Dialysis Meter is intended for use by hemodialysis professionals to measure the conductivity, temperature, pH, and pressure of the dialysate solution associated with hemodialysis delivery systems. Water purification specialists may also use the 90XL instrumentation system to verify proper characteristics of water used in hemodialysis. These parameters are key indicators of system performance that require periodic monitoring and verification in hemodialysis systems.
Device Story
90XL Instrumentation System measures conductivity, temperature, pH, and pressure of dialysate and water used in hemodialysis delivery systems; operated by hemodialysis professionals and water purification specialists; provides quantitative measurements to verify system performance; periodic monitoring of these parameters ensures proper function of hemodialysis equipment; aids clinical decision-making regarding system maintenance and patient safety.
Technological Characteristics
Instrumentation system for measuring conductivity, temperature, pH, and pressure; intended for use in hemodialysis and water purification environments.
Indications for Use
Indicated for use by hemodialysis professionals and water purification specialists to monitor and verify conductivity, temperature, pH, and pressure parameters in hemodialysis delivery systems and water purification systems used in hemodialysis.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.